Search Results
Found 1 results
510(k) Data Aggregation
(227 days)
Powder Free Chloroprene Examination Glove Coloured (Blue, Green)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner
It is a powder-free chloroprene patient examination gloves made by on-line polymer-coating and mild on-line chlorination process. The process modifies the surface characteristics without the use of any dusting or donning powder. The gloves are non-sterile, disposable and for single use only.
Characteristics:
- Ambidextrous with beaded cuff
- Blended synthetic latex (Chloroprene and Nitrile Latex)
- Blue or green Coloured ●
- Five sizes: X-Small, Small, Medium, Large and X-Large. ●
The provided document is a 510(k) Premarket Notification for Chloroprene Examination Gloves Powder Free (Blue, Green). This document focuses on demonstrating substantial equivalence to a predicate device, which involves comparing material, design, and performance characteristics for a medical device (examination gloves), not an AI/ML-driven device.
Therefore, the requested information regarding AI/ML device acceptance criteria and study details (such as sample size for test/training sets, data provenance, expert adjudication, MRMC studies, standalone performance, and ground truth establishment) is not applicable to this document.
What the document does provide is a detailed comparison of the physical and biological characteristics of the proposed gloves against established ASTM standards and the predicate device.
Here's a summary of the acceptance criteria and reported performance for the Chloroprene Examination Gloves Powder Free (Blue, Green), as extracted from the provided text:
1. A table of acceptance criteria and the reported device performance:
Characteristic | Standard | Acceptance Criteria | Reported Device Performance (Subject Device K213993) |
---|---|---|---|
Product Code | - | LZA (Same as predicate) | LZA |
Intended Use | - | Intended for medical purposes, worn on examiner's hand/finger to prevent contamination between patient and examiner (Same as predicate) | Intended for medical purposes, worn on examiner's hand/finger to prevent contamination between patient and examiner |
Design | - | Powder Free, Non-Sterile, Ambidextrous, Beaded Cuff (Same as predicate) | Powder Free, Non-Sterile, Ambidextrous, Beaded Cuff |
Indications for Use | - | A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner (Same as predicate) | A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner |
Construction | - | Ambidextrous, Polymer Coated or Chlorinated, Powder Free Nitrile (Similar to predicate's Ambidextrous, Polyacrylic Polymer Inner Coating, Powder Free Nitrile) | Ambidextrous, Polymer Coated or Chlorinated, Powder Free Nitrile |
Single Use | - | Yes (Same as predicate) | Yes |
Packaging | - | Packed in Dispenser Boxes (Same as predicate) | Packed in Dispenser Boxes |
Dimension (Length) | ASTM D6977-19 | XSmall, Small: 220 mm min | |
Medium, Large, XLarge: 230 mm min | Meet 220mm min (XSmall, Small) | ||
Meet 230mm min (Medium, Large, XLarge) | |||
Dimension (Thickness) | ASTM D6977-19 | Palm: 0.05 mm min | |
Finger: 0.05 mm min | Meet 0.05mm min (Palm) | ||
Meet 0.05mm min (Finger) | |||
Dimension (Width) | ASTM D6977-19 | XSmall: 70 ± 10 mm | |
Small: 80 ± 10 mm | |||
Medium: 95 ± 10 mm | |||
Large: 110 ± 10 mm | |||
XLarge: 120 ± 10 mm | Meet 70 ± 10 mm (XSmall) | ||
Meet 80 ± 10 mm (Small) | |||
Meet 95 ± 10 mm (Medium) | |||
Meet 110 ± 10 mm (Large) | |||
Meet 120 ± 10 mm (XLarge) | |||
Physical Properties (Before Aging) | ASTM D6977-19 | Tensile Strength: Min. 14 MPa | |
Ultimate Elongation: Min. 500% | Meet 14MPa min | ||
Meet 500% min | |||
Physical Properties (After Aging) | ASTM D6977-19 | Tensile Strength: Min. 14 MPa | |
Ultimate Elongation: Min. 400% | Meets 14MPa min | ||
Meet 400% min. | |||
Water Leak Test (Before Aging) | ASTM D6977-19 | G-I, AQL 2.5 (FDA GII, AQL 2.5) | Meet AQL 1.5 |
Water Leak Test (After Aging) | ASTM D5151-19 | (Assumed same as Before Aging in this context - AQL 2.5 for general examination gloves) | Meet AQL 2.5 |
Powder Free Residue | ASTM D6977-19 | ||
ASTM D6124-06 | Max. 2 mg/glove | Meet 2mg/glove max. | |
Biocompatibility: Primary Skin Irritation Test | ISO 10993-10 | No Animal Irritation | Passes. Conclusion: Did not cause an irritant response. |
Biocompatibility: Skin Sensitization Test | ISO 10993-10 | No Animal Irritation (No sensitization effect) | Passes. Conclusion: Did not produce a skin sensitization effect. |
Biocompatibility: In Vitro Cytotoxicity | ISO 10993-5 | No cytotoxicity reactivity at 100.0% extract concentrations under the conditions of this test | Exhibited cytotoxicity response at 100% of extract concentration. (Remarked as "Different" compared to predicate which had "No Information") – It still passed the overall safety conclusion. |
Biocompatibility: Acute Systemic Toxicity | ISO 10993-11 | No adverse biological reaction | Passed. Conclusion: Showed no adverse biological reaction. (Remarked as "Different" compared to predicate which had "No Information") |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated as a single "test set" sample size for a clinical study. The data presented are results of various laboratory and biocompatibility tests on the gloves. These tests typically involve a specific number of gloves or samples per test as per the ASTM and ISO standards referenced (e.g., AQL for water leak tests, specific batches for physical property tests). The exact number of gloves or samples for each test is not detailed in this summary, but the results confirm they "Meet" or "Pass" the specified standards.
- Data Provenance: The manufacturer is Careglove Global SDN BHD, located in Malaysia. The testing was conducted to comply with international standards (ASTM and ISO). The document does not specify whether the data is retrospective or prospective in the context of a clinical trial, as this is a premarket notification for a Class I general medical device based on engineering and biocompatibility testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. This is a physical medical device (examination gloves) where "ground truth" is established through adherence to standardized laboratory test methods (ASTM, ISO) for physical properties and biocompatibility, not by expert interpretation of medical images or patient outcomes.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. As per point 3, this is a physical device tested against objective standards, not an AI/ML system requiring human expert adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is not an AI/ML device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for this device's performance is based on objective measurements defined by international standards (ASTM D6977-19, ASTM D5151-19, ASTM D6124-06, ISO 10993-10, ISO 10993-5, ISO 10993-11) for physical properties (e.g., length, thickness, tensile strength, elongation, freedom from holes, powder residue) and biocompatibility (e.g., skin irritation, sensitization, cytotoxicity, systemic toxicity).
8. The sample size for the training set
- Not Applicable. This is not an AI/ML device; there is no "training set."
9. How the ground truth for the training set was established
- Not Applicable. This is not an AI/ML device; there is no "training set" or establishment of ground truth in that context.
Ask a specific question about this device
Page 1 of 1