(322 days)
Not Found
No
The summary describes standard electrocardiograph functionality and hardware components. There is no mention of AI, ML, or related concepts in the device description, intended use, or performance studies. The "interpreted ECG with measurements is offered to clinicians on an advisory basis only" suggests a rule-based or algorithmic interpretation, not necessarily AI/ML.
No.
This device is an electrocardiograph used for diagnosis, not for therapy or treatment.
Yes
The device, an electrocardiograph, acquires ECG signals which "can help users to analyze and diagnose heart disease." Additionally, one model includes a stress ECG function "which will allow to diagnose concealed coronary heart disease and atypical angina pectoris." These statements indicate its use in diagnosing medical conditions.
No
The device description explicitly mentions hardware components such as a thermal recorder, 32-bit processor, large-capacity memorizer, LCD screen, and battery, indicating it is a physical device with integrated software, not a software-only medical device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. These tests are performed outside the body (in vitro).
- Device Function: The SE-12 series electrocardiograph acquires ECG signals from the body surface using electrodes. This is a non-invasive procedure performed on the patient's body, not on a sample taken from the body.
- Intended Use: The intended use is to acquire ECG signals and help users analyze and diagnose heart disease based on these signals. This is a diagnostic process based on physiological measurements from the body, not on laboratory testing of biological samples.
Therefore, the SE-12 series electrocardiograph falls under the category of a medical device used for physiological monitoring and diagnosis, but it does not meet the definition of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The intended use of SE-12 series electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements is offered to clinicians on an advisory basis only.
Product codes (comma separated list FDA assigned to the subject device)
DPS
Device Description
The SE-12/12 Express/1200/1200 Express Electrocardiograph gathers ECG signals of 12 leads simultaneously. It displays the operation menu, ECG parameters as well as electrocardiograms, and be powered by the mains supply or battery. And the system has advanced performance and high reliability due to high resolution thermal recorder, 32-bit processor and a large-capacity memorizer. Design of the system took much consideration on ergonomics so the size is suitable for clinic and hospital uses. There are four selectable modes in the system, including manual, auto, rhythm, R-R analysis or VCG, and VCG is configured with SE-12 Express & SE-1200 Express. SE-12 and SE-1200 share the same single color LCD screen of which the resolution is 320x240 dot; and LCD screen of SE-12 Express and SE-1200 Express is 800x600 multicolor LCD screen. Moreover, SE-12 Express is configured with stress ECG function, which will allow to diagnose concealed coronary heart disease and atypical angina pectoris, prescribe the workload for patients with myocardial infarction before leaving hospital, and assess the effect of the treatment.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
body surface
Indicated Patient Age Range
adult and pediatric patients
Intended User / Care Setting
hospitals or healthcare facilities by doctors and trained healthcare professionals.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-clinical data:
Biocompatibility testing: Conducted in accordance with ISO 10993-1. Included Cytotoxicity, Skin Sensitization, and Skin Irritation tests.
Electrical safety and electromagnetic compatibility (EMC): Complies with IEC 60601-1:2005/A1: 2012 for safety and IEC 60601-1-2: 2007 for EMC.
Bench Testing: Conducted per IEC 60601-2-25: 2011, all results passed.
Software Verification and Validation Testing: Conducted as recommended by FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Software considered "moderate" level of concern.
Battery Testing: Conducted per UL 2054, UL60950-1:2007, and IEC 61233:2012, all results passed.
Clinical data: Not applicable.
Key results: The non-clinical data support the safety of the device and software verification and validation demonstrate that SE-12 series electrocardiograph device should perform as intended in the specified use conditions. All data demonstrates that the subject devices perform comparably to the predicate device.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
K102830, K131503, K060167, K152427
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.2340 Electrocardiograph.
(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 15, 2017
Edan Instruments, Inc. % Doug Worth Sr. Dir. US RA/QA Edan Medical 1200 Crossman Ave. Suite 200 Sunnyvale, California 94089
Re: K160876 Trade/Device Name: Electrocardiograph, Models: SE-12, SE-12 Express, SE-1200, and SE-1200 Express Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: January 5, 2017 Received: January 9, 2017
Dear Doug Worth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
1
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mude Yellen
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K160876
Device Name
Device Name: Electrocardiograph Model: SE-12, SE-12 Express, SE-1200, and SE-1200 Express
Indications for Use (Describe)
The intended use of SE-12 series electrocardiograph is to acquire ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements is offered to clinicians on an advisory basis only.
Type of Use (Select one or both, as applicable) | |
---|---|
☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(K) Summary
Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc
3/F - B, Nanshan Medical
Equipments Park, Nanhai Rd 1019#,
Shekou, Nanshan Shenzhen,
518067 P.R. China
Tel: +86(0755) 26858736
Fax: +1 (408) 418-4059 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person:
Preparing date: | Alice Yan
January 5, 2016 |
| 2. Device name and
classification: | Device Name: Electrocardiograph
Model: SE-12, SE-12 Express, SE-1200, and SE-1200 Express
Classification Name/ Product code:
870.2340 Electrocardiograph/DPS
Regulatory Class: Class II |
| 3. Predicate Device(s): | 1) EDAN Instrument, Electrocardiograph, model SE-12, SE-12
Express, SE-1200, and SE-1200 Express, K102830
2) EDAN Instrument, Electrocardiograph, model SE-601, K131503
3) Cardiac Science Corporation, model Atria 6100, K060167
4) EDAN Instrument, PC ECG, model SE-1515, K152427 |
| 4. Reason for
Submission | Technology change |
| 5. Pre-Submission,
IDE | Not applicable, there is no prior submission. |
| 6. Device Description: | The SE-12/12 Express/1200/1200 Express Electrocardiograph
gathers ECG signals of 12 leads simultaneously. It displays the
operation menu, ECG parameters as well as electrocardiograms, and
be powered by the mains supply or battery. And the system has
advanced performance and high reliability due to high resolution
thermal recorder, 32-bit processor and a large-capacity memorizer.
Design of the system took much consideration on ergonomics so the
size is suitable for clinic and hospital uses.
There are four selectable modes in the system, including manual, |
4
auto, rhythm, R-R analysis or VCG, and VCG is configured with SE-12 Express & SE-1200 Express. SE-12 and SE-1200 share the same single color LCD screen of which the resolution is 320×240 dot; and LCD screen of SE-12 Express and SE-1200 Express is 800×600 multicolor LCD screen. Moreover, SE-12 Express is configured with stress ECG function, which will allow to diagnose concealed coronary heart disease and atypical angina pectoris, prescribe the workload for patients with myocardial infarction before leaving hospital, and assess the effect of the treatment. The intended use of SE-12 series electrocardiograph is to acquire 7. Intended Use: ECG signals from adult and pediatric patients through body surface ECG electrodes. The electrocardiograph is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by the electrocardiograph can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements is offered to clinicians on an advisory basis only.
8. Predicate Device Comparison
The subject devices share the same characteristics in all items with the predicate device, concluding from using the same technology and principle. All the technological differences existed between the subject and predicate devices are only some performance parameters items, which are shown in the following tables in details.
| Item | Predicate device
(SE-12, SE-12 Express,
SE-1200, and SE-1200 Express) | Proposed device
(SE-12, SE-12 Express,
SE-1200, and SE-1200 Express) | Comparison
Result |
|--------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| K# | K102830 | Current Submission | — |
| Intended Use | The intended use of the
12-channel electrocardiograph is
to acquire ECG signals from
adult and pediatric patients
through body surface ECG
electrodes. The
electrocardiograph is only
intended to be used in hospitals
or healthcare facilities by doctors
and trained healthcare
professionals. The cardiogram
recorded by the | The intended use of SE-12 series
electrocardiograph is to acquire
ECG signals from adult and
pediatric patients through body
surface ECG electrodes. The
electrocardiograph is only
intended to be used in hospitals
or healthcare facilities by doctors
and trained healthcare
professionals. The cardiogram
recorded by the
electrocardiograph can help users | Same |
Table 1: Comparison between the current SE-12 series and previous cleared system | |||||||
---|---|---|---|---|---|---|---|
electrocardiograph can help users | |||||||
to analyze and diagnose heart | |||||||
disease. However, the interpreted | |||||||
ECG with measurements and | |||||||
interpretive statements is offered | |||||||
to clinicians on an advisory basis | |||||||
only. | to analyze and diagnose heart | ||||||
disease. However, the interpreted | |||||||
ECG with measurements and | |||||||
interpretive statements is offered | |||||||
to clinicians on an advisory basis | |||||||
only. | |||||||
Safety Specifications | |||||||
Safety Standards | IEC 60601-1:1988+A1+A2, | ||||||
EN 60601-1:1990+A1+A2, | |||||||
IEC/EN 60601-1-2: 2001+A1, | |||||||
IEC/EN60601-2-25 | |||||||
ANSI/AAMI EC11 | |||||||
IEC/EN 60601-2-51 | IEC 60601-1:2005 | ||||||
EN 60601-1:2006 | |||||||
EN 60601-1-2:2007 | |||||||
IEC 60601-1-2:2007 | |||||||
IEC/EN60601-2-25:2011 | Different | ||||||
Anti-electric-shock type: | Class I with internal power supply | Class I with internal power supply | Same | ||||
Anti-electric-shock degree: | Type CF | Type CF | Same | ||||
Degree of protection against harmful ingress of water: | Ordinary equipment (Sealed equipment without liquid proof) | Ordinary equipment (Sealed equipment without liquid proof) | Same | ||||
Disinfection/sterilization method: | Refer to the user manual for details | Refer to the user manual for details | Same | ||||
Degree of safety of application in the presence of flammable gas: | Equipment not suitable for use in the presence of flammable gas | Equipment not suitable for use in the presence of flammable gas | Same | ||||
Working mode: | Continuous operation | Continuous operation | Same | ||||
EMC: | CISPR 11 Group 1, Class A | CISPR 11 Group 1, Class A | Same | ||||
Ingress rating | IPX0 | IPX0 | Same | ||||
Environment Specifications | |||||||
Temperature | |||||||
Transport & Storage | -20°C (-4°F) ~ +55°C (+131°F) | -20°C (-4°F) ~ +55°C (+131°F) | Same | ||||
Working | +5°C (+41°F) ~ +40°C (+104°F) | +5°C (+41°F) ~ +40°C (+104°F) | Same | ||||
Relative Humidity: | |||||||
Transport & Storage | 25%~93% | ||||||
Non-Condensing | 25%~93% | ||||||
Non-Condensing | Same | ||||||
Working | 25% RH ~ 80% RH | ||||||
Non-Condensing | 25% RH ~ 80% RH | ||||||
Non-Condensing | Same | ||||||
Atmospheric Pressure: | |||||||
Transport & Storage | 700 hPa ~ 1060 hPa | Operating Voltage = | 700 hPa ~ 1060 hPa | Operating Voltage = | Same | ||
Working | Mains Supply: | 860 hPa ~ 1060 hPa | 100V-240V~ | 860 hPa ~ 1060 hPa | 100V-240V~ | Same | Same |
Operating frequency = 50 Hz / | |||||||
60 Hz | Operating frequency = 50 Hz / 60 Hz | Same | |||||
Input Power = 70VA | Input Current = 0.9-0.4A | ||||||
Or Input power = 96VA | Different | ||||||
Rated voltage = 14.8 V | Rated voltage = 14.8 V | Same | |||||
SE-12 Express&SE-1200 | |||||||
Express: Rated capacity = | |||||||
4400mAh | SE-12 Express&SE-1200 | ||||||
Express: Rated capacity = | |||||||
5000mAh | Different | ||||||
SE-12& SE-1200: Rated | |||||||
capacity =2200mAh | SE-12& SE-1200: Rated capacity | ||||||
= 2500mAh | |||||||
Built-in Lithium | |||||||
Battery Pack: | When the battery is fully | ||||||
charged, SE-12&SE-1200 can | |||||||
work | |||||||
normally about 4 hours, and it | |||||||
can continually print about 1.5 | |||||||
hours in the manual mode or | |||||||
print about 300 ECG reports of | |||||||
$3×4+1R$ in the auto mode; SE-12 | |||||||
Express &SE-1200 Express can | |||||||
work | |||||||
normally about 5 hours, and it | |||||||
can continually print about 2.5 | |||||||
hours in the manual mode or | |||||||
print about 350 ECG reports of | |||||||
$3×4+1R$ in the auto mode. | When the battery is fully | ||||||
charged, SE-12& SE-1200 can | |||||||
work normally about 4 hours, and | |||||||
it can continually print about 1.5 | |||||||
hours in the manual mode or | |||||||
print about 300 ECG reports of | |||||||
$3×4+1R$ in the auto mode; SE-12 | |||||||
Express& SE-1200 Express can | |||||||
work normally about 5 hours, and | |||||||
it can continually print about 2.5 | |||||||
hours in the manual mode or | |||||||
print about 350 ECG reports of | |||||||
$3×4+1R$ in the auto mode. | Same | ||||||
Performance Specifications | |||||||
Recording | |||||||
Recorder: | Thermal dot-matrix recorder | Thermal dot-matrix recorder | Same | ||||
8 dots per mm / 200 dots per inch | |||||||
(amplitude axes) | 8 dots per mm / 200 dots per inch | ||||||
(amplitude axes) | |||||||
Printing Density | 40 dots per mm / 1000 dots per | ||||||
inch (time axes, @ 25 mm/s) | 40 dots per mm / 1000 dots per | ||||||
inch (time axes, @ 25 mm/s) | Same | ||||||
Recorder Paper: | Folded thermal paper: | ||||||
210mm×295mm×100pages | Folded thermal paper: | ||||||
210mm×295mm×100pages | |||||||
Folded thermal paper: | |||||||
215mm×280mm×100pages | |||||||
(Optional) | Folded thermal paper: | ||||||
215mm×280mm×100pages | |||||||
(Optional) | Same | ||||||
Rolled thermal paper: | |||||||
210mm×30m (Optional) | Rolled thermal paper: | ||||||
210mm×30m (Optional) | |||||||
Effective Width: | 203 mm | 203mm | Same |
5
6
7
5mm/s, 6.25mm/s, 10mm/s, | 5mm/s, 6.25mm/s, 10mm/s, | ||
---|---|---|---|
Paper Speed: | 12.5mm/s, 25mm/s, 50mm/s | ||
(±3%) | 12.5mm/s, 25mm/s, 50mm/s | ||
(±3%) | Same | ||
Accuracy of data: | ±5% (x-axis), ±5% (y-axis) | ±5% (x-axis), ±5% (y-axis) | Same |
HR Recognition | |||
Technique: | Peak-Peak Detection | Peak-Peak Detection | Same |
HR Range: | 30 BPM ~ 300 BPM | 30 BPM ~ 300 BPM | Same |
Accuracy: | ±1 BPM | ±1 BPM | Same |
ECG Unit | |||
Leads: | Standard 12 leads | Standard 12 leads | Same |
Acquisition | |||
Mode: | Simultaneously 12 leads | Simultaneously 12 leads | Same |
A/D | 24 bits | 24 bits | Same |
Resolution: | 2.52uV/LSB | 2.52uV/LSB | Same |
Time Constant: | ≥ 3.2 s | ≥ 3.2 s | Same |
Frequency | |||
Response: | 0.05 Hz ~ 150 Hz (-3 dB) | 0.05 Hz ~ 150 Hz (-3 dB) | Same |
Gain: | 2.5, 5, 10, 20, 10/5, AGC | ||
(mm/mV) | 2.5, 5, 10, 20, 10/5, AGC | ||
(mm/mV) | Same | ||
Input Impedance: | ≥50 MΩ(10 Hz) | ≥50 MΩ(10 Hz) | Same |
Input Circuit | |||
Current: | ≤0.01μA | ≤0.01μA | Same |
Input Voltage | |||
Range | ≤±5 mVpp | ≤±5 mVpp | Same |
Calibration | |||
Voltage: | 1 mV±3% | 1 mV±3% | Same |
DC Offset | |||
Voltage: | ±600 mV | ±600 mV | Same |
Noise: | ≤12.5 µVp-p | ≤12.5 µVp-p | Same |
Multi-channel | |||
Crosstalk | ≤0.5 mm | ≤0.5 mm | Same |
Filter | AC Filter: On / Off | AC Filter: On / Off | Same |
DFT Filter: 0.05Hz / 0.15Hz / | |||
0.25Hz / 0.32Hz / 0.5Hz / | |||
0.67Hz | DFT Filter: 0.05Hz / 0.15Hz / | ||
0.25Hz / 0.32Hz / 0.5Hz / 0.67Hz | Same | ||
EMG Filter: 25Hz / 35Hz / 45Hz | |||
/ OFF | EMG Filter: 25Hz / 35Hz / 45Hz | ||
/ OFF | Same | ||
LOWPASS Filter: 150Hz / | |||
100Hz / 75Hz | LOWPASS Filter: 150Hz / 100Hz | ||
/ 75Hz | Same | ||
CMRR | ≥115 dB | ≥115 dB | |
Sampling | |||
Frequency | 1000 Hz | 1000 Hz | |
Pacemaker Detection | |||
Amplitude | ±2 to ±700 mV | ±2 to ±700 mV | Same |
Width | 0.1 to 2.0 ms | 0.1 to 2.0 ms | Same |
Sampling | |||
Frequency | 10,000/sec/channel | 10,000/sec/channel | Same |
External Input/ Output | |||
Input | ≥100 kΩ; Sensitivity 10 | ||
mm/V±5%; | |||
Single ended | ≥100 kΩ; Sensitivity 10 | ||
mm/V±5%; | |||
Single ended | Same | ||
Output | ≤100Ω; Sensitivity 1 V/mV ±5%; | ||
Single ended | ≤100Ω; Sensitivity 1 V/mV ±5%; | ||
Single ended | Same |
8
Table 2: Comparison between SE-12 series and SE-601 series
| Item | Predicate device
(SE-601C) | Proposed device
(SE-12, SE-12 Express,
SE-1200, and SE-1200 Express) | Comparison
Result |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| K# | K131503 | Current Submission | — |
| Intended Use | The intended use of SE-601 is to
acquire ECG signals from adult
and pediatric patients through
body surface ECG electrodes.
The electrocardiograph is only
intended to be used in hospitals or
healthcare facilities by doctors
and trained healthcare
professionals. The cardiogram
recorded by the
electrocardiograph can help users
to analyze and diagnose heart
disease. However, the interpreted
ECG with measurements and
interpretive statements is only
intended to be used on adult
patients and is offered to
clinicians on an advisory basis
only. | The intended use of SE-12 series
electrocardiograph is to acquire
ECG signals from adult and
pediatric patients through body
surface ECG electrodes. The
electrocardiograph is only
intended to be used in hospitals or
healthcare facilities by doctors
and trained healthcare
professionals. The cardiogram
recorded by the
electrocardiograph can help users
to analyze and diagnose heart
disease. However, the interpreted
ECG with measurements and
interpretive statements is offered
to clinicians on an advisory basis
only. | Same |
| Transmitting
Frequency | 2412-2497MHz | 2400-2497MHz | |
| Frequency Band | 2412-2497MHz | 2400-2497MHz | |
| Modulation Type | DSSS, CCK | DSSS, CCK,OFDM | Different |
| Transmitting
Power | 18dBm | 6-17dBm | |
| Effective | 18dBm | 6-17dBm | |
9
Radiated Power | ||
---|---|---|
Table 3: Comparison between predicate device Atria 6100 and the proposed system | ||||
---|---|---|---|---|
| Item | Predicate device
(Atria 6100) | Proposed device
(SE-12, SE-12 Express,
SE-1200, and SE-1200 Express) | Comparison
Result |
|--------------|----------------------------------|----------------------------------------------------------------------------|----------------------|
| K# | K060167 | Current Submission | — |
| Screen Size | 640480 | 320240 (SE-12, SE-1200)
800*600(SE-12Express,
SE-1200Express) | Different |
| SW Algorithm | Glasgow | Glasgow | Same |
Table 2-4: Comparison between predicate device SE-1515 (K152427) and SE-12, SE-12 Express, SE-1200. | |||
---|---|---|---|
and SE-1200 Express |
| Item | Predicate device
(SE-1515) | Proposed device
(SE-12, SE-12 Express,
SE-1200, and SE-1200 Express) | Comparison
Result |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| K# | K152427 | Current Submission | |
| Intended Use | SE-1515 PC ECG is intended to
acquire, process and store ECG
signals from adult and pediatric
patients undergoing stress exercise
test or resting test. The SE-1515 PC
ECG is intended to be used only in
hospitals and healthcare facilities by
doctors and trained healthcare
professionals. The cardiogram
recorded by the SE-1515 PC ECG
can help users to analyze and
diagnose heart diseases. However,
the ECG with measurements and
interpretive statements is offered to
clinicians on an advisory basis only.
It is mainly used in the ECG
Outpatient Department and Physical
Examination Department. | The intended use of SE-12 series
electrocardiograph is to acquire ECG
signals from adult and pediatric
patients through body surface ECG
electrodes. The electrocardiograph is
only intended to be used in hospitals
or healthcare facilities by doctors and
trained healthcare professionals. The
cardiogram recorded by the
electrocardiograph can help users to
analyze and diagnose heart disease.
However, the interpreted ECG with
measurements and interpretive
statements is offered to clinicians on
an advisory basis only. | different |
| SW Algorithm | SEMIP | SEMIP | Same |
As seen in the comparison tables, the subject and predicate devices have similar design features and performance specifications. The main technological differences between the subject and predicate devices are minor differences, and do not raise different questions of safety or effectiveness. As
10
demonstrated in the non-clinical testing, the different technological characteristics do not affect the safety and effectiveness of the Edan SE-12 series electrocardiograph.
9. Performance Data:
Non-clinical data:
The following performance data were provided in support of the substantial equivalence determination.
Biocompatibility testing
The biocompatibility evaluation for SE-12 series electrocardiograph is conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered surface contacting for duration of less than 24 hours. And the battery of testing included the following tests:
- Cytotoxicity .
- Skin Sensitization .
- . Skin Irritation
Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the SE-12 series electrocardiograph, consisting of all the modules and accessories in the system. The system complies with the IEC 60601-1:2005/A1: 2012 standard for safety and the IEC 60601-1-2: 2007 standard for EMC.
Bench Testing
Bench testing was conducted per IEC 60601-2-25: 2011, and all the results show pass.
Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a failure or latent flaw in the software will not directly result in serious injury or death to the patient or operator, and the software device is an accessory to a medical device that has a Moderate Level of Concern.
Battery Testing
Testing on battery are conducted per UL 2054, UL60950-1:2007 and IEC 61233:2012, and results show pass for all testing items.
Clinical data: Not applicable.
Summary
Based on the non-clinical and clinical performance as documented in the system development, the subject devices were found to have a safety and effectiveness profile that is similar to the predicate
11
device.
10. Conclusion
The non-clinical data support the safety of the device and software verification and validation demonstrate that SE-12 series electrocardiograph device should perform as intended in the specified use conditions, and all the data demonstrate that the subject devices perform comparably to the predicate device that is currently marketed for the same intended use. In other words, the subject SE-12 series electrocardiograph device is substantially equivalent to the predicate devices.