Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K111772
    Date Cleared
    2011-10-20

    (119 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    POWDER FREE NITRILE PATIENT EXAMINATION GLOVE, BLUE COLORED AND WHITE (NON-COLORED). NON-STERILE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Powder Free Nitrile Patient Examination Glove, Blue Colored, and White (Non Colored). Non-Sterile, meets all of the requirements of ASTM D 6319-10.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the "Powder Free Nitrile Patient Examination Glove, Blue Colored, and White (Non-Colored). Non-Sterile" device:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criteria (Standards)Device Performance
    DimensionsASTM D 6319-10Meets
    Physical PropertiesASTM D 6319-10Meets
    Freedom from pin-holesASTM D 5151-06
    ASTM D 6319-10Meets
    Powder Free ResidueASTM D 6124-06
    ASTM D 6319-10Meets
    Biocompatibility:
    Dermal SensitizationISO 10993-10:2010Not a contact skin sensitizer
    Primary Skin Irritation16 CFR Part 1500Not a primary skin irritant

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample sizes used for each individual test (Dimensions, Physical Properties, Freedom from pin-holes, Powder Free Residue, Biocompatibility). It only states that the device "meets" the referenced ASTM and ISO standards. These standards themselves would define the required sample sizes for their respective tests.

    The data provenance is not explicitly stated in terms of country of origin of the data or whether it was retrospective or prospective. Given that Kossan Latex Industries is a Malaysian company and they are claiming compliance with international standards, it's highly probable the testing was conducted by or overseen by the manufacturer, potentially in Malaysia or by certified testing labs. The document implies these are prospective tests conducted to demonstrate compliance for the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is not applicable to this submission. The device is a patient examination glove, and its performance is evaluated against objective, measurable, and standardized technical specifications (ASTM, ISO, CFR standards). Ground truth, in the sense of expert consensus, is typically relevant for interpretative tasks, such as diagnostic imaging analysis or clinical assessments, which is not the case for this product.

    4. Adjudication Method for the Test Set

    This is not applicable. The performance evaluations are based on direct measurement and laboratory testing against pre-defined, quantitative standards, not on subjective interpretations requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This submission is for a physical medical device (patient examination glove), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study and analysis of human reader improvement with AI assistance are irrelevant.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable for the same reasons as point 5. There is no algorithm involved with this device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is the objective measurement and testing results against the established quantitative criteria defined in the referenced standards (ASTM D 6319-10, ASTM D 5151-06, ASTM D 6124-06, ISO 10993-10:2010, and 16 CFR Part 1500). For example:

    • Dimensions: Measured directly against the specified ranges in ASTM D 6319-10.
    • Physical Properties: Tensile strength, elongation, etc., measured according to methods outlined in ASTM D 6319-10.
    • Freedom from pin-holes: Water leak test (ASTM D 5151-06) or air leak test methods.
    • Powder Free Residue: Measured gravimetrically as per ASTM D 6124-06.
    • Biocompatibility: Standardized biological assays (dermal sensitization, primary skin irritation) with predefined endpoints for "not a sensitizer" or "not an irritant."

    8. The Sample Size for the Training Set

    This is not applicable. This is a physical device, and there is no "training set" in the context of machine learning or AI. Performance is based on physical testing of manufactured samples.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1