Search Results
Found 9 results
510(k) Data Aggregation
(67 days)
POWDER FREE VINYL PATIENT EXAMINATION GLOVES, COLORED (YELLOW)
Powder Free Vinyl Patient Examination Gloves, Colored (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free Vinyl Patient Examination Gloves, Colored (Yellow) that meets all of the requirements of ASTM standard D 5250-06(Reaffirmation 2011).
The provided document describes the acceptance criteria and study for the HEBEI LEADER PLASTIC CO., LTD's "Powder Free Vinyl Patient Examination Gloves, Colored (Yellow)." This is a medical device, and the evaluation focuses on its conformance to established standards for patient examination gloves.
Here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
Feature/Test | Acceptance Criteria (Standard) | Reported Device Performance (Subject Device) |
---|---|---|
Intended Use | A disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner (as per Medical Glove Guidance Manual 1661) | Is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. |
Device Description & Specifications | Meets ASTM D5250 (If vinyl: meet all current specifications listed under ASTM Specification D5250 or equivalent consensus standard) | Meets ASTM D5250 |
Materials | PVC (If the glove is made of a polymer or other type of material, identify the material.) | PVC |
Dusting or Donning Powder (Lubricant) | PU (If a donning lubricant is used, state the composition and include biocompatibility data for the lubricant in an identified attachment; also state the name, manufacturer, and address below) | PU (Surface Coating Agent) |
Product Specifications | Meets ASTM D5250 standard or an equivalent standard for vinyl | Meets ASTM D5250 |
Performance Data (Holes) | ASTM D5151 (Detection of Holes in Medical Gloves) | Meets ASTM D5151 |
Performance Data (Residual Powder) | ASTM D6124 (Residual Powder on Medical Gloves) | Meets ASTM D6124 |
Single Patient Use | Single Patient Use | Single Patient Use |
Biocompatibility | SKIN IRRITATION DERMAL and SENSITIZATION STUDIES (Meets ISO 10993-10) | SKIN IRRITATION DERMAL and SENSITIZATION STUDIES Meets ISO 10993-10 |
Waterleak Test on Pinhole AQL | FDA requirements for waterleak test on pinhole AQL | Meets FDA requirements for waterleak test on pinhole AQL |
General Requirements | Meets 21 CFR 800.20 | Meets 21 CFR 800.20 |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for each specific test (e.g., ASTM D5151, D6124, biocompatibility). However, it implies that the device "meets" these standards, which typically involve specific sampling plans and test methods outlined within the respective standard.
- Data Provenance: The manufacturing company is HEBEI LEADER PLASTIC CO., LTD, located in Shijiazhuang, China. The testing likely occurred in China, although the specific laboratories are not identified.
- Retrospective or Prospective: The testing appears to be prospective as it was conducted to demonstrate substantial equivalence for the premarket notification (510(k)) of a new device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
This type of information is not applicable (N/A) for this device. The "ground truth" for medical gloves is established by adherence to recognized consensus standards (like ASTM and ISO) for physical properties and biocompatibility. It does not involve subjective human interpretation or diagnostic decision-making that would require expert consensus on a test set.
4. Adjudication Method for the Test Set
This is not applicable (N/A) for this device. Adjudication methods like 2+1 or 3+1 are used in studies where multiple human readers evaluate data (e.g., medical images) and their disagreements need to be resolved to establish a definitive ground truth. For medical gloves, performance is measured against objective physical and chemical standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This type of study is relevant for diagnostic devices (often AI-powered) where human performance can be augmented or improved by AI assistance. This device is a patient examination glove, not an AI diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
No, a standalone (algorithm only) performance assessment was not done. This concept is applicable to AI/ML algorithms. The device being reviewed is a physical medical glove.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device is based on recognized consensus standards (ASTM D5250, ASTM D5151, ASTM D6124) and regulatory requirements (21 CFR 800.20, ISO 10993-10 for biocompatibility). These standards define objective pass/fail criteria for the physical and chemical properties of the gloves.
8. The Sample Size for the Training Set
This is not applicable (N/A). Training sets are used for machine learning algorithms. The evaluation of this medical glove relies on physical and chemical testing against established standards, not on training an AI model.
9. How the Ground Truth for the Training Set Was Established
This is not applicable (N/A) for the same reason as above. There is no AI training set involved.
Ask a specific question about this device
(232 days)
POWDER FREE VINYL PATIENT EXAMINATION GLOVES, COLORED (YELLOW)
Powder Free Vinyl Patient Examination Gloves, Colored (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free Vinyl Patient Examination Gloves, Colored (Yellow) are disposable devices which made of poly(vinyl chloride) polymer compound, intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between pattent and examiner and they meets all of the requirements of ASTM standard D 5250-06 el .
1. Acceptance Criteria and Reported Device Performance
Characteristics | Acceptance Criteria (Standard) | Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-06e1 | Meets |
Physical Properties | ASTM standard D 5250-06e1 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Residual | ASTM standard D 5250-06e1 and D6124-01 | Meets ( |
Ask a specific question about this device
(44 days)
POWDER FREE VINYL PATIENT EXAMINATION GLOVES, COLORED (YELLOW)
Powder Free Vinyl Patient Examination Gloves, Colored (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free Vinyl Patient Examination Gloves, Colored (Yellow) that meets all of the requirements of ASTM standard D 5250-06.
Here's an analysis of the acceptance criteria and study information for the K102570 submission, formatted as requested:
Acceptance Criteria and Device Performance Study for K102570
The device in question is "Powder free Vinyl Patient Examination Gloves, Colored (Yellow)." The submission focuses on demonstrating substantial equivalence to a predicate device by meeting established ASTM standards and FDA regulations for patient examination gloves.
1. Table of Acceptance Criteria and Reported Device Performance
Characteristic | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-06 | Meets |
Physical Properties | ASTM standard D 5250-06 | Meets |
Freedom from pinholes | 21 CFR 800.20 (Waterleak test on pinhole AQL) | Meets |
Powder Residual | ASTM standard D 5250-06 and D6124-01 |
Ask a specific question about this device
(44 days)
POWDER FREE VINYL PATIENT EXAMINATION GLOVES, COLORED (YELLOW)
Powder Free Vinyl Patient Examination Gloves, Colored (Yellow), is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free Vinyl Patient Examination Gloves, Colored (Yellow) that meets all of the requirements of ASTM standard D 5250-06.
The provided text describes a 510(k) summary for "Powder Free Vinyl Patient Examination Gloves, Colored (Yellow)". This is a medical device and not an AI/ML powered device, therefore most of the requested information (e.g., sample size for test/training set, number of experts, adjudication method, MRMC study, standalone performance) is not applicable or not found in the document.
Here's an attempt to answer the questions based on the provided text, indicating where information is not applicable or not found:
1. Table of Acceptance Criteria and Reported Device Performance
Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-06 | Meets |
Physical Properties | ASTM standard D 5250-06 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Residual | ASTM standard D 5250-06 and D6124-01 | Meets ( |
Ask a specific question about this device
(44 days)
POWDER FREE VINYL PATIENT EXAMINATION GLOVES, COLORED (YELLOW)
Powder Free Vinyl Patient Examination Gloves, Colored (Yellow) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder free Vinyl Patient Examination Gloves, Colored (Yellow) that meets all of the requirements of ASTM standard D 5250-06.
Here's a breakdown of the acceptance criteria and related study information for the device described in the provided 510(k) summary:
This device is a Powder free Vinyl Patient Examination Glove. The 510(k) summary focuses on demonstrating substantial equivalence to a predicate device, which primarily involves meeting recognized standards for physical properties, safety, and performance.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are generally established by recognized ASTM standards and FDA regulations for patient examination gloves. The reported device performance states that the device "Meets" these standards.
Characteristic | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-06 | Meets |
Physical Properties | ASTM standard D 5250-06 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Residual | ASTM standard D 5250-06 and D6124-01 |
Ask a specific question about this device
(181 days)
POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, COLORED (YELLOW)
Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder-free Vinyl Patient Examination Gloves, Colored (Yellow) that meets all of the requirements of ASTM standard D 5250-06.
Here's a breakdown of the requested information based on the provided 510(k) summary for the Powder-free Vinyl Patient Examination Gloves, Colored (Yellow).
It's important to note that this document is for a Class I medical device (patient examination gloves). For devices of this classification, the regulatory requirements are generally less stringent compared to higher-risk devices, and often the "studies" primarily involve adherence to recognized standards rather than complex clinical trials or AI performance evaluations.
Here is the requested information:
1. Table of acceptance criteria and the reported device performance
Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-06 | Meets |
Physical Properties | ASTM standard D 5250-06 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Residual | ASTM standard D 5250-06 and D6124-01 |
Ask a specific question about this device
(34 days)
POWDER-FREE VINYL PATIENT EXAMINATION GLOVES, COLORED (YELLOW)
Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powder free vinyl patient examination gloves, Colored (Yellow) that meets all of the requirements of ASTM standard D 5250-06.
Here's a breakdown of the acceptance criteria and the study information based on the provided text, formatted to your requests:
Acceptance Criteria and Device Performance Study
The device reviewed is "Powder Free Vinyl Patient Examination Gloves, Colored (Yellow)" from Hebei Handform Plastic Products Co.,Ltd.
1. Table of Acceptance Criteria and Reported Device Performance
Characteristics | Acceptance Criteria (Standard) | Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-06- | Meets |
Physical Properties | ASTM standard D 5250-06- | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Residual | ASTM standard D 5250-06 and D6124-06 |
Ask a specific question about this device
(56 days)
POWDER FREE VINYL PATIENT EXAMINATION GLOVES, COLORED (YELLOW)
Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powder free vinyl patient examination gloves, Colored (Yellow) that meets all of the requirements of ASTM standard D 5250-06.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Powder Free Vinyl Patient Examination Gloves, Colored (Yellow):
Acceptance Criteria and Device Performance Study
The information provided describes the non-clinical testing performed to establish substantial equivalence for the Powder Free Vinyl Patient Examination Gloves, Colored (Yellow).
1. Table of Acceptance Criteria and Reported Device Performance
Characteristic | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Physical Properties | ||
Dimension | ASTM standard D 5250-06 | Meets |
Physical Properties (e.g., tensile strength, elongation) | ASTM standard D 5250-06 | Meets |
Freedom from Pinhole | 21 CFR 800.20 (Water Leak Test AQL) | Meets |
Powder Residual | ASTM standard D 5250-06 and D6124-06 |
Ask a specific question about this device
(46 days)
POWDER FREE VINYL PATIENT EXAMINATION GLOVES, COLORED (YELLOW)
Powder free vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination' between patient and examiner.
Powder-free Vinyl Patient Examination Gloves, Colored (Yellow) that meets all of the requirements of ASTM standard D 5250-06.
This document describes the safety and effectiveness information for the Powder-free Vinyl Patient Examination Gloves, Colored (Yellow) manufactured by Tangshan Jiteng Plastic Products Co., Ltd.
1. Table of Acceptance Criteria and Reported Device Performance
Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-06 | Meets |
Physical Properties | ASTM standard D 5250-06 | Meets |
Freedom from pinholes | 21 CFR 800.20 | Meets |
Powder Residual | ASTM standard D 5250-06 and D6124-06 | Meets |
Biocompatibility (Primary Skin Irritation) | ISO10993-10 (Primary Skin Irritation in rabbits) | Passes (Not a Primary Skin Irritation) |
Biocompatibility (Dermal Sensitization) | ISO10993-10 (Dermal sensitization in the guinea pig) | Passes (Not a Dermal sensitization) |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the exact sample sizes used for each specific test (e.g., dimension, physical properties, pinholes, powder residual, biocompatibility tests). However, it implies that the device was tested to meet the requirements of the specified ASTM and CFR standards, which typically involve specific sampling plans.
- Data Provenance: The testing was conducted by or for Tangshan Jiteng Plastic Products Co., Ltd., based in China. The data is reported as part of a premarket notification for a medical device, indicating it's retrospective testing performed to demonstrate compliance with established standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
This information is not applicable as the device is a patient examination glove, and the objective safety and performance characteristics are determined by standardized physical and chemical tests rather than expert interpretation of medical images or data. The "ground truth" for these tests is the quantitative measurement against the established limits in the respective standards (ASTM D 5250-06, D6124-06, 21 CFR 800.20, ISO10993-10).
4. Adjudication Method for the Test Set
This is not applicable as the testing involves objective measurements against established standards, not judgmental evaluations requiring adjudication.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human readers interpret medical images or data. The device in question is a medical glove, for which such a study is not relevant.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
This is not applicable as the device is a physical medical glove and does not involve any algorithms or artificial intelligence.
7. The Type of Ground Truth Used
The ground truth used for performance evaluation is based on objective measurement against established regulatory and consensus standards. This includes:
- ASTM standard D 5250-06: For dimensions and physical properties (e.g., tensile strength, elongation).
- 21 CFR 800.20: For freedom from pinholes (water leak test).
- ASTM standard D 6124-06: For powder residual.
- ISO10993-10: For biocompatibility (Primary Skin Irritation and Dermal Sensitization).
8. The Sample Size for the Training Set
This is not applicable. The device is a physical product and does not involve any machine learning algorithms that would require a training set.
9. How the Ground Truth for the Training Set Was Established
This is not applicable, as there is no training set for this device.
Ask a specific question about this device
Page 1 of 1