Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K232292
    Date Cleared
    2023-11-07

    (98 days)

    Product Code
    Regulation Number
    888.3080
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Peridot-EX Expandable Intervertebral Body Fusion System

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Peridot-EX Expandable Intervertebral body fusion system is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). This device is to be used with autogenous bone graft and/or allogenic bone graft composed of cancellous and/or corticocancellous bone graft.

    The device is intended to be used with supplemental fixation systems that have been cleared for use in the lumbosacral spine (e.g., posterior pedicle screw and rod systems and anterior screw and anterior screw and rod systems). The Peridot-EX Expandable Intervertebral body fusion system is to be used with supplemental fixation. Patients should have at least six (6) months of nonoperative treatment with an intervertebral cage. Hyperlordotic interbody devices (> 20° lordosis) must be used with at least anterior supplemental fixation.

    Device Description

    The Peridot-EX Expandable Intervertebral body fusion system is various lengths and heights, which can be inserted between two lumbar or lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. It is manufactured from titanium alloy (Ti-6AI-4V ELI) as described by ASTM F136 and 3D Printing manufactured from titanium power (Ti-6AI-4V ELI) as described by ASTM F3001.

    The Peridot-EX Expandable Intervertebral body fusion system expands for adjustable lordosis and height to match patient anatomy. The hollow geometry of the implants allows them to be packed with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone.

    The devices are intended to be used with supplemental spinal fixation, either applied anterior or posterior (e.g. using posterior pedicle screws, anterior plate system or anterior screw and rod system).

    The devices are to be sterilized by gamma radiation per ISO 11137 or provided non-sterile. The non-sterile devices are to be sterilized by steam sterilization per ISO 17665.

    All implants are intended for SINGLE USE ONLY and should not be reused under any circumstances.

    The device is supplied with their specific instrument. The device must be used in combination with the dedicated instrument supplied. The specific instrument is supplied dedicated tray and non-sterile.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called the Peridot-EX Expandable Intervertebral body fusion system. It focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria in a clinical study. Therefore, much of the requested information about clinical study design, sample sizes, expert involvement, and ground truth establishment is not available in this regulatory submission.

    Here's a breakdown of the available information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not present a table of specific acceptance criteria. Instead, it lists various "Performance Testing" conducted according to ASTM standards for mechanical integrity. The conclusion states that the device is "substantially equivalent to predicate device," implying that its mechanical performance met the expected standards for such devices, likely by demonstrating similar or superior results to the predicate devices in these tests.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Not Explicitly Stated)Reported Device Performance
    Mechanical PerformanceImplied: Meet or exceed mechanical performance of predicate devices as per ASTM standards.- Static Compression (ASTM F2077): Tested
    • Static Compression-shear (ASTM F2077): Tested
    • Static Torsion (ASTM F2077): Tested
    • Fatigue Compression (ASTM F2077): Tested
    • Fatigue Compression-shear (ASTM F2077): Tested
    • Subsidence (ASTM F2267): Tested
    • Expulsion (ASTM F-04.25.02.02): Tested |

    2. Sample size used for the test set and the data provenance

    The document only mentions "The worst-case devices were tested" for mechanical performance. It does not provide specific sample sizes (e.g., number of devices) for any of these tests.

    The data provenance is from mechanical testing in a laboratory setting, not from human subjects or clinical data. There is no mention of country of origin of this specific test data, but the submitter is GBS Commonwealth Co., Ltd. from South Korea.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable as the study described is mechanical performance testing, not a study involving human interpretation or clinical diagnosis.

    4. Adjudication method for the test set

    This information is not applicable for mechanical performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was conducted or mentioned. The device is a physical intervertebral body fusion system, not an AI-assisted diagnostic or therapeutic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is a physical implant, not an algorithm.

    7. The type of ground truth used

    For the mechanical performance tests, the "ground truth" would be the physical properties and failure limits defined by the ASTM standards applied and potentially the performance data of the predicate devices for comparison.

    8. The sample size for the training set

    This information is not applicable. This document describes mechanical testing of a medical device, not a machine learning model.

    9. How the ground truth for the training set was established

    This information is not applicable.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1