Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K102958
    Date Cleared
    2011-03-18

    (165 days)

    Product Code
    Regulation Number
    884.2740
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Avalon Fetal Monitor FM20: Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, non-invasive blood pressure, pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.

    Avalon Fetal Monitor FM30: Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.

    Avalon Fetal Monitor FM40: Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas,

    Avalon Fetal Monitor FM50: Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.

    Device Description

    The Philips Avalon Fetal/Maternal Monitors FM20, FM30, FM40 and FM50 offer monitoring of fetal and maternal heart rates, uterine activity, maternal ECG, maternal non-invasive blood pressure (NIBP) and oxygen saturation (Sp02) with pulse rate, during antepartum testing and labor and delivery.

    The Avalon Fetal/Maternal Monitor FM20 provides the following external measurement parameters:

    • Up to three Fetal Heart Rates (FHR) via ultrasound (US)
    • Fetal Movement Profile
    • Uterine activity via external Toco
    • Maternal Pulse Rate
    • Maternal Heart Rate (MHR) via maternal ECG
    • Non-invasive blood pressure (NIBP)

    The Avalon FM30 shares all the features and capabilities of the Avalon FM20. In addition, the Avalon FM30 provides the following external and internal measurement parameters:

    • Single Fetal Heart Rate via direct ECG (DECG)
    • Twin Fetal Heart Rates via direct ECG (DECG) and/or ultrasound
    • Uterine activity via intrauterine pressure (IUP)
    • Maternal ECG (MECG)
    • Pulse oximetry (maternal Sp02).
      *) Note: maximum three FHR's can be monitored.

    The Avalon Fetal/Maternal Monitor FM40 provides the following external measurement parameters:

    • Up to three Fetal Heart Rates (FHR) via ultrasound (US)
    • Fetal Movement Profile
    • Uterine activity via external Toco
    • Maternal Pulse Rate
    • Maternal Heart Rate (MHR) via maternal ECG
    • Non-invasive blood pressure (NIBP).
    • Pulse oximetry (maternal Sp02).

    The Avalon FM50 shares all the features and capabilities of the Avalon FM40. In addition, the Avalon FM50 provides the following external and internal measurement parameters:

    • Single Fetal Heart Rate via direct ECG *) (DECG)
    • Twin Fetal Heart Rates , via direct ECG (DECG) and/or ultrasound "
    • Uterine activity via intrauterine pressure , (IUP)
    • Maternal ECG (MECG) ,
      " Note: maximum three FHR's can be monitored.
    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary (K102958) does not explicitly state specific quantitative acceptance criteria or detailed device performance metrics in a readily extractable table format.

    Instead, it refers to the existing specifications cleared for the predicate devices as the basis for Pass/Fail criteria. The study concluded that the modified algorithm met these unstated criteria.

    Acceptance Criteria (Explicitly Stated)Reported Device Performance (Explicitly Stated)
    "Pass/Fail criteria were based on the specifications cleared for the predicate devices" (K071800 and K092028)."test results showed substantial equivalence."
    "The results demonstrate that the Philips Avalon Fetal Monitors meet all reliability requirements and performance claims."The improved ultrasound performance algorithm addressed "problems with halving and doubling of the fetal heart rate and mistaken display of maternal heart rate for fetal heart rate."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated. The document mentions "Some tests included the use of previously recorded patient traces," but doesn't quantify the number of traces or patients.
    • Data Provenance: Not explicitly stated. The term "previously recorded patient traces" suggests retrospective data, but the country of origin is not mentioned.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

    • Number of Experts: Not mentioned.
    • Qualifications of Experts: Not mentioned.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not mentioned. It's unclear how ground truth was established from the "previously recorded patient traces" or if any adjudication was performed for the test set.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No, there is no mention of a multi-reader, multi-case comparative effectiveness study involving human readers with and without AI assistance.
    • Effect Size: Not applicable, as no MRMC study was reported.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Was a standalone study done? Yes, the document describes "system level and regression tests" and testing of "the improved ultrasound performance algorithm." This implies a standalone evaluation of the algorithm's performance in addressing the identified issues of heart rate inaccuracies (halving, doubling, maternal heart rate mistaken for fetal). The focus of the 510(k) is on the algorithm modification itself.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Not explicitly stated. The document mentions using "previously recorded patient traces." For evaluating the accuracy of fetal heart rate detection and differentiation from maternal heart rate, an accurate reference (e.g., from a different, highly accurate monitoring method, or expert review of the original traces) would be required, but this is not detailed.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. The document discusses an "improved ultrasound performance algorithm" which was a modification to existing devices. There is no mention of a "training set" in the context of machine learning or AI development in this summary. The algorithm improvement seems to be a programmatic enhancement rather than a machine learning model trained on data.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth Was Established for Training Set: Not applicable, as no training set for an AI/ML model is indicated. The improvements described are algorithm modifications to address specific known problems with heart rate detection.
    Ask a Question

    Ask a specific question about this device

    K Number
    K092028
    Date Cleared
    2010-02-17

    (226 days)

    Product Code
    Regulation Number
    884.2740
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Philips Avalon FM20, FM30, FM40 and FM50 Fetal/Maternal Monitors are intended for non-invasive monitoring of the physiological parameters of pregnant women during antepartum testing and labor and delivery. All monitors are intended for monitoring fetal heart rates, fetal movement profiles, maternal heart rates, uterine activity, maternal noninvasive blood pressure, and additionally for the FM30, FM40 and FM50 maternal oxygen saturation (Sp02). The FM30 and FM50 are additionally intended for maternal ECG and for invasive monitoring of fetal Direct ECG and intrauterine pressure. All monitors are intended for generating alarms from fetal and maternal parameters, for displaying, storing and recording patient data and related waves, transmitting patient data to a patient information and surveillance system on a network, and for postpartum monitoring of the mother. All monitors are intended for use by trained health care professionals. They are intended for use in labor and delivery rooms, antepartum testing areas and during postpartum recovery in the hospital environment. They are not intended for use in intensive care units or operating rooms. The FM20 and FM30 are additionally intended for use in healthcare and for example in doctors' offices, and in private households.

    Device Description

    The subject devices Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 are a modification of the legally marketed Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 and offer monitoring of fetal heart rates, fetal movement profiles, fetal Direct ECG, maternal uterine activity, maternal heart rates, maternal pulse rates, maternal ECG, maternal non-invasive blood pressure (NIBP) and maternal oxygen saturation (Sp02) during antepartum testing and labor and delivery. The modification of the Philips Avalon Fetal Monitors FM20, FM30, FM40 and FM50 only permits specific Sp02 sensors, which are already cleared for the Philips picoSATIIplus SpO2 pulse oximetry module, the Philips Intellivue X2 (M3002A) multi measurement module and the Philips M1020B Sp02 plug-in module, to be also used with the Philips Avalon Fetal monitors FM30, FM40 and FM50.

    AI/ML Overview

    The Philips Avalon Fetal Monitors (FM20, FM30, FM40, and FM50) are a modification of previously cleared devices, primarily to permit the use of specific SpO2 sensors already cleared for other Philips modules. The submission focuses on demonstrating substantial equivalence to these predicate devices.

    Acceptance Criteria and Device Performance:

    The document explicitly states that "hazard criteria were based on the Pass/Fail analysis. specifications cleared for the predicate devices and test results showed substantial equivalence." This implies that the acceptance criteria for the modified devices were the same performance specifications as those established for the predicate devices. The study concludes that the "results demonstrate that the Philips Avalon Fetal Monitors meet all reliability requirements and performance claims."

    Acceptance Criteria (Implied)Reported Device Performance
    Performance, functionality, and reliability characteristicsMeets all reliability requirements and performance claims, demonstrating substantial equivalence to predicate devices.
    Compliance with specifications cleared for predicate devicesTest results showed substantial equivalence to predicate device specifications.
    Pass/Fail analysis based on hazard criteriaTesting involved regression and system-level tests; hazard criteria met.

    Study Details:

    1. Sample Size and Data Provenance (Test Set):

      • The document does not specify the sample size used for the test set.
      • The data provenance (country of origin, retrospective/prospective) is not mentioned.
    2. Number of Experts and Qualifications (Ground Truth for Test Set):

      • The document does not provide information on the number of experts or their qualifications used to establish ground truth for the test set.
    3. Adjudication Method (Test Set):

      • The document does not specify any adjudication method (e.g., 2+1, 3+1, none) for the test set.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No MRMC comparative effectiveness study is mentioned. The submission focuses on device performance and substantial equivalence, not human reader improvement with AI assistance.
    5. Standalone Performance Study:

      • A standalone performance study was implicitly done, as "Verification and Validation testing activities" were conducted to "establish the performance, functionality, and reliability characteristics of the modified devices." This testing was against predefined specifications, indicating an algorithm-only (device-only) performance assessment.
    6. Type of Ground Truth Used:

      • The document does not explicitly state the type of ground truth used (e.g., expert consensus, pathology, outcomes data). The ground truth for the verification and validation appears to be the "specifications cleared for the predicate devices."
    7. Sample Size for Training Set:

      • The document does not mention a training set, as the evaluation method described is primarily focused on verification and validation against pre-established specifications for the modified device, rather than training a new algorithm from data.
    8. How Ground Truth for Training Set was Established:

      • Not applicable, as no training set is mentioned in the context of this submission. The device is a modification of an existing, cleared device, implying its core algorithms were already established and validated. The current submission focuses on verifying continued performance after a component modification.
    Ask a Question

    Ask a specific question about this device

    K Number
    K071800
    Date Cleared
    2007-09-27

    (87 days)

    Product Code
    Regulation Number
    884.2740
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Avalon Fetal Monitor FM20:
    Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, non-invasive blood pressure, pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.

    Avalon Fetal Monitor FM30:
    Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms, in antepartum testing areas and in private households.

    Device Description

    The Philips Avalon Fetal Monitors FM20 and FM30 are intended for noninvasive monitoring of parameters of pregnant women during antepartum testing and labor and delivery. The FM30 is additionally intended for invasive monitoring. All monitors are intended for monitoring fetal and maternal heart rates, uterine activity, maternal noninvasive blood pressure, and additionally for the FM30, oxygen saturation (SpO2). All monitors are intended for generating alarms for displaying, storing and recording patient data and related waves, transmitting patient data to a patient information and surveillance system on a network, and for postpartum monitoring of the mother.

    AI/ML Overview

    The provided 510(k) summary (K071800) for the Philips Avalon Fetal Monitors FM20 and FM30 does not contain information related to specific acceptance criteria or details of a study proving device performance against such criteria.

    This 510(k) is a premarket notification for a modification to an already cleared device: the Philips Avalon Fetal Monitors FM20 and FM30 (cleared under K052795 and K062137). The modification only introduces the capability for healthcare professionals to monitor pregnant women in private households.

    Therefore, the submission relies on the substantial equivalence to the previously cleared devices and does not present new performance study data. The document focuses on regulatory classification, intended use, and substantial equivalence, rather than detailed performance metrics.

    Based on the provided text, I cannot complete the requested information for the acceptance criteria and the study that proves the device meets them. The pertinent sections of the request that cannot be addressed due to the lack of information in the document are:

    1. A table of acceptance criteria and the reported device performance: Not available.
    2. Sample sized used for the test set and the data provenance: Not available.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not available.
    4. Adjudication method for the test set: Not available.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable/Available. (The device is not AI-assisted).
    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable/Available. (The device does not involve a standalone algorithm for diagnostic performance).
    7. The type of ground truth used: Not available.
    8. The sample size for the training set: Not available.
    9. How the ground truth for the training set was established: Not available.
    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Avalon Fetal Monitor FM30: Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rate in labor and delivery rooms and in antepartum testing areas.

    Avalon Fetal Monitor FM20: Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, non-invasive blood pressure, pulse rate of pregnant women and the fetal heart rate in labor and delivery rooms and in antepartum testing areas.

    Device Description

    The subject devices Philips Avalon Fetal Monitors FM20 and FM30 are modification of the legally marketed Philips Series 50XM (M1350B) Fetal/Maternal Monitor (K954351) and offer monitoring of fetal and maternal heart rates, uterine activity, maternal ECG, maternal non-invasive blood pressure (NIBP) and oxygen saturation (Sp02) during antepartum testing and labor and delivery. The Avalon FM20 and FM30 are intended for non-invasive and invasive monitoring of the physiological parameters of pregnant women during antepartum testing and labor and delivery. The Avalon FM20 and FM30 are intended for monitoring fetal and maternal heart rates, uterine activity, maternal non-invasive blood pressure (NIBP) and oxygen saturation (SpO2). The Avalon FM20 and FM30 are intended for generating alarms, for displaying, storing and recording patient data and related waves.

    AI/ML Overview

    Here's an analysis of the provided text regarding acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The provided text does not explicitly list specific quantitative acceptance criteria or detailed performance metrics. Instead, it makes a general statement:

    Acceptance Criteria CategoryReported Device Performance
    Performance, Functionality, Reliability"meet all reliability requirements and performance claims." "Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence."

    Explanation: The submission is for a modified device (Philips Avalon Fetal Monitors FM20 and FM30) that is stated to have the "same technological characteristics" and "same intended use" as legally marketed predicate devices. The "acceptance criteria" here are implicitly tied to demonstrating that the modified device performs equivalently to the predicate devices and meets their established specifications. The text confirms that this equivalence was met.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The submission describes verification, validation, and testing activities against predicate device specifications, not a clinical study involving experts establishing ground truth for diagnostic accuracy.

    4. Adjudication method for the test set

    Not applicable. The submission describes engineering and system-level testing, not a clinical study requiring adjudication of expert interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-powered diagnostic device, and no MRMC comparative effectiveness study is mentioned. The device is a fetal monitor.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    The device itself is a standalone fetal monitor designed to provide physiological data. Its "standalone" performance was assessed through system-level, performance, and safety tests against predicate device specifications. There is no "algorithm only" performance reported in the context of AI.

    7. The type of ground truth used

    The "ground truth" for the device's performance was established by the specifications cleared for the predicate device. The testing demonstrated that the modified device met these existing specifications.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1