Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K110287
    Date Cleared
    2011-02-15

    (15 days)

    Product Code
    Regulation Number
    870.2900
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PHILIPS ECG LEADWIRE SET

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Philips ECG leadsets are indicated for use in the monitoring of cardiac signals for both diagnostic and monitoring purposes. Use is limited by the indications for use of the connected monitoring or diagnostic equipment.

    Device Description

    The Philips ECG Leadwire Set is a single patient electrode cable system used to transfer signals from patient electrodes to various electrocardiograph recorders and monitors. The system is designed to provide a family of lead wires that will link the patient and the compatible patient trunk cable system.

    AI/ML Overview

    This document describes the 510(k) premarket notification for the Philips ECG Leadwire Set (K110287). This submission is for a medical device (leadwire set) and not an AI/ML powered device, therefore much of the requested information (sample sizes, ground truth, expert qualifications, adjudication methods, MRMC studies) is not applicable.

    Here's the breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this device, the "acceptance criteria" are compliance with established electrical and biocompatibility standards, and the "reported device performance" is the claim of substantial equivalence to the predicate device based on meeting these standards and having similar technological characteristics.

    Acceptance CriteriaReported Device Performance
    Compliance with recognized standards:Philips ECG Leadwire Set is designed and tested to conform to the following FDA-recognized standards:
    - IEC 60601-1:1998 (Medical electrical equipment)Achieved through verification testing.
    - UL 60601-1:2003 (National deviations for IEC 60601-1)Achieved through verification testing.
    - ANSI/AAMI ES 60601-1:2005 (National deviations for IEC 60601-1)Achieved through verification testing.
    - AAMI/ANSI EC 13:2002/(R)2007 (Cardiac monitors, heart rate meters and alarms)Achieved through verification testing.
    - AAMI/ANSI EC53:1995/(R) 2008 (ECG cables and leadwires, including Amendment 1)Achieved through verification testing.
    - AAMI/ANSI/ISO 10993-5:2009 (Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity)Achieved through verification testing.
    - ISO 10993-10:2002 (Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity)Achieved through verification testing.
    - Product Inspection for non-measurable characteristicsCompliance achieved for characteristics such as color and intrinsic design.
    - Substantial Equivalence to Predicate Device (K102430)The new device has equivalent technological characteristics related to safety and effectiveness as the predicate device. The primary difference (grabber vs. snap connector) is considered substantially equivalent. Indications for use are identical.

    2. Sample Size Used for the Test Set and Data Provenance

    This is not applicable as this is a traditional medical device submission (ECG leadwire set) and not an AI/ML device that would involve a "test set" of data in the sense of patient data for model evaluation. The "testing" referred to is engineering verification testing against recognized standards, not performance evaluation on a patient dataset.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not applicable. Ground truth, in the context of AI/ML, refers to reference labels for a dataset. For this device, the "ground truth" is defined by compliance with established engineering and biocompatibility standards. Expert review would be part of the internal design and testing process, but not in the format of "ground truth" for a dataset.

    4. Adjudication Method for the Test Set

    Not applicable. There is no "test set" in the AI/ML context described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML-powered diagnostic or screening devices to assess how the AI impacts human reader performance, which is not applicable to a passive leadwire set.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone performance study was not done. This is also relevant for AI/ML algorithms and not for an ECG leadwire set, which is a physical component.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance evaluation is compliance with recognized international and national standards for medical electrical equipment, cardiac monitors, ECG cables and leadwires, and biological evaluation of medical devices. This involves objective measurements and assessments against predefined criteria specified in these standards.

    8. The Sample Size for the Training Set

    Not applicable. There is no AI/ML algorithm requiring a "training set" for this device.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1