Search Filters

Search Results

Found 13 results

510(k) Data Aggregation

    K Number
    K160876
    Device Name
    PC ECG
    Date Cleared
    2017-02-15

    (322 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PC ECG

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of SE-12 series electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements is offered to clinicians on an advisory basis only.

    Device Description

    The SE-12/12 Express/1200/1200 Express Electrocardiograph gathers ECG signals of 12 leads simultaneously. It displays the operation menu, ECG parameters as well as electrocardiograms, and be powered by the mains supply or battery. And the system has advanced performance and high reliability due to high resolution thermal recorder, 32-bit processor and a large-capacity memorizer. Design of the system took much consideration on ergonomics so the size is suitable for clinic and hospital uses. There are four selectable modes in the system, including manual, auto, rhythm, R-R analysis or VCG, and VCG is configured with SE-12 Express & SE-1200 Express. SE-12 and SE-1200 share the same single color LCD screen of which the resolution is 320×240 dot; and LCD screen of SE-12 Express and SE-1200 Express is 800×600 multicolor LCD screen. Moreover, SE-12 Express is configured with stress ECG function, which will allow to diagnose concealed coronary heart disease and atypical angina pectoris, prescribe the workload for patients with myocardial infarction before leaving hospital, and assess the effect of the treatment.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for Electrocardiograph Models SE-12, SE-12 Express, SE-1200, and SE-1200 Express by Edan Instruments, Inc. It details the device comparison and performance data, but it does not describe specific acceptance criteria in terms of diagnostic performance metrics (e.g., sensitivity, specificity) for the ECG interpretation algorithm, nor does it present a study proving the device meets such criteria.

    Instead, the document focuses on demonstrating substantial equivalence to predicate devices through:

    • Comparison of intended use, safety specifications, environmental specifications, and performance specifications.
    • Non-clinical testing, including biocompatibility, electrical safety/EMC, bench testing against IEC standards, and software verification/validation.

    Here's an analysis of the requested information based on the provided text, highlighting where information is not available:


    1. Table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative diagnostic acceptance criteria (e.g., minimum sensitivity or specificity for detecting heart conditions) for its ECG interpretation capabilities. The closest information is comparisons of general performance specifications with predicate devices and compliance with safety and electrical standards.

    Acceptance Criteria CategorySpecific Criteria (from document)Reported Device Performance (from document)
    Safety Standards ComplianceIEC 60601-1:2005, EN 60601-1:2006, EN 60601-1-2:2007, IEC 60601-1-2:2007, IEC/EN60601-2-25:2011"The system complies with the IEC 60601-1:2005/A1: 2012 standard for safety and the IEC 60601-1-2: 2007 standard for EMC." "Bench testing was conducted per IEC 60601-2-25: 2011, and all the results show pass."
    BiocompatibilityISO 10993-1 (Cytotoxicity, Skin Sensitization, Skin Irritation)"The biocompatibility evaluation... is conducted in accordance with the International Standard ISO 10993-1... The battery of testing included... Cytotoxicity, Skin Sensitization, Skin Irritation." (Implied pass, as it supports substantial equivalence).
    Electrical Safety & EMCIEC 60601-1:2005/A1: 2012 standard for safety and IEC 60601-1-2: 2007 standard for EMC. Also CISPR 11 Group 1, Class A for EMC."Electrical safety and EMC testing were conducted on the SE-12 series electrocardiograph... The system complies with the IEC 60601-1:2005/A1: 2012 standard for safety and the IEC 60601-1-2: 2007 standard for EMC." "EMC: CISPR 11 Group 1, Class A" (Stated as a "Same" characteristic as predicate, implying compliance).
    Software QualityCompliance with "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices""Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance... The software for this device was considered as a 'moderate' level of concern..." (Implied pass, as it supports substantial equivalence).
    Battery PerformanceUL 2054, UL60950-1:2007 and IEC 61233:2012 standards"Testing on battery are conducted per UL 2054, UL60950-1:2007 and IEC 61233:2012, and results show pass for all testing items." (Regarding battery capacity, the proposed device has increased capacity compared to the predicate, but this is a change, not a specific "acceptance criterion" met).
    ECG Interpretation (Advisory Basis)Equivalent performance to predicate devices (K102830, K131503, K060167, K152427) using "Glasgow" or "SEMIP" algorithms.The document states, "The subject devices share the same characteristics in all items with the predicate device, concluding from using the same technology and principle." For ECG interpretation, the models use the "Glasgow" or "SEMIP" algorithms, which are stated to be the "Same" as their respective predicates. The interpreted ECG results are explicitly "offered to clinicians on an advisory basis only," indicating that the device's interpretation is not a definitive diagnosis and clinical data for diagnostic performance was "Not applicable."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not specified for any diagnostic performance evaluation. The document states "Clinical data: Not applicable," which implies no specific clinical test set for evaluating diagnostic accuracy was used or required for this 510(k) submission.
    • Data Provenance: Not applicable, as no clinical test set data is described. The device is manufactured in China (Shenzhen, P.R. China).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. No clinical test set data or expert-established ground truth is described. The device's integrated interpretation uses specific algorithms (Glasgow, SEMIP) which are compared to those in predicate devices. The interpreted ECG with measurements and interpretive statements is "offered to clinicians on an advisory basis only," suggesting that the device's output is not intended as a definitive ground truth in itself.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. No clinical test set and thus no adjudication method for ground truth establishment is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study is mentioned. The submission states, "Clinical data: Not applicable."

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • While the device includes an ECG interpretation algorithm (Glasgow/SEMIP), the document does not present a standalone performance study in terms of diagnostic accuracy (e.g., sensitivity, specificity, accuracy against a clinical gold standard). The focus is on the device's technical specifications and substantial equivalence to previously cleared devices that also include such algorithms, with the understanding that the interpretation is advisory. The "Bench Testing" (IEC 60601-2-25: 2011) primarily assesses the technical performance of the electrocardiograph (e.g., frequency response, noise, gain, etc.), not the diagnostic accuracy of its interpretive algorithm against clinical outcomes or expert diagnoses.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable for diagnostic performance. For the technical and safety performance, the "ground truth" is adherence to international and national standards (e.g., IEC 60601 series, ISO 10993-1, UL standards).

    8. The sample size for the training set

    • Not specified. The document describes the device as incorporating existing algorithms (Glasgow, SEMIP) found in predicate devices. Details about the training set for these algorithms (if they are considered "AI" in the modern sense) are not provided within this 510(k) summary.

    9. How the ground truth for the training set was established

    • Not specified. As noted above, details about the training of the Glasgow or SEMIP algorithms are not part of this submission.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161302
    Device Name
    PC ECG
    Date Cleared
    2016-07-01

    (53 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PC ECG

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of PADECG is to acquire resting ECG signals from adult and pediatric patients through body surface ECG electrodes. It is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by PADECG can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only. It is mainly used in ECG inpatient department of hospitals or healthcare facilities.

    Device Description

    PADECG is iPad-Based ECG work station. PADECG System primarily composed of DX12(iOS) Transmitter and PADECG Analysis Software, the product is designed to collect and analyzes 12-Lead resting ECG. The DX12(iOS) Transmitter contains lead wires and sends the ECG data to PADECG Analysis Software through Bluetooth. The PADECG Analysis Software will be uploaded to the App Store, so that customers can download it to their own iPad and install it. The PADECG Analysis Software which is installed in the iPad can display the ECG data. The PADECG Analysis Software can analysis the ECG data, and provide an advisory diagnostic result.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the Edan Instruments, Inc. PC ECG, model PADECG. The information focuses on demonstrating substantial equivalence to a predicate device (Edan Instruments, Inc., SE-1515, K152427) rather than providing detailed acceptance criteria and a study proving the device meets those criteria for its interpretive statements.

    However, based on the information available and common regulatory practices for ECG devices with interpretive capabilities, we can infer some details regarding the "ECG interpretation feature validation by database testing" mentioned.

    Here's an attempt to answer your questions based on the provided text, and where information is not explicitly stated, common practice for similar devices is used as a reasonable inference:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state a table of acceptance criteria for the interpretive algorithm's performance (e.g., sensitivity, specificity for specific diagnoses). Instead, it mentions that "ECG interpretation features were also validated by database testing." The "Comparison between PADECG and Predicate Device SE-1515" table (pages 4-5) details technical specifications of the device itself and explicitly states:

    ItemPADECGSE-1515 (Predicate)
    AlgorithmAlgorithm of the Smart ECG Measurement and Interpretation Programs (SEMIP), version 1.8Algorithm of the Smart ECG Measurement and Interpretation Programs (SEMIP), version 1.8
    Heart Rate Meter30 BPM ~300 BPM, ±1BPM30 BPM ~300 BPM, ±1BPM
    Measuring the ECGAmplitudes (mV), intervals (ms), and slopes (mV/s) can be ensured on all ECG waveformsAmplitudes (mV), intervals (ms), and slopes (mV/s) can be ensured on all ECG waveforms
    ReanalysisManually change the measurement marks of medians and the interpretation resultManually change the measurement marks of medians and the interpretation result
    Interpretation libraryHas Interpretation library and can editHas Interpretation library and can edit
    Resolution2.52uV/LSB2.52uV/LSB
    Time Constant≥3.2 s≥3.2 s
    Frequency Response0.05 Hz ~ 150 Hz (-3 dB)0.05 Hz ~ 150 Hz (-3 dB)
    Input Impedance≥20MΩ (10Hz)≥20MΩ (10Hz)
    System Noise85% or >90%) for various diagnostic categories (normal, abnormal, specific arrhythmias, ST-T changes, etc.).
    • Agreement with measurements: High correlation (e.g., R-squared > 0.95) and minimal bias in automated measurements of intervals (PR, QRS, QT), amplitudes (P, QRS, T), and axes compared to expert measurements.
    • Clinical Significance: Demonstrating that the advice provided by the algorithm does not lead to significant misdiagnosis or harm when used as an advisory tool.

    Given that the PADECG uses the "Algorithm of the Smart ECG Measurement and Interpretation Programs (SEMIP), version 1.8," which is the same algorithm as its predicate device (SE-1515), the primary evidence for meeting acceptance criteria for interpretation relies on the predicate device's prior validation and the demonstration that the PADECG implements this identical, validated algorithm correctly. The statement "ECG interpretation features were also validated by database testing" for the PADECG suggests that this algorithm's performance was confirmed on a test set, likely to ensure its faithful implementation on the new platform.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified. The document only states "database testing."
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). This is a common omission in 510(k) summaries, which often focus on the type of testing rather than granular details of the dataset.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts: Not specified.
    • Qualifications of Experts: Not specified. For ECG interpretation, ground truth is typically established by board-certified cardiologists or electrophysiologists, often with significant experience in ECG reading.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not specified. For ECG interpretation, consensus or adjudication by multiple expert readers is standard practice to establish ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Comparative Effectiveness Study: No, the document does not mention an MRMC comparative effectiveness study where human readers' performance with and without AI assistance was evaluated. The device's interpretation is clearly stated as "advisory basis only," indicating it's not a standalone diagnostic device but a tool that assists a professional.
    • Effect Size of Improvement: Not applicable, as no such study was reported.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    • Standalone Performance: Yes, implicitly. The "ECG interpretation features were also validated by database testing" refers to the performance of the algorithm itself against a pre-established ground truth. While the device is not standalone in general clinical use (it advises clinicians), the algorithm's performance in interpreting ECGs from the database is a standalone evaluation of its output against ground truth. The device is cleared as an "advisory" tool, meaning its standalone diagnostic accuracy is assessed but not necessarily expected to be perfect or replace human judgment.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Type of Ground Truth: Not explicitly stated but heavily implied to be expert consensus or reference standard ECG interpretations derived from cardiologists. For ECG interpretation, this is the gold standard for validating algorithms. Pathology or outcomes data are generally not the direct ground truth for ECG interpretation algorithms, though they might be used in broader clinical studies of the overall diagnostic pathway.

    8. The sample size for the training set

    • Training Set Sample Size: Not specified. The document focuses on the validation of the interpretation features via "database testing" for the specific device being cleared. Details of the training set used for the original development of the "SEMIP, version 1.8" algorithm are not provided in this 510(k) summary.

    9. How the ground truth for the training set was established

    • Training Set Ground Truth Establishment: Not specified in this document. For the original development of the SEMIP algorithm, it would likely have involved a large dataset of ECGs interpreted by multiple cardiologists, with discrepancies resolved through an adjudication process to establish the most accurate ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K152427
    Device Name
    PC ECG
    Date Cleared
    2016-02-05

    (163 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PC ECG

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SE-1515 PC ECG is intended to acquire, process and store ECG signals from adult and pediatric patients undergoing stress exercise test or resting test. The SE-1515 PC ECG is intended to be used only in hospitals and healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the SE-1515 PC ECG can help users to analyze and diagnose heart diseases. However, the ECG with measurements is offered to clinicians on an advisory basis only. It is mainly used in the ECG Outpatient Department and Physical Examination Department.

    Device Description

    SE-1515 is a PC-based ECG designed to acquire, process and store ECG signals and analyze and diagnose heart disease. SE-1515 consists of analysis software and sampling boxes, and has four configurations: Wired 18-lead ECG system, Wired 16-lead ECG system, Wired 12-lead ECG system and Wireless 12-lead ECG system. Wired 18-lead ECG system consists of analysis software and wired DE18 sampling box. Wired 16-lead ECG system consists of analysis software and wired DE15 sampling box. Wired 12-lead ECG system consists of analysis software and wired DP12 sampling box. Wireless 12-lead ECG system contains analysis software, wireless DX12 sampling boxes (DX12 transmitter and DX12 receiver). The communication between DX12 transmitter and DX12 receiver is based on Bluetooth technology. For stress exercise testing, the user should connect a BP monitor and a treadmill or an ergo meter to the PC.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the EDAN Instruments, Inc. PC ECG, model SE-1515, based on the provided text:

    Important Note: The provided document is a 510(k) summary, which focuses on demonstrating
    substantial equivalence to predicate devices, not necessarily a detailed clinical
    effectiveness study itself. Therefore, specific details about acceptance criteria, ground truth
    establishment (especially for a new algorithm), and effect sizes from MRMC studies are often
    not explicitly stated in this type of document. The information below is extracted directly
    from the text; if a point is not mentioned, it's because the document does not provide it.


    Acceptance Criteria and Reported Device Performance

    The document does not explicitly list "acceptance criteria" but rather compares the performance parameters of the SE-1515 with its predicate devices (SE-1010, Philips DXL 12/16-lead ECG Algorithm, Philips TC70). The implicit acceptance criteria are that the new device's performance should be similar or comparable to the predicate devices, or that any differences do not raise new questions of safety or effectiveness.

    Below is a table summarizing the performance comparison, highlighting where quantitative parameters are presented. Since "Remark" column indicates "Same," "Different," or "---", for the purpose of showing "acceptance criteria" and "reported device performance", the "predicate value" can be considered as the implicit acceptance criteria that the new device needs to meet to establish substantial equivalence.

    Item (Device Component)Acceptance Criteria (Predicate Value, where applicable)Reported Device Performance (SE-1515)
    DE15 Sampling Box
    DFT Filterweak/strong0.01 Hz/0.05 Hz/0.32 Hz/0.67 Hz
    Low Pass/EMG Filter25 Hz/35 Hz/45 Hz/75 Hz/100 Hz/150 Hz300 Hz/270 Hz/150 Hz/100 Hz/75 Hz (Low Pass Filter); 25 Hz/35 Hz/45 Hz/Off (EMG Filter)
    Lead number12-lead16-lead
    Sampling Rate1 kHz /channel16 kHz /channel
    Resolution2.52 µV/LSB @ 1 kHz0.1575 µV/LSB @ 1 kHz
    Frequency Response0.05 Hz ~ 150 Hz0.01 Hz~300 Hz
    Input Impedance>50 MΩ≥100 MΩ
    Common Mode Rejection>110 dB≥140 dB(AC ON) / ≥123 dB(AC OFF)
    Pace Detect±2 to ±700 mV @ 0.1 ms to 2.0 ms±750 uV~±700 mV @ 50 us ~2.0 ms
    DE18 Sampling Box
    DFT Filterweak/strong0.01 Hz/0.05 Hz/0.32 Hz/0.67 Hz
    Low Pass/EMG Filter25 Hz/35 Hz/45 Hz/75 Hz/100 Hz/150 Hz300 Hz/270 Hz/150 Hz/100 Hz/75 Hz (Low Pass Filter); 25 Hz/35 Hz/45 Hz/Off (EMG Filter)
    Lead number12-lead18-lead
    Sampling Rate1 kHz /channel16 kHz /channel
    Resolution2.52 µV/LSB @ 1 kHz0.1575 µV/LSB @ 1 kHz
    Frequency Response0.05 Hz ~ 150 Hz0.01 Hz~300 Hz
    Input Impedance>50 MΩ≥100 MΩ
    Common Mode Rejection>110 dB≥140 dB (AC ON) / ≥123 dB (AC OFF)
    Pace Detect±2 to ±700 mV @ 0.1 ms to 2.0 ms±750 uV~±700 mV @ 50 us ~2.0 ms
    DP12 Sampling Box
    Sampling Rate1,000 /sec/channel1,000 /sec/channel
    Resolution2.52 µV/LSB @ 1 kHz2.52 µV/LSB @ 1 kHz
    Time Constant≥3.2 s≥3.2 s
    Frequency Response0.05 Hz ~ 150 Hz (-3 dB)0.05 Hz ~ 150 Hz (-3 dB)
    Input Impedance>50 MΩ @ 10 Hz, defibrillator protected>50 MΩ @ 10 Hz, defibrillator protected
    Dynamic RangeAC Differential ± 5 mV, DC offset ±600 mVAC Differential ± 5 mV, DC offset ±600 mV
    System Noise≤12.5 µVp-p≤12.5 µVp-p
    Common Mode Rejection>110 dB>110 dB
    Pace Detect±2 to ±700 mV@0.1 to 2.0 ms±2 to ±700 mV@0.1 to 2.0 ms
    ESD Sensitivity±6 kV contact / ±8 kV air±6 kV contact / ±8 kV air
    Heart Rate Meter30 BPM ~300 BPM, ±1 BPM30 BPM ~300 BPM, ±1 BPM
    DX12 Sampling Box
    All listed parameters (Sampling Rate, Resolution, Time Constant, Frequency Response, Input Impedance, Dynamic Range, System Noise, Common Mode Rejection, Pace Detect, ESD Sensitivity, Heart Rate Meter, Transmitting Frequency, Frequency Band, Modulation Type, Transmitting Power)Matches predicateMatches predicate
    Analysis Software
    Operating systemWindows XP SP2/SP3, Vista, Windows 7Windows XP SP2/SP3 (32/64 bit), Windows 7 SP1 (32/64 bit) or Windows 8 (32/64 bit)

    Study Information:

    1. Sample size used for the test set and the data provenance:

      • The document mentions "database testing" for ECG interpretation features but does not specify the sample size for this test set or the provenance (country of origin, retrospective/prospective nature) of the data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document. The document refers to "ECG interpretation features... validated by database testing," but details on the ground truth establishment for this validation are absent.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This information is not provided in the document.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • An MRMC comparative effectiveness study, particularly one measuring human reader improvement with AI assistance, was not indicated. The comparison is mainly against predicate devices, focusing on technical specifications and substantial equivalence, not algorithm-assisted human reading.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • The document states, "ECG interpretation features were also validated by database testing." This implies a standalone performance evaluation of the algorithm against a database, though details are missing. The device's indications for use also state that the "ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only," suggesting a standalone interpretive function.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the "ECG interpretation features," the document mentions "database testing." However, the specific type of ground truth used within this database (e.g., expert consensus diagnoses, correlation with patient outcomes, or other clinical data) is not specified.
    7. The sample size for the training set:

      • The document does not specify a training set and its size. The validation focuses on the device's performance and ECG interpretation features using "database testing."
    8. How the ground truth for the training set was established:

      • Since no training set is explicitly mentioned or described, the method for establishing its ground truth is also not provided.
    Ask a Question

    Ask a specific question about this device

    K Number
    K131819
    Device Name
    PC ECG
    Date Cleared
    2013-11-22

    (155 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PC ECG

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SE-1010 PC ECG is a PC-based diagnostic tool intended to acquire, process and store ECG Signal from adult and pediatric patients undergoing stress exercise test or resting test. SE-1010 PC ECG is intended to be used only in hospitals and healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by SE-1010 PC ECG can help users to analyze and diagnose heart disease. However the ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.

    Device Description

    PC ECG is a kind of resting and exercising electrocardiograph based on PC. PC ECG including data sampling modules and software which can be installed on desktop or laptop PC. PC ECG can acquire 12 channel waveforms simultaneously, which can also print out 3/6/12 channel electrocardiograph wave simultaneously by a 210mm (A4) wide printer. Advanced digital filter technique has been used in PC ECG including baseline anti-drift filter AC noise (50/60Hz) filter; EMG Filter and Low pass Filter which can help the user to record the ECG more clearly. Moreover. the ECG Measurement and interpretation program(SEMIP) is included in this machine, this program can provide more detail information to diagnose heart disease.

    AI/ML Overview

    The provided text
    {0} to {4} gives very limited information about the acceptance criteria and study proving device performance as it's a 510(k) summary, not a full study report. The available details are almost exclusively for the device itself and its general features, not for specific algorithm performance.

    Here's an analysis based on the information provided, highlighting what can be answered and what cannot:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Not explicitly stated in the document. The document mentions "Software testing," "Risk analysis," "Safety testing," and "Environment test" were applied. However, no specific performance metrics (e.g., accuracy, sensitivity, specificity for interpretation) are provided as acceptance criteria or results.
    The device "can acquire 12 channel waveforms simultaneously.""3/6/12-channel ECG wave display and printing simultaneously" is a feature.
    The device "can also print out 3/6/12 channel electrocardiograph wave simultaneously by a 210mm (A4) wide printer.""Report printing with PDF format, word format or JPG format" is a feature.
    Advanced digital filter technique has been used, including baseline anti-drift filter AC noise (50/60Hz) filter, EMG Filter and Low pass Filter.These filters are listed as being included.
    The ECG Measurement and interpretation program (SEMIP) is included.This program "can provide more detail information to diagnose heart disease." "Supporting auto measurement and diagnosing" is a feature.
    The device is intended to acquire, process and store ECG signals from adult and pediatric patients.No specific performance metrics for this are given.
    The device recorded cardiogram can help users to analyze and diagnose heart disease.The interpretive statements are "offered to clinicians on an advisory basis only."

    Study Details

    1. Sample size used for the test set and the data provenance: Not available. The document only states "Verification and validation testing was done on PC ECG" but provides no details on the test set, its size, or its origin.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not available. There is no information regarding ground truth establishment for a test set.

    3. Adjudication method for the test set: Not available. No details about a test set or its adjudication are provided.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not available. The submission does not mention an MRMC study or any comparison of human readers with vs. without AI assistance. The interpretation program "is offered to clinicians on an advisory basis only," suggesting it's a tool, not a replacement or direct comparison to human reading in this context.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not explicitly stated/detailed. While the device includes an "ECG Measurement and interpretation program (SEMIP)" that supports "auto measurement and diagnosing," no standalone performance metrics for this algorithm are provided (e.g., diagnostic accuracy, sensitivity, specificity for specific conditions). The statement that these interpretations are "advisory" further suggests that rigorous standalone performance was not the primary focus of the 510(k) summary for interpretative accuracy.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): Not available. Information on how ground truth was established for any performance evaluation is not provided.

    7. The sample size for the training set: Not available. There is no information about a training set.

    8. How the ground truth for the training set was established: Not available. There is no information about a training set or its ground truth establishment.

    Summary of Limitations:

    This 510(k) summary focuses on the safety and effectiveness (substantial equivalence) of the PC ECG device, rather than detailed performance metrics of its interpretive algorithms. It largely describes the device's features and general testing (software, risk, safety, environment). For an interpretive ECG device, a full study report would typically include detailed performance characteristics like sensitivity, specificity, and accuracy for various ECG findings, often compared against a gold standard established by expert cardiologists. This information is absent from the provided document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102854
    Device Name
    PC ECG
    Date Cleared
    2010-12-03

    (65 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PC ECG

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SE-1010 PC ECG is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. SE-1010 PC ECG is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by SE-1010 PC ECG can help users to analyze and diagnose heart disease. However the ECG with measurements and interpretive statements is offered to clinician on an advisory basis only.
    SE-1010 PC ECG is a PC-based diagnostic tool intended to acquire, process and store ECG signals from adult and pediatric patients undergoing stress exercise test or resting test. SE-1010 PC ECG is intended to be used only in hospitals and healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by SE-1010 PC ECG can help users to analyze and diagnose heart disease. However the ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.

    Device Description

    PC ECG is a kind of resting and exercising electrocardiograph based on PC. PC ECG including data sampling modules and software which can be installed on desktop or laptop PC. PC ECG can acquire 12 channel waveforms simultaneously, which can also print out 3/6/12 channel electrocardiograph wave simultaneously by a 210mm (A4) wide printer. Advanced digital filter technique has been used in PC ECG including baseline anti-drift filter AC noise (50/60Hz) filter EMG Filter and Low pass Filter, which can help the user to record the ECG more clearly. Measurement Moreover, the ECG and interpretation program(SEMIP) is included in this machine, this program can provide more detail information to diagnose heart disease. PC ECG has the features as follows: 3/6/12-channel ECG wave display and printing simultaneously ECG wave frozen and review Measurement point adjustment and re-analyzing, manual measurement with high precision electronic ruler data management and processing function Report printing with PDF format, word format or JPG format Multi-language supporting Supporting auto measurement and diagnosing Automatic baseline adjustment for optimal printing

    AI/ML Overview

    The provided text is a 510(k) summary for the Edan Instruments, Inc. PC ECG device. While it mentions “Software testing”, “Risk analysis”, “Safety testing”, and “Environment test” as quality assurance measures, it does not provide detailed acceptance criteria or a study that specifically proves the device meets such criteria in terms of diagnostic performance (e.g., sensitivity, specificity, accuracy for disease detection).

    The document is primarily a regulatory submission summarizing the device, its intended use, and its substantial equivalence to predicate devices (Cardiosoft/case cardiac testing system, K031561; and PC ECG, MODEL SE-1010, K092010).

    Therefore, based only on the provided text, I cannot complete most of the requested information regarding acceptance criteria and performance studies.

    Here's what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    • Cannot be provided. The document states "Verification and validation testing was done on PC ECG" but does not define specific acceptance criteria (e.g., minimum sensitivity/specificity for particular conditions) or report detailed performance metrics against such criteria. The mention of "auto measurement and diagnosing" implies some diagnostic capability, but no performance figures are given.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Cannot be provided. The document does not mention any specific test set, its size, or provenance for evaluating diagnostic performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Cannot be provided. No information about a test set or ground truth establishment is present.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Cannot be provided. No information about a test set or adjudication is present.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Cannot be provided. The document does not describe any MRMC study or an "AI assistance" component in the context of improving human reader performance. The device includes an "interpretation program (SEMIP)" for diagnosis, but its impact on human readers is not studied.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Cannot be provided definitively for diagnostic performance. The document states "Supporting auto measurement and diagnosing" and "can provide more detail information to diagnose heart disease" regarding the SEMIP. It also states "However the ECG with measurements and interpretive statements is offered to clinician on an advisory basis only." This suggests a standalone interpretive function, but no performance metrics (e.g., accuracy against a gold standard) for this standalone function are provided in the text.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Cannot be provided. No information about ground truth for performance evaluation is present.

    8. The sample size for the training set

    • Cannot be provided. There is no mention of a training set or its size.

    9. How the ground truth for the training set was established

    • Cannot be provided. There is no mention of a training set or its ground truth establishment.

    Summary of Available Information from the Provided Text:

    The document describes the PC ECG device as a PC-based electrocardiograph that acquires and processes 12-channel ECG waveforms. It includes an "ECG and interpretation program (SEMIP)" which can "provide more detail information to diagnose heart disease" and offers "auto measurement and diagnosing". However, the interpretive statements are "offered to clinician on an advisory basis only."

    The regulatory submission focuses on demonstrating substantial equivalence to predicate devices (Cardiosoft/case cardiac testing system K031561 and PC ECG, MODEL SE-1010 K092010), rather than presenting a detailed clinical performance study for its interpretive function. The "Test Summary" only lists general quality assurance measures like "Software testing", "Risk analysis", "Safety testing", and "Environment test," without specifying performance-based acceptance criteria or results.

    Ask a Question

    Ask a specific question about this device

    K Number
    K092010
    Date Cleared
    2009-10-22

    (108 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PC ECG, MODEL SE-1010

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SE-1010 PC ECG is intended to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. SE-1010 PC ECG is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by SE-1010 PC ECG can help users to analyze and diagnose heart disease. However the ECG with measurements and interpretive statements is offered to clinician on an advisory basis only.

    Device Description

    PC ECG including model SE-1010 has the similar functions of an ordinary electrocardiograph. ECG data can be sampled, displayed and stored in PC machine, and they can be printed with several kinds of printing types, including PDF format, word format and JPG format. ECG wave can be frozen and reviewed. Auto measurement and diagnosing is available, and Diagnose Template can be edited. PC ECG has the features as follows: - 3/6/12-channel ECG wave display and printing simultaneously - ECG wave frozen and review - Measurement point adjustment and re-analyzing, manual measurement with high precision electronic ruler - data management and processing function - Report printing with PDF format, word format or JPG format - -Multi-language supporting - Supporting auto measurement and diagnosing -- -Automatic baseline adjustment for optimal printing

    AI/ML Overview

    The provided text is a 510(k) summary for the Edan Instruments PC ECG (model SE-1010). However, it does not contain a detailed study proving the device meets specific acceptance criteria for performance, especially regarding its "auto measurement and diagnosing" capabilities.

    The document primarily focuses on:

    • Substantial equivalence to a predicate device (Cardiosoft/case cardiac testing system, K031561).
    • General device description and intended use.
    • Mention of general quality assurance measures (software testing, risk analysis, safety testing, environment test, verification and validation testing) to demonstrate substantial equivalence.

    Therefore, many of the requested details cannot be extracted from the provided text.

    Here's an attempt to answer the questions based only on the provided information, noting where information is absent:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Specific performance criteria (e.g., sensitivity, specificity for diagnostic functions)Not provided in the document. The document states "Auto measurement and diagnosing is available" but offers no performance metrics.
    Ability to acquire ECG signals from adult and pediatric patients.Stated in "Intended Use" and "Device Description". No specific performance metrics (e.g., signal quality, artifact rejection) are given.
    Display and storage of ECG data in PC machine.Stated in "Device Description". No specific performance metrics are given.
    Printing with several kinds of printing types (PDF, Word, JPG).Stated in "Device Description". No specific performance metrics are given.
    ECG wave frozen and reviewed.Stated in "Device Description". No specific performance metrics are given.
    Auto measurement and diagnosing.Stated in "Device Description". No performance data is provided to demonstrate accuracy or reliability of these functions. The document explicitly states "the ECG with measurements and interpretive statements is offered to clinician on an advisory basis only."
    Measurement point adjustment and re-analyzing, manual measurement.Stated in "Device Description". No specific performance metrics are given.
    Data management and processing.Stated in "Device Description". No specific performance metrics are given.
    Multi-language support.Stated in "Device Description". No specific performance metrics are given.
    Automatic baseline adjustment.Stated in "Device Description". No specific performance metrics are given.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified for any specific performance study related to the "auto measurement and diagnosing" feature. The document only mentions "Verification and validation testing was done on PC ECG." without detailing the scope or data used.
    • Data Provenance: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not specified.

    4. Adjudication method for the test set

    • Not specified.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study is mentioned or detailed in the document. The device's diagnostic interpretation is stated to be "advisory basis only," which suggests a standalone or assistive role rather than a replacement for human interpretation based on a proven MRMC study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • A standalone performance study for the interpretive algorithm is not detailed in the document. While "Auto measurement and diagnosing is available," no performance metrics (e.g., sensitivity, specificity, accuracy) are provided for this function.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not specified for any analytical or diagnostic performance studies.

    8. The sample size for the training set

    • Not applicable/Not specified. There is no mention of a machine learning or AI model being "trained" in this document. The "auto measurement and diagnosing" feature likely refers to rule-based algorithms or established clinical criteria, rather than a modern AI/ML model requiring a distinct "training set."

    9. How the ground truth for the training set was established

    • Not applicable/Not specified, as no training set is mentioned.

    Summary of what is present:
    The 510(k) summary primarily asserts substantial equivalence to a predicate device based on similar functions and general quality assurance processes. It lists capabilities of the device but does not provide any specific quantitative performance data or details of studies validating the accuracy of its diagnostic or measurement features. The diagnostic statements are explicitly advisory.

    Ask a Question

    Ask a specific question about this device

    K Number
    K080141
    Manufacturer
    Date Cleared
    2008-04-25

    (94 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PC ECG 1200W SYSTEM

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ECG is intended to disclose either normal condition or patterns of arrhythmia, myocardial ischemia, rate abnormalities, or features of prognostic value in patients:
    l ) suspected of cardiac abnormalities, or
    2) in populations of patients at an age or period in which a routine baseline evaluation of ECG characteristics are desired.
    3) OT Analysis is useful in the assessment of long QT syndrome (LQTS). In some instances, LQTS can be corrected by pharmacologic therary. QT analysis is also used to measure QT dispersion, the difference between maximal and minimal QT values. QT dispersion is a measure of the inhomogeneity of ventricular repolarization. 4) The PC ECG 1200W System has been tested to measure Heart Rate Variability within 1 millisecond tolcrance. The clinical significance of Heart Rate Variability measures should be determined by a physician.
    5) The PC ECG 1200W System has been tested to measure Latc Potential within 1 millisecond tolerance in the time domain, and I microvolt tolerance in voltage. The clinical significance of Heart Rate Variability measures should be determined by a physician.

    Stress testing intended use:
    Angina pectoris (chest pain) is a clinical syndrome resulting from myocardial ischemia, indicative of a reduced blood supply to the cardiac muscle. The electrocardiogram may establish the diagnosis of ischemic heart disease if characteristic changes are present. Stress testing is the most widely used method to decide whether this chest pain is related to myocardial ischemia, and thereby coronary artery disease. In stress testing, the contractile capability of the heart muscle is monitored via ECG during patient exercise. Patients are exercised by bicycle, treadmill, or other means while continuously monitoring the ECG. Exercise loads are determined by predefined protocols. The ECG signals are recorded for the resting, exercise and recovery portions of the exercise protocol. The changes in ECG waveforms are compared to the resting ECG records. Although not necessary, most of the commercial stress test systems control the bicycle or treadmill automatically according to the requirements of the chosen protocol.
    ST segment monitoring is intended as an aid in the evaluation of myocardial ischemia in patients with known or suspected coronary artery disease. The ST segment algorithm has been tested for accuracy of the ST segment data, and a database has been used as a tool for performance testing. The significance of the ST segment changes must be determined by a physician.

    Device Description

    The PC ECG 1200W System is designed to disclose either normal condition or patterns of arrhythmia, myocardial ischemia, rate abnormalities, or features of prognostic value. The System comprises:

    • · PCECG 1200W data acquisition unit
    • · PCECG 1200WR RF transceiver
    • · USB communications cable
    • · Software application
    • · Software security lock (dongle) for access to stress testing functions (optional)
    AI/ML Overview

    The NORAV PC ECG 1200W System is substantially equivalent to the predicate device, Norav's PC ECG 1200 K000404, with the primary difference being the addition of wireless data communication capabilities. The submission primarily focuses on the device's technical specifications and intended use, asserting its safety and reliability based on similarity to the predicate and adherence to recognized standards.

    Here's an analysis of the provided information regarding acceptance criteria and performance studies:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document describes the device's intended functionality and accuracy in certain measurements, but it does not explicitly state specific pass/fail acceptance criteria in a quantitative manner for most functions. Instead, it makes claims of the device's capabilities and how it has been tested.

    Acceptance Criteria (from text)Reported Device Performance (from text)
    ECG intended to disclose normal condition or patterns of arrhythmia, myocardial ischemia, rate abnormalities, or features of prognostic value."ECG is intended to disclose either normal condition or patterns of arrhythmia, myocardial ischemia, rate abnormalities, or features of prognostic value..." (Indication of intended capability rather than a performance metric against a specific threshold)
    Heart Rate Variability analysis accuracy"The PC ECG 1200W System has been tested to measure Heart Rate Variability within 1 millisecond tolerance."
    Late Potential analysis accuracy (time domain)"The PC ECG 1200W System has been tested to measure Late Potential within 1 millisecond tolerance in the time domain..."
    Late Potential analysis accuracy (voltage)"...and 1 microvolt tolerance in voltage."
    ST segment algorithm accuracy for myocardial ischemia evaluation"The ST segment algorithm has been tested for accuracy of the ST segment data, and a database has been used as a tool for performance testing." (No specific numerical accuracy or tolerance is given, only that it has been tested for accuracy.)
    Safety and reliability"PC ECG 1200W System constitutes a safe and reliable medical device. Similarly to the predicate devices, the System operation presents no adverse health effect or safety risks to patients when used as intended." (Qualitative statement based on adherence to standards and similarity to predicate, rather than specific performance metrics from a study.)
    Compliance with wireless communication standards (Wireless capability is the primary difference from predicate)The device uses wireless data communication. The document references several EMC and Radio spectrum standards (e.g., EN301 489-1, EN301 489-3, EN300 440) which would imply compliance with these as acceptance criteria for its wireless functionality, but no specific performance results related to wireless communication (e.g., range, data integrity, latency) are provided, only that the capability exists and the standards are relevant.
    Functional equivalence to predicate device (K000404)"PC ECG 1200W System is identical to the predicate device in all features, functions and specifications except its capability of wireless data communication." and "The clearance for the PC ECG 1200W System is sought on the grounds of its claimed substantial equivalence (SE) to Norav's PC ECG 1200 K000404 for complete physical and functional identity except the capability of wireless data communication." (This is a foundational claim for the 510(k) process, implying verification through comparison and testing, though specific comparative test results or criteria are not detailed in this summary.)

    2. Sample Size Used for the Test Set and Data Provenance:

    • Test Set Sample Size: The document mentions that for the ST segment algorithm, "a database has been used as a tool for performance testing." However, no specific sample size for this database or any other test set is provided.
    • Data Provenance: Not specified. It does not mention the country of origin, nor whether the data was retrospective or prospective.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • This information is not provided in the document. The document states that "The significance of the ST segment changes must be determined by a physician," but it does not detail how ground truth was established for the performance testing database.

    4. Adjudication Method for the Test Set:

    • This information is not provided in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • A MRMC comparative effectiveness study was not mentioned or implied in the provided summary. The submission focuses on device equivalence and performance claims rather than comparing human reader performance with and without AI assistance.

    6. Standalone (Algorithm Only) Performance:

    • The claims regarding Heart Rate Variability ("within 1 millisecond tolerance") and Late Potential analysis ("within 1 millisecond tolerance in the time domain, and 1 microvolt tolerance in voltage") appear to relate to the standalone performance of the device's algorithms or measurement capabilities.
    • Similarly, the statement that "The ST segment algorithm has been tested for accuracy of the ST segment data, and a database has been used as a tool for performance testing" indicates a standalone algorithmic test.
    • However, specific metrics or a detailed study of standalone performance results (e.g., sensitivity, specificity, AUC) are not explicitly provided beyond the tolerance statements for HRV and Late Potential.

    7. Type of Ground Truth Used:

    • For Heart Rate Variability and Late Potential analysis, the "tolerance" statements imply the device's measurements were compared against a reference standard for accuracy, which would constitute the ground truth. However, the nature of this reference standard (e.g., a gold standard measurement, expert-derived ground truth) is not specified.
    • For the ST segment algorithm, it states a "database has been used as a tool for performance testing." The ground truth for this database to assess "accuracy" of ST segment data is not explicitly stated (e.g., expert consensus, pathology, follow-up outcomes). The "significance of the ST segment changes must be determined by a physician" suggests clinical interpretation is ultimately required.

    8. Sample Size for the Training Set:

    • This information is not provided in the document. The document refers to "testing" with a database but does not distinguish between training and test sets.

    9. How the Ground Truth for the Training Set Was Established:

    • As no information about a training set is provided, the method for establishing its ground truth is also not mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K052295
    Manufacturer
    Date Cleared
    2006-01-13

    (143 days)

    Product Code
    Regulation Number
    870.2340
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    CARDIOSOFT PC ECG, USB VERSION

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The CardioSoft PC ECG is intended for use in acquiring, processing, recording, archiving and displaying electrocardiographic information. It can provide serial comparison of ECG information to facilitate review of current and previous records and is intended for use by trained operators. The system offers no diagnostic opinion to the user, but may be used by a physician to render his or her own medical opinion.

    Device Description

    The CardioSoft PC ECG is an electrocardiograph that detects signals associated with cardiac activity and produces an ECG graphical record. The device operates on a personal computer through a USB hardware module. The device can acquire, process, record, archive and display 3-, 6-, or 12-lead ECG and vector cardiograms. The device provides portable ECG with important advantages compared to traditional machines.

    AI/ML Overview

    The presented document is a 510(k) summary for the CardioSoft PC ECG device, a substantially equivalent submission to the predicate device CARDIAX (K924544). The document does not describe specific acceptance criteria or a dedicated study to prove the device meets performance criteria through clinical data or comparative effectiveness. Instead, it relies on demonstrating substantial equivalence to a predicate device and compliance with recognized standards.

    Here's a breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document explicitly states: "The FDA has not established performance standards for this device." Therefore, there are no specific FDA-mandated acceptance criteria or direct reported device performance metrics (e.g., sensitivity, specificity, accuracy) derived from a clinical study for this 510(k) submission.

    The "performance" section instead focuses on demonstrating equivalence through non-clinical testing and adherence to quality assurance measures.

    Acceptance CriteriaReported Device Performance
    None explicitly defined by FDA in this document.Non-clinical testing demonstrated "equivalence with the predicate device CARDIAX and the results show similar performance with this device."
    Compliance with established voluntary/consensus standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-25, IEC 60601-2-27, ECS53, EC38, UL 2601-1, ANSI/UL 2601-1 ad CAN/CSA C22.2)Device "complies with the following recognized voluntary and consensus standards".
    Conformance to product development procedures and plans"Assured by the application of system tests, design reviews, and product verification and validation testing performed prior to product release."
    Quality assurance measures applied during developmentRequirements specification reviews, Code inspections, Software and hardware evaluation and testing, Safety testing, Environment testing, Final validation.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a clinical "test set" with a specific sample size. The performance assessment relies on non-clinical testing and comparison to a predicate device. Therefore, no information on data provenance (country of origin, retrospective/prospective) for a clinical test set is provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. As no clinical test set or ground truth establishment process for diagnostic performance is described, there's no mention of experts or their qualifications in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no clinical test set described in the submission, therefore no adjudication method is mentioned.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The document makes no mention of a multi-reader multi-case (MRMC) comparative effectiveness study. The device, an electrocardiograph, "offers no diagnostic opinion to the user" and there is no AI component mentioned that would assist human readers in a diagnostic setting.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    The device is an "electrocardiograph that detects signals associated with cardiac activity and produces an ECG graphical record." It is not an algorithm providing diagnostic output without human interpretation. It generates a record "that may be utilized only by a qualified healthcare professional for review and interpretation when based upon patient history, medical examination, and other clinical findings." Therefore, a standalone algorithm-only performance study as typically understood for AI-driven diagnostic tools was not conducted or described. The "performance" refers to the device's ability to accurately acquire, process, record, archive, and display ECG information.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Given the nature of the device (ECG acquisition and display) and the absence of a clinical diagnostic performance study, the concept of "ground truth" as typically applied to diagnostic algorithms (e.g., expert consensus for disease presence or absence, pathology results) is not relevant or mentioned in this submission. The "ground truth" for demonstrating performance would likely involve the accuracy of signal acquisition, processing, and display compared to known electrical signals or established ECG benchmarks.

    8. The sample size for the training set

    Not applicable. The CardioSoft PC ECG is described as an upgrade/modification of an existing device, preserving the "software processes of the predicate device." It's not a machine learning or AI-driven device that typically requires a 'training set' in the conventional sense for a 510(k). The focus is on hardware and interface improvements while maintaining functional equivalence.

    9. How the ground truth for the training set was established

    Not applicable, as there is no mention of a training set for a machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K000404
    Device Name
    PC ECG 1200
    Manufacturer
    Date Cleared
    2000-11-22

    (289 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    PC ECG 1200

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K992584
    Manufacturer
    Date Cleared
    2000-01-13

    (164 days)

    Product Code
    Regulation Number
    870.2800
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PE 501 PC-ECG MONITOR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PE 501 PC-ECG is indicated for acquiring and displaying 2 channel ECG signals through a personal computer and transferring the recorded ECG signals to a remote computer. It is designed for stationary use and is not for ambulatory use. It is intended for use primarily in out-patient clinics. The PE 501 PC-ECG device should be operated by a licensed healthcare practitioner.

    Device Description

    The PE 501 PC-ECG Monitor is a battery-powered, personal computer based electrocardiograph monitoring device. It is designed to acquire, display and record 2 channel ECG signals. The ECG signals are sent to the computer through a serial port and displayed on the computer screen. The ECG signals associated with personal information can be stored into the computer hard disk or floppy disk. Desired segments of the recorded ECG signals can be printed on any Windows compatible printers. The recorded ECG signals can be easily retrieved and can also be transferred to a remote computer with floppy disk, direct machine to machine or E-mail.

    AI/ML Overview

    The provided text for K992584 for the PE 501 PC-ECG Monitor does not contain information about acceptance criteria or a study proving its performance against such criteria. The document is primarily a 510(k) summary and an FDA clearance letter, which focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed performance study results against specific acceptance criteria.

    Therefore, I cannot fulfill your request for the detailed table and study information as it is not present in the provided document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2