Search Results
Found 1 results
510(k) Data Aggregation
(109 days)
The PTA-Plus PTA Balloon Catheter is indicated for use within synthetic arteriovenous dialysis fistulae to remove embolic material (thrombus/debris) and dilate stenosis for treatment of obstructive lesions.
The proposed PTA-Plus PTA Balloon Catheter is designed for de-clotting and treating stenosis in synthetic dialysis grafts. It is an .035" guide-wire compatible, PTA balloon catheter with a proprietary valve system which allows injection of contrast and an embolectomy coil for mechanical removal of thrombus.
Here's a breakdown of the acceptance criteria and study information for the Hotspur Technologies, Inc. PTA-Plus PTA Balloon Catheter, based on the provided document:
1. Acceptance Criteria and Reported Device Performance
The document states that the PTA-Plus PTA Balloon Catheter met "all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance documents, test protocols, and/or customer inputs." However, it does not provide a specific table or list of quantitative acceptance criteria values alongside the device's reported performance for each criterion. Instead, it broadly lists the types of tests performed and concludes that the device successfully met the established specifications.
Therefore, the table below reflects the categories of tests conducted and the general performance outcome as stated in the submission. Specific numerical values for acceptance criteria or performance are not detailed in the provided text.
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Biocompatibility Testing | Materials are non-toxic and non-sensitizing, consistent with intended use. |
| - ISO MEM Elution Assay | Met established specifications. |
| - Materials Mediated Rabbit Pyrogen | Met established specifications. |
| - ASTM Hemolysis Assay | Met established specifications. |
| - Complement Activation C3a and SC5b-9 Assay | Met established specifications. |
| - Partial Thromboplastin Time | Met established specifications. |
| - Four Hour Thromboresistance Evaluation | Met established specifications. |
| - ISO Guinea Pig Maximization Sensitization | Met established specifications. |
| - ISO Acute Systemic Injection Test | Met established specifications. |
| - ISO Intracutaneous Reactivity Test | Met established specifications. |
| - Pyrogen (LAL) Chromogenic | Met established specifications. |
| In-vitro Performance Bench Testing | The device confirmed performance characteristics compared to predicate devices and met established specifications. |
| - Balloon Crossing Profile | Met established specifications. |
| - Catheter Shaft Diameter | Met established specifications. |
| - Guidewire Lumen Diameter | Met established specifications. |
| - Catheter Shaft Markings | Met established specifications. |
| - Balloon Working Length | Met established specifications. |
| - Balloon Rated Burst Pressure | Met established specifications. |
| - Balloon Compliance/Outer Diameter | Met established specifications. |
| - Balloon Inflation/Deflation Time | Met established specifications. |
| - Balloon Fatigue | Met established specifications. |
| - Catheter Body Burst Strength | Met established specifications. |
| - Catheter Bond and Tip Pull Strength | Met established specifications. |
| - Catheter Torque Strength | Met established specifications. |
| - Contrast Media Flow Rate | Met established specifications. |
| - Adherent Clot Removal | Met established specifications. |
| - Simulated Use/Flexibility/Kink | Met established specifications. |
| - Radiopacity | Met established specifications. |
| In-vivo Animal Model Testing | The device showed vessel and organ response comparable to control device treatments, meeting established specifications. |
2. Sample Size for Test Set and Data Provenance
- Sample Size for Test Set: The document does not specify a numerical sample size for the in-vitro or in-vivo tests. For the in-vivo testing, it mentions "ovine specimens" and "test and control comparator groups" but no specific number of animals.
- Data Provenance:
- In-vitro testing: Performed internally ("in-vitro").
- In-vivo testing: Conducted using an animal model (ovine specimens). The study was performed "in accordance with 21 CFR Part 58 'Good Laboratory Practices for Nonclinical Laboratory Studies'," which suggests a controlled, prospective study design. There's no mention of the country of origin, but given the submission to the FDA, it's likely U.S.-based or conducted under U.S. regulatory standards.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This information is not provided in the document. The device's performance was evaluated through physical, chemical, and animal testing rather than human interpretation or expert-established ground truth in the context of diagnostic performance.
4. Adjudication Method for the Test Set
This information is not applicable as the evaluation involved physical/chemical bench testing and animal studies, not human adjudication of results.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance
This information is not applicable. The device is a medical instrument (balloon catheter), not an AI-assisted diagnostic tool.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
This information is not applicable. The device is a medical instrument, not an algorithm. The "standalone" performance here would refer to its physical and functional attributes, which were tested in-vitro and in-vivo.
7. The type of ground truth used
- Biocompatibility: Established by adherence to ISO and ASTM standards and specific assay results (e.g., non-toxic, non-sensitizing, non-pyrogenic).
- In-vitro Performance: Established by engineering specifications, comparative performance to predicate devices (which themselves met established safety and efficacy profiles), and adherence to predefined test protocols (e.g., balloon burst pressure, inflation/deflation times).
- In-vivo Performance: Established by surgical implantation outcomes, observation of animal specimens, and post-survival vessel and organ histology, comparing responses to test and control devices. The ground truth here is the physiological response and histological findings in the ovine model.
8. The sample size for the training set
This information is not provided. The concept of a "training set" is typically associated with machine learning or AI models, which is not applicable to this physical medical device. The "training" for this device would refer to its design and manufacturing processes, which are not detailed in terms of a data set.
9. How the ground truth for the training set was established
This information is not applicable for the same reasons as point 8.
Ask a specific question about this device
Page 1 of 1