Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K234074
    Manufacturer
    Date Cleared
    2024-07-02

    (193 days)

    Product Code
    Regulation Number
    870.1250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Next Generation Access Catheter

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Next Generation Access Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.

    Device Description

    The Next Generation Access Catheter is a single-lumen, flexible, variable stiffness composite catheter. The distal tip of the catheter shaft includes a marker band while the proximal end of the catheter has a luer fitting to allow attachment of accessories and infusion of liguids through the system. The Next Generation Access Catheter distal shaft has an external hydrophilic coating which provides a lubricious surface during use. The Next Generation Access Catheter is visible under fluoroscopy. The Next Generation Access Catheter dimensions are included on the individual device label. The Next Generation Access Catheter is offered in various sizes to accommodate physician preferences and anatomical variations. A peelable split introducer sheath is provided in the package to provide support and facilitate the introduction of the distal tip of the Next Generation Access Catheter into an appropriate vascular sheath. The catheter and introducer sheath are provided sterile, non-pyrogenic, and are intended for single use only.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the "Next Generation Access Catheter" (K234074). This documentation focuses on demonstrating substantial equivalence to a predicate device through non-clinical performance testing. It does not contain information about a study involving human readers with or without AI assistance, or any specific AI component that would require a ground truth established by experts. Therefore, many of the requested sections related to AI performance, sample sizes for test/training sets, expert qualifications, and adjudication methods are not applicable here.

    Here's a summary of the available information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document doesn't explicitly list "acceptance criteria" for each test but rather states whether the device "Passed" or "Met requirements." The "Test Method Summary" implicitly defines the criteria being evaluated.

    TestTest Method SummaryReported Device Performance (Results)
    Biocompatibility
    Cytotoxicity - MEM ElutionTested in accordance with ISO 10993-5Pass (Considered non-cytotoxic)
    Hemocompatibility - Hemolysis (Direct Contact & Extract Method)Tested in accordance with ISO 10993-4Pass (Considered non-hemolytic)
    Hemocompatibility - Complement Activation SC5b-9 AssayTested in accordance with ISO 10993-4Pass (Requirements met)
    Hemocompatibility - Partial Thromboplastin Time (PTT)Tested in accordance with ASTM F2382-18Pass (Requirements met)
    Hemocompatibility – Blood Platelet and Leukocyte Count (PLC)Tested in accordance with ASTM F2888-19Pass (Requirements met)
    Hemocompatibility - Comparative Surface AssessmentVisually inspected at more than 40x magnification.Pass (Requirements met)
    Hemocompatibility - Thrombogenicity in a Canine ModelTested in accordance with ISO 10993-4Pass (Performed similarly to comparator)
    Pyrogenicity - Material-Mediated Rabbit PyrogenTested in accordance with USP 151Pass (Requirements met)
    Sensitization - Guinea Pig Maximization SensitizationTested in accordance with ISO 10993-10Pass (Did not elicit a sensitization response)
    Systemic Toxicity - Acute Systemic InjectionTested in accordance with ISO 10993-11Pass (Requirements met)
    Irritation - Intracutaneous ReactivityTested in accordance with ISO 10993-23Pass (Requirements met)
    Physical and Mechanical Performance
    Dimensional VerificationCatheter outer diameter, inner diameter, usable length, tip length, and coating length were measured.Pass
    Surface ContaminationVisual inspection completed for surface defects.Pass
    Tensile StrengthPeak tensile force evaluated per ISO 10555-1 after preconditioning in a simulated use model.Pass
    Kink ResistanceEvaluated after preconditioning in a simulated use model.Pass
    Liquid LeakageExposed to liquid pressure for 30 seconds; inspected for leakage per ISO 10555-1.Pass
    Air LeakageTested for air leakage into the hub per ISO 10555-1.Pass
    Dynamic BurstVerified ability to withstand internal liquid pressure under dynamic flow conditions with distal end open.Pass
    Torque StrengthMeasured number of catheter rotations until failure after preconditioning in a simulated use model and compared to predicate.Pass
    Hub Validation TestingMet established acceptance criteria per ISO 80369-7.Pass
    Particulate MatterUnderwent simulated use testing; particulate testing conducted including a reference device for comparison.Pass
    Tip BucklingMaximum force to cause catheter tip buckling while constrained at varying distances was measured.The tip stiffness was comparable to the predicate and other cleared catheters.
    CorrosionTested per ISO 10555-1.No evidence of corrosion and met requirements per ISO 10555-1.
    Static BurstDistal tip blocked, fluid injected at increasing pressure until burst (per ISO 10555-1); compared static burst pressure with maximum pressure generated with manual syringe injection.Pass
    Coating IntegrityInspected before and after preconditioning through a simulated use model.No evidence of surface damage or coating defects.
    Saline and Contrast ExposureAfter device use to deliver saline and contrast media, inspected for damage, and dimensional attributes measured.No visual evidence of damage or dimensional changes.
    Radiopacity (Visibility)Tested to demonstrate acceptable radiopacity.Marker radiopacity is comparable to the predicate.
    Design Validation / UsabilitySubject and predicate devices prepared per IFU and tested for usability in a clinically relevant anatomical model.Device preparation, introduction, trackability, and retrieval were comparable to the predicate.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify exact sample sizes for each bench test conducted. The provenance of the data is from Balt USA, LLC and describes non-clinical bench testing. There is no mention of human data, so concepts like "country of origin of the data" or "retrospective or prospective" do not apply.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This section is not applicable. The tests performed are engineering and biocompatibility tests on a physical medical device, not a diagnostic algorithm requiring ground truth established by medical experts.

    4. Adjudication Method for the Test Set

    This section is not applicable. There was no human interpretation or subjective assessment that would require an adjudication method.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This section is not applicable. This submission is for a physical medical device (catheter), not an AI-powered diagnostic tool, and therefore no MRMC study or AI assistance evaluation was conducted.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This section is not applicable. This submission is for a physical medical device, not a standalone algorithm.

    7. The Type of Ground Truth Used

    For the non-clinical performance tests, the "ground truth" is defined by established engineering standards (e.g., ISO, ASTM, USP) and the physical properties and performance characteristics of the device itself (e.g., directly measured dimensions, observed leakage, visual inspection for defects, comparison to a predicate device's performance).

    8. The Sample Size for the Training Set

    This section is not applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the Ground Truth for the Training Set Was Established

    This section is not applicable. There is no "training set" as this is not an AI/machine learning device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1