Search Results
Found 4 results
510(k) Data Aggregation
(86 days)
NewTom 5G XL
The NewTom 5G XL is cone beam computed tomography x-ray imaging system that acquires sequences of images of the head, including ear, nose and throat (ENT), of dento-maxillofacial complex, teeth, mandible and jaw, temporo-mandbular joint (TMI), other areas of the human skull and neck with sections of upper cervical spine and lower extremities for use in diagnostic support.
The NewTom 5G XL is a volumetric Cone Beam Computed Tomography (CBCT) system intended for diagnostic use. It obtains geometrical and x-ray information through digital 2D and 3D images of the objects in the area being examined.
The NewTom 5G XL is able to analyze the regions of the head, including the dental and maxillofacial and the otorhinolaryngology areas, the cervical spinal column and upper and lower limbs.
Particularly, the NewTom 5G XL can be used in the following fields:
- maxillofacial surgery, orthodontics, implants, endodontics, otorhinolaryngology, prosthetics, and orthopedics.
The NewTom 5G XL CBCT device acquires a sequence of x-ray images by performing a rotation around the patient's head then, starting from these images, reconstructs a 3D matrix of the examined volume and produces 2D views of it.
From these views it is possible to reconstruct other 2D images on which 3D images, angle and distance measurements can be taken. Each image can be printed and/or stored in a magnetic/optical media (CD, hard disk, network, USB pen or other mass storage device).
NewTom 5G XL consists of three main components: the scanner unit, the motorized patient support and the main workstation (secondary workstations could be also available as optional) that supports the dedicated software, enabling data acquisition.
The software requires a Microsoft Windows OS (version 7 or later).
I am sorry, but the provided text does not contain the detailed information needed to construct the complete table of acceptance criteria and reported device performance, nor does it provide a full study description as requested.
The document discusses the substantial equivalence of the NewTom 5G XL device to predicate devices (NewTom 5G and NewTom VGi Evo) based on safety, non-clinical performance, and clinical performance tests. However, it does not explicitly state acceptance criteria as specific numerical thresholds that the device was required to meet for each performance metric. It generally states that the device's performance is "comparable" or "able to fulfill the main standard reference."
Here's a breakdown of what can be extracted and what is missing:
1. A table of acceptance criteria and the reported device performance:
Metric | Acceptance Criteria (Not explicitly stated as numerical thresholds in the document) | Reported Device Performance (Comparative statements) |
---|---|---|
SNR (Signal-to-Noise Ratio) | Comparable to predicate device NewTom 5G (K142222) and reference device NewTom VGi Evo (K151612) | NewTom 5G XL has SNR comparable to the values given by the predicate device NewTom 5G (K142222) and by the reference device NewTom VGi Evo (K151612). |
Spatial Resolution | Comparable to predicate device NewTom 5G (K142222) and reference device NewTom VGi Evo (K151612) | NewTom 5G XL has spatial resolution comparable to the values given by the predicate device NewTom 5G (K142222) and by the reference device NewTom VGi Evo (K151612). |
MTF (Modulation Transfer Function) | Comparable to predicate device NewTom 5G (K142222) | NewTom 5G XL has modulation transfer function (MTF) comparable to the values obtained using the predicate device NewTom 5G (K142222). Additionally, specific MTF values are listed in the comparison table: MTF50 > 6, MTF10 > 12. |
NPS (Noise Power Spectrum) | Comparable to predicate device NewTom 5G (K142222) | NewTom 5G XL has noise power spectrum (NPS) comparable to the values obtained using the predicate device NewTom 5G (K142222). |
CTDIw (Weighted Computed Tomography Dose Index) | Comparable to predicate device NewTom 5G (K142222) for similar FOVs, or lower | The CTDIw dose value given by NewTom 5G XL are comparable to the values obtained using the predicate device NewTom 5G (K142222). Generally speaking, considering similar FOVs, the CTDIw value measured with the NewTom 5GXL is lower than the CDTIw value measured with the NewTom 5G. |
Image Quality (DIN 6868-161) | "Quality level" evaluated according to German standard DIN6868-161 | Test report according to standard DIN 6868-161 has been performed to evaluate the quality level of images obtained using the device NewTom 5G XL, according to the German standard DIN6868-161. (Implicitly, the results met the standard's requirements for image quality). |
Image Quality (NEMA XR 29-2013 / MITA Smart Dose) | Diagnostic quality images while optimizing radiation use | Test report according to standard NEMA XR 29-2013 (MITA Smart Dose) has been performed to evaluate that patient undergoing CT exams by means of the proposed device NewTom 5G XL, are subjected to procedures that produces images having diagnostic quality expected while optimizing the use of ionizing radiation. (Implicitly, the device produces such images). |
Image Quality (IEC 61223-3-5) | Performance parameters (image quality, patient dose, positioning) evaluated and met | With this performance test, the essential parameters which describe the performance of the CT scanners with regard to image quality, patient dose and positioning have been evaluated. (Implicitly, the device met the requirements of this standard). |
Clinical Image Resolution and Quality | Diagnostic reliability and utility based on clinical experience | On the subject device have been performed clinical evaluation in order to judge the images resolution and quality during CBCT acquisition. The quality evaluation has been done considering different anatomical sites and different patient population. Based on clinical experience, the test results can show the diagnostic reliability and utility of the radiographic images produced by the NewTom 5G XL CBCT scanner. |
Missing information for a complete study description:
- Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- The document mentions "clinical evaluation" and "different anatomical sites and different patient population" but does not specify the number of cases/patients, the origin of the data, or if it was retrospective or prospective.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- The document mentions "clinical evaluation" and "clinical experience" but does not specify the number of experts or their qualifications.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- No information provided.
- If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- This device is a CBCT imaging system, not an AI-powered diagnostic algorithm. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable in this context (and not mentioned). The studies described are assessing the performance of the imaging device itself rather than a CAD/AI component for interpretation.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable as it's an imaging device, not a standalone algorithm. The "clinical evaluation" mentioned is of the images generated by the device, not an algorithm's output.
- The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the clinical evaluation, the ground truth seems to be based on "clinical experience" and the ability of the images to show "diagnostic reliability and utility." It's not explicitly stated as expert consensus, pathology, or outcomes data. For the non-clinical tests, physical phantoms were used as the reference.
- The sample size for the training set:
- Not applicable; this is an imaging device, not an algorithm that requires a training set in the typical machine learning sense. The software updates and functionalities are compared to previous versions, but no "training set" for an AI algorithm is referenced.
- How the ground truth for the training set was established:
- Not applicable.
In summary: The provided text focuses on demonstrating the substantial equivalence of a new CBCT imaging device (NewTom 5G XL) to existing predicate devices by comparing technical specifications and performance metrics using phantoms and general clinical evaluation. It does not describe an AI/CAD algorithm validation study with specific acceptance criteria, sample sizes, and expert adjudication details typical for such submissions.
Ask a specific question about this device
(166 days)
NewTom 5G
The NewTom 5G is a cone beam computed tomography x-ray imaging system that acquires of the head including the ear, nose and throat (ENT), dento-maxillofacial complex, teeth, mandible and jaw, temporo-mandibularjoint (TMJ), other areas of human skull and neck with sections of upper cervical spine, and lower extremities for use in diagnostic support. The device accomplishes this task by reconstructing a three dimensional matrix of the examined volume and producing two dimensional views of this volume, displaying both two and three dimensional images.
The NewTom 5G is a cone beam computed tomography x-ray imaging system that acquires of the head including the ear, nose and throat (ENT), dento-maxillofacial complex, teeth, mandible and jaw, temporo-mandibularjoint (TMJ), other areas of human skull and neck with sections of upper cervical spine, and lower extremities for use in diagnostic support. The device accomplishes this task by reconstructing a three dimensional matrix of the examined volume and producing two dimensional views of this volume, displaying both two and three dimensional images.
I am sorry, but the provided text does not contain information about acceptance criteria, device performance, or any studies conducted on the "NewTom 5G" device. The document is primarily a 510(k) clearance letter from the FDA, stating that the device is substantially equivalent to legally marketed predicate devices, and outlining regulatory requirements. It specifies the "Indications for Use" for the device but does not detail any performance studies, data provenance, expert qualifications, or ground truth methodologies.
Ask a specific question about this device
(132 days)
NEWTOM VGI AND NEWTOM 5G
NewTom VGi is a cone beam computed tomography x-ray imaging system that acquires a 360 degree rotational sequences of the head including the ENT, dento-maxillofacial complex, temporo-mandibular-joint (TMJ), other areas of human skull and neck with sections of upper cervical spine for use in diagnostic support. The device accomplishes this task by reconstructing a three dimensional matrix of the examined volume and producing two dimensional views of this volume, displaying both two and three dimensional images. The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals.
NewTom 5G is a cone beam computed tomography x-ray imaging system that acquires a 360 derree rotational sequences of the head including the ENT, dento-maxillofacial complex, temporo-mandibular-joint (TMJ), other areas of human skull and neck with sections of upper cervical spine for use in diagnostic support. The device accomplishes this task by reconstructing a three dimensional matrix of the examined volume and producing two dimensional views of this volume, displaying both two and three dimensional images. The device is operated and used by physicians, dentists, x-ray technologists and other legally qualified professionals.
NewTom 5G is especially designed for:
- dento-maxillo-facial complex imaging;
- teeth, mandible and jaw imaging for implant planning;
- temporal-mandibular joint (TMJ) imaging;
- ear, nose and throat (ENT) analysis;
- sections of upper cervical-spine imaging;
NewTom VGi and NewTom 5G are cone beam computed tomography x-ray imaging systems that acquire a 360 degree rotational sequences of the head. The devices reconstruct a three dimensional matrix of the examined volume and produce two dimensional views of this volume, displaying both two and three dimensional images.
Based on the provided documents for K130442, which describe the NewTom VGi and NewTom 5G computed tomography x-ray systems, the information requested to describe acceptance criteria and associated studies is not available.
These documents are a 510(k) clearance letter and Indications for Use statements. They specify the regulatory clearance of the devices for diagnostic support in various anatomical regions of the head and neck. They do not contain clinical study data, performance metrics, acceptance criteria, sample sizes, or details about ground truth establishment.
To answer your request, detailed performance data from the 510(k) submission itself would be needed, which is not part of these publicly available summary documents.
Therefore, I cannot provide the requested table and study details. If you have access to the full 510(k) submission, that would be the source for this information.
Ask a specific question about this device
(68 days)
NEWTOM 5G / NEWTOM 5G VERSION FP
The NewTom 5G Computed Tomography X-Ray System is a dedicated X-ray imaging device that acquires a 360-degree rotational X-ray sequence of images for use as diagnostic support in radiology of the dento-maxillo-facial complex and in the field of maxillofacial surgery.
The NewTom 5G accomplishes this task by reconstructing a three-dimensional matrix of the examined volume, producing two-dimensional views of this volume and displaying both two-dimensional images and three-dimensional renderings.
The NewTom 5G is a dedicated X-ray imaging device that acquires a 360-degree rotational X-ray sequence of images. It reconstructs a three-dimensional matrix of the examined volume and produces two-dimensional views of such volume, displaying both two- and three-dimensional images. The NewTom 5G can measure distances and thickness on two-dimensional images. Such images can be printed or exported on magnetic and optical media.
The NewTom 5G hardware, including a scanner unit (comprised of the X-ray source, flat panel detector and the motorized arm) and a motorized patient support, facilitates the acquisition of a full X-ray sequence by the device software. The NewTom 5G software runs on an x86 architecture based workstation. The NewTom 5G reconstructs a three-dimensional model of X-ray images similar to the three-dimensional model obtained using the parent NewTom VG Computed Tomography X-Ray System.
This document describes the NewTom 5G Computed Tomography X-Ray System. The provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than providing detailed acceptance criteria and a standalone study for a new AI/software component. Therefore, much of the requested information regarding AI device performance, sample sizes for test/training sets, expert consensus for ground truth, and MRMC studies is not available in the provided text.
The submission highlights minor technological differences between the NewTom 5G and its predicate device, the NewTom VG, primarily concerning patient positioning and image quality/reliability improvements through hardware and software modifications. The core imaging principle (360-degree rotational X-ray sequence to reconstruct 3D models for diagnostic support in dento-maxillofacial radiology) remains the same.
Here's the information that can be extracted or inferred from the provided text, along with a clear indication of what is not available:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Implied) | Reported Device Performance |
---|---|
Safety and Effectiveness | The NewTom 5G computed tomography X-ray system met all requirements, and functioned as intended and is therefore safe and effective for its intended use. (This is a general statement about meeting regulatory requirements rather than specific performance metrics.) |
Technological Equivalence | The NewTom 5G and the predicate NewTom VG are substantially equivalent. Both use a 360-degree rotational X-ray sequence to reconstruct 3D models and produce 2D views for diagnostic support in dento-maxillofacial radiology and maxillofacial surgery. Minor hardware/software changes for ergonomics and improved image quality/reliability do not affect overall safety or effectiveness. |
Electrical Safety, EMC/EMI | Electrical safety, EMC/EMI testing was performed. The system met all requirements. |
Verification and Validation | Verification and validation testing was performed to support hardware and software modifications. The system met all requirements and functioned as intended. |
2. Sample Sizes for Test Set and Data Provenance
- Sample Size for Test Set: Not specified. The document primarily focuses on demonstrating substantial equivalence through technological comparison and general safety/effectiveness testing, not a detailed clinical performance study with a specific test set for AI.
- Data Provenance: Not specified.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Not applicable/not specified. The provided text does not describe a clinical study involving experts establishing ground truth for a test set. This type of information is typically part of a performance study for AI/software, which is not the primary focus of this 510(k) summary.
- Qualifications of Experts: Not applicable/not specified.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable/not specified.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- MRMC Study Done? No. The document does not describe an MRMC study comparing human readers with and without AI assistance. This device is an imaging system, and while it produces images that humans interpret, it doesn't describe an AI-powered diagnostic aid that would typically be evaluated with an MRMC study in this context.
- Effect Size of Human Readers with/without AI: Not applicable, as no such study was described.
6. Standalone (Algorithm Only) Performance Study
- Standalone Study Done? No. This submission describes an imaging device, not a standalone algorithm. The "software" mentioned is for image acquisition and reconstruction, not autonomous diagnostic interpretation.
7. Type of Ground Truth Used
- Type of Ground Truth: Not applicable/not specified in the context of a performance study. The "ground truth" for this device's evaluation is primarily its ability to produce diagnostically acceptable images and function safely and effectively, demonstrating substantial equivalence to its predicate device. This is assessed through engineering tests, V&V, and comparison of technical characteristics, not by comparing its outputs against independent clinical ground truth like pathology for specific disease detection in a clinical study.
8. Sample Size for the Training Set
- Sample Size for Training Set: Not applicable/not specified. The device is not described as using machine learning models that require a "training set" in the conventional sense for AI clinical applications. Its software performs image reconstruction, not learned pattern recognition from a large dataset.
9. How the Ground Truth for the Training Set Was Established
- How Ground Truth for Training Set Was Established: Not applicable/not specified, as there is no mention of a training set for an AI algorithm.
Ask a specific question about this device
Page 1 of 1