K Number
K173799
Device Name
NaviCross 0.018
Manufacturer
Date Cleared
2018-03-29

(105 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NaviCross 0.018" is intended to guide wire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline or diagnostic contrast agents.

Device Description

The NaviCross 0.018" is a single use, ethylene oxide sterilized device that is intended to guide and support a guide wire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline or diagnostic contrast agents.

NaviCross 0.018" features a three-layer construction, which consists of a stainless steel mesh braid sandwiched between an outer layer of polyamide and an inner layer of polytetrafluoroethylene. It has a distal tip that is comprised of a polyamide. It also has a hydrophilic coating over the distal 40 cm of the catheter.

The device is offered in effective lengths of 65, 90, 135 and 150 cm.

AI/ML Overview

The provided text describes a 510(k) submission for the NaviCross 0.018" catheter. It details the device's technical specifications, comparison to predicate and reference devices, and non-clinical performance testing. However, it explicitly states that no clinical tests were performed or included in this 510(k) submission.

Therefore, I cannot provide information regarding acceptance criteria, device performance, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone algorithm performance, or ground truth establishment based on the provided text, as these typically pertain to clinical studies.

The document focuses on demonstrating substantial equivalence through non-clinical performance testing and biocompatibility testing.

Here's a breakdown of what is available in the document regarding non-clinical performance:

1. A table of acceptance criteria and the reported device performance:

The document lists various performance tests conducted. While it states that "Performance testing met the predetermined acceptance criteria and is acceptable for clinical use throughout its shelf life," the specific numerical acceptance criteria and the exact reported performance values for each test are not detailed in the provided text. The table (Table 5.4) only lists the Test Items.

Test ItemReported Device Performance against Acceptance Criteria
Radio-detectabilityMet predetermined acceptance criteria.
SurfaceMet predetermined acceptance criteria.
Peak tensile forceMet predetermined acceptance criteria.
Freedom from leakageMet predetermined acceptance criteria.
Power injectionMet predetermined acceptance criteria.
Distal tipMet predetermined acceptance criteria.
Fluid leakage (Hub)Met predetermined acceptance criteria.
Sub-atmospheric pressure air leakage (Hub)Met predetermined acceptance criteria.
Stress cracking (Hub)Met predetermined acceptance criteria.
Resistance to separation from axial load (Hub)Met predetermined acceptance criteria.
Resistance to separation from unscrewing (Hub)Met predetermined acceptance criteria.
Resistance to overriding (Hub)Met predetermined acceptance criteria.
Particulate evaluationMet predetermined acceptance criteria.
Coating integrityMet predetermined acceptance criteria.
Torque strengthMet predetermined acceptance criteria.
Distal tip strengthMet predetermined acceptance criteria.
Flow rateMet predetermined acceptance criteria.
Product dimensionMet predetermined acceptance criteria.
Bending stiffness of catheter shaftMet predetermined acceptance criteria.
Flexibility and kink testMet predetermined acceptance criteria.
Torque transmission propertyMet predetermined acceptance criteria.
Interior sliding characteristicsMet predetermined acceptance criteria.
Exterior sliding characteristicsMet predetermined acceptance criteria.
Wire support characteristicsMet predetermined acceptance criteria.
Simulated Use - ReachabilityMet predetermined acceptance criteria.
Simulated Use - Usability TestMet predetermined acceptance criteria.
Biocompatibility Tests
Cytotoxicity (Non-aged, accelerated-aged)Device is biocompatible.
SensitizationDevice is biocompatible.
Intracutaneous ReactivityDevice is biocompatible.
Acute Systemic ToxicityDevice is biocompatible.
PyrogenicityDevice is biocompatible.
Hemolysis (Non-aged, accelerated-aged)Device is biocompatible.
ThrombogenicityDevice is biocompatible.
Complement Activation (Immunology)Device is biocompatible.
Physicochemical Profile (Non-aged, accelerated-aged)Device is biocompatible.

2. Sample sized used for the test set and the data provenance:

  • Sample Size: The document does not specify exact sample sizes for each non-clinical test. It states that tests were performed on "non-aged and accelerated aged samples."
  • Data Provenance: The device manufacturer and sterilization facility are the Ashitaka Factory of Terumo Corporation in Japan (Fujinomiya, Shizuoka 418-0015, Japan). The testing was conducted as part of their 510(k) submission for the US market. The testing is pre-market, laboratory-based performance and biocompatibility testing, not human data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This is not applicable as the testing described is non-clinical performance and biocompatibility testing, not clinical data requiring expert ground truth or interpretation of images/patient outcomes.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable for non-clinical performance testing.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. No clinical studies, especially MRMC studies comparing human performance with and without AI assistance, were performed or referenced in this document.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a medical catheter, not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

For the non-clinical tests, the "ground truth" or reference for acceptance is established by predefined engineering and material science standards (e.g., ISO standards, internal specifications for dimensions, strength, leakage, etc.). It's based on physical and chemical measurements rather than clinical ground truth types.

8. The sample size for the training set:

Not applicable. This device is a physical medical device, not an AI algorithm requiring a training set.

9. How the ground truth for the training set was established:

Not applicable. No training set for an algorithm is mentioned.

In summary, the provided document details non-clinical performance and biocompatibility testing for a medical device (catheter) to establish substantial equivalence. It explicitly states that no clinical tests were included in this 510(k) submission. Therefore, most of the questions relating to clinical study design, expert involvement, and AI performance are not applicable to the information given.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

March 29, 2018

Terumo Medical Corporation Ms. Yuko Watanabe Sr. Regulatory Affairs Specialist 265 Davidson Ave., Suite 320 Somerset, New Jersey 08873

Re: K173799

Trade/Device Name: NaviCross 0.018" Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: March 16, 2018 Received: March 16, 2018

Dear Ms. Watanabe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Ms. Yuko Watanabe

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

M.A. Willhemen

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173799

Device Name NaviCross 0.018"

Indications for Use (Describe)

The NaviCross 0.018" is intended to guide wire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline or diagnostic contrast agents.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY

A. SUBMITTER INFORMATION (807.92(a)(1))

Prepared by:Yuko Watanabe
Senior Regulatory Affairs Specialist
Terumo Medical Corporation
Tel. (732) 302-4900
Fax (410) 398-6079

Prepared for: Owner/Operator

Terumo Corporation 44-1, 2-Chome, Hatagaya Shibuya-Ku, Tokyo Japan 151-0072 Owner/Operator Number: 801 002 6

Manufacturer and Sterilization Facility (Applicant)

Ashitaka Factory of Terumo Corporation 150 Maimaigi-cho Fujinomiya, Shizuoka 418-0015, Japan Registration Number: 968 183 4

Contact Person: Yuko Watanabe Senior Regulatory Affairs Specialist Terumo Medical Corporation 265 Davidson Avenue, Suite 320 Somerset, NJ 08873 Tel. (732) 302-4900 Fax (410) 398-6079 E-mail: yuko.watanabe@terumomedical.com

December 13th, 2017 Date prepared:

{4}------------------------------------------------

B. DEVICE NAME (807.92(a)(2))

Proprietary Name:NaviCross 0.018"
Common Name:Support Catheter
Classification Name:Percutaneous Catheter
Classification Panel:Cardiovascular
Regulation:21 CFR 870.1250
Product Code:DQY
Classification:Class II

C. PREDICATE DEVICE (807.92(a)(3))

The legally marketed device to which substantial equivalence is claimed is:

Predicate Device: K110540 – TERUMO SUPPORT CATHETER, manufactured by Ashitaka Factory of Terumo Corporation.

Reference Devices:

    1. K160884: CXI 0.018" support catheter, Cook, Inc.
    1. K033678: Quick-Cross 0.018" support catheter, Spectranetics, Inc.
    1. K122394: Rubicon 0.018" support catheter, Boston Scientific Corp.

D. REASON FOR 510(k) SUBMISSION

This traditional 510(k) is being submitted for the NaviCross 0.018", which will be a line extension of the currently marketed NaviCross 0.035" (K110540, cleared under the name "Terumo Support Catheter"), for the purposes of establishing substantial equivalence to a legally marketed predicate device.

E. DEVICE DESCRIPTION (807.92(a)(4))

Principle of Operation Technology

The NaviCross 0.018" submitted in this 510(k) and its predicate (K110540) are operated by a manual process.

{5}------------------------------------------------

Design/Construction

The NaviCross 0.018" is a single use, ethylene oxide sterilized device that is intended to guide and support a guide wire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline or diagnostic contrast agents.

NaviCross 0.018" features a three-layer construction, which consists of a stainless steel mesh braid sandwiched between an outer layer of polyamide and an inner layer of polytetrafluoroethylene. It has a distal tip that is comprised of a polyamide. It also has a hydrophilic coating over the distal 40 cm of the catheter.

The device is offered in effective lengths of 65, 90, 135 and 150 cm. French size and shaft inner diameter are as follows:

French SizeShaft Inner Diameter (mm)
2.6Fr$0.55 \pm 0.02$

Materials

The materials for the NaviCross 0.018" are provided in the table below.

No.Name of ComponentRaw material
1*ShaftOuter layerPolyamide
Pigment
2Braid†Stainless steel
3*CatheterInner layerPolytetrafluoroethylene
4*Distal tip†Polyamide
Pigment
5Radiopaque markerTungsten
6*Hydrophilic polymer coatingDimethyl acrylamide –glycidyl methacrylate copolymer
7Quick-drying glueCyanoacrylate
8*HubPolyamide
9Anti-kink protectorPolyester elastomer
Pigment

Table 5.1: List of Materials

*Blood contacting material.

*Distal-tip is not braided.

{6}------------------------------------------------

Specifications

The specifications for the NaviCross 0.018" are provided in the table below.

PartSpecification
Catheter Size2.6 Fr.
Catheter ID/OD$0.55 \pm 0.02$ mm/ $0.85 \pm 0.03$ mm
Catheter Effective Lengths*65 $\pm$ 1.5 cm, 90 $\pm$ 1.5 cm,135 $\pm$ 1.5 cm, 150 $\pm$ 1.5 cm
Hydrophilic Coating LengthDistal 40 $\pm$ 3cm
Maximum guidewire outer diameter0.018"

Table 5.2: NaviCross 0.018" Specifications

*The length from the proximal anti-kink protector to the catheter distal tip.

F. INDICATIONS FOR USE (807.92(a)(5))

The NaviCross 0.018" is intended to guide and support a guide wire during access of the peripheral vasculature, allow for wire exchanges and provide a conduit for the delivery of saline or diagnostic contrast agents.

The indications for use are equivalent to the predicate (K110540).

G. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6))

The NaviCross 0.018", the subject of this Traditional 510(k), is substantially equivalent in its intended use, technology/principle of operation, materials, and performance to the predicate, K110540 - TERUMO SUPPORT CATHETER, manufactured by Ashitaka Factory of Terumo Corporation.

In addition to the above-listed primary predicate, Terumo has identified the following reference devices. These are market leading devices with the same intended use and basic design as the subject device. Because these devices are frequently used in clinical practice, Terumo felt it was appropriate to use them as references when setting the acceptance criteria for NaviCross 0.018" performance testing.

    1. Cook Inc. CXI 0.018" support catheter (K160884)
    1. Spectranetics, Inc. Quick-Cross 0.018" support catheter (K033678)
    1. Boston Scientific Corp. Rubicon 0.018" support catheter (K122394)

{7}------------------------------------------------

A comparison of the technological characteristics is summarized in the table below.

{8}------------------------------------------------

Terumo Corporation
Traditional 510(k) – NaviCross 0.01
Section 5 510(k) Summary

DeviceCharacteristicSubject Device:NaviCross 0.018"Predicate Device:TERUMO SUPPORTCATHETER (NaviCross0.035") (K110540)Reference Device #1:CXI 0.018" supportcatheter(K160884)Reference Device #2:Quick-Cross 0.018"support catheter(K033678)Reference Device #3:Rubicon 0.018"support catheter(K122394)
ManufacturerAshitaka Factory ofTerumo CorporationSameCook Inc.Spectranetics, Inc.Boston Scientific Corp.
Intended UseIndications forUseIt is intended to guide andsupport a guide wireduring access of theperipheral vasculature,allow for wire exchangesand provide a conduit forthe delivery of saline ordiagnostic contrast agents.It is intended to guide andsupport a guide wireduring access of thevasculature, allow forwire exchanges andprovide a conduit for thedelivery of saline ordiagnostic contrast agents.It is intended for use insmall vessel orsuperselectiveanatomy for diagnosticand interventionalprocedures, includingperipheral use.It is designed foruse in the vascularsystem. The cathetersare intended to support aguidewire during accessof the vasculature, allowfor exchange ofguidewires, and providea conduit for thedelivery of salinesolutions or diagnosticcontrast agents.It is intended tofacilitate placement andsupport of guidewiresand other Interventionaldevices within theperipheral vasculatureand to allow forexchange of guidewires,and provide a conduitfor the delivery of salineor contrast solutions.
OperationPrincipleManualSameSameSameSame
Design/ConstructionThree-layer constructioncatheter shaft withhydrophilic coating, distaltip and hubSameMulti layers withhydrophilic coating,distal tip and hubOne layer withhydrophilic coating,distal tip and hubMulti layers withhydrophilic coating,distal tip and hub
DeviceCharacteristicSubject Device:NaviCross 0.018"Predicate Device:TERUMO SUPPORTCATHETER (NaviCross0.035") (K110540)Reference Device #1:CXI 0.018" supportcatheter(K160884)Reference Device #2:Quick-Cross 0.018"support catheter(K033678)Reference Device #3:Rubicon 0.018"support catheter(K122394)
Catheter shaftOuter layer*:Polyamide w/pigmentCatheter shaftOuter layer*:PolyesterInformation notpublicly available.Information not publiclyavailable.Information not publiclyavailable.
Mesh braid: StainlesssteelMesh braid: Stainlesssteel
Inner layer*:PolytetrafluoroethyleneInner layer*:Polyesterelastomer/tungsten
MaterialsHydrophilic coating*:Dimethyl acrylamide-glycidyl methacrylatecopolymerThree radiopaquemarkers: TungstenHydrophilic coating*:Dimethyl acrylamide-glycidyl methacrylatecopolymerThree radiopaquemarkers*: Platinumalloy
Distal Tip*:Polyamide w/pigmentHub*: PolyamideAnti-kink protector:Polyester elastomerw/ pigmentAdhesive:Cyanoacrylate*: blood contactingmaterialHub*: PolyamideAnti-kink protector:Polyester elastomerw/ pigmentAdhesive:Cyanoacrylate*: blood contactingmaterial
DeviceCharacteristicSubject Device:NaviCross 0.018"Predicate Device:TERUMO SUPPORTCATHETER (NaviCross0.035") (K110540)Reference Device #1:CXI 0.018" supportcatheter(K160884)Reference Device #2:Quick-Cross 0.018"support catheter(K033678)Reference Device #3:Rubicon 0.018"support catheter(K122394)
PackageIndividual package onwhich the productlabel and the peel-offlabels are attached 1 unit per packageSameSameSameSame
SpecificationsEffective lengths: 65,90, 135 and 150 cm French size: 2.6Fr(0.85 mm) O.D.: 0.85 mm I.D.: 0.55 mm Maximum guidewireouter diameter: 0.018" Distal tip shape:straight/angled Maximum injectionpressure: 300 psiEffective lengths: 65,90, 135 and 150 cm French size: 4Fr(1.39 mm) O.D.: 1.39 mm I.D.:1.05 mm Maximum guidewireouter diameter: 0.035" Distal tip shape:straight/angled Maximum injectionpressure: 750 psi90, 150 cm 2.6 Fr O.D.: 0.87 mm I.D.: unknown Same (0.018") Same(straight/angled) 1200 psi90, 135 and 150 cm 3.4 Fr O.D.: 1.12 mm(Proximal shaft) I.D.: unknown Same (0.018") Straight only Same (300 psi)90, 135 and 150 cm 4.0 Fr O.D.: 1.35 mm(Proximal shaft) I.D.: unknown Same (0.018") Straight only Same (300 psi)
SterilizationEthylene oxideSameSameRadiationSame
Shelf life36 monthsSameUnknownUnknownUnknown

{9}------------------------------------------------

Terumo Corporation
Traditional 510(k) — NaviCross 0.01
Section 5 510(k) Summary

{10}------------------------------------------------

Terumo Corporation
Traditional 510(k) — NaviCross 0.01
Section 5 510(k) Summary

Page 8 of 11

{11}------------------------------------------------

H. NON CLINICAL TESTS (807.92(b)(1))

Performance Testing

Performance testing was conducted to ensure the safety and effectiveness of the NaviCross 0.018" throughout the shelf life, verify conformity to the applicable external and internal standards, and demonstrate substantial equivalence to the predicate device. With the exception of the Radio-detectability test, the following performance tests were performed on non-aged and accelerated aged samples. The following table provides a list of performance tests that were performed on the NaviCross 0.018".

Test Item
Radio-detectability
Surface
Peak tensile force
Freedom from leakage
Power injection
Distal tip
Fluid leakage (Hub)
Sub-atmospheric pressure air leakage (Hub)
Stress cracking (Hub)
Resistance to separation from axial load (Hub)
Resistance to separation from unscrewing (Hub)
Resistance to overriding (Hub)
Particulate evaluation
Coating integrity
Torque strength
Distal tip strength
Flow rate
Product dimension
Bending stiffness of catheter shaft
Flexibility and kink test
Torque transmission property
Interior sliding characteristics
Exterior sliding characteristics
Wire support characteristics
Simulated Use - Reachability
Simulated Use - Usability Test
Table 5.4: Summary of Performance Testing
-------------------------------------------------

1 Only non-aged sample was tested since the amount of contrast media contained in the product would not change over time.

{12}------------------------------------------------

Performance testing met the predetermined acceptance criteria and is acceptable for clinical use throughout its shelf life.

Biocompatibility

In accordance with ISO 10993-1, the NaviCross 0.018" is classified as: Externally Communicating Device, Circulating Blood, Limited Contact (<24 hours). The finished device's patient contacting parts were tested in accordance with the tests recommended in the FDA Guidance for Industry and Food and Drug Administration Staff - Use of International Standard ISO-10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." Screening tests were performed on accelerated aged devices to show that the biocompatibility is maintained throughout the shelf life of the product. The table below provides a list of biocompatibility tests conducted on the NaviCross 0.018".

Non-aged, sterile, whole device
Cytotoxicity
Sensitization
Intracutaneous Reactivity
Acute Systemic Toxicity
Pyrogenicity
Hemolysis
Thrombogenicity
Complement Activation (Immunology)
Physicochemical Profile (Physicochemical and FT-IR)
Accelerated-aged (3 years), sterile, whole device
Cytotoxicity
Hemolysis
Physicochemical Profile (Physicochemical and FT-IR)

Table 5.5: Summary of ISO 10993 Biocompatibility Testing

Results of the testing demonstrate that the device is biocompatible throughout the shelf life of the product.

{13}------------------------------------------------

Sterilization

The sterility of the device is assured using a sterilization method validated in accordance with ISO 11135:2014, Sterilization of Health Care Products - Ethylene Oxide – Requirements for the development, validation and routine control of a sterilization process for medical devices, to provide a Sterility Assurance Level (SAL) of 10-6.

I. CLINICAL TESTS (807.92(b)(2))

This 510(k) does not include data from clinical tests.

J. CONCLUSION (807.92(b)(3))

In summary, the NaviCross 0.018", the subject of this 510(k), is substantially equivalent in its intended use, technology/principle of operation, materials, and performance to the predicate, K110540 - TERUMO SUPPORT CATHETER, manufactured by Ashitaka Factory of Terumo Corporation.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).