Search Results
Found 1 results
510(k) Data Aggregation
(111 days)
Purism Non-Sterile Powder Free Nitrile Examination Gloves
The Purism Non-sterile Powder Free Nitrile Examination Gloves is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
The Purism Non-sterile Powder Free Nitrile Examination Gloves are a non-sterile, powder free, single use, disposable item. The gloves are made from nitrile compound and are blue in color. The gloves are ambidextrous and offered in Small, Medium, Large and X-Large sizes.
The provided document describes the FDA's 510(k) clearance for the Purism Non-sterile Powder Free Nitrile Examination Gloves. It outlines the non-clinical tests performed to demonstrate that the device meets acceptance criteria.
Here's the breakdown of the information requested:
1. Table of Acceptance Criteria and Reported Device Performance
Test Method | Purpose | Acceptance Criteria | Reported Device Performance | Pass/Fail |
---|---|---|---|---|
ASTM D6319-19 (Dimensions) | To determine glove length | Min Length (mm): Size S: 230, Size M: 230, Size L: 230, Size XL: 230 | Avg Min Length (mm): Size S: 236, Size M: 239, Size L: 241, Size XL: 238 | Pass |
To determine glove palm width | Palm Width (mm): Size S: 80±10, Size M: 95±10, Size L: 110±10, Size XL: ≥110 | Avg Palm Width (mm): Size S: 84, Size M: 95, Size L: 103, Size XL: 111 | Pass | |
To determine glove finger thickness | Finger Thickness (mm): Size S: 0.100±0.02, Size M: 0.100±0.02, Size L: 0.100±0.02, Size XL: 0.100±0.02 | Avg Finger Thickness (mm): Size S: 0.096, Size M: 0.111, Size L: 0.108, Size XL: 0.111 | Pass | |
To determine glove palm thickness | Palm Thickness (mm): Size S: 0.060±0.02, Size M: 0.060±0.02, Size L: 0.060±0.02, Size XL: 0.060±0.02 | Avg Palm Thickness (mm): Size S: 0.058, Size M: 0.066, Size L: 0.074, Size XL: 0.080 | Pass | |
ASTM D6319-19 (Physical Properties) | To determine glove physical properties before and after ageing | Before Ageing: Tensile Strength min 14 MPa, Ultimate Elongation min 500% | Before Ageing: Tensile Strength min 18.0 MPa, Ultimate Elongation min 694% | Pass |
After Ageing: Tensile Strength min 14 MPa, Ultimate Elongation min 400% | After Ageing: Tensile Strength min 18.2 MPa, Ultimate Elongation min 664% | Pass | ||
ASTM D5151-19 | To detect holes and check for water leaks | As per ISO 2859-1. GI Level II, Lot Size 3000, Sample Size 125, AQL 1.5, Ac=5, Re=6 | Pass at AQL 1.5 | Pass |
ASTM D6124 | To determine residual powder | ≤2 mg/glove | 0.06 mg/glove | Pass |
ISO 10993-10:2010 (Primary Skin Irritation) | To determine if device is a skin irritant | The device must be a non-irritant | Under the study conditions not an irritant | Pass |
ISO 10993-10:2010 (Dermal Sensitization) | To determine if device is a dermal sensitizer | The device must be a non-sensitizer | Under the study conditions not a sensitizer | Pass |
ISO 10993-5:2009 (In vitro Cytotoxicity) | To determine if device extract is cytotoxic | The device must be non-cytotoxic | Under the study conditions cytotoxic | Fail |
ISO 10993-11:2017 (Acute Systemic Toxicity) | To determine if device induces systemic toxicity | The device must not induce systemic toxicity | Under the study conditions did not induce systemic toxicity | Pass |
Note on Cytotoxicity: While the device showed cytotoxic properties in the in vitro test, the document states that additional acute systemic toxicity testing (ISO 10993-11:2017) supported that the device was not systemically toxic. The primary skin irritation and dermal sensitization tests also indicated the device was not an irritant or skin sensitizer. This suggests that the manufacturer provided sufficient evidence to mitigate concerns arising from the in vitro cytotoxicity result in the context of the device's intended use.
2. Sample size used for the test set and the data provenance
The document specifies "Sample Size 125" for the ASTM D5151-19 (Freedom from Holes) test, which is based on ISO 2859-1, GI Level II, and Lot Size 3000. For other tests like dimensions, physical properties, and biocompatibility, specific sample sizes are not explicitly stated in this summary, but the results are reported as average values, indicating that multiple samples were tested.
The provenance of the data (country of origin, retrospective/prospective) is not explicitly stated for each test. However, the manufacturer is "Dezhou Purism Medical Technology Co., Ltd." located in China, suggesting the testing was likely conducted in or overseen from China. The studies are non-clinical (laboratory-based) tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The tests performed are objective, standardized laboratory tests according to ASTM and ISO standards for medical gloves. They do not involve expert interpretation or ground truth establishment in the way clinical studies with human assessors would.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. As these are objective laboratory tests, there is no adjudication method involved. The results are quantitative measurements against predefined criteria.
5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a nitrile examination glove, not an AI-powered diagnostic tool, and therefore, MRMC studies are not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical medical glove. The performance evaluation is based on its material properties and manufacturing quality, not an algorithm.
7. The type of ground truth used
The "ground truth" for the performance tests conducted (dimensions, physical properties, freedom from holes, residual powder, biocompatibility) is established by adherence to recognized international standards (ASTM D6319-19, ASTM D5151-19, ASTM D6124, ISO 10993-10:2010, ISO 10993-5:2009, ISO 10993-11:2017) and their specified acceptance criteria. These standards define the measurable properties that a glove must meet for its intended use.
8. The sample size for the training set
Not applicable. This device is a physical product and does not involve AI or machine learning, thus there is no "training set."
9. How the ground truth for the training set was established
Not applicable. There is no training set for this device.
Ask a specific question about this device
Page 1 of 1