Search Results
Found 1 results
510(k) Data Aggregation
(257 days)
PACSEAL PIT & FISSURE SEALANT
PacSeal™ Pit & Fissure Sealant is intended for use as:
- Seal the pits and fissures in teeth
PacSeal™ Pit & Fissure Sealant is a fluoride releasing, light-cured acrylate resin designed to fill and seal the pits and fissures of teeth.
Here's a breakdown of the acceptance criteria and study information for the PacSeal™ Pit & Fissure Sealant as derived from the provided document:
This document is a 510(k) summary, which typically focuses on demonstrating substantial equivalence to a predicate device rather than presenting a full clinical study with specific acceptance criteria in the manner of a novel device. Therefore, the "acceptance criteria" are implied by the comparison to the predicate device's performance characteristics.
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state "acceptance criteria" as pass/fail thresholds. Instead, it compares the subject device's performance to that of predicate devices, implicitly suggesting that performance comparable to or better than the predicate is acceptable.
Performance Characteristic | Acceptance Criteria (Implied by Predicate) | Reported Device Performance (PacSeal™ Pit & Fissure Sealant) |
---|---|---|
Depth of Cure (ISO 6874-2005) | Comparable to or better than 1.86 mm (3M™ Clinpro™ Sealant) | Complies (with ISO 6874-2005) |
Compressive Strength | Comparable to or better than 210 MPa (3M™ Clinpro™ Sealant) | 181 MPa |
Diametral Tensile Strength | Comparable to or better than 37 MPa (3M™ Clinpro™ Sealant) | Subject device is equivalent to the predicate device |
Film Thickness | Comparable to or better than 10 µm (3M™ Clinpro™ Sealant) | Subject device is equivalent to the predicate device |
Flexural Strength (ISO 4049-2009) | Comparable to or better than 110 MPa (3M™ Clinpro™ Sealant) | Complies (with ISO 4049-2009) |
Light Curing Time (ISO 6874-2005) | 30 sec (matches 3M™ Clinpro™ Sealant) | 30 sec |
Shear Bond Strength to Etched Enamel | Comparable to or better than 8.12 MPa (3M™ Clinpro™ Sealant) | Subject device is equivalent to the predicate device |
Note: For "Compressive Strength," the subject device (181 MPa) is lower than the predicate (210 MPa). The document still concludes substantial equivalence, suggesting that this difference was deemed acceptable in the context of overall performance and "equivalent to the predicate" claim for other properties.
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the sample sizes used for the performance tests (e.g., number of specimens for depth of cure, strength tests).
The data provenance is not explicitly stated beyond being presented in a 510(k) summary by Pac-Dent International, Inc. These types of tests are typically in-vitro laboratory tests, not clinical studies with patient data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of information is not applicable to the data provided. The "ground truth" for these physical property tests is based on standardized laboratory measurements (e.g., ISO standards) and not on expert consensus from human readers or clinicians.
4. Adjudication Method for the Test Set
This is not applicable as the tests are for physical properties measured in a lab, not for diagnostic or interpretive tasks requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, an MRMC comparative effectiveness study was not done. This document describes the physical and chemical characteristics of a dental sealant, not an AI or imaging device that would typically undergo MRMC studies.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
No, this is not relevant to the device described. The device is a physical dental sealant, not an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for the performance characteristics (e.g., depth of cure, strength, film thickness) is based on standardized laboratory test methods (e.g., ISO 6874-2005, ISO 4049-2009).
8. The Sample Size for the Training Set
This is not applicable. The device is a physical product, not an algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This is not applicable for the same reason as above.
Ask a specific question about this device
Page 1 of 1