NemoCast
NemoCast is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned dectives. It can also be applied during the treatment to inspect and analyze the progress of the treatment.
The use of the NemoCast requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
NemoCast is a software system used for the management of 3D scanned orthodontic models of the patients, orthodontic diagnosis by measuring, analyzing, inspecting and visualize 3D scanned orthodontic models, virtual planning of orthodontic treatments by simulating tooth movements, virtual placement of orthodontic brackets on the 3D models and design of orthodontic appliances based on 3D scanned orthodontic models, including transfer methods for indirect bonding of brackets. Output includes STL Models (also called dental casts) for thermoforming aligners, STL files for direct printing aligners and Indirect Bonding Transfer Trays (also called orthodontic bracket placement trays). The device has no patient contact.
This document (K232549) is a 510(k) premarket notification for the device "NemoCast," an orthodontic software. It establishes substantial equivalence to existing predicate devices, particularly 3Shape Ortho System™ (K152086) and a previous version of NemoCast (K193003).
The key takeaway is that this is primarily a software validation and substantial equivalency claim, rather than a study proving new clinical performance. The manufacturer is demonstrating that their current software performs similarly to a previously cleared version and a predicate device.
Here's an analysis of the provided information regarding acceptance criteria and performance study:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly present a "table of acceptance criteria" with corresponding "reported device performance" in the typical format of a clinical study summary. Instead, the acceptance criteria are implicitly met by demonstrating substantial equivalence to predicate devices. The "reported device performance" is the functionality of NemoCast itself, and the "study" is the comparison against the predicate devices.
The "Comparison of Intended Use and Technological Characteristics with the reference Device" table (pages 7-8) serves as the primary evidence of meeting "acceptance criteria" for substantial equivalence. It lists various features and functions, and for each, it aims to show "None" under "Differences" or a difference that does not affect safety and effectiveness.
Here's a condensed version of how that table functions as a de facto "acceptance criteria" and "performance report":
Feature | Acceptance Criteria (Implied by Predicate) | NemoCast Performance (Reported) |
---|---|---|
Product Code | PNN, LLZ | PNN, LLZ |
Common Name | Orthodontic Software | Orthodontic Software |
Classification Name | Orthodontic Plastic Bracket | Orthodontic Plastic Bracket |
Regulation Number | 21 CFR 872.5470 | 21 CFR 872.5470 |
Supported anatomic areas | Maxilla and Mandible | Maxilla and Mandible |
Use by dental professionals in orthodontic treatment planning | Yes (NemoCast K193003: "only before treatment") | Yes (NemoCast K232549: "before, during, after treatment") |
Difference noted: broadened scope compared to reference device, but aligned with primary predicate. | ||
Management of patients and models | Yes | Yes |
Inspection, measurement and analysis of orthodontic models | Yes | Yes |
Treatment simulation | Yes | Yes |
Virtual appliance preparation, handling and export | Yes | Yes |
Provide digital file and device output | Yes (STL files for dental casts) | Yes (STL files for dental casts, and additionally indirect Bonding Transfer Media) |
Supported PC formats | Windows | Windows |
Creating, editing, deleting and copying patient data | Yes | Yes |
Creating, editing, deleting and copying case data | Yes | Yes |
Surface scan from intra-oral scanner | Yes | Yes |
Surface scan from STL, PLY, OBJ file | Yes (Predicate: STL only) | Yes (STL, PLY, OBJ) |
Difference noted: Broader import formats, stated not to affect security/safety. | ||
CT image data | DICOM | DICOM |
2D overlay | PNG, JPG, BMP | PNG, JPG, BMP |
Aligning surface scan and CT image | Yes | Yes |
Aligning cephalometric images | Yes | Yes |
Alignment of surface scan with 2D overlays | Yes | Yes |
Ability to check/adjust DICOM visibility | Yes | Yes |
DICOM scan segmentation | Yes | Yes |
Occlusal Orientation | Yes (Reference Device) / No (Predicate Device) | Yes |
Difference noted: Feature present in subject and reference, not in primary predicate. | ||
Segmenting teeth roots | Yes (Reference Device) / No (Predicate Device) | Yes |
Difference noted: Feature present in subject and reference, not in primary predicate. | ||
DICOM orientation | Yes | Yes |
2D measurement toolbox | Yes | Yes |
3D measurement toolbox | Yes | Yes |
Arch shape analysis | Yes | Yes |
Wire length analysis | Yes | Yes |
Tooth width analysis | Yes | Yes |
Bolton analysis | Yes (Predicate Device) / No (Reference Device) | Yes |
Space analysis | Yes | Yes |
Overjet/overbite analysis | Yes (Predicate Device) / No (Reference Device) | Yes |
Occlusion map | Yes | Yes |
Treatment analysis and report generation | Yes | Yes |
2D & 3D simulation | Yes | Yes |
Orthodontic appliance search | Yes | Yes |
Orthodontic appliance virtual preparation | Yes | Yes |
Orthodontic appliance design | Yes | Yes |
Orthodontic appliance export | Yes | Yes |
Virtual articulator | Yes | Yes |
Intended User | Dental Professionals | Dental Professionals |
Intended Patient Population | Patients requiring Orthodontic Treatment (Predicate) / Adults requiring Orthodontic Treatment (Reference) | Patients requiring Orthodontic Treatment |
Difference noted: Aligned with primary predicate. |
2. Sample size used for the test set and the data provenance
The document states: "The performance testing remains unchanged from the company's own reference device submission, NemoCast K193003. The performance testing for the subject device is being leveraged from the company's own reference device including: design verification and validation testing."
This implies that the sample size and data provenance for the current 510(k) submission are not new. They are relying on previous testing. The document does not explicitly state the sample size (number of cases/patients) or the data provenance (e.g., country of origin, retrospective/prospective) for the test set. This information would typically be found in the original K193003 submission or internal validation reports, which are not detailed here.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not provided in the document. As this is a software substantial equivalence submission leveraging previous testing, details about the ground truth establishment for the test set (number and qualifications of experts) are not specified here.
4. Adjudication method for the test set
This information is not provided in the document.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a MRMC comparative effectiveness study demonstrating improved human reader performance with AI assistance was not done for this submission. The "NemoCast" device is described as a "medical front-end device providing tools for management... analysis, treatment simulation, and virtual appliance design." It is a planning and design software, not an AI-assisted diagnostic tool that would typically involve a human-in-the-loop MRMC study for assessing reader improvement.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document mentions "The software is thoroughly tested in accordance with a documented test plan. This test plan is derived from the specifications and ensures that all controls and features are functioning properly. The software is validated together with end-users." This general statement indicates functional and validation testing that would assess standalone performance (software functioning as intended). However, specific quantifiable metrics of "algorithm-only" performance (e.g., accuracy of automatic measurements) are not reported in this summary. The focus is on the functional equivalence of the tools provided within the software.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The document does not explicitly state the type of "ground truth" used for the underlying validation of the software's measurements or simulations. Given the nature of orthodontic planning software, ground truth for features like measurements or segmentations would typically be established through:
- Expert consensus: Manual measurements or segmentations performed by experienced orthodontists on 3D models.
- Physical measurements/phantom data: Validation against known physical dimensions or phantom models.
However, these details are not provided in this 510(k) summary.
8. The sample size for the training set
This information is not provided. As this is a software product, not necessarily one relying on a large deep learning model needing a defined "training set" in the context of AI development for image interpretation, this detail might not be applicable or explicitly stated. If there are features utilizing machine learning (which is not explicitly detailed but possible for functions like segmentation), the training set details are not included. The submission is focused on demonstrating functional equivalence.
9. How the ground truth for the training set was established
This information is not provided and is likely not applicable in the context of this 510(k) submission, as it focuses on demonstrating substantial equivalence of a software tool rather than a novel AI algorithm with a distinct training phase. If machine learning components exist, the process for establishing their "training ground truth" is not disclosed here.
In summary, the provided text details a 510(k) submission for substantial equivalence. It does not contain the detailed clinical study information (like sample sizes, expert qualifications, adjudication methods, or MRMC study results) that would typically accompany a submission for a novel diagnostic AI device where independent performance validation is the primary focus. Instead, it relies on demonstrating that the updated software (NemoCast K232549) is functionally equivalent to its predicate devices (NemoCast K193003, 3Shape Ortho System™ K152086) and that any differences do not impact safety or effectiveness.
Ask a specific question about this device
(221 days)
NemoCast
NemoCast is intended for supporting the diagnostic and treatment plan processes for orthodontic procedures related to minor anterior tooth movements; it provides tools for management of orthodontic models, systemic inspection, detailed analysis, treatment simulation and virtual appliance design options (export of models) based on 3D models of the patient's dentition before the start of an orthodonic treatment. It can also be applied to evaluate if the outcome is consistent with the planned/desired treatment objectives.
NemoCast requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
Not Found
I apologize, but the provided text from the FDA 510(k) clearance letter for NemoCast does not contain the information necessary to describe the acceptance criteria and the study that proves the device meets the acceptance criteria in the level of detail requested.
The letter is a formal notification of substantial equivalence, and while it references the device and its indications for use, it does not include:
- A table of acceptance criteria and reported device performance.
- Sample sizes used for test sets or training sets.
- Data provenance (country, retrospective/prospective).
- Details about experts for ground truth establishment (number, qualifications, adjudication methods).
- Information about MRMC studies or effect sizes.
- Standalone algorithm performance data.
- Specifics on ground truth types or how it was established.
The document primarily focuses on the regulatory aspects of the clearance (e.g., product code, regulation number, substantial equivalence determination, general controls, and compliance requirements).
To answer your request thoroughly, I would need access to the full 510(k) submission, which typically includes detailed validation studies, performance data, and methodology. This information is not present in the provided FDA clearance letter.