Search Results
Found 6 results
510(k) Data Aggregation
(51 days)
Mini Ridge Filter and auxiliary functions for PROBEAT-V
The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment.
The beam delivery system is composed of the following components
- Gantry Room o
- . Rotating Gantry
- Scanning Nozzle .
- . Patient Positioning System
- Cone Beam CT / X-ray Imaging System .
- Fixed Beam Room o
- . Scanning Nozzle
- Patient Positioning System 트
- . Cone Beam CT / X-ray Imaging System
The beam production system is composed of the following components
- Accelerator system (LINAC, Synchrotron) o
- O Beam transport system (Low/High Energy Beam Transport systems)
The system incorporates several optional features and accessories, namely:
- The Mini Ridge Filter ("mRF") is an optional accessory to modify the beam of the o PROBEAT-V system. The mRF is installed manually and may be used in conjunction with the range shifters inside the nozzle or extended range shifter. The mRF can be added to the cleared PROBEAT-V nozzle to spread out the Bragg peak along the axis of the beam in order to reduce the amount of beam energy in the delivery of proton radiation to defined target volumes.
- Beam gating function allowing for interface with cleared external gating systems to o control the beam delivery for treatment such as to synchronize irradiation with respiration. Although the overall treatment time tends to be longer than the treatment time without gating, the extension of the treatment time will not affect irradiation performance to the target treatment site. Instead, the gating functionality may limit radiation exposure to regions outside of the target treatment volume.
- Allows for use of fluoroscopy during proton irradiation at the physician's discretion. o Fluoroscopy may be used for observation of treatment site during treatment, which could be used for interruption of the treatment or analysis for treatment planning.
The provided text describes a 510(k) premarket notification for a medical device called PROBEAT-V, a proton beam therapy system. It outlines additional optional features and the performance data for these features.
However, the document does not contain the following information typically found in a comprehensive acceptance criteria and study report:
- A table of acceptance criteria and reported device performance: While it states "All tests were successful and confirmed the performance of these additional optional features," it does not specify quantitative acceptance criteria or the numerical performance metrics.
- Sample size used for the test set and data provenance.
- Number of experts used to establish the ground truth for the test set and their qualifications.
- Adjudication method for the test set.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size of human readers improve with AI vs without AI assistance. (This is not an AI device, so an MRMC study related to AI assistance would not be applicable here).
- If a standalone performance study (algorithm only without human-in-the-loop performance) was done. (Again, not an AI device).
- The type of ground truth used.
- The sample size for the training set. (Not an AI/machine learning device that typically requires a training set in that context).
- How the ground truth for the training set was established. (Not an AI/machine learning device).
Summary of available information regarding acceptance criteria and study:
The device's additional optional features are the Mini Ridge Filter (mRF), an external beam gating function, and the ability to use fluoroscopy during proton irradiation.
1. Acceptance Criteria and Reported Device Performance:
Feature | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Mini Ridge Filter (mRF) | To evaluate range loss, distal dose falloff, and beam spot size. | "All tests were successful and confirmed the performance of these additional optional features." |
External Beam Gating | To ensure appropriate signal reception and adequate gating function. | "All tests were successful and confirmed the performance of these additional optional features." |
Fluoroscopy function | To ensure fluoroscopy can be enabled/disabled without compromising proton dose monitoring. | "All tests were successful and confirmed the performance of these additional optional features." |
2. Sample size used for the test set and the data provenance:
- Not specified. The document mentions "testing" but does not detail the number of tests performed or the data used.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable / not specified. For a physical device like a proton therapy system, "ground truth" relates to physical measurements and engineering specifications, not expert interpretation of outputs in the same way it would for diagnostic imaging devices.
4. Adjudication method for the test set:
- Not applicable / not specified.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. This is not an AI-assisted device, so an MRMC study for AI assistance is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes (for the device's functions). The "Performance Data" section describes testing conducted on the device's features (mRF, gating, fluoroscopy) to confirm their intended function. This is a standalone performance evaluation of the device itself rather than an AI algorithm.
7. The type of ground truth used:
- Physical measurements and engineering specifications. For a proton therapy system, "ground truth" would be established through calibrated instruments measuring beam properties (range, dose falloff, spot size) and system responses to ensure they meet pre-defined engineering tolerances and physical laws.
8. The sample size for the training set:
- Not applicable. This device is not an AI/machine learning system that requires a "training set" in the conventional sense. Its functionality is based on established physics and engineering.
9. How the ground truth for the training set was established:
- Not applicable. As above, no training set in the context of machine learning.
Ask a specific question about this device
(79 days)
PROBEAT-V Proton Beam Therapy System with X-ray Limiting Accessory
The PROBEAT-V with X-ray Limiting Accessory is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
The PROBEAT-V with X-ray Limiting Accessory is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment.
The X-ray Beam Limiting accessory for use in conjunction with the cleared PROBEAT-V system is designed to limit the patient's X-ray exposure during for patient positioning with the Cone Beam CT (CBCT). The device has 4 motorized leaves which are controlled by the PROBEAT-V imaging software to position the leaves in order to limit the patient's radiation exposure during patient positioning.
The provided text describes a 510(k) premarket notification for the PROBEAT-V Proton Beam Therapy System with X-ray Limiting Accessory. The document focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study report with specific acceptance criteria and detailed performance data on the X-ray Limiting Accessory itself.
Based on the provided information, I can extract the following:
1. A table of acceptance criteria and the reported device performance
The document does not provide a specific table of acceptance criteria with numerical targets. Instead, it states that "In all cases, the device performed as intended and met all pre-specified success criteria." The performance evaluation focused on the integration of the X-ray Beam Limiting Device with the PROBEAT-V system.
Acceptance Criteria Category | Reported Device Performance |
---|---|
Slice Pitch Accuracy | Met all pre-specified success criteria and performed as intended. |
Spatial Accuracy | Met all pre-specified success criteria and performed as intended. |
Spatial Resolution | Met all pre-specified success criteria and performed as intended. |
Isocenter Accuracy | Met all pre-specified success criteria and performed as intended. |
Performance with Proton Beam Therapy Positioning System | Met all pre-specified success criteria and performed as intended. |
Electrical Safety | Evaluated and met all pre-specified success criteria and performed as intended. |
Electromagnetic Compatibility | Performed and met all pre-specified success criteria and performed as intended. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not provide details on sample sizes for the test set or data provenance (country of origin, retrospective/prospective). The testing appears to be a validation of the integrated system in an engineering and performance testing context rather than a clinical study with patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. The testing described appears to be technical performance testing of the device itself (e.g., accuracy, resolution, safety) and not a diagnostic or therapeutic output requiring interpretation by medical experts to establish ground truth.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. The testing described appears to be technical performance testing of the device itself and not a diagnostic or therapeutic output requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document describes a proton beam therapy system with an X-ray limiting accessory, not an AI-based diagnostic or therapeutic assistance system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This refers to the performance of the device's X-ray limiting accessory and its integration with the PROBEAT-V system. The "Performance Data" section describes this standalone performance testing for the integrated system, including slice pitch accuracy, spatial accuracy, spatial resolution, isocenter accuracy, and performance with the proton beam therapy positioning system, as well as electrical safety and electromagnetic compatibility. The device performed as intended and met all pre-specified success criteria in these tests.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
For the technical performance tests mentioned (slice pitch accuracy, spatial accuracy, etc.), the "ground truth" would be established by validated measurement standards and reference devices used in engineering and quality control, rather than clinical ground truth like pathology or expert consensus. For electrical safety and electromagnetic compatibility, the ground truth would be compliance with relevant international and national standards.
8. The sample size for the training set
Not applicable for this type of device and submission. This is not an AI/machine learning device requiring a training set.
9. How the ground truth for the training set was established
Not applicable for this type of device and submission. This is not an AI/machine learning device requiring a training set.
Ask a specific question about this device
(126 days)
PROBEAT-V and accessories
The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
The PROBEAT-V includes optional accessories to assist the radiation oncologist in the delivery of proton radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment.
This document describes the PROBEAT-V and its accessories, focusing on modifications to the minimum MU and the introduction of a high-accuracy camera system for patient positioning. The document states that testing was conducted to confirm the device meets pre-defined acceptance criteria. However, it does not provide specific details about the acceptance criteria or the study that proves the device meets them.
Therefore, I cannot answer your request based on the provided text. The document states that "All the tests were successfully conducted and confirmed that the device meets the pre-defined acceptance criteria," but it does not elaborate on:
- A table of acceptance criteria and the reported device performance: No table or specific performance metrics are provided.
- Sample size used for the test set and the data provenance: No information on sample size or data origin is given.
- Number of experts and their qualifications: No details about experts are mentioned.
- Adjudication method: Not described.
- MRMC comparative effectiveness study: No indication of such a study.
- Standalone performance: Not explicitly addressed in terms of a separate study.
- Type of ground truth: Not specified.
- Sample size for the training set: Not applicable as this is not a description of an AI/ML algorithm development.
- How ground truth for the training set was established: Not applicable.
Ask a specific question about this device
(53 days)
PROBEAT-V
The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment.
The beam delivery system is composed of the following components:
- Gantry Room
- Rotating Gantry
- Scanning Nozzle
- Patient Positioning System
- Cone Beam CT / X-ray Imaging System
- Fixed Beam Room
- Scanning Nozzle
- Patient Positioning System
- Cone Beam CT
The beam production system is composed of the following components:
- Accelerator system (LINAC, Synchrotron)
- Beam transport system (Low/High Energy Beam Transport systems)
The provided text is a 510(k) summary for the PROBEAT-V proton beam therapy system. It describes the device, its intended use, and performance data. However, it does not contain the detailed information required to answer all the questions about specific acceptance criteria and a study proving the device meets those criteria, particularly regarding AI performance, expert ground truth, sample sizes for training/test sets, and adjudication methods, as this is neither an AI device nor does it have relevant information about AI device testing.
Here's a breakdown of what can be extracted from the document:
1. A table of acceptance criteria and the reported device performance
The document provides a general statement about performance testing without specific quantitative acceptance criteria. It broadly states that the device "functioned as intended and met its specifications."
Acceptance Criteria (Stated broadly) | Reported Device Performance |
---|---|
Mechanical performance of rotating gantry and patient couch | Functioned as intended and met specifications |
Beam dose shape and beam dose | Functioned as intended and met specifications |
Beam stop control, dose monitor, area safety, and mechanical interlocks | Functioned as intended and met specifications |
Electrical safety (IEC 60601-1) | Met specifications |
Electromagnetic compatibility (IEC 60601-1-2) | Met specifications |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
This information is not provided in the document. The document describes general performance testing but does not specify sample sizes or data provenance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This information is not applicable/provided. The PROBEAT-V is a therapeutic device (proton beam therapy system), not a diagnostic or AI-driven decision support system that typically requires expert-established ground truth for image interpretation or diagnosis. The performance testing focuses on the physical and functional aspects of the hardware and beam delivery.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This information is not applicable/provided for the same reasons as the previous point. Adjudication methods are typically used in studies involving subjective assessments or classifications by human readers, which is not the nature of the described performance testing for a proton beam therapy system.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
There is no indication that an MRMC comparative effectiveness study was done, nor is AI assistance mentioned. The device is a proton beam therapy system, not an AI-driven image analysis or decision support tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This information is not applicable/provided. The PROBEAT-V is a physical medical device for radiation therapy, not an algorithm. Its performance testing relates to its mechanical, electrical, and beam delivery functionalities.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The concept of "ground truth" as typically applied to diagnostic studies (e.g., pathology for a tumor diagnosis) is not directly applicable here. The performance testing described (mechanical, beam, safety interfaces, electrical safety, EMC) likely uses objective, measurable physical parameters and engineering standards as its "ground truth" or reference for meeting specifications. For example, a "beam dose shape" would be compared against a pre-defined expected profile, or electrical safety against a standard like IEC 60601-1.
8. The sample size for the training set
This information is not applicable/provided. The PROBEAT-V is a hardware medical device; it does not involve a "training set" in the context of machine learning. Its design and manufacturing are based on engineering principles and established medical device standards.
9. How the ground truth for the training set was established
This information is not applicable/provided for the same reasons as the previous point.
Ask a specific question about this device
(57 days)
Extended Range Shifter for PROBEAT-V
The Extended Range Shifter is an accessory to the PROBEAT-V system that is intended to assist the radiation oncologist in the delivery of proton radiation to defined target volumes while sparing surrounding normal tissue and critical organs from excess radiation.
This accessory is added to the nozzle of the cleared PROBEAT-V to suppress the beam spreading further compared to use of the standard range shifter.
The provided text is a 510(k) summary for the "Extended Range Shifter" device. It describes the device, its intended use, and its technological characteristics. However, it does not contain the type of information typically found in a study proving a device meets specific acceptance criteria, especially for AI/ML-based medical devices.
Specifically, the text is about a physical accessory for a proton therapy system, not a software or AI device that would require detailed performance metrics like sensitivity, specificity, or human reader improvement with AI assistance. The "Performance Data" section simply states that "All the tests were successfully conducted and confirmed the performance of the ERS." without providing any quantitative data or details about the sample size, ground truth, or expert evaluation methods.
Therefore, I cannot fulfill your request for:
- A table of acceptance criteria and reported device performance: The document only generically states tests were "successfully conducted" but provides no specific criteria or performance numbers.
- Sample size and data provenance: Not mentioned.
- Number of experts and qualifications: Not relevant for this type of physical device's performance testing described here.
- Adjudication method: Not applicable.
- MRMC comparative effectiveness study: Not applicable.
- Standalone performance: Not applicable.
- Type of ground truth: Not specified beyond physical measurements.
- Training set sample size and ground truth establishment: Not applicable, as this is not an AI/ML device.
In summary, the provided document does not contain the information required to answer your questions about acceptance criteria and study details for an AI/ML device demonstrating performance. It is a premarket notification for a physical medical device accessory.
Ask a specific question about this device
(34 days)
PROBEAT-V
The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment.
The beam delivery system is composed of the following components"
- o Gantry Room
- · Rotating Gantry
- . Scanning Nozzle
- . Patient Couch (or Patient Positioning System)
- . X-ray Imaging System
- o Fixed Beam Room
- Scanning Nozzle .
- . Patient Couch (or Patient Positioning System)
- . X-ray Imaging System
The beam production system is composed of the following components"
- Accelerator system (LINAC, Synchrotron) O
- Beam transport system (Low/High Energy Beam Transport systems) O
The provided text describes the 510(k) premarket notification for the Hitachi PROBEAT-V, a medical charged-particle radiation therapy system. It outlines performance data but does not contain information about specific acceptance criteria, study methodologies, or AI performance metrics as requested.
Therefore, I cannot provide a table of acceptance criteria and reported device performance or information about sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, or training set details as these are not present in the provided document.
The document indicates that Hitachi performed validation testing to evaluate:
- Mechanical performance: Specifically for the rotating gantry and patient couch.
- Beam performance: To assess beam dose shape and beam dose.
- Safety interface testing: To evaluate beam stop control, dose monitor, area safety, and mechanical interlocks.
- Electrical safety and electromagnetic compatibility: In accordance with IEC 60601-1 and IEC 60601-1-2.
The conclusion of these tests was that "In all instances, the PROBEAT-V functioned as intended and met its specifications." This statement broadly implies that the device met its internal acceptance criteria for these functional and safety aspects, but the specific numerical or qualitative criteria for "met its specifications" are not detailed.
Ask a specific question about this device
Page 1 of 1