Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K962497
    Date Cleared
    1996-10-08

    (104 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Faichney Medical Company and Becton-digital thermometers are intended for measuring body temperature using oral, rectal, or axillary sites.

    Device Description

    The Faichney Medical Company and Becton-digital thermometers are intended for measuring body temperature using oral, rectal, or axillary sites. There are no technological differences between these thermometers.

    AI/ML Overview

    This 510(k) summary for the Faichney Medical Clinical Electronic Thermometer provides some, but not all, of the requested information. Here's a breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device references compliance with ASTM standards instead of setting specific acceptance criteria within the 510(k) itself. The reported performance is in comparison to a mercury-in-glass thermometer.

    Acceptance Criteria (ASTM E1112-86 or comparison to mercury-in-glass)Reported Device Performance (Faichney Digital Thermometer)
    Accuracy: (Implicitly, to be comparable or better than mercury-in-glass)Better than mercury-in-glass thermometer's accuracy across temperature range (Mean & Standard Deviation).
    Within 0.10°F of mercury-in-glass accuracy (with 89% confidence interval) at every range except 96.40°F to 98.00°F.
    Response Time: (Implicitly, to be comparable enough for clinical use)Average 26 seconds to register peak temperature (without probe covers).
    Average 28 seconds to register peak temperature (with probe covers).
    (Not explicitly compared to mercury-in-glass for this metric, but noted as a "slight increase").

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size for the test set used in the "Summary of Bench Data." It refers to "the product" and "the same units" without specifying numbers.

    • Sample Size (Test Set): Not explicitly stated.
    • Data Provenance: The tests were conducted by an "independent testing house." The country of origin is not specified, but it's likely within the US given the FDA submission. The study was prospective bench testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not relevant or provided for a clinical electronic thermometer's bench testing. Accuracy is determined against a known standard or a reference thermometer (mercury-in-glass), not by expert interpretation.

    • Number of Experts: Not applicable.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    Not applicable for a device that measures a physical property against a reference. The "investigator" made conclusions, but this is not an adjudication method in the context of clinical interpretation or diagnostic agreement.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC study was not done. This type of study is typically for diagnostic imaging or interpretation where multiple human readers assess cases. For a thermometer, the measurement is objective.

    6. If a Standalone (Algorithm Only Without Human-in-the Loop Performance) Was Done

    Yes, the "Summary of Bench Data" describes standalone performance testing of the electronic thermometer. Its accuracy and response time were measured directly.

    7. The Type of Ground Truth Used

    The ground truth for accuracy was based on comparison to a "mercury-in-glass thermometer" which served as a reference standard. While not explicitly stated as the absolute ground truth, it was the comparator for assessing the electronic thermometer's performance. Response time was measured objectively.

    • Ground Truth Type: Comparison to a reference standard (mercury-in-glass thermometer).

    8. The Sample Size for the Training Set

    This device does not use machine learning or AI, so there is no "training set."

    • Sample Size (Training Set): Not applicable.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1