Search Results
Found 2 results
510(k) Data Aggregation
(154 days)
QardioArm 2
The QardioArm 2 is a fully automatic, wireless blood pressure monitor intended to measure the systolic and diastolic blood pressures and pulse rate of adult individuals. It utilizes an inflatable cuff that is wrapped around the upper arm. The arm circumference is limited to 8.7 - 18.5 inches (22 - 47 cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. It is not intended to be used as a diagnostic device and users are instructed to contact their physician if hypertensive values are indicated.
The QardioArm 2 is an over-the-counter (OTC) non-invasive blood pressure monitor intended for spot-checking in adults. It uses a detachable inflatable cuff that is wrapped around the upper arm to determine blood pressure based on the oscillometric method. Two cuff sizes are available: standard size for an arm circumference of 8.7 to 14.6 inches or large size for an arm circumference of 14.6 to 18.5 inches.
The QardioArm 2 is a non-invasive blood pressure measurement system. The acceptance criteria and the study that proves the device meets them are detailed below.
1. A table of acceptance criteria and the reported device performance
Unfortunately, the provided text does not contain a specific table of acceptance criteria or reported device performance in the format of a direct comparison table. However, it states that the device "met all the requirements of the standard" and "passed all tests" for various criteria.
Based on the general information provided, we can infer some key performance criteria and the reported performance:
Acceptance Criteria Category | Specific Criteria (Inferred from standards) | Reported Device Performance |
---|---|---|
Accuracy (Blood Pressure) | As per ISO 81060-2:2018 (Non-invasive sphygmomanometers — Part 2: Clinical investigation of automated measurement type), which is a recognized standard for blood pressure monitor accuracy. This standard typically requires a mean difference between the device and reference standard of ≤ ±5 mmHg, and a standard deviation of differences ≤ 8 mmHg. (Implicit) | "The device met all the requirements of the standard [ISO 81060-2:2018]." This implies the accuracy criteria of ISO 81060-2 were met. The technological comparison section explicitly states an accuracy of "+/-3mmHg or 2% reading for blood pressure." |
Accuracy (Pulse) | As per ISO 81060-2:2018 or similar standards for pulse measurement. (Implicit) | The technological comparison section explicitly states an accuracy of "+/-5% for pulse." |
Electrical Safety | Compliance with IEC 60601-1:2005 (3rd ed) + CORR. 1:2006 + CORR.2:2007+A1:2012 Medical electrical equipment: Part 1: General requirements for basic safety and essential performance. | "The device passed all tests, including the US deviations." |
Electromagnetic Compatibility (EMC) | Compliance with IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: electromagnetic disturbances – Requirements and tests. | "The QardioArm 2 passed all tests." |
Battery Safety | Compliance with IEC 62133-2:2017, Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems. | "found to comply with all relevant sections." |
Wireless Coexistence | Compliance with AAMI TIR69:2017 and ANSI C63.27:2017. | "All tests passed." |
Software Verification & Validation | Adherence to FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," for a "Moderate" level of concern. | "Software verification and validation testing were conducted, and documentation was provided as recommended... The software for this device was considered as a 'Moderate' level of concern." While not explicitly stating "passed," meeting the guidance for a moderate level of concern implies successful validation. |
Home Healthcare Environment | Compliance with IEC 60601-1-11:2015, Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. | "The device passed all tests." |
Automated Non-Invasive Sphygmomanometer Performance | Compliance with IEC 80601-2-30:2018, Medical electrical equipment – Part 2: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers. | "The device passed all tests." |
Human Factors | Adherence to FDA guidance "Applying Human Factors and Usability Engineering to Medical Devices," demonstrating the device is safe and effective for its intended users, uses, and use environments. | "The QardioArm 2 was found to be safe and effective for its intended users, uses, and use environments." |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Clinical Data (Test Set): 86 subjects.
- Data Provenance: The study was a "prospective, single arm paired comparison study." The document does not specify the country of origin of the data, but it was conducted to demonstrate performance in accordance with ISO 81060-2:2018 and FDA's requirements for 510(k) submissions, suggesting a standard clinical study environment.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
The study description mentions a "same arm sequential method with dual auscultators." This implies that the ground truth for blood pressure measurements was established by two human observers (auscultators) using a reference sphygmomanometer. The document does not specify the number of experts beyond "dual auscultators" or their specific qualifications (e.g., years of experience, medical role). However, for such studies, these individuals are typically trained and validated healthcare professionals.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
The method used was a "same arm sequential method with dual auscultators." While not explicitly stated as "2+1" or "3+1" (which are typically used for imaging assessments), the use of dual auscultators suggests a comparison or consensus method between the two observers for establishing reference measurements. Differences between the dual auscultators would typically be adjudicated or the reading discarded according to the ISO standard's protocol, though the specific adjudication method is not described in detail.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This device is a non-invasive blood pressure monitor, not an AI-assisted diagnostic imaging device that involves human interpretation of images. The study focused on the device's accuracy compared to a reference standard, not on human reader performance with or without AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, in essence, a standalone performance assessment was done. The QardioArm 2 measures blood pressure automatically (oscillometric method) and provides a reading. The clinical study directly compared these device-generated readings against reference measurements obtained by human auscultators. The device operates autonomously to produce the blood pressure and pulse rate values.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for blood pressure measurements was established using a reference sphygmomanometer via a same arm sequential method with dual auscultators. This constitutes an expert consensus or expert-derived reference standard for physiological measurements, rather than pathology or outcomes data.
8. The sample size for the training set
The document does not provide information about a separate "training set" or its sample size. This type of device's core functionality (oscillometric measurement) is based on established principles and algorithms, often developed and refined internally. For a 510(k) submission of a blood pressure monitor, the primary focus is on the performance of the final device as compared to a reference standard in a clinical test set, rather than detailing the training data for an AI algorithm.
9. How the ground truth for the training set was established
As no specific training set data is mentioned for an AI algorithm, the method for establishing ground truth for a training set is not applicable or detailed in this document. The device's operation is based on physical measurement principles rather than a deep learning model requiring a large labeled training dataset in the same way an image analysis AI would.
Ask a specific question about this device
(159 days)
QARDIOARM
QardioArm, model A100 is a fully automatic, non-invasive, wireless blood pressure monitor. QardioArm, model A100 is a blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual. QardioArm, model A100 utilizes an inflatable cuff that is wrapped around the upper arm.
This device is not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated. The cuff circumference is limited to 22cm-37cm (8.7in -14.6in)
QardioArm, model A100 measures both systolic and diastolic blood pressure and heart pulse rate via a standard oscillometric method. The oscillometric method senses the vibrating signal via the closed air pipe system and utilizes a microcomputer to automatically sense the characteristics of the pulse signal. Unlike with the traditional measuring method, based on the Korotkov sound, with the oscillometric method the use of a stethoscope is not required. Through simple calculations, this method provides accurate blood pressure readings: the systolic pressure is defined as the blood pressure when the cuff pressure oscillating amplitude begins to increase, while the diastolic blood pressure is defined as the pressure when the cuff pressure oscillating amplitude stops decreasing.
The device includes a plastic enclosure and an integrated wraparound cuff, and it requires an external device (e.g. a smartphone) to display results and perform user interaction.
This document describes the QardioArm, model A100, a non-invasive blood pressure monitoring system. It references regulatory standards, predicate devices, and performance claims but does not contain a detailed study proving the device meets specific acceptance criteria in the format requested. The document is a 510(k) summary, which focuses on demonstrating substantial equivalence to already cleared devices rather than providing a standalone clinical study report.
Thus, the following information is not present in the provided text:
- A table of acceptance criteria and the reported device performance with specific metrics like sensitivity, specificity, accuracy for a given disease or condition. The document primarily reports pressure and pulse accuracy as engineering specifications.
- Sample size used for a test set, data provenance, number of experts for ground truth, adjudication method, or details of a multi-reader multi-case study.
- Details of a standalone performance study as would be conducted if the device were completely novel.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for any clinical testing beyond the general statement of applying clinical test standards.
- Sample size for the training set or how ground truth was established for a training set (as this device is not an AI/ML device in the context of the question).
However, based on the provided text, here is an extraction of relevant information, focusing on the available performance specifications and the clinical testing mentioned:
1. A table of acceptance criteria and the reported device performance
The document mentions meeting specific accuracy criteria for blood pressure and pulse measurements. While not explicitly stated as "acceptance criteria" for a study, these are performance specifications that the device is stated to achieve and would likely be part of the product's design and validation requirements.
Performance Metric | Acceptance Criteria (as reported specification) | Reported Device Performance |
---|---|---|
Pressure Accuracy | ±3mmHg or ±2% of readout value | ±3mmHg or ±2% of readout value |
Pulse Accuracy | ±5% of reading value | ±5% of reading value |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document states under "3. PERFORMANCE & CLINICAL TEST" that the device was tested against "ANSI/AAMI ISO 81060-2:2009" and "EN 1060-4:2004". These are international standards for non-invasive sphygmomanometers. ISO 81060-2:2009 specifies requirements for the clinical investigation of non-invasive blood pressure measuring equipment. These standards typically define the population (e.g., age, sex, blood pressure ranges) and the number of subjects required for clinical validation (e.g., at least 85 subjects with specific distributions). However, the specific sample size, data provenance, or whether the study was retrospective or prospective are not detailed in this summary.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the summary. For blood pressure monitors, "ground truth" (or reference measurements) would typically be established using a auscultatory method by trained observers following a specific protocol, often with double or triple readings. However, details of observers' qualifications or number are not present.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not provided in the summary. For clinical validation of blood pressure devices against a reference method, specific protocols for averaging or adjudicating reference readings are standard (e.g., two observers taking simultaneous readings, with a third if there is a significant discrepancy). However, the summary does not detail the methods used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable to the QardioArm device as described. The QardioArm is a blood pressure monitor, not an AI-assisted diagnostic imaging or health analytics device that would typically undergo an MRMC study comparing human reader performance with and without AI assistance. The document does not suggest any AI components for interpretation or assistance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This is not applicable in the context of "algorithm only" as typically applied to AI/ML devices. The device's performance is inherently "standalone" in the sense that it measures blood pressure and pulse rate automatically using an oscillometric method without human intervention in the measurement process itself, though it requires an external device for display and user interaction. The "Performance & Clinical Test" section implies validation of this automated measurement process.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For a blood pressure monitor, the "ground truth" would almost certainly be simultaneous or sequential auscultatory blood pressure measurements taken by trained observers using a calibrated sphygmomanometer, following a standard protocol (e.g., as defined by ISO 81060-2). The document implicitly references this by stating conformity to ISO 81060-2.
8. The sample size for the training set
This information is not provided. The concept of a separate "training set" for an algorithm is typically associated with AI/Machine Learning models. The QardioArm uses a fundamental oscillometric method, which is a well-established algorithm, not typically "trained" in the modern AI sense. Its development would involve engineering design and calibration, followed by validation.
9. How the ground truth for the training set was established
As there is no mention of a training set in the AI/ML context, this information is not applicable to the provided summary.
Ask a specific question about this device
Page 1 of 1