(29 days)
Not Found
No
The provided text does not mention AI, ML, or any related terms, and the device description focuses on standard physiological monitoring and data handling.
No
The device is described as a "Fetal Monitor" and its intended use is for "monitoring of physiological parameters," which are diagnostic functions, not therapeutic.
No.
The device monitors physiological parameters and generates alarms, but it is not described as providing a diagnosis or diagnostic information.
No
The device description explicitly states that the devices are "Philips Avalon Fetal Monitors FM20 and FM30" and are intended for "non-invasive and invasive monitoring of the physiological parameters". This indicates a hardware component is involved in the monitoring process, not just software. The summary also mentions "Verification, validation, and testing activities were conducted to establish the performance, functionality, and reliability characteristics of the modified device", which would typically include hardware testing for a physical medical device.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to diagnose, monitor, or screen for various health issues.
- Device Function: The Avalon Fetal Monitor FM20 and FM30 are used for monitoring physiological parameters of pregnant women and fetuses. This involves using sensors (like ultrasound transducers for fetal heart rate, and cuffs for blood pressure) that are applied externally to the body.
- No Sample Analysis: The device does not analyze samples taken from the body. It directly measures physiological signals from the body.
Therefore, based on the provided information, the Avalon Fetal Monitor FM20 and FM30 fall under the category of medical devices used for monitoring, not in vitro diagnostics.
N/A
Intended Use / Indications for Use
Avalon Fetal Monitor FM30:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.
Avalon Fetal Monitor FM20:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, non-invasive blood pressure, pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.
Product codes
MAA, HEL, HEK, KNG, HGP, HGS, KXO, HFO, HFN, HFM, HGM, HGL, DSJ, DSK, DXN, DRQ, DRT, DPS, DRJ, DQA, DSF, DSA
Device Description
The modification of the Philips Avalon Fetal Monitors FM20 and FM30 introduces the capability of monitoring fetal heart rates of triplets. The Philips Avalon Fetal Monitors FM20 and FM30 are intended for non-invasive and invasive monitoring of the physiological parameters of pregnant women during antepartum testing and labor and delivery. The Avalon FM20 and FM30 are intended for monitoring fetal and maternal heart rates, uterine activity, maternal non-invasive blood pressure (NIBP) and maternal oxygen saturation (SpO2). The Avalon FM20 and FM30 are intended for generating alarms, for displaying, storing and recording patient data and related waves.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
intended for use by trained health care professionals in labor and delivery rooms and in antepartum testing areas. They are not intended for use in intensive care units, operating rooms or outside the health care facilities.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Verification, validation, and testing activities were conducted to establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicates. Testing involved performance tests, regression tests, and testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence. The results demonstrate that the Philips Avalon Fetal Monitors FM20 and FM30 meet all reliability requirements and performance claims.
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 884.2740 Perinatal monitoring system and accessories.
(a)
Identification. A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.(b)
Classification. Class II (performance standards).
0
510 (k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
-
- The submitter of this premarket notification is: Markus Stacha Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany
Tel: ++49 7031 463-2840 Fax: ++49 7031 463-2442 Email: markus.stacha@philips.com
- The submitter of this premarket notification is: Markus Stacha Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany
This summary was prepared on July 21, 2006.
-
- The name of the devices is the Philips Avalon Fetal Monitors FM20 and FM30. Classification names are as follows:
Device Panel | Classification | ProCode | Description |
---|---|---|---|
Obstetrical and | |||
Gynecological | |||
Monitoring | |||
Devices | $884.2660, II | MAA | |
HEL | |||
HEK | |||
KNG | Fetal ultrasonic monitor and | ||
accessories | |||
$884.2675, II | HGP | Fetal scalp circular (spiral) | |
electrode and applicator | |||
$884.2700, II | HGS | ||
KXO | |||
HFO | |||
HFN | Intrauterine pressure monitor | ||
and accessories | |||
$884.2720, II | HFM | External uterine contraction | |
monitor and accessories | |||
$884.2740, II | HGM | Perinatal monitoring system and | |
accessories | |||
Circulatory | |||
System Devices | $884.2960, II | HGL | Obstetric ultrasonic transducer |
and accessories | |||
$870.1100, II | DSJ | Alarm, Blood Pressure | |
$870.1110, II | DSK | Computer, Blood Pressure | |
$870.1130, II | DXN | System, Measurement, Blood- | |
Pressure, Non-Invasive | |||
$870.2060, II | DRQ | Amplifier and Signal | |
Conditioner, Transducer Signal | |||
$870.2300, II | DRT | Monitor, Cardiac (incl. | |
Cardiotachometer & Rate Alarm) | |||
$870.2340, II | DPS | Electrocardiograph | |
$870.2600, I | DRJ | System, Signal Isolation | |
$870.2700, II | DQA | Oximeter | |
$870.2810, I | DSF | Recorder, Paper Chart | |
$870.2900, I | DSA | Cable, Transducer and | |
Electrode, incl. Patient | |||
Connector |
-
- The modified devices Philips Avalon Fetal Monitors FM20 and FM30 are substantially equivalent to previously cleared
1
Philips Avalon Fetal Monitors FM20 and FM30 marketed pursuant to K052795.
-
- The modification of the Philips Avalon Fetal Monitors FM20 and FM30 introduces the capability of monitoring fetal heart rates of triplets.
-
- The modified devices Philips Avalon Fetal Monitors FM20 and FM30 have the same intended use as the legally marketed predicate device Philips Avalon Fetal Monitors FM20 and FM30. The Philips Avalon Fetal Monitors FM20 and FM30 are intended for non-invasive and invasive monitoring of the physiological parameters of pregnant women during antepartum testing and labor and delivery. The Avalon FM20 and FM30 are intended for monitoring fetal and maternal heart rates, uterine activity, maternal non-invasive blood pressure (NIBP) and maternal oxygen saturation (SpO2). The Avalon FM20 and FM30 are intended for generating alarms, for displaying, storing and recording patient data and related waves. The Avalon FM20 and FM30 are intended for use by trained health care professionals in labor and delivery rooms and in antepartum testing areas. They are not intended for use in
-
- The modified devices Philips Avalon Fetal Monitors FM20 and FM30 have the same technological characteristics as the legally marketed predicate devices Avalon Fetal Monitors FM20 and FM30.
intensive care units, operating rooms or outside the health
care facilities.
-
- Verification, validation, and testing activities were conducted to establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicates. Testing involved performance tests, regression tests, and testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence. The results demonstrate that the Philips Avalon Fetal Monitors FM20 and FM30 meet all reliability requirements and performance claims.
-2-
- Verification, validation, and testing activities were conducted to establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicates. Testing involved performance tests, regression tests, and testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence. The results demonstrate that the Philips Avalon Fetal Monitors FM20 and FM30 meet all reliability requirements and performance claims.
2
Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health and Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human figures, possibly representing health, family, and community.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 4 2006
Mr. Markus Stacha Senior Regulatory Affairs Engineer Philips Medizin Systeme Böblingen GmbH Hewlett-Packard-Str. 1 71034 Böblingen GERMANY
Re: K062137
Trade/Device Name: Philips Avalon Fetal Monitors FM20 and FM30 Regulation Number: 21 CFR 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: II Product Code: HGM and DXN Dated: July 21, 2006 Received: July 26, 2006
Dear Mr. Stacha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following ultrasound transducer intended for use with the Philips Avalon Fetal Monitors FM20 and FM30, as described in your premarket notification:
Transducer Model Number
M2736A, 1 MHz Ultrasound Transducer
3
Page 2 - Mr. Markus Stacha
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G. (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small
4
Page 3 -- Mr. Markus Stacha
Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Mr. Glenn Bell at (301) 594-1180.
Sincerely yours,
David A. Leppsm
for
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
5
Indications for Use
510(k) Number (if known):_KOG2 137
Device Name: Philips Avalon Fetal Monitors FM20 and FM30.
Indications for Use:
Avalon Fetal Monitor FM30:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, ECG, oxygen saturation, non-invasive blood pressure, and pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.
Avalon Fetal Monitor FM20:
Indicated for use by trained health care professionals whenever there is a need for monitoring of the physiological parameters uterine activity, heart rate, non-invasive blood pressure, pulse rate of pregnant women and the fetal heart rates of single fetuses, twins, and triplets in labor and delivery rooms and in antepartum testing areas.
Prescription Use Yes Over-The-Counter Use No AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seymour
(Division Sign-Off) Division of Reproductive, Abdomi 510(k) Number