Search Results
Found 3 results
510(k) Data Aggregation
(161 days)
Nitrile Examination Gloves (Blue, Violet)
The Nitrile Examination Gloves is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
The proposed device, Nitrile Examination Gloves (Blue, Violet) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size and color. The color of the proposed device is Blue and Violet. The proposed device is not provided as sterilized. The proposed device is made of Nitrile.
This document describes the performance testing for Nitrile Examination Gloves (Blue, Violet). Here's a breakdown of the acceptance criteria and study details:
Acceptance Criteria and Reported Device Performance
Device: Nitrile Examination Gloves (Blue, Violet)
Purpose: Non-sterile disposable device worn to prevent contamination between patient and examiner.
Test Method | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
ASTM D5151 | Testing for Freedom from Holes | Freedom from holes AQL 2.5 | Blue Color: No water leakage inspected from 200 samples |
Violet Color: No water leakage inspected from 200 samples | |||
ASTM D6124 | Determine the powder residue for powder free gloves |
Ask a specific question about this device
(195 days)
Powder free Nitrile Examination Gloves (Blue, Violet Blue, White, Cobalt Blue)
The Powder free Nitrile Examination Gloves (Blue, White, Cobalt Blue) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed device, Powder Free Nitrile Examination Gloves (Blue, Violet Blue, White, Cobalt Blue) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed devices are Powder Free Nitrile Examination Gloves and includes variations of different size and color. The colors of the proposed device are Blue, Violet Blue, White, and Cobalt Blue.
The proposed device is not provided as sterilized
This document is a 510(k) Pre-Market Notification for Powder free Nitrile Examination Gloves. The acceptance criteria and the study proving the device meets these criteria are detailed in the "Performance and Physical Specifications" table (Table 2) on page 5, and the "Performance Comparison" table (Table 3) on pages 6-7, supported by non-clinical bench tests.
1. A table of acceptance criteria and the reported device performance
Acceptance Criteria (ASTM D6319-15) | Reported Device Performance (Powder free Nitrile Examination Gloves - Proposed Device) |
---|---|
Pinhole AQL | 1.5 |
Before Aging | |
Tensile Strength | 14 MPa, min |
Ultimate Elongation | 500 % min |
After Aging | |
Tensile Strength | 14 MPa, min |
Ultimate Elongation | 400 % min |
Fulfill ASTM D6319 | Comply with ASTM D6319 |
Freedom from Holes | Be free from holes when tested in accordance with ASTM D5151 |
Powder Content | Less than 2 mg per glove when tested in accordance with ASTM D6124 |
Biocompatibility | |
Irritation and Sensitization | Under the conditions of the study, not an irritant and not a sensitizer |
Cytotoxicity | Under conditions of the study, not cytotoxic |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document does not explicitly state the sample size used for the test set for each specific test (e.g., tensile strength, pinholes). However, the testing was conducted according to recognized ASTM and ISO standards, which typically specify sampling plans. The data provenance is from non-clinical bench tests performed to verify the device's compliance with design specifications and relevant standards. The country of origin of the data is implied to be related to the manufacturer in Vietnam or the testing facilities they utilized. The studies are prospective as they were specifically conducted for this 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This is not applicable for this type of device and study. The "ground truth" for performance of examination gloves is established by quantitative measurements according to standardized test methods (ASTM and ISO standards), not by expert opinion.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. As described above, performance is determined by objective measurements using standardized test methods.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an examination glove, not an AI-powered diagnostic tool requiring human reader studies.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is an examination glove, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the performance of the examination gloves is based on established industry standards and test methods (ASTM D6319-15, ASTM D5151-06, ASTM D6124-17, ISO 10993-10:2010, ISO 10993-5:2009). These standards define the acceptable physical properties, barrier integrity, and biocompatibility of medical examination gloves.
8. The sample size for the training set
Not applicable. This device is an examination glove, not a machine learning model that requires a training set.
9. How the ground truth for the training set was established
Not applicable. There is no training set for this device.
Ask a specific question about this device
(192 days)
Nitrile Examination Gloves (Blue, Violet Blue, White)
The NITRILE EXAMINATION GLOVES (Blue, Violet Blue, \Vhite) is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed device, NITRILE EXAMINATION GLOVES (Blue, Violet Blue, White) are disposable devices intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. The proposed devices are NITRILE EXAMINATION GLOVES (Blue, Violet Blue, White) includes variations of different size and color. The colors of the proposed device are Blue, and White.
The document describes the K202402 Nitrile Examination Gloves. Here's a breakdown of the acceptance criteria and the study details:
1. Table of Acceptance Criteria and the Reported Device Performance
Acceptance Criteria (Standard) | Test Parameter | Proposed Device Performance (Minimum) | Predicate Device AQL (or value) |
---|---|---|---|
ASTM D6319-15 | Pinhole AQL | 1.5 | N/A (implied 1.5 by ASTM) |
Tensile Strength (Before Aging) | 15 MPa | 15 MPa | |
Ultimate Elongation (Before Aging) | 500 % | 500 % | |
Tensile Strength (After Aging) | 14 MPa | 14 MPa | |
Ultimate Elongation (After Aging) | 500 % | 400 % | |
ASTM D5151-06 | Freedom from Holes | Free from holes | Free from holes |
ASTM D6124-17 | Powder Content | Less than 2 mg per glove | Meet ASTM D6319 requirements |
ISO 10993-10:2010 | Irritation | Not an irritant | Not an irritant |
Sensitization | Not a sensitizer | Not a sensitizer | |
ISO 10993-5:2009 | In vitro cytotoxicity | N/A (implied non-cytotoxic) | N/A (implied non-cytotoxic) |
Biocompatibility | Acute Systemic Toxicity | Not toxic (for proposed device) | Not Available (for predicate) |
Physical (Dimensions) | Length (XS, S, M) | 230 mm min | 230 mm min |
Length (L, XL) | 235 mm min | 230 mm min | |
Width | 70, 80, 95, 110, 120 mm (±10 mm) | 70, 80, 95, 110, 120 mm (±10 mm) | |
Finger Thickness | 0.05 mm min | 0.10-0.12 mm (±0.03 mm) | |
Palm Thickness | 0.05 mm min | 0.08-0.10 mm (±0.03 mm) | |
Cuff Thickness | 0.05 mm min | 0.06-0.09 mm (±0.03 mm) |
Note: The document explicitly states that "The above data of size, performance, and physical specifications of proposed gloves meet all the current specifications listed in the ASTM standard D6319." and that the bench tests "demonstrated that the proposed device complies with the following standards".
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes for each test that was performed. It generally refers to "bench tests" for physical properties and "biocompatibility tests".
- Provenance: No specific country of origin is mentioned for the test data itself, beyond the manufacturer being in China. The study is non-clinical bench testing, not human data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a medical glove, and the "ground truth" for its performance is established by adherence to recognized international standards (ASTM, ISO) through objective physical and biocompatibility testing, not expert consensus on interpretations of complex medical images or outcomes.
4. Adjudication method for the test set
Not applicable, as the tests are objective measurements against standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical glove, not an AI-powered diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is a medical glove, not an algorithm.
7. The type of ground truth used
The ground truth used for performance evaluation is defined by international consensus standards for medical gloves (ASTM D6319-15, ASTM D5151-06, ASTM D6124-17) and biocompatibility testing standards (ISO 10993-10:2010, ISO 10993-5:2009). The device is tested against the specified parameters and thresholds outlined in these standards.
8. The sample size for the training set
Not applicable. This is a physical medical device (gloves), not an AI algorithm requiring a training set.
9. How the ground truth for the training set was established
Not applicable. This is a physical medical device (gloves), not an AI algorithm.
Ask a specific question about this device
Page 1 of 1