    Device Name :

    NOVOPORE DRESSING

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NovoPore™ is intended to be used as part of a Negative Pressure Wound Therapy (NPWT) system. When used in conjunction with a NPWT system, it is intended to remove excess exudate, infectious material, and tissue debris from wounds which may promote wound healing. NPWT systems are indicated for use on chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), explored fistulas, flaps, and grafts.

    Device Description

    The NovoPore™ Dressing is a polyurethane foam wound dressing used as an accessory to powered suction pumps. The single-use dressing is packaged in a peelable oriented polyamide / polyethylene bag and is sterilized by gamma irradiation. The NovoPore™ Dressing is available in one size: medium.

    AI/ML Overview

    The NovoPore™ Dressing is a polyurethane foam wound dressing intended for use with Negative Pressure Wound Therapy (NPWT) systems. The study provided focuses on establishing substantial equivalence to predicate devices (KCI GranuFoam® and Genadyne A4-XLR8 Foam Dressing) through non-clinical performance data and does not involve human clinical trials for efficacy or effectiveness.

    Here’s a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally implied by demonstrating performance that is "substantially equivalent" or comparable to the predicate devices. For mechanical properties, the device aimed to show similar or better performance compared to the predicates.

    Acceptance Criteria (Implied by Predicate Equivalence)NovoPore™ Dressing PerformancePredicate (GranuFoam™) PerformancePredicate (A4-XLR8 Foam) Performance
    Material Composition
    Dressing CompositionPorous, open-celled polyurethane foamPorous, open-celled polyurethane foamPorous, open-celled polyurethane foam
    MaterialPolyurethanePolyurethanePolyurethane
    Physical Characteristics
    FormFoamFoamFoam
    Method to achieve Open-Celled FoamCrushing the foamReticulating the foamReticulating the foam
    ColorWhiteBlackGreen
    Dimensions (medium), cm18 x 12.5 x 318 x 12.5 x 3.218 x 12.5 x 3.3
    Foam Density66.1 kg/m³23.0 kg/m³21.1 kg/m³
    Open Pore Size100 - 600 microns400 - 600 micronsNot available
    Mechanical Properties Under Pressure (Wet)
    Cyclic Compression - Mean Upper Compressive Stress (0 min)20.7 ± 3.2 kPa13.4 ± 0.4 kPa12.8 ± 1.3 kPa
    Cyclic Compression - Mean Upper Compressive Stress (1 hr)21.4 ± 3.4 kPa12.5 ± 0.6 kPa11.5 ± 0.9 kPa
    Cyclic Compression - Mean Upper Compressive Stress (2 hrs)20.3 ± 3.4 kPa12.8 ± 0.9 kPa11.6 ± 0.9 kPa
    Cyclic Compression - Mean Upper Compressive Stress (4 hrs)18.0 ± 2.7 kPa13.4 ± 1.0 kPa12.0 ± 1.0 kPa
    Cyclic Compression - Mean Upper Compressive Stress (6 hrs)17.2 ± 2.3 kPa13.9 ± 1.1 kPa12.6 ± 1.2 kPa
    Resilience Change (0 min)0.0 ± 0 %0.0 ± 0 %0.0 ± 0 %
    Resilience Change (1 hr)2.0 ± 0 %8.9 ± 1.6 %10.3 ± 1.2 %
    Resilience Change (2 hrs)2.0 ± 0 %10.3 ± 1.6 %11.9 ± 1.4 %
    Resilience Change (4 hrs)2.0 ± 0 %12.3 ± 2.1 %14.4 ± 2.1 %
    Resilience Change (6 hrs)2.0 ± 0 %14.2 ± 2.2 %16.7 ± 3.1 %
    Mean Change in Thickness after 6 hours (1350 cycles)Decrease by avg of 1.5 ± 0.5%Decrease by avg of 7.6 ± 2.6%Decrease by avg of 8.0 ± 2.4%
    Durability after 1350 compressive cycles, wetNo DamageNo DamageNo Damage
    Vacuum Bench Test
    Pressure DistributionEvenly Distributed at each time pointEvenly Distributed at each time pointNot Tested
    Fluid Flow (at T=0, 24h, 48h, 72h)25 ml fluid drawn through within 30 seconds at every time point25 ml fluid drawn through within 30 seconds at every time pointNot Tested
    Dimensional change after 72h