Search Results
Found 1 results
510(k) Data Aggregation
(105 days)
NaviCross 0.018
The NaviCross 0.018" is intended to guide wire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline or diagnostic contrast agents.
The NaviCross 0.018" is a single use, ethylene oxide sterilized device that is intended to guide and support a guide wire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline or diagnostic contrast agents.
NaviCross 0.018" features a three-layer construction, which consists of a stainless steel mesh braid sandwiched between an outer layer of polyamide and an inner layer of polytetrafluoroethylene. It has a distal tip that is comprised of a polyamide. It also has a hydrophilic coating over the distal 40 cm of the catheter.
The device is offered in effective lengths of 65, 90, 135 and 150 cm.
The provided text describes a 510(k) submission for the NaviCross 0.018" catheter. It details the device's technical specifications, comparison to predicate and reference devices, and non-clinical performance testing. However, it explicitly states that no clinical tests were performed or included in this 510(k) submission.
Therefore, I cannot provide information regarding acceptance criteria, device performance, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone algorithm performance, or ground truth establishment based on the provided text, as these typically pertain to clinical studies.
The document focuses on demonstrating substantial equivalence through non-clinical performance testing and biocompatibility testing.
Here's a breakdown of what is available in the document regarding non-clinical performance:
1. A table of acceptance criteria and the reported device performance:
The document lists various performance tests conducted. While it states that "Performance testing met the predetermined acceptance criteria and is acceptable for clinical use throughout its shelf life," the specific numerical acceptance criteria and the exact reported performance values for each test are not detailed in the provided text. The table (Table 5.4) only lists the Test Items.
Test Item | Reported Device Performance against Acceptance Criteria |
---|---|
Radio-detectability | Met predetermined acceptance criteria. |
Surface | Met predetermined acceptance criteria. |
Peak tensile force | Met predetermined acceptance criteria. |
Freedom from leakage | Met predetermined acceptance criteria. |
Power injection | Met predetermined acceptance criteria. |
Distal tip | Met predetermined acceptance criteria. |
Fluid leakage (Hub) | Met predetermined acceptance criteria. |
Sub-atmospheric pressure air leakage (Hub) | Met predetermined acceptance criteria. |
Stress cracking (Hub) | Met predetermined acceptance criteria. |
Resistance to separation from axial load (Hub) | Met predetermined acceptance criteria. |
Resistance to separation from unscrewing (Hub) | Met predetermined acceptance criteria. |
Resistance to overriding (Hub) | Met predetermined acceptance criteria. |
Particulate evaluation | Met predetermined acceptance criteria. |
Coating integrity | Met predetermined acceptance criteria. |
Torque strength | Met predetermined acceptance criteria. |
Distal tip strength | Met predetermined acceptance criteria. |
Flow rate | Met predetermined acceptance criteria. |
Product dimension | Met predetermined acceptance criteria. |
Bending stiffness of catheter shaft | Met predetermined acceptance criteria. |
Flexibility and kink test | Met predetermined acceptance criteria. |
Torque transmission property | Met predetermined acceptance criteria. |
Interior sliding characteristics | Met predetermined acceptance criteria. |
Exterior sliding characteristics | Met predetermined acceptance criteria. |
Wire support characteristics | Met predetermined acceptance criteria. |
Simulated Use - Reachability | Met predetermined acceptance criteria. |
Simulated Use - Usability Test | Met predetermined acceptance criteria. |
Biocompatibility Tests | |
Cytotoxicity (Non-aged, accelerated-aged) | Device is biocompatible. |
Sensitization | Device is biocompatible. |
Intracutaneous Reactivity | Device is biocompatible. |
Acute Systemic Toxicity | Device is biocompatible. |
Pyrogenicity | Device is biocompatible. |
Hemolysis (Non-aged, accelerated-aged) | Device is biocompatible. |
Thrombogenicity | Device is biocompatible. |
Complement Activation (Immunology) | Device is biocompatible. |
Physicochemical Profile (Non-aged, accelerated-aged) | Device is biocompatible. |
2. Sample sized used for the test set and the data provenance:
- Sample Size: The document does not specify exact sample sizes for each non-clinical test. It states that tests were performed on "non-aged and accelerated aged samples."
- Data Provenance: The device manufacturer and sterilization facility are the Ashitaka Factory of Terumo Corporation in Japan (Fujinomiya, Shizuoka 418-0015, Japan). The testing was conducted as part of their 510(k) submission for the US market. The testing is pre-market, laboratory-based performance and biocompatibility testing, not human data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This is not applicable as the testing described is non-clinical performance and biocompatibility testing, not clinical data requiring expert ground truth or interpretation of images/patient outcomes.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
Not applicable for non-clinical performance testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. No clinical studies, especially MRMC studies comparing human performance with and without AI assistance, were performed or referenced in this document.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This device is a medical catheter, not an AI algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
For the non-clinical tests, the "ground truth" or reference for acceptance is established by predefined engineering and material science standards (e.g., ISO standards, internal specifications for dimensions, strength, leakage, etc.). It's based on physical and chemical measurements rather than clinical ground truth types.
8. The sample size for the training set:
Not applicable. This device is a physical medical device, not an AI algorithm requiring a training set.
9. How the ground truth for the training set was established:
Not applicable. No training set for an algorithm is mentioned.
In summary, the provided document details non-clinical performance and biocompatibility testing for a medical device (catheter) to establish substantial equivalence. It explicitly states that no clinical tests were included in this 510(k) submission. Therefore, most of the questions relating to clinical study design, expert involvement, and AI performance are not applicable to the information given.
Ask a specific question about this device
Page 1 of 1