Search Results
Found 3 results
510(k) Data Aggregation
(246 days)
Neurovascular Access System Family
The Neurovascular Access System Family is indicated for the introduction of interventional devices into the peripheral and neuro vasculature. It is not intended for use in coronary arteries.
The subject Neurovascular Access System Family consists of sterile, single-use catheters and accessories that are used together to facilitate the insertion and guidance of an appropriately sized interventional device into a selected vessel in the peripheral and neurovasculature. The Neurovascular Access System is comprised of the Neurovascular Access Catheter, Access Tool, Peel Away Introducer and Dilator. The Neurovascular Access System Family is offered in various lengths to accommodate physician preferences and anatomical variations. The distal end of the catheter is coated with hydrophilic coating to reduce friction during use and a radiopaque marker on the distal tip that is visible under fluoroscopy.
This document is a 510(k) Summary for a medical device called the "Neurovascular Access System Family." It details the device's characteristics, intended use, and compares it to legally marketed predicate devices to demonstrate substantial equivalence.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
Test Name | Objective | Test Method / Standard or Guidance | Acceptance Criteria (Implied by "Pass") | Reported Device Performance |
---|---|---|---|---|
Visual Surface Requirements | To demonstrate that the device meets the visual surface requirements. | ISO 10555-1 2013, Section 4.4 | Device meets visual surface requirements. | Pass |
Dimensional Verification | To demonstrate that the device meets the dimensional requirements. | ISO 10555-1 2013, Section 4.5 | Device meets dimensional requirements. | Pass |
Liquid Leakage Under Pressure | To demonstrate that the device passes the liquid leakage under pressure test. | ISO 10555-1 2013, Section 4.7.1, Annex C | Device passes liquid leakage test. | Pass |
Hub Aspiration Air Leakage | To demonstrate that the device passes the hub aspiration air leakage test. | ISO 10555-1 2013, Section 4.7.2, Annex D | Device passes hub aspiration test. | Pass |
Simulated Use and Usability | To demonstrate that the device passes testing specified in the simulated use test protocol. Simulated use testing includes usability assessment with multiple physicians. | FDA guidance: "Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions", April 2023 | Device passes simulated use and usability protocols. | Pass |
Flex Fatigue | To demonstrate that the device passes the flex fatigue test. | FDA guidance: "Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions", April 2023 | Device passes flex fatigue test. | Pass |
Torque Test | To demonstrate the torque strength of the device. | FDA guidance: "Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions" April 2023 | Device demonstrates adequate torque strength. | Pass |
Tip Deflection | To demonstrate that the tip deflection is comparable to the predicate device. | FDA Guidance: "Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions" April 2023 | Tip deflection is comparable to predicate. | Pass |
Device Removal Forces | To demonstrate that the removal forces are comparable to the predicate device. | FDA Guidance: "Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions" April 2023 | Removal forces are comparable to predicate. | Pass |
Peak Tensile Strength Testing | To demonstrate that the device passes the peak tensile strength testing including all bonds and joints. | ISO 10555-1 2013, Section 4.6, Annex B | Device passes peak tensile strength testing. | Pass |
Flow Rate | To demonstrate that the flow rate is comparable to the predicate device. | ISO 10555-1 2013: Section 4.9 | Flow rate is comparable to predicate. | Pass |
Corrosion Resistance | To demonstrate that the device has no visual evidence of corrosion. | ISO 10555-1 2013, Section 4.5, Annex A | No visual evidence of corrosion. | Pass |
Radiopacity | To demonstrate that the distal marker band is clearly visible under typical fluoroscopic imaging conditions. | ISO 10555-1 2013, Section 4.2; ASTM F640-12 | Distal marker band is clearly visible. | Pass |
Kink Resistance | To demonstrate that shaft kink resistance is comparable to the predicate device and clinically relevant for the intended anatomical locations for use. | FDA Guidance: "Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions" | Shaft kink resistance is comparable to predicate and clinically relevant. | Pass |
Particulates and Coating Integrity | To demonstrate the quantity and size of particles generated during simulated use are comparable to the predicates and reference devices. | FDA Guidance: "Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions" | Particulates and coating integrity are comparable to predicates. | Pass |
Dynamic Burst | To demonstrate that the device withstands dynamic burst strength. | ISO 10555-1 2103, Annex G | Device withstands dynamic burst strength. | Pass |
Static Burst | To demonstrate that the device passes static burst pressure as specified in the test protocol. | ISO 10555-1 2013, Section F | Device passes static burst pressure. | Pass |
Shelf Life | Repeated bench tests after accelerated aging to demonstrate that the device performance is maintained over the proposed shelf-life (6 months). | ASTM D4332-22, ASTM D4169-22, ASTM F1980-21, ASTM F1886 / F1866M-16, ASTM 2096-11 (2019), ASTM F88/F88M:21 | Device performance maintained over 6-month shelf-life. | Pass |
Biocompatibility Tests:
Test Name | Objective (Implied) | Test Method / Standard or Guidance | Acceptance Criteria (Implied by "Results") | Reported Device Performance |
---|---|---|---|---|
Cytotoxicity | Assess cell toxicity. | ISO 10993-5 | No reactivity observed. | Non-cytotoxic |
Sensitization | Assess potential for allergic reactions. | ISO10993-10 | No evidence of delayed dermal contact sensitization. | Non-sensitizing |
Intracutaneous Reactivity | Assess localized irritation. | ISO 10993-23 | Scores from test article extracts were acceptable ( |
Ask a specific question about this device
(136 days)
SelectFlex Neurovascular Access System Family
The SelectFlex Neurovascular Access System Family is indicated for the introduction of interventional devices into the peripheral and neurovasculature.
The subject SelectFlex Neurovascular Access System Family is a line extension to expand the previously cleared device (K211893) by offering a new SelectFlex III 064 Neurovascular Access Catheter with a diameter of 6F, a 6F dilator, an introducer sheath, and a 4.5F Access Tool that can be used with the catheter to access the desired anatomy. The 4.5F Access Tool is an accessory provided with the SelectFlex III 064 Neurovascular Access Catheter and is packaged in the same sterile pouch with the SelectFlex III 064 Neurovascular Access Catheter and accessories as part of the SelectFlex Neurovascular Access System Family.
The provided text is a 510(k) Summary for a medical device (SelectFlex Neurovascular Access System Family) and does not describe a study that proves the device meets specific acceptance criteria in the context of an AI/human-in-the-loop diagnostic system. Instead, it describes nonclinical performance testing (bench testing) and biocompatibility testing conducted to demonstrate substantial equivalence to a legally marketed predicate device.
Therefore, many of the requested items (e.g., sample sized for the test set in an AI study, number of experts for ground truth, MRMC study, standalone performance, training set details) are not applicable to this document.
However, I can extract the acceptance criteria (goals) and reported performance (results) from the nonclinical performance data section.
Here's the information based on the provided text:
1. A table of acceptance criteria and the reported device performance
For the SelectFlex Neurovascular Access System Family (including the SelectFlex III 064 Neurovascular Access Catheter, 6F Dilator, and 4.5F Access Tool where applicable):
Test Name | Acceptance Criteria (Goal) | Reported Device Performance (Result) |
---|---|---|
Visual Surface Requirements | To demonstrate that the device meets the visual surface requirements. | Pass: All samples met the predetermined acceptance criteria. |
Dimensional Verification | To demonstrate that the device meets the dimensional requirements. | Pass: All samples met the predetermined acceptance criteria. |
Liquid Leakage Under Pressure | To demonstrate that the device passes the liquid leakage under pressure test. | Pass: All samples met the predetermined acceptance criteria. |
Hub Aspiration Air Leakage | To demonstrate that the device passes the hub aspiration air leakage test. | Pass: All samples met the predetermined acceptance criteria. |
Simulated Use/Usability | To demonstrate that the device passes testing specified in the simulated use test protocol. Simulated use testing includes usability assessment with multiple physicians. | Pass: All samples met the predetermined acceptance criteria. |
Flex Fatigue | To demonstrate that the device passes the flex fatigue test. | Pass: All samples met the predetermined acceptance criteria. |
Inflation Fatigue | To demonstrate that the device passes the inflation fatigue test. | Pass: All samples met the predetermined acceptance criteria. |
Burst Volume | To demonstrate that the device passes the burst volume test – tested to 2x the inflation volume. | Pass: All samples met the predetermined acceptance criteria. |
Torque Test | To demonstrate the device's ability to rotate 720 degrees (2 full revolutions) at the proximal end. | Pass: All samples met the predetermined acceptance criteria. |
Tip Deflection | To demonstrate that the tip deflection is comparable to the predicate device. | Pass: All samples met the predetermined acceptance criteria. |
Device Removal in Support and Tracking Modes | To demonstrate that the forces in both support and tracking modes are comparable to the predicate device. | Pass: All samples met the predetermined acceptance criteria. |
Peak Tensile Testing | To demonstrate that the device passes the peak tensile strength testing including all bonds and joints. | Pass: All samples met the predetermined acceptance criteria. |
Flow Rate | To demonstrate that the flow rate is comparable to the predicate device. | Pass: All samples met the predetermined acceptance criteria. |
Corrosion Resistance | To demonstrate that the device has no visual evidence of corrosion. | Pass: All samples met the predetermined acceptance criteria. |
Radiopacity | To demonstrate that the marker band is positioned at the distal tip of the catheter and is clearly visible under typical fluoroscopic imaging conditions. | Pass: All samples met the predetermined acceptance criteria. |
Particulates, Coating Integrity, Lubricity, Durability | To demonstrate the quantity and size of particles generated during simulated use are comparable to the predicate and reference devices. | Pass: All samples met the predetermined acceptance criteria. |
Static Burst | To demonstrate that the device passes static burst as specified in the test protocol. | Pass: All samples met the predetermined acceptance criteria. |
Shelf Life | To demonstrate that the device performance is maintained over the proposed shelf-life (6 months). | Pass: All samples met the predetermined acceptance criteria. |
Biocompatibility Testing:
Test Name | Acceptance Criteria (Goal) | Reported Device Performance (Result) | Conclusion |
---|---|---|---|
Cytotoxicity | Absence of reactivity | No reactivity was observed with the test article at 24 and 48 hours. | Non-cytotoxic |
Sensitization | No evidence of sensitization | The test article extracts showed no evidence of delayed dermal contact sensitization in the guinea pig maximization test. | Non-sensitizing |
Intracutaneous Reactivity | No irritation | The scores from test article extracts were 0 from the saline extract and 0 from the sesame seed oil extract. | Non-irritant |
Acute Systemic Toxicity | No abnormal clinical signs/toxicity | No abnormal clinical signs indicative of toxicity were observed for 72 hours. All animals were alive at the end of 72 hours and body weight changes were within acceptable parameters. | Non-toxic |
Material Mediated Pyrogenicity | No significant temperature rise | No rabbit temperature rise ≥ 0.5 °C. | Non-pyrogenic |
Hemolysis - Direct Contact and Extract Method | Non-hemolytic | Blank corrected hemolytic index: 0.15, 0.13. | Non-hemolytic |
Complement Activation | Results within acceptable range | Results within acceptable range as compared to the controls. | Not a Sc5b-9 complement activator |
Thrombogenicity | No adverse effects/thrombus | No adverse effects or clinical signs during test period and no thrombus score ≥ 2 for either test or control device. | Non-thrombogenic |
2. Sample sized used for the test set and the data provenance
The document does not specify sample sizes for each particular test, only stating "All samples" or providing a conclusion based on the samples tested. The data provenance is "bench testing" and "biocompatibility testing," which are laboratory-based tests of the device itself, not clinical data or data from human subjects.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a medical device performance and biocompatibility study, not an AI diagnostic study relying on expert ground truth. However, for "Simulated Use/Usability," it states "usability assessment with multiple physicians," but specific numbers or qualifications are not provided.
4. Adjudication method for the test set
Not applicable.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. No AI component is mentioned.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. No AI component is mentioned.
7. The type of ground truth used
For nonclinical performance, the "ground truth" is defined by established engineering and material science standards (e.g., ISO 10555-1, FDA guidance documents, ASTM D4332, AAMI TIR42, ISO 10993 series for biocompatibility). The device's physical properties and interactions are measured against these objective criteria.
8. The sample size for the training set
Not applicable. There is no AI training set as this is a device performance study.
9. How the ground truth for the training set was established
Not applicable. There is no AI training set.
Ask a specific question about this device
(45 days)
SelectFlex Neurovascular Access System Family
The SelectFlex Neurovascular Access System Family is indicated for the introduction of interventional devices into the peripheral and neurovasculature.
The SelectFlex Neurovascular Access System Family consists of sterile, single-use intravascular catheters used to facilitate access to target vasculature during interventional procedures. The system is composed of a SelectFlex Neurovascular Access Catheter, a 3cc Mode Control Syringe, a 7Fr Peel Away Introducer, Luer Activated Valve and a Dilator (only for versions with a Long Hydrophilic coating to ease dermal entry). The SelectFlex Neurovascular Access Catheter comes in usable lengths of 95, 105 or 115cm. The SelectFlex Neurovascular Access Catheter has variable stiffness along its length and has a dual mode stiffness mechanism on the distal portion of the catheter that is activated by the user allowing the device to transition between tracking and support modes. The distal end of the SelectFlex Neurovascular Access Catheter comes in two options of a short hydrophilic coating length of 11.5cm or a long hydrophilic coating length of 30cm.
The provided FDA 510(k) summary for the SelectFlex Neurovascular Access System Family describes non-clinical performance data (bench testing) to demonstrate substantial equivalence to a predicate device, rather than a study involving human subjects or AI performance metrics. Therefore, many of the requested elements pertaining to AI studies, ground truth establishment, expert adjudication, and MRMC studies are not applicable or cannot be extracted from this document.
Here's the information that can be extracted and a clear indication where the requested information is not available in the document:
1. Table of Acceptance Criteria and Reported Device Performance
Test Description | Acceptance Criteria (Implied) | Reported Device Performance |
---|---|---|
Visual Surface Requirements (catheter and dilator) | All samples meet pre-determined acceptance criteria | PASS (All samples met the pre-determined acceptance criteria) |
Dimensional Verification (catheter and dilator) | All samples meet pre-determined acceptance criteria | PASS (All samples met the pre-determined acceptance criteria) |
Simulated Use/Usability | All samples meet pre-determined acceptance criteria | PASS (All samples met the pre-determined acceptance criteria) |
Flexion Fatigue | All samples meet pre-determined acceptance criteria | PASS (All samples met the pre-determined acceptance criteria) |
Inflation Fatigue | All samples meet pre-determined acceptance criteria | PASS (All samples met the pre-determined acceptance criteria) |
Burst Volume | All samples meet pre-determined acceptance criteria | PASS (All samples met the pre-determined acceptance criteria) |
ISO 80369-7: Small Bore Connectors for Hypodermic Applications | All samples meet pre-determined acceptance criteria | PASS (All samples met the pre-determined acceptance criteria) |
Particulate Count | All samples meet pre-determined acceptance criteria | PASS (All samples met the pre-determined acceptance criteria) |
Peak Tensile Testing (catheter and dilator) | All samples meet pre-determined acceptance criteria | PASS (All samples met the pre-determined acceptance criteria) |
Note: The document states "All samples met the pre-determined acceptance criteria" for each test, implying that the acceptance criteria were defined prior to testing, but the specific quantitative or qualitative criteria for each test (e.g., maximum allowable particulate count, specific tensile strength values, specific pass/fail conditions for visual inspection) are not explicitly detailed in this summary.
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample size for each bench test beyond "All samples." It states "Simulated use testing includes a usability assessment with multiple physicians" but does not give a number for the physicians or the number of simulated cases.
Data Provenance: The data is from non-clinical bench testing and simulated use conditions, not from human clinical data or retrospective/prospective studies on patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is not an AI performance study requiring ground truth established by experts. For the "Simulated Use/Usability" test, it mentions "multiple physicians," but their specific number or qualifications are not provided as they are assessing usability, not establishing ground truth on patient data.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is not an AI performance study requiring adjudication of expert readings.
5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was done. This document describes the substantial equivalence of a physical medical device (catheter system) based on bench testing, not an AI software.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This document does not describe an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable. For the non-clinical bench tests, "ground truth" would be objective measurements against engineering specifications and industry standards (e.g., ISO 10555-1: 2013, ISO 80369-7) rather than clinical ground truth types like pathology or expert consensus.
8. The sample size for the training set
Not applicable. This document does not describe an AI system that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This document does not describe an AI system.
Ask a specific question about this device
Page 1 of 1