Search Filters

Search Results

Found 13 results

510(k) Data Aggregation

    Why did this record match?
    Device Name :

    Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl, is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. The tested chemotherapy drugs are: Carmustine ( 3.3mg/mL 3,300 ppm) 11.1 min Cisplatin ( 1.0 mg/mL 1,000 ppm) >240 min Cyclophosphamide (20 mg/mL 20,000 ppm) >240 min Cytarabine (Cytosine) (100 mg/mL 100,000 ppm) >240 min Dacarbazine (10 mg/mL 10,000 ppm) >240 min Doxorubicin (Adriamycin) (2.0 mg/mL 2,000 ppm) >240 min Etoposide (Toposar) (20.0 mg/mL 20,000 ppm) >240 min Fluorouracil (5 Flu) (50.0 mg/mL 50,000 ppm) >240 min Ifosfamide (50.0 mg/mL 50,000 ppm) >240 min Methotrexate (25.0 mg/mL 25,000 ppm) >240 min Mitomycin C (0.5 mg/mL 500 ppm) >240 min Mitoxantrone HCL (2.0 mg/mL 2,000 ppm >240 min Paclitaxel (6.0 mg/mL 6,000 ppm) >240 min Thiotepa (10.0 mg/mL 10.000 ppm) 34.8 min Vincristine Sulfate (1.0 mg/mL 1,000 ppm >240 min Fentanyl Citrate Injection (100 mcg/2mL) >240 min Note: Carmustine and Thiotepa have extremely low permeation times of 11.1 and 34.8 minutes respectively. Warning: Do Not Use with Carmustine. Thiotepa

    Device Description

    The Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl are non-sterile, single use only, disposable examination gloves intended for medical purposes to be worn by examiners to prevent contamination between the patient and the examiner. The gloves are blue color, powder free, nitrile ambidextrous gloves are offered in five sizes, extra small, small, medium, large, and extra-large, packed in a paper box. The gloves are designed and manufactured in accordance with the ASTM D6319 standard and are tested for use with chemotherapy drugs and Fentanyl per ASTM D6978 standard.

    AI/ML Overview

    The provided document describes the acceptance criteria and results for the "Powder free, Non-sterile, Blue Nitrile Examination Gloves, Tested for Use with Chemotherapy Drugs and Fentanyl" device.

    Here's an analysis of the requested information:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance (Proposed Device)
    DimensionsPass
    Length (XS, S)220mm min
    Length (M, L, XL)230mm min
    Width (XS)70±10mm
    Width (S)80±10mm
    Width (M)95±10mm
    Width (L)110±10mm
    Width (XL)120±10mm
    Thickness (Finger)≥0.05mm
    Thickness (Palm)≥0.05mm
    Physical properties (ASTM D6319-19)Pass
    Tensile Strength (Before Aging)≥14 MPa, min.
    Tensile Strength (After Aging)≥14 MPa, min.
    Elongation (Before Aging)500%, min.
    Elongation (After Aging)400%, min.
    Freedom from holes (ASTM D5151-19)Pass
    No leakage at sampling level of G-1, AQL 2.5Complies
    Residual Powder (ASTM D6124-06)Pass
    240 min
    Cyclophosphamide (20 mg/mL 20,000 ppm)>240 min
    Cytarabine (Cytosine) (100 mg/mL 100,000 ppm)>240 min
    Dacarbazine (10 mg/mL 10,000 ppm)>240 min
    Doxorubicin (Adriamycin) (2.0 mg/mL 2,000 ppm)>240 min
    Etoposide (Toposar) (20.0 mg/mL 20,000 ppm)>240 min
    Fluorouracil (5 Flu) (50.0 mg/mL 50,000 ppm)>240 min
    Ifosfamide (50.0 mg/mL 50,000 ppm)>240 min
    Methotrexate (25.0 mg/mL 25,000 ppm)>240 min
    Mitomycin C (0.5 mg/mL 500 ppm)>240 min
    Mitoxantrone HCL (2.0 mg/mL 2,000 ppm)>240 min
    Paclitaxel (6.0 mg/mL 6,000 ppm)>240 min
    Thiotepa (10.0 mg/mL 10,000 ppm)34.8 min
    Vincristine Sulfate (1.0 mg/mL 1,000 ppm)>240 min
    Fentanyl Citrate Injection (100 mcg/2mL)>240 min
    Biocompatibility (ISO 10993-1, -5, -10, -11, -23)Pass
    Irritation (ISO 10993-23)Not an irritant
    Sensitization (ISO 10993-10)Not a sensitizer
    In Vitro Cytotoxicity (ISO 10993-5)Showed potential toxicity (Note: See point 7 for context)
    Acute Systemic Toxicity (ISO 10993-11)Does not induce acute systemic toxicity response

    Note on Biocompatibility: The document states that despite the in vitro cytotoxicity showing "potential toxicity," the overall biocompatibility assessment was deemed acceptable, asserting it "does not raise different safety or performance questions since the subject device has acceptable biocompatibility per the biocompatibility endpoint assessment."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample sizes used for each performance test (e.g., how many gloves were tested for dimensions, holes, tensile strength, or chemotherapy permeation). It references standards like ASTM D6319-19, ASTM D5151-19 (G-1, AQL 2.5), ASTM D6124-06, and ASTM D6978-05, which would prescribe the methodology and sample sizes.

    The data provenance regarding country of origin or retrospective/prospective nature is not explicitly mentioned. The applicant, Anhui Powerguard Technology Co., Ltd., is based in China, suggesting the testing may have been conducted there or through contract labs globally.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This device is a physical medical device (examination gloves), not an AI/software device that requires expert interpretation of images or data to establish ground truth. Therefore, the concept of "experts used to establish ground truth" in the context of medical imaging or AI algorithms is not applicable here. The "ground truth" for this device is established by physical and chemical testing against established engineering and safety standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Similarly, as this is a physical device and not an AI/software for diagnostic interpretation, adjudication methods like 2+1 or 3+1 are not applicable. Performance is measured against quantitative criteria defined by international standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI-assisted diagnostic tools, not for medical gloves.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, this question is not applicable. The device is a physical product (gloves), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the device's performance is established by:

    • International Standards: ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves for Medical Application), ASTM D6124-06 (Standard Test Method for Residual Powder on Medical Gloves), ASTM D5151-19 (Standard Test Method for Detection of Holes in Medical Gloves), and ASTM D6978-05 (Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs).
    • Biocompatibility Standards: ISO 10993-1, ISO 10993-5 (In Vitro Cytotoxicity), ISO 10993-10 (Dermal Sensitization), ISO 10993-11 (Acute Systemic Toxicity), and ISO 10993-23 (Primary Skin Irritation).

    These standards define the methodologies and acceptance limits for evaluating the physical, chemical, and biological properties of the gloves.

    Regarding In Vitro Cytotoxicity: The report notes that the "test article extract showed potential toxicity." However, the analysis in the 510(K) Summary explicitly states: "Analysis: In vitro cytotoxicity and acute systemic toxicity information for the predicate device is not publicly available. This does not raise different safety or performance questions since the subject device has acceptable biocompatibility per the biocompatibility endpoint assessment." This indicates that while an in vitro test showed some indication, the overall biocompatibility assessment, likely considering other in vivo tests like skin irritation and sensitization which passed, led to the conclusion of acceptable biocompatibility. This implicitly uses the wider expert consensus and risk assessment methodology embedded in the ISO 10993 series of standards as the 'ground truth' for biocompatibility.

    8. The sample size for the training set

    This question is not applicable as this is a physical medical device, not an AI/software device that requires a training set.

    9. How the ground truth for the training set was established

    This question is not applicable as this is a physical medical device, not an AI/software device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220209
    Date Cleared
    2022-04-11

    (75 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    Powder Free Nitrile Examination Glove, Non-sterile (Blue)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A Nitrile powder free examination glove is a disposable device intended for medical purposes, worn on the hand or finger to prevent contamination between examiner and patient.

    Device Description

    Powder Free Nitrile Examination Glove, Non-sterile (Blue)

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for Powder Free Nitrile Examination Gloves. It does not contain any information about an AI/ML medical device, acceptance criteria, or performance study data.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets the acceptance criteria based on the provided text. The information requested (e.g., sample sizes, expert qualifications, ground truth, MRMC studies) is relevant to the evaluation of AI/ML devices, not to a simple medical device like examination gloves.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220232
    Date Cleared
    2022-02-20

    (24 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    Nitrile Powder Free Examination Gloves, Non-Sterile - Blue, Black, Orange

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Nitrile Powder Free Examination Gloves, Non-Sterile - Blue, Black, Orange

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for Nitrile Powder Free Examination Gloves. This type of document does not contain the information requested regarding acceptance criteria and a study proving a device meets those criteria for software or AI-powered medical devices.

    The letter pertains to a Class I medical device (non-powdered patient examination glove), which is a physical product, not a software or AI-driven diagnostic or therapeutic tool. As such, the clearance process for such a device typically involves demonstrating substantial equivalence to a predicate device based on material properties, performance standards (e.g., tensile strength, barrier integrity), and biocompatibility, rather than algorithms, ground truth establishment, or human-in-the-loop studies.

    Therefore, I cannot extract the requested information from the provided text because it describes the clearance of examination gloves, not an AI/software medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211310
    Date Cleared
    2021-06-16

    (47 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    AMADEX - Nitrile Powder Free Examination Gloves Non-Sterile (Blue, Black, Orange)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    AMADEX - Nitrile Powder Free Examination Gloves Non-Sterile, (Blue, Black, Orange) are Class I patient examination gloves bearing the product code Nitrile - LZA (2 ICFR880.6250). AMADEX - Nitrile Powder Free Examnation Gloves Non-Sterile,(Blue, Black, Orange) meet all the requirements of ASTM standard D 6319-10 and FDA 21 CFR 880.6250. The AMADEX - Nitrile Powder Free Examination Gloves Non-Sterile, (Blue, Black, Orange) is single-use device to prevent contamination between patient and examiner.

    AI/ML Overview

    This document describes the acceptance criteria and study results for the AMADEX Nitrile Powder Free Examination Gloves Non-Sterile (Blue, Black, Orange) to demonstrate substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsStandardPurposeAcceptance CriteriaReported Device PerformanceRemarks
    DimensionsASTM D 6319-10
    LengthTo determine the length of the glovesMin 230 mm (All Sizes)>230 mm (Passes)Same
    Width (XS)70 ± 10 mm76 ± 3 mmSame
    Width (S)80 ± 10 mm84 ± 3 mmSame
    Width (M)95 ± 10 mm94 ± 3 mm (Also reported as 98 ± 2 mm, which passes 95 ± 10 mm)Same
    Width (L)110 ± 10 mm105 ± 3 mmSame
    Width (XL)120 ± 10 mm113 ± 3 mmSame
    Physical PropertiesASTM D 6319-10To determine the physical properties (Tensile strength and Elongation)
    Tensile Strength (Before Aging)14 Mpa min24.5 Mpa (Passes) (Also reported as 14 Mpa in comparison table)Same
    Ultimate Elongation (Before Aging)500% min500% (Implied, as result is 14 Mpa and 500% in comparison table)Same
    Tensile Strength (After Aging)14 Mpa min14 Mpa (Implied)Same
    Ultimate Elongation (After Aging)400% min400% (Implied)Same
    ThicknessASTM D 6319-10To determine the thickness of the glovesPalm 0.05 mm min, Finger 0.05 mm minPalm: 0.09 mm (Passes), Finger: 0.12 mm (Passes)Same
    Powder FreeASTM D 6319-10, D 6124-06To determine the amount of residual powder and non-powder solids found on gloves
    Ask a Question

    Ask a specific question about this device

    K Number
    K210755
    Date Cleared
    2021-05-07

    (53 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Powder Free Nitrile Patient Examination Gloves, Non-Sterile (Blue)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Powder Free Nitrile Patient Examination Glove, Non-Sterile (Blue) meet all the requirements of ASTM standard D6319-19 and FDA 21 CFR 880.6250.

    The powder free nitrile examination glove is manufactured from synthetic rubber latex. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that assists the user in donning the gloves with ease without using any lubricant such as powder on the glove surface. The glove is ambidextrous, i.e., can be worn on right hand or left hand.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for Powder Free Nitrile Patient Examination Gloves, Non-Sterile (Blue). This is a medical device of Class I and the submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving the performance of an AI/ML algorithm.

    Therefore, the requested information about AI model acceptance criteria, test sets, ground truth establishment, expert adjudication, MRMC studies, and training sets is not applicable to this document.

    The document describes the acceptance criteria and device performance as they relate to the physical and biocompatibility characteristics of the examination gloves. Here's a summary based on the provided text:


    1. Acceptance Criteria and Reported Device Performance for Powder Free Nitrile Patient Examination Gloves

    The acceptance criteria for these medical gloves are primarily based on established ASTM (American Society for Testing and Materials) and ISO standards. The reported device performance is compared against these standards and the predicate device.

    Table of Acceptance Criteria and Reported Device Performance

    CharacteristicStandard/Test MethodAcceptance CriteriaDevice Performance (Current Blue)StatusCompared to Predicate
    MaterialASTM D6319-10NitrileNitrileSameSame
    Color-BlueBlueSameSame
    Texture-Finger texturedFinger texturedSameSame
    Physical Properties
    Before AgingASTM D6319-10 (Tensile Test: ASTM D412-16)Tensile Strength: ≥ 14MPa; Ultimate Elongation: ≥ 500%Tensile Strength: 15.3 MPa (Pass); Ultimate Elongation: 500% (Pass)PassSame
    After AgingASTM D6319-10 (Tensile Test: ASTM D412-16)Tensile Strength: ≥ 14MPa; Ultimate Elongation: ≥ 400%Tensile Strength: 14.8 MPa (Pass); Ultimate Elongation: 460% (Pass)PassSame
    ThicknessASTM D6319-19 (Measurement: ASTM D3767-03)Finger: 0.07-0.10mm; Palm: 0.06-0.09mm; Cuff: 0.06-0.08mmFinger: 0.07-0.10mm; Palm: 0.06-0.09mm; Cuff: 0.06-0.08mm (Within ranges)PassSimilar
    Powder FreeASTM D6124-06≤ 2 mg/glove1.5 mg/glovePassSame
    Biocompatibility
    Primary Skin IrritationISO 10993-10:2010Passes (Not a primary skin irritant); PII = 0Passes (Not a primary skin irritant); PII = 0PassSimilar
    Dermal SensitizationISO 10993-10:2010Passes (Not a contact sensitizer); No allergic reaction observedPasses (Not a contact sensitizer); No allergic reaction observedPassSimilar
    CytotoxicityISO 10993-5:2009 (MEM Elution)No cytotoxicity reactivity at specific extract concentrations (e.g., 12.5%, 6.25%, 3.125%)No cytotoxicity reactivity at 12.5%, 6.25%, 3.125% extract concentrationsPassSimilar
    Acute Systemic ToxicityISO 10993-11:2017 (E)Passes (no adverse biological reaction); No mortality observedPasses (no adverse biological reaction); No mortality observedPassNot Available for Predicate
    Watertight (Hole Detection)ASTM D5151-19Inspection Level 1, AQL 1.5; Acceptance No: 10 (for a sample size of 315)6 pieces found with leaks (within acceptance criteria)PassSame
    DimensionASTM D3767-03 (2020)Length: Min 240mm; Width (XS: 70±10mm, S: 80±10mm, M: 95±10mm, L: 110±10mm, XL: >110mm); Thickness: Finger ≥ 0.05mm, Palm ≥ 0.05mmLength: Min 240mm (Pass); Width (Ave: 72mm-113mm for sizes, all Pass); Thickness: Finger 0.10mm, Palm 0.06mm (Pass)PassPart of "Similar" Thickness

    2. Sample Sizes Used for Testing and Data Provenance

    The sample sizes for testing are defined by the specific ASTM standards followed.

    • Physical Properties (Tensile Strength, Elongation): Not explicitly stated, but typically these tests involve a specified number of specimens per batch or test run according to ASTM D412.
    • Watertight (Hole Detection): Sample size of 315 pieces for a batch sampling of 150,001 to 500,000 gloves, based on Inspection Level GI and AQL 1.5. This is a prospective test conducted on manufactured gloves.
    • Powder Free (Residual Powder): Sample size of 5 pieces.
    • Biocompatibility Tests (Skin Irritation, Sensitization, Cytotoxicity, Systemic Toxicity): These tests are typically conducted on animal models (e.g., mice, rabbits) and cell cultures. The counts of animals (e.g., number of test group animals vs. control group animals) are not explicitly stated in this summary, but the results (e.g., "no mortality," "no erythema") indicate biological testing. This data would be prospective from laboratory testing.
    • Data Provenance: The manufacturing company is Meditech Gloves Sdn Bhd, Malaysia. The tests described are laboratory and quality control tests conducted as part of the manufacturing and regulatory submission process.

    3. Number of Experts and Qualifications for Ground Truth

    This type of medical device submission (examination gloves) does not involve human expert interpretation of images or complex diagnostic data like an AI/ML device would. Therefore, there are no "experts establishing ground truth" in the diagnostic sense. The "ground truth" for the performance of these gloves is defined by the objective, quantifiable results of the standardized physical and biocompatibility tests (e.g., direct measurements of tensile strength, elongation, amount of powder, number of leaks, observed biological reactions in laboratory tests).

    4. Adjudication Method for the Test Set

    Not applicable. There is no human interpretation or subjective assessment that would require an adjudication method. The tests are objective and quantitative.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not an AI/ML diagnostic device that interacts with human readers or interprets medical cases.

    6. Standalone (Algorithm Only) Performance

    Not applicable. The device is a physical examination glove, not an algorithm.

    7. Type of Ground Truth Used

    The ground truth used for this device is based on objective, quantitative measurements and laboratory test results against established international standards (ASTM, ISO) for material properties, physical dimensions, leak detection, and biocompatibility.

    8. Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for an AI/ML algorithm, the concept of establishing ground truth for it does not apply here.


    In summary, the provided document is a regulatory submission for a physical medical device (examination gloves) where "acceptance criteria" relate to meeting specific physical and biological performance standards, not to the performance of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101820
    Date Cleared
    2011-03-03

    (246 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    NON-STERILE, BLUE & WHITE POWDERED NITRILE EXAMINATION GLOVES

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K033887
    Date Cleared
    2004-03-02

    (78 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    NON-STERILE BLUE & WHITE COLOUR POWDER FREE BARRIER PRO POLYBUTADIENE COPOLYMER EXAMINATION GLOVES

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.

    Device Description

    Blue & White Colour Powder Free Barrier Pro™ Polybutadiene Copolymer Examination Gloves meets all the current specifications listed under the ASTM Specification D6319-00a563 Standard Specification for Nitrile Examination Gloves for Medical Application. They are made from BarrierPro™ synthetic copolymer rubber latex, a polybutadiene based copolymer. They are blue in colour and are powder free.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance of Non-Sterile Blue & White Colour Powder Free Barrier Pro™ Polybutadiene Copolymer Examination Gloves. This device is a medical glove, not an AI/ML powered medical device, and therefore the acceptance criteria and study detailed below are for the physical and chemical properties of the gloves, not an algorithm's performance.

    Here's the information based on the provided text, formatted to your request where applicable:

    1. Table of Acceptance Criteria and the Reported Device Performance

    TESTASTM D6319-00a3 (Acceptance Criteria)POWDER FREE BARRIER PROTM EXAM. GLOVES (Reported Device Performance)
    1. Watertight (1000ml)Multiple Normal GI AQL = 2.5Pass GI AQL = 2.5
    2. Length (mm)240 mm minimum for all sizes
    Size XSMin 220
    Size SMin 220
    Size MMin 230
    Size LMin 230
    Size XL-
    3. Palm width (mm)
    Size XS$70 \pm 10$73 - 78
    Size S$80 \pm 10$83 - 88
    Size M$95 \pm 10$93 - 98
    Size L$111 \pm 10$103 - 107
    Size XL-
    4. Thickness (mm) (Single Layer)
    FingerMin 0.05Min 0.08
    PalmMin 0.05Min 0.08
    5. Physical Properties
    Before Aging
    - Tensile Strength (MPa)Min 1417-21
    - Ultimate Elongation (%)Min 500630 - 710
    After Aging
    - Tensile Strength (MPa)Min 1416-20
    - Ultimate Elongation (%)Min 400610-670
    6. Powder ContentMax 2.0mg/glove (FDA Requirement quoted below table)Below 2 mg/glove

    In addition, the device successfully passed the Bio-compatibility Test, which included a Primary Dermal Irritation Test and a Guinea Pig Sensitization (Buehler) test.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample size used for the test set for each of the physical property tests. The data provenance is also not specified beyond the fact that the manufacturer is YTY Industry (Manjung) Sdn Bhd (Malaysia). There is no indication whether the testing was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable as the "ground truth" here refers to objective physical and chemical properties measured according to established ASTM standards, not subjective interpretations by human experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable as the testing involves objective measurements against predefined standards (ASTM D6319-00a3 and FDA watertight test for physical properties, and bio-compatibility tests) and does not involve human adjudication for "ground truth".

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable as the device is a medical glove, not an AI/ML powered medical device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is a medical glove, not an AI/ML powered medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context is based on established industry standards and objective measurements. Specifically:

    • ASTM D6319-00a3 Standard Specification for Nitrile Examination Gloves for Medical Application for physical properties (e.g., length, width, thickness, tensile strength, elongation).
    • FDA 1000ML watertight test for barrier integrity.
    • Bio-compatibility tests (Primary Dermal Irritation Test and Guinea Pig Sensitization (Buehler) test) for biological safety.

    8. The sample size for the training set

    This question is not applicable as the device is a medical glove, not an AI/ML powered medical device that requires a training set.

    9. How the ground truth for the training set was established

    This question is not applicable as the device is a medical glove, not an AI/ML powered medical device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K002691
    Date Cleared
    2000-12-08

    (101 days)

    Product Code
    Regulation Number
    878.4370
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    TOWEL - NON-STERILE, BLUE, GREEN, WHITE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K003233
    Date Cleared
    2000-11-03

    (17 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    RS NITRILSTST MEDICAL EXAMINATION GLOVES, POWDERED (NON STERILE) BLUE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K001497
    Manufacturer
    Date Cleared
    2000-07-26

    (72 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    MULTISAFE-POWDERED NITRILE GLOVES, NON-STERILE (BLUE)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2