Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K073487
    Date Cleared
    2008-02-11

    (61 days)

    Product Code
    Regulation Number
    866.5510
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    OLYMPUS IGM REAGENT

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    System reagent for the quantitative determination of IgM immunoglobulins in human serum and plasma on OLYMPUS analyzers. For in vitro diagnostic use.

    Device Description

    In this Olympus procedure:

    • When a sample is mixed with R1 buffer and R2 antiserum solution, human IgM reacts specifically with anti-human IgM antibodies to yield insoluble aggregates.
    • Immune complexes formed in solution scatter light in proportion to their size, shape and concentration.
    • Turbidimeters measure the reduction of incidence light due to reflection, absorption or scatter.
    • In the Olympus procedure, the decrease in intensity of light transmitted (increase in absorbance) through particles suspended in solution is as a result of complexes formed during the antigen-antibody reaction.
    AI/ML Overview

    The provided 510(k) summary describes a new Olympus IgM reagent (OSR6X173) and compares its performance to a predicate device. This document is a premarket notification for a diagnostic reagent, which falls under IVD (in vitro diagnostic) devices. For IVD devices, the "study" typically refers to analytical and clinical performance studies, not AI-specific studies like MRMC or standalone algorithm performance. Therefore, many of the requested items related to AI device performance (e.g., number of experts, adjudication methods, MRMC studies, standalone performance, training set details) are not directly applicable or reported in this type of submission.

    Here's an analysis based on the provided text, focusing on the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in a separate section. Instead, it presents performance characteristics of the new device and the predicate device side-by-side, implying that the new device's performance is acceptable if it is comparable to or better than the predicate. For the purpose of this analysis, I will treat the performance values of the predicate device and the new device as demonstrating acceptable performance through substantial equivalence.

    Performance CharacteristicAcceptance Criteria (Predicate Device)Reported Device Performance (New Olympus IgM reagent OSR6X173)
    Precision (Total CV%)
    AU400/400eSample 1: 1.74%
    Sample 2: 1.18%
    Sample 3: 1.49%Sample 1: 4.03%
    Sample 2: 2.95%
    Sample 3: 2.60%
    AU600/640/640eSample 1: 2.16%
    Sample 2: 1.48%
    Sample 3: 1.35% (for AU600)
    Sample 1: 1.5%
    Sample 2: 1.3% (for AU640/640e)Sample 1: 3.44%
    Sample 2: 3.29%
    Sample 3: 4.08%
    AU2700/5400Sample 1: 2.49%
    Sample 2: 2.57%
    Sample 3: 3.07%Sample 1: 3.79%
    Sample 2: 3.31%
    Sample 3: 3.52%
    Assay RangeNot explicitly stated, inferred from Method Comparison20-500 mg/dL
    Hook Effect/ProzoneMay occur with highly elevated IgM samples > 3,500 mg/dL polyclonalMay occur with highly elevated IgM samples > 10,000 mg/dL polyclonal
    Method Comparison (Linear Regression)Slope: 0.968
    Intercept: 0.6
    R: 0.998
    N: Not clearly stated in predicate table (appears truncated)Slope: 1.006
    Intercept: 2.8
    R: 1.000
    N: 107
    Interfering Substances (Bilirubin)
    AU400/400eInterference less than 2% up to 40 mg/dLInterference less than 4% up to 40 mg/dL
    AU600/640/640eInterference less than 5% up to 40 mg/dLInterference less than 3% up to 40 mg/dL
    AU2700/5400Interference less than 10% up to 40 mg/dLInterference less than 8% up to 40 mg/dL
    Interfering Substances (Hemolysis)
    AU400/400eInterference less than 2% up to 500 mg/dLInterference less than 4% up to 500 mg/dL
    AU600/640/640eInterference less than 10% up to 500 mg/dLInterference less than 3% up to 500 mg/dL
    AU2700/5400Interference less than 5% up to 500 mg/dLInterference less than 3% up to 500 mg/dL
    Interfering Substances (Lipemia)
    AU400/400eInterference less than 10% up to 300 mg/dLInterference less than 10% up to 300 mg/dL
    AU600/640/640eInterference less than 10% up to 400 mg/dLInterference less than 10% up to 300 mg/dL
    AU2700/5400Interference less than 10% up to 400 mg/dLInterference less than 10% up to 200 mg/dL
    Interfering Substances (Ascorbic Acid)
    AU400/400eInterference less than 2% up to 20 mg/dLNot Tested
    AU600/640/640eInterference less than 1% up to 20 mg/dLNot Tested
    AU2700/5400Interference less than 3% up to 20 mg/dLNot Tested
    Functional SensitivityNot Specified
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1