Search Results
Found 2 results
510(k) Data Aggregation
(30 days)
POWERFLEX PRO PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY CATHETER
The Powerflex™ Pro PTA catheter is intended to dilate stenoses in illiac, femoral, popliteal, infra popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The device is also indicated for post-difatation of balloon-expandable and self-expanding stents in the peripheral vasculature.
The Powerflex™ Pro Percutaneous Transluminal Angioplasty (PTA) catheter is a catheter with a distal inflatable balloon. The Powerflex™ Pro PTA Catheter is designed for use with a 0.035″ guide wire and a catheter sheath introducer and is available in a variety of diameters and lengths. Two radiopaque marker bands indicate the dilating section of the balloon and aid in balloon placement. The catheter tip is tapered to ease entry into peripheral arteries and to facilitate the crossing of tight stenoses.
Here's a breakdown of the acceptance criteria and study information based on the provided text, though it's important to note that this document is a 510(k) summary for a medical device (a catheter) and not for a software or AI/ML device. Therefore, many of the requested fields related to software performance evaluation (like multi-reader studies, standalone algorithm performance, number of experts for ground truth, etc.) are not applicable in this context.
The document describes the process for obtaining 510(k) clearance for additional sizes of an existing Percutaneous Transluminal Angioplasty (PTA) catheter, demonstrating substantial equivalence to a predicate device.
1. Table of Acceptance Criteria and the Reported Device Performance:
The document lists performance tests conducted, implying these tests have associated acceptance criteria that the device met. However, the specific numerical acceptance criteria for each test are not provided in this summary. It only states that the tests were "completed" and "no new safety or performance issues were raised."
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Marker Band Spacing | Met requirements |
Balloon Diameter (Nominal, RBP) | Met requirements |
System Burst (Inflation Lumen) | Met requirements |
System Burst (Guidewire Lumen) | Met requirements |
Hub to Shaft Pull Strength | Met requirements |
CSI Insertion | Met requirements |
Kink Diameter | Met requirements |
Catheter Shaft | Met requirements |
Inflation/Deflation Time | Met requirements |
Multiple Inflation (System Fatigue) | Met requirements |
Torque Testing | Met requirements |
Particle Free | Met requirements |
Hemolysis | Met requirements |
Balloon Working Length | Met requirements |
Balloon Burst | Met requirements |
Marker Band Placement | Met requirements |
Proximal Pull Strength | Met requirements |
Tip to Balloon/Inner Body Pull Strength | Met requirements |
CSI Withdrawal Force | Met requirements |
Guidewire Compatibility | Met requirements |
Usable Catheter Length | Met requirements |
Rated Burst Pressure | Met requirements |
Preconditioning | Met requirements |
Tensile Testing | Met requirements |
Cytotoxicity | Met requirements (met ISO 10993-1:2009) |
Physicochemical Aqueous Extraction | Met requirements |
2. Sample Size Used for the Test Set and the Data Provenance:
- Sample Size: The document does not specify the sample size for each performance test. It generically states that "non-clinical design verification" and "in-vitro performance tests" were completed.
- Data Provenance: The data is from "non-clinical design verification" and "in-vitro performance tests" conducted by the manufacturer (Cordis Corporation). This is retrospective, in-house testing. Country of origin is not explicitly stated for the testing, but the company is based in Bridgewater, NJ, USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:
Not Applicable. This is a medical device (catheter) 510(k) summary, not a software or AI/ML device. Ground truth, in this context, would be established by engineering specifications and objective measurements, not by expert interpretation for diagnostic accuracy.
4. Adjudication Method for the Test Set:
Not Applicable. As above, this pertains to physical device performance and conformity to engineering specifications, not subjective interpretation requiring expert adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not Applicable. This is not a software or AI/ML device that assists human readers.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done:
Not Applicable. This is not an algorithm.
7. The Type of Ground Truth Used:
The "ground truth" for the device's performance tests would be based on:
- Engineering Specifications: Predetermined design parameters and performance thresholds (e.g., specific balloon diameters, burst pressures, tensile strengths).
- Standardized Test Methods: Adherence to established test protocols (likely internal company protocols and potentially relevant ISO standards, as Bio-compatibility refers to ISO 10993-1:2009).
- Objective Measurements: Direct physical measurements and observations of the device components and entire system during in-vitro testing.
8. The Sample Size for the Training Set:
Not Applicable. This is a physical medical device, not a software or AI/ML model that requires a training set.
9. How the Ground Truth for the Training Set was Established:
Not Applicable. (See point 8).
Ask a specific question about this device
(221 days)
POWERFLEX PRO PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY CATHETER
The Powerflex™ Pro Percutaneous Transluminal Angioplasty (PTA) catheter is intended to dilate stenoses in iliac, femoral, ilio-femoral, popliteal, infra popliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. The device is also indicated for post-dilatation of balloon-expandable and self-expanding stents in the peripheral vasculature.
The Powerflex™ Pro Percutaneous Transluminal Angioplasty (PTA) catheter with a distal inflatable balloon. The Powerflex™ Pro PTA Catheter is designed for use with a 0.035" guide wire and a catheter sheath introducer and is available in a variety of diameters and lengths. Two radiopaque marker bands indicate the dilating section of the balloon and aid in balloon placement. The catheter tip is tapered to ease entry into peripheral arteries and to facilitate the crossing of tight stenoses.
The provided text describes the Cordis Powerflex™ Pro Percutaneous Transluminal Angioplasty Catheter. It outlines the device description, indications for use, and a summary of performance testing to establish substantial equivalence to predicate devices.
Here's an analysis based on your requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The submission does not explicitly list "acceptance criteria" with specific quantitative thresholds. Instead, it states that the conducted in-vitro performance tests provide "reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use." The performance is demonstrated by the completion of these tests aligning with engineering standards and existing predicate devices.
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Device Design and Functionality: | The Powerflex™ Pro PTA Catheter is designed for use with a 0.035" guide wire and a catheter sheath introducer. It features two radiopaque marker bands for balloon placement and a tapered tip for ease of entry and crossing stenoses. |
Material Biocompatibility: | All materials used in the proposed device are similar to the predicate devices and meet the requirements of ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and Testing within a risk management process. |
Mechanical/Physical Performance (In-vitro tests): | All listed in-vitro performance tests (Marker Band Spacing, Balloon Diameter (Nominal, RBP), System Burst (Inflation Lumen & Guidewire Lumen), Hub to Shaft Pull Strength, CSI Insertion Kink Diameter, Catheter Shaft, Inflation/Deflation Time, Multiple Inflation (System Fatigue), Torque Testing, Particle Free, Hemolysis, Balloon Working Length, Balloon Burst, Marker Band Placement, Proximal Pull Strength, Tip to Balloon/Inner Body Pull Strength, CSI Withdrawal Force, Guide Wire Compatibility, Useable Catheter Length, Rated Burst Pressure, PreConditioning, Tensile Testing, Cytotoxicity, Physicochemical Aqueous Extraction) were completed. The submission states, "The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing." This implies acceptable performance across all these metrics relative to the predicate devices and industry standards. |
Substantial Equivalence to Predicate Devices: | Comparison with Cordis Opta Pro/Powerflex™ P3 PTA Catheter and ClearStream Technologies, Savvy Long PTA Catheter (for dimensions) showed "technological characteristics such as materials, components, biocompatibility, performance properties, dimensions (size range), method of delivery, fundamental technology (operating principle), packaging configuration and packaging materials, labeling, manufacturing and sterilization processes featured with the Powerflex Pro PTA Catheter are substantially equivalent." The intended use is also identical, with the exception of the additional post-dilatation indication, for which the Cordis Corp., Aviator Plus PTA Catheter serves as a predicate. |
Safety and Effectiveness: | Demonstrated via non-clinical design verification and validation tests, analyses, and post-market surveillance data analyses of the predicate devices. The document concludes that the device "has been shown to be appropriate for its intended use and is considered to be substantially equivalent." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact sample sizes used for each of the in-vitro performance tests. It merely lists the tests that were completed.
- Data Provenance: The studies were "non-clinical design verification and validation tests and analyses," meaning they were laboratory/bench tests. The origin of the data is therefore laboratory-based, and not patient-specific (i.e., not a clinical study involving human subjects). It also mentions "post-market surveillance data analyses of the predicate devices," which would be retrospective data from existing devices.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts
Not applicable. This device's safety and effectiveness were demonstrated through non-clinical, bench testing and comparison to predicate devices, not through a diagnostic accuracy study requiring expert consensus on a test set to establish "ground truth."
4. Adjudication Method for the Test Set
Not applicable, as there was no test set requiring human adjudication in the context of diagnostic accuracy. The testing was laboratory-based.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device (catheter), not an AI diagnostic software.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a medical device (catheter), not an AI algorithm. The performance described is "standalone" in the sense that the device itself was tested in a lab setting, but not in the context of an AI algorithm's performance.
7. The Type of Ground Truth Used
For the in-vitro performance tests, the "ground truth" would be established by:
- Engineering specifications and design requirements.
- Industry standards (e.g., ISO for biocompatibility).
- Performance characteristics of the legally marketed predicate devices.
- Accepted scientific and engineering principles for material properties and mechanical function.
8. The Sample Size for the Training Set
Not applicable. This is a medical device (catheter), not an AI algorithm requiring a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable. This is a medical device (catheter), not an AI algorithm requiring a training set.
Ask a specific question about this device
Page 1 of 1