Search Results
Found 54 results
510(k) Data Aggregation
(260 days)
Nitrile Powder Free Examination Gloves with Low Dermatitis Potential, Tested for Use with Chemotherapy
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs, fentanyl citrate and gastric acid as per ASTM D6978-05 (2023) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs:
The following drugs and concentration had NO breakthrough detected up to 240 minutes:
- Bendamustine HCl (5 mg/mL)
- Bleomycin Sulfate (15 mg/mL)
- Busulfan (6 mg/mL)
- Carboplatin (10 mg/mL)
- Carfilzomib (2 mg/mL)
- Cetuximab (Erbitux) (2 mg/mL)
- Chloroquine (50 mg/mL)
- Cisplatin (1 mg/mL)
- Cladribine (1 mg/mL)
- Cyclophosphamide (20 mg/mL)
- Cyclosporin A (100 mg/mL)
- Cytarabine HCl (100 mg/mL)
- Cytovene (10 mg/mL)
- Dacarbazine (10 mg/mL)
- Daunorubicin HCl (5 mg/mL)
- Decitabine (5 mg/mL)
- Docetaxel (10 mg/mL)
- Doxorubicin HCl (2 mg/mL)
- Epirubicin HCl (2 mg/mL)
- Etoposide (20 mg/mL)
- Fludarabine Phosphate (25 mg/mL)
- Fluorouracil (50 mg/mL)
- Fulvestrant (50 mg/mL)
- Gemcitabine HCl (38 mg/mL)
- Idarubicin HCl (1 mg/mL)
- Ifosfamide (50 mg/mL)
- Irinotecan HCl (20 mg/mL)
- Mechlorethamine HCl (1 mg/mL)
- Melphalan HCl (5 mg/mL)
- MESNA (100 mg/mL)
- Methotrexate (25 mg/mL)
- Mitomycin C (0.5 mg/mL)
- Mitoxantrone HCl (2 mg/mL)
- Oxaliplatin (2 mg/mL)
- Paclitaxel (6 mg/mL)
- Pemetrexed (25 mg/mL)
- Propofol (10 mg/mL)
- Raltitrexed (0.5 mg/mL)
- Retrovir (10 mg/mL)
- Rituximab (10 mg/mL)
- Temsirolimus (25 mg/mL)
- Topotecan HCl (1 mg/mL)
- Triclosan (1 mg/mL)
- Trisenox (Arsenic Trioxide) (1 mg/mL)
- Velcade (Bortezomib) (1 mg/mL)
- Vidaza (Azacitidine) (25 mg/mL)
- Vinblastine Sulfate (1 mg/mL)
- Vincristine Sulfate (1 mg/mL)
- Vinorelbine Tartrate (10 mg/mL)
- Zoledronic Acid (0.8 mg/mL)
The following chemotherapy drugs and concentration showed breakthrough detected in less than 60 minutes:
- Carmustine (3.3 mg/mL), breakthrough detected at 55.5 minutes
- Thiotepa (10 mg/mL), breakthrough detected at 50.8 minutes
Warning: Not recommended for use with Carmustine and Thiotepa
The following hazardous drugs (opioids) and concentration had NO breakthrough detected up to 240 minutes:
- Fentanyl Citrate Injection (100mcg/2mL)
- Simulated Gastric Acid Fluid/Fentanyl Citrate Injection Mix 50/50 Solution
The subject device, Nitrile Powder Free Examination Gloves with Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid is a single use, disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. This product demonstrated reduced potential for sensitizing users to chemical additives, supported by a negative skin sensitization test (Modified Draize-95 Test) and tested for use with chemotherapy drugs, fentanyl citrate and gastric acid.
The gloves are made of nitrile rubber, powder free, ambidextrous with beaded cuff. Inner surface of gloves undergoes surface treatment process to produce a smooth surface that facilitates the user in donning the gloves without using lubricant and donning powder on the glove surface. These gloves are offered in six sizes (XS, S, M, L, XL, XXL), and supplied in non-sterile state.
The provided FDA 510(k) clearance letter and summary concern a medical device, "Nitrile Powder Free Examination Gloves with Low Dermatitis Potential, Tested for Use with Chemotherapy Drugs, Fentanyl Citrate and Gastric Acid." This document details the product's attributes, testing, and comparison to a predicate device for regulatory clearance.
However, the questions you've asked (acceptance criteria, study details, sample sizes, expert qualifications, adjudication, MRMC studies, standalone performance, ground truth, and training set details) are highly specific to the performance of an AI/ML medical device rather than a physical product like examination gloves. The provided document describes the physical and chemical performance of the gloves (e.g., resistance to permeation by chemotherapy drugs, physical properties, biocompatibility, and skin sensitization), not the performance of an artificial intelligence or machine learning algorithm.
Therefore, many of your questions are not applicable to the information contained within this 510(k) clearance for medical gloves. I will answer the applicable questions based on the provided text, and explicitly state when a question is not relevant to this type of device.
Acceptance Criteria and Device Performance for Nitrile Examination Gloves
The "acceptance criteria" for these gloves are primarily physical, chemical, and biological performance standards, as demonstrated through various ASTM and ISO tests. The "reported device performance" indicates whether the gloves met these pre-defined standards.
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria (Requirement) | Reported Device Performance (Result) |
---|---|
Physical Properties (ASTM D6319-19): | |
- Tensile strength (Before Aging): min 14 MPa | Pass |
- Ultimate elongation (Before Aging): min 500% | Pass |
- Tensile strength (After Aging): min 14 MPa | Pass |
- Ultimate elongation (After Aging): min 400% | Pass |
Dimensions (ASTM D6319-19): | |
- Length: XS: min 220mm, S: min 220mm, M: min 230mm, L: min 230mm, XL: min 230mm, XXL: min 230mm | Pass |
- Palm Width: XS: 70 ± 10mm, S: 80 ± 10mm, M: 95 ± 10mm, L: 110 ± 10mm, XL: 120 ± 10mm, XXL: 130 ± 10mm | Pass |
- Thickness (Finger): min 0.05mm | Pass |
- Thickness (Palm): min 0.05mm | Pass |
Watertight Test (Freedom from Holes - ASTM D5151-19): | Pass Inspection Level G1, AQL 2.5 |
Residual Powder Content (ASTM D6124-06): | Residual powder |
Ask a specific question about this device
(268 days)
Nitrile Powder Free Examination Gloves Biodegradable (Green)
The Nitrile Powder Free Examination Gloves Biodegradable (Green) is a medical glove which is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between examiner and patient.
Nitrile Powder Free Examination Gloves Biodegradable (Green) meet all the requirements of ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application. The gloves are single use, disposable, and non-sterile. The gloves are offered in five sizes: S, M, L, XL, XXL.
The provided document is a 510(k) Premarket Notification for Nitrile Powder Free Examination Gloves Biodegradable (Green). It describes the device, its intended use, and comparative testing against a predicate device to demonstrate substantial equivalence.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
Important Note: This document pertains to a medical device (examination gloves) that is a physical product, not an AI/software-based device. Therefore, several of the requested categories (e.g., number of experts for ground truth, MRMC study, training set ground truth) are not applicable in this context, as they relate specifically to the validation of AI/ML systems or diagnostic aids, not physical medical devices like gloves.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for these gloves are primarily based on established ASTM and ISO standards for medical examination gloves and biocompatibility.
Characteristic | Test Methodology / Standard | Acceptance Criteria (Specification) | Reported Device Performance (Results) |
---|---|---|---|
Dimensions | ASTM D6319-19 | (S) Length: Min 230mm, Width: 80±10mm | |
(M) Length: Min 230mm, Width: 95±10mm | |||
(L) Length: Min 230mm, Width: 105±5mm | |||
(XL) Length: Min 230mm, Width: 115±5mm | |||
(XXL) Length: Min 230mm, Width: >120mm | (S) Average: Length 246mm, Width 86mm | ||
(M) Average: Length 246mm, Width 99mm | |||
(L) Average: Length 246mm, Width 106mm | |||
(XL) Average: Length 250mm, Width 115mm | |||
(XXL) Average: Length 254mm, Width 125mm | |||
Material Thickness | ASTM D6319-19 | Palm & Finger: Min 0.05mm | (S) Palm 0.13mm, Finger 0.19mm |
(M) Palm 0.13mm, Finger 0.19mm | |||
(L) Palm 0.13mm, Finger 0.18mm | |||
(XL) Palm 0.13mm, Finger 0.17mm | |||
(XXL) Palm 0.12mm, Finger 0.19mm | |||
Physical Properties (Tensile Strength) | ASTM D6319-19 | Before Aging: ≥14MPa | |
After Aging: ≥14MPa | M: Before Aging 29.95MPa, After Aging 29.66MPa (Only M size reported) | ||
Physical Properties (Ultimate Elongation) | ASTM D6319-19 | Before Aging: ≥500% | |
After Aging: ≥400% | M: Before Aging 561%, After Aging 496% (Only M size reported) | ||
Hole Detection (Watertight) | ASTM D5151-19, 21 CFR 800.20 | AQL 1.5 | All sizes (S, M, L, XL, XXL) Pass AQL 1.5 |
Residual Powder | ASTM D6124-06(2022) | Max - 2 mg/glove | 0.36 mg/glove |
Biodegradation | ASTM D5511-18 | Tested (Specific % is not an 'acceptance criterion' but a 'result' of testing) | Adjusted Percent Biodegraded (%): 7.2 |
Biocompatibility - Cytotoxicity | ISO 10993-5 | Not explicitly stated as a numeric criterion, but expectation of low/no reactivity | Mild cytotoxicity reactivity result (2) with the neat extract (100%). (This indicates a mild reaction but is often acceptable depending on the product and risk assessment) |
Biocompatibility - Skin Sensitization | ISO 10993-10:2021 | Not explicitly stated, but expectation of non-sensitizing | Extracts are not sensitizing to the animal model. |
Biocompatibility - Irritation | ISO 10993-23:2021 | Not explicitly stated, but expectation of non-irritating | Device is non-irritating, extracts are not irritating to the animal model. |
Biocompatibility - Systemic Toxicity | ISO 10993-11:2017 | Not explicitly stated, but expectation of no acute toxic reaction | Did not demonstrate any acute toxic reaction. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the numerical sample size for each specific test (e.g., how many gloves were tested for dimensions, tensile strength, or holes). However, the tests are conducted according to recognized standards (ASTM, ISO), which inherently specify sampling plans and test methods (e.g., AQL 1.5 for hole detection typically implies a specific sampling size based on lot size).
- Data Provenance: The tests were conducted by the manufacturer, Comfort Rubber Gloves Industries Sdn. Bhd., located in Matang, Perak, Malaysia. The data appears to be prospective as it directly supports the 510(k) submission for this specific device.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not Applicable. For a physical product like medical gloves, "ground truth" is established through standardized physical, chemical, and biological testing methods (e.g., measuring dimensions, tensile strength, or observing biological reactions in animal models), not through expert consensus on interpretations like with AI/imaging data. The standards themselves define the "ground truth" for compliance.
4. Adjudication Method for the Test Set
Not Applicable. Adjudication methods (e.g., 2+1, 3+1) are relevant for subjective human assessments, particularly in image interpretation or clinical diagnosis. For physical and chemical property testing of gloves, the results are typically objective measurements or observations by trained laboratory personnel following the specified test protocols.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of Human Reader Improvement
Not Applicable. MRMC studies are used to evaluate the diagnostic performance of a system (often AI-assisted) by comparing human reader performance with and without the system. This is not relevant for a physical product like examination gloves.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done
Not Applicable. "Standalone" performance refers to the diagnostic capability of an algorithm or AI without human interaction. This concept does not apply to a physical medical device. The device's performance is intrinsically linked to its physical and material properties.
7. The Type of Ground Truth Used
The "ground truth" for this device is based on objective measurements and observations derived from:
- Standardized Test Methods: Adherence to established and recognized international standards (e.g., ASTM D6319-19, ASTM D5151-19, ASTM D5511-18, ASTM D6124-06(2022), ISO 10993 series). These standards define the test procedures and acceptance criteria for various physical, chemical, and biological properties.
- Direct Physical and Chemical Analysis: Measurements of dimensions, tensile strength, elongation, residual powder, and observations of biodegradation and biocompatibility reactions.
- Biological Endpoints: For biocompatibility, the ground truth is based on the observed biological response (e.g., presence/absence of irritation, sensitization, or systemic toxicity) in established animal or in-vitro models as per ISO 10993.
8. The Sample Size for the Training Set
Not Applicable. The concept of a "training set" (for machine learning models) does not apply to the development or validation of a physical medical device like these gloves. The manufacturing process is controlled by quality management systems (e.g., 21 CFR Part 820 Quality System Regulation), and product performance is validated through testing of manufactured lots against specifications, not through machine learning.
9. How the Ground Truth for the Training Set Was Established
Not Applicable. As mentioned above, there is no "training set" in the context of this device. The physical and performance characteristics are inherent to the material and manufacturing process, validated through testing against pre-defined standards.
Ask a specific question about this device
(68 days)
Nitrile Powder Free Examination Gloves (Blue), Non-Sterile Low Dermatitis Potential
The Nitrile Powder Free Examination Gloves (Blue), Non-Sterile Low Dermatitis Potential is a medical glove which is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between examiner and patient.
Nitrile Powder Free Examination Gloves (Blue), Non-Sterile Low Dermatitis Potential meet ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.
This document pertains to the 510(k) submission for Nitrile Powder Free Examination Gloves (Blue), Non-Sterile Low Dermatitis Potential (K241528).
1. A table of acceptance criteria and the reported device performance:
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Physical Dimensions - Thickness | ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application | |
ISO 2859-1/S2/AQL 4.0 | ||
Length - Size XS, S, M, L, XL – Min. 230 mm | ||
Width - XS - 70±10mm, S – 80±10mm, M – 95±10mm, L – 110±10mm, XL – 120±10mm | ||
Thickness at Finger – min. 0.05mm | ||
Thickness at Palm – min. 0.05mm | Pass | |
Physical Properties | ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application & ASTM D412-16(2021) Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers-Tension | |
Before aging: Tensile Strength: ≥14 MPa, Ultimate elongation: ≥500% | ||
After aging: Tensile Strength: ≥14 MPa, Ultimate elongation: ≥400% | Pass | |
Powder Residue | ASTM D6124-06(2022) Standard Test Method for Residual Powder on Medical Gloves | |
≤ 2 mg/glove | Pass | |
Freedom from Holes | ASTM D5151-19 Standard Test Method for Detection of Holes in Medical Gloves | |
AQL 1.5 | Pass | |
Irritation | ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests for irritation | |
Under the conditions of the study, the device is not an irritant. | Under the conditions of the study, the device was not an irritant. | |
Sensitization | ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization | |
Under the conditions of the study, the device is not a sensitizer. | Under the conditions of the study, the device was not a sensitizer. | |
Cytotoxicity | ISO 10993-5 Biological evaluation of medical devices-Part 5 Tests for in vitro cytotoxicity | |
Under the conditions of the study, the subject device extract does not exhibit cytotoxicity reactivity. | Under the conditions of the study, the subject device extract exhibits cytotoxicity from 100.0% extract concentrations to 50.0% extract concentrations and no cytotoxicity reactivity from 25.0% extract concentrations to 3.125% extract concentrations. (Note: This is a deviation from the acceptance criteria, as the device exhibited cytotoxicity at higher concentrations). | |
Acute systemic toxicity | ISO 10993-11 Biological evaluation on medical device Part 11 - Test for systemic toxicity | |
Under the conditions of this study, the test article does not induce acute systemic toxicity. | Under the conditions of this study, the test article did not induce acute systemic toxicity. | |
Low Dermatitis Potential (Modified Draize) | Modified Draize ASTM 6355-07 Standard Test Method for Human Repeat Insult Patch Testing of Medical Gloves. | |
No clinical evidence that the glove may induce Type IV allergy in the unsensitized general user population. | There was no clinical evidence that the glove may induce Type IV allergy in the unsensitized general user population. |
2. Sample size used for the test set and the data provenance:
- Physical Dimensions, Physical Properties, Powder Residue, Freedom from Holes: The specific sample sizes are not explicitly stated within the provided text, but the tests were conducted according to ASTM and ISO standards (e.g., ISO 2859-1/S2/AQL 4.0 for dimensions, AQL 1.5 for freedom from holes), which prescribe sampling plans. The data provenance is from non-clinical tests conducted by the manufacturer, Comfort Rubber Gloves Industries Sdn. Bhd. No country of origin for the specific testing labs is provided, nor is it specified if the data is retrospective or prospective, though it's likely prospective for the submission.
- Irritation, Sensitization, Cytotoxicity, Acute systemic toxicity: The specific sample sizes for these biocompatibility tests are not explicitly stated in the document. These tests are likely performed ex vivo or in vivo with animal models (e.g., mice for acute systemic toxicity) or in vitro (e.g., for cytotoxicity). The data provenance is from non-clinical tests conducted by the manufacturer.
- Low Dermatitis Potential (Modified Draize): The specific sample size for the human repeat insult patch test (HRIPT) is not explicitly stated. The Modified Draize test (ASTM 6355-07) is a clinical test involving human subjects, therefore, the data provenance is clinical, likely from a prospective study.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided in the document. The tests performed are primarily laboratory-based or standardized clinical tests with defined acceptance criteria, rather than requiring expert interpretation for ground truth establishment in the traditional sense of medical image analysis or complex diagnostic tasks. The "ground truth" for these tests is the quantitative or qualitative outcome against a standard.
4. Adjudication method for the test set:
Not applicable. The tests performed are objective, standardized tests with predefined acceptance criteria. There is no mention or implication of a need for adjudication by multiple human experts for the results of these physical, chemical, or biological tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a medical glove, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is irrelevant to this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable. This device is a medical glove, not an algorithm or software.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The ground truth for most of these tests is based on objective measurements and established standards:
- Physical Properties (Dimensions, Tensile Strength, Elongation, Powder Residue, Freedom from Holes): Ground truth is based on quantitative measurements against the physical requirements specified in ASTM D6319-19, ASTM D412-16(2021), ASTM D6124-06(2022), and ASTM D5151-19.
- Biocompatibility (Irritation, Sensitization, Acute Systemic Toxicity): Ground truth is based on observations and measurements from standardized biological evaluations (ISO 10993 series) against set criteria for biological responses.
- Cytotoxicity: Ground truth is based on observable cellular responses in in vitro tests (ISO 10993-5).
- Low Dermatitis Potential (Modified Draize): Ground truth is based on the absence of clinical evidence of Type IV allergy in human subjects, as observed and reported in the HRIPT study, following ASTM 6355-07. This involves clinical observation in human subjects.
8. The sample size for the training set:
Not applicable. This is a physical medical device (gloves) and not an AI/ML-based algorithm that requires a training set.
9. How the ground truth for the training set was established:
Not applicable. As this device is a medical glove, it does not involve a training set or associated ground truth establishment for an AI/ML model.
Ask a specific question about this device
(202 days)
Nephron Nitrile, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs
Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl) is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.
In addition, these gloves were tested for use with Chemotherapy drugs and Fentanyl in accordance with ASTM D6978-05 (2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
The following chemotherapy drugs and concentration had NO breakthrough detected up to 240 minutes:
Bleomycin Sulfate (15.0 mg/ml) Busulfan (6.0 mg/ml) Carboplatin (10.0 mg/ml) Cisplatin (1.0 mg/ml) Cyclophosphamide (20.0 mg/ml) Cytarabine (100.0 mg/ml) Dacarbazine (10.0 mg/ml) Daunorubicin HCl (5.0 mg/ml) Docetaxel (10.0 mg/ml) Doxorubicin HCl (2.0 mg/ml) Epirubicin HCl (2.0 mg/ml) Etoposide (20.0 mg/ml) Fludarabine (25.0 mg/ml) Fluorouracil (50.0 mg/ml) Gemcitabine (38.0 mg/ml) Idarubicin HCl (1.0 mg/ml) Ifosfamide (50.0 mg/ml) Irinotecan (20.0 mg/ml) Mechlorethamine HCl (1.0 mg/ml) Melphalan (5.0 mg/ml) Methotrexate (25.0 mg/ml) Mitomycin C (0.5 mg/ml) Mitoxantrone HCl (2.0 mg/ml) Paclitaxel (6.0 mg/ml) Rituximab (10.0 mg/ml) Trisenox (1.0 mg/ml) Vincristine Sulfate (1.0 mg/ml)
The tested Opioid is: Fentanyl Citrate Injection (100 mcg/2 mL) Permeation: no breakthrough up to 240 minutes
The following chemotherapy drugs have low permeation times: Carmustine (3.3 mg/ml) : 26.3 minutes Thiotepa (10.0 mg/ml) : 49.1 minutes Warning: Not for Use with: Carmustine, Thiotepa
Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl) are Class I patient examination gloves, bearing the product codes LZA, LZC, OPJ, QDO (21 CFR 880.6250). They meet all the current specifications listed under the ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application and also complies with requirements for Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs as per ASTM D6978-05 (2019). The powder-free gloves are made from Nitrile (NBR) latex and are black in color. The product is non-sterile, fingertip textured, ambidextrous with beaded cuff, and single use only.
Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl) include glove sizes: X-Small, Medium, Large, X-Large and XX-Large.
The provided text describes the acceptance criteria and the results of a study for the "Nephron Nitrile™, Nitrile Powder-Free Examination Gloves Black (Tested For Use With Chemotherapy Drugs and Fentanyl)".
Here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
Test Method (Characteristic) | Acceptance Criteria | Reported Device Performance (Results) |
---|---|---|
Physical Properties (ASTM D6319-19) | ||
Length | 220mm minimum for size XS and S; 230mm minimum for size M, L, XL and XXL | Pass |
Width | XS: 70 ± 10 mm; S: 80 ± 10 mm; M: 95 ± 10 mm; L: 110 ± 10 mm; XL: 120 ± 10 mm; XXL: 130 ± 10 mm | Pass |
Thickness (Palm & Finger) | Palm: 0.05 mm min for all sizes; Finger: 0.05 mm min for all sizes | Pass |
Tensile Strength (Before Ageing) | 14MPa min for all sizes | Pass |
Tensile Strength (After Ageing) | 14MPa min for all sizes | Pass |
Ultimate Elongation (Before Ageing) | 500% min for all sizes | Pass |
Ultimate Elongation (After Ageing) | 400% min for all sizes | Pass |
Freedom from Holes (ASTM D5151-19) | G-I, AQL 2.5 | Pass |
Powder Free Residue (ASTM D6124-06) | ≤ 2 mg/glove | Pass |
Resistance to Permeation by Chemotherapy Drugs (ASTM D6978-05) | ||
Bleomycin Sulfate (15 mg/ml) | >240 Minutes | >240 Minutes |
Busulfan (6 mg/ml) | >240 Minutes | >240 Minutes |
Carboplatin (10 mg/ml) | >240 Minutes | >240 Minutes |
Carmustine (3.3 mg/ml) | 33.8 Minutes (Predicate) | 26.3 Minutes (Subject Device - Note: This value is lower than the predicate's reported value, but both are below 240 mins and thus subject to warning. The specific acceptance criteria for "low permeation" drugs is that they are identified and warned against.) |
Cisplatin (1 mg/ml) | >240 Minutes | >240 Minutes |
Cyclophosphamide (20 mg/ml) | >240 Minutes | >240 Minutes |
Cytarabine (100 mg/ml) | >240 Minutes | >240 Minutes |
Dacarbazine (10 mg/ml) | >240 Minutes | >240 Minutes |
Daunorubicin HCl (5 mg/ml) | >240 Minutes | >240 Minutes |
Docetaxel (10 mg/ml) | >240 Minutes | >240 Minutes |
Doxorubicin HCl (2 mg/ml) | >240 Minutes | >240 Minutes |
Epirubicin HCl (2 mg/ml) | >240 Minutes | >240 Minutes |
Etoposide (20 mg/ml) | >240 Minutes | >240 Minutes |
Fludarabine (25 mg/ml) | >240 Minutes | >240 Minutes |
Fluorouracil (50 mg/ml) | >240 Minutes | >240 Minutes |
Gemcitabine (38 mg/ml) | >240 Minutes | >240 Minutes |
Idarubicin HCl (1 mg/ml) | >240 Minutes | >240 Minutes |
Ifosfamide (50 mg/ml) | >240 Minutes | >240 Minutes |
Irinotecan (20 mg/ml) | >240 Minutes | >240 Minutes |
Mechlorethamine HCl (1 mg/ml) | >240 Minutes | >240 Minutes |
Melphalan (5 mg/ml) | >240 Minutes | >240 Minutes |
Methotrexate (25 mg/ml) | >240 Minutes | >240 Minutes |
Mitomycin C (0.5 mg/ml) | >240 Minutes | >240 Minutes |
Mitoxantrone HCl (2 mg/ml) | >240 Minutes | >240 Minutes |
Paclitaxel (6 mg/ml) | >240 Minutes | >240 Minutes |
Rituximab (10 mg/ml) | >240 Minutes | >240 Minutes |
Thiotepa (10 mg/ml) | 128.1 Minutes (Predicate) | 49.1 Minutes (Subject Device - Note: This value is lower than the predicate's reported value, but both are below 240 mins and thus subject to warning. The specific acceptance criteria for "low permeation" drugs is that they are identified and warned against.) |
Trisenox (1 mg/ml) | >240 Minutes | >240 Minutes |
Vincristine Sulfate (1 mg/ml) | >240 Minutes | >240 Minutes |
Resistance to Permeation by Opioid Drugs (ASTM D6978-05) | ||
Fentanyl Citrate Injection (100mcg/2mL) | >240 Minutes | >240 Minutes |
Biocompatibility (ISO 10993-23: Irritation) | Under the condition of study, not an irritant | Under the conditions of the study, the test article met the requirements of the test. |
Biocompatibility (ISO 10993-10: Dermal Sensitization) | Under the conditions of the study, not a sensitizer | Under the conditions of the study, the test article was not considered a sensitizer. |
Biocompatibility (ISO 10993-11: Acute Systemic Toxicity) | Under the conditions of study, the device extracts do not pose an acute systemic toxicity concern | Under the conditions of study, there was no mortality or evidence of acute systemic toxicity. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document lists "Bench Data" and "Biocompatibility Data" for the non-clinical tests. For these types of tests, specific sample sizes beyond what's typically required by the ASTM or ISO standards are not detailed in this summary. The data provenance is not specified, but these are standard laboratory tests. It is implied these are prospective tests conducted on the subject device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is Not Applicable (N/A). The device is a medical glove, and the evaluation relies on established standardized physical, chemical permeation, and biocompatibility tests, not subjective expert assessment of images or clinical outcomes that would require ground truth establishment by human experts.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This section is Not Applicable (N/A). Adjudication methods like 2+1 or 3+1 are used for resolving disagreements in expert assessments (e.g., in medical image interpretation). The tests performed are objective, standardized laboratory tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is Not Applicable (N/A). This device is a medical glove, not an AI-assisted diagnostic tool. Therefore, MRMC studies and AI assistance are irrelevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This section is Not Applicable (N/A). This device is a medical glove, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the performance of the gloves is established by objective measurements against the specified ASTM and ISO standards. For instance:
- Physical properties (length, width, thickness, tensile strength, elongation) are measured directly.
- Freedom from holes is determined by leakage tests per ASTM D5151-19.
- Powder residue is measured per ASTM D6124-06.
- Resistance to permeation by chemotherapy and opioid drugs is determined by measuring breakthrough time according to ASTM D6978-05.
- Biocompatibility is assessed through standardized biological tests (irritation, sensitization, systemic toxicity) as per ISO 10993 series.
8. The sample size for the training set
This section is Not Applicable (N/A). The device is a physical product (medical glove) and does not involve machine learning models that require training sets.
9. How the ground truth for the training set was established
This section is Not Applicable (N/A), as it refers to a training set for machine learning, which is irrelevant for this medical device.
Ask a specific question about this device
(123 days)
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (
Black), Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Black), Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Orange) are made from synthetic rubber latex. It is single use and powder-free. The subject device is a patient examination glove made from nitrile latex compound, Black and Orange color, powder free and non-sterile (Per 21 CFR 880.6250, class I). The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. The available sizes are Small, Medium, Large ,X-Large and XX-Large.
Here's a breakdown of the acceptance criteria and the study information based on the provided text, formatted as requested:
1. A table of acceptance criteria and the reported device performance
The provided document lists various performance criteria for the Nitrile Powder Free Examination Gloves (Black and Orange) and reports the results against these criteria.
Characteristic / Test | Acceptance Criteria | Reported Device Performance (Black Gloves) | Reported Device Performance (Orange Gloves) |
---|---|---|---|
Freedom From Holes | Meet requirement inspection level G-1, AQL 2.5 Sampling size 125pcs (Ac: 7, Re: 8) | Pass (≤3pcs) | Pass (≤4pcs) |
Dimension: Length | 220 mm min (small) | ||
230 mm min (other sizes) | 241 mm min | 240 mm min | |
Dimension: Width (S) | 80 ± 10 mm | average 85.5mm | average 85.8mm |
Dimension: Width (M) | 95 ± 10 mm | average 96.3mm | average 96.1mm |
Dimension: Width (L) | 110 ± 10 mm | average 106.4mm | average 106.8mm |
Dimension: Width (XL) | 120 ± 10 mm | average 115.5mm | average 115.9mm |
Dimension: Width (XXL) | 130 ± 10 mm | average 125.1mm | average 125.6mm |
Dimension: Thickness (Palm) | Minimum 0.05 mm | 0.178mm min. | 0.181mm min. |
Dimension: Thickness (Finger) | Minimum 0.05 mm | 0.199mm min. | 0.212mm min. |
Physical Properties (Before Aging): Tensile Strength | 14 MPa, min. | 15.1 MPa, min. | 15.3 MPa, min. |
Physical Properties (Before Aging): Elongation | 500%, min. | 518%, min. | 502%, min. |
Physical Properties (After Aging): Tensile Strength | 14 MPa, min. | 15.0 MPa, min. | 14.1 MPa, min. |
Physical Properties (After Aging): Elongation | 400%, min. | 460%, min. | 462%, min. |
Residual Powder Content | Not more than 2 mg per glove | average 0.32mg | average 0.30mg |
Biocompatibility: Primary Skin Irritation | Not a primary skin irritant | Pass (Conclusion: The test result showed that the response of the test article was categorized as negligible under the test condition) | Pass (Conclusion: The test result showed that the response of the test article was categorized as negligible under the test condition - implied similar from the text for both colors) |
Biocompatibility: Skin Sensitization | Not a contact sensitizer | Pass (Conclusion: Under the conditions of this study, the test article showed no significant evidence of causing skin sensitization in the guinea pig) | Pass (Conclusion: Under the conditions of this study, the test article showed no significant evidence of causing skin sensitization in the guinea pig - implied similar from the text for both colors) |
Biocompatibility: Acute Systemic Toxicity | No systemic toxicity | Pass (Conclusion: Under the conditions of this study, there was no evidence of acute systemic toxicity from the extract. The test article extract met the requirements of the study.) | Pass (Conclusion: Under the conditions of this study, there was no evidence of acute systemic toxicity from the extract. The test article extract met the requirements of the study. - implied similar from the text for both colors) |
Chemotherapy Drugs & Fentanyl Citrate Permeation (Minimum Breakthrough Detection Time in minutes) | As per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. (Specific breakthrough times vary by drug) | ||
Carmustine (BCNU) | Not explicitly stated as a general minimum, but the document highlights extremely low permeation times for this drug and warns against its use. | 21.6 minutes | 22.7 minutes |
Cyclophosphamide | >240 min. | >240 min. | >240 min. |
Doxorubicin HCl | >240 min. | >240 min. | >240 min. |
Etoposide | >240 min. | >240 min. | >240 min. |
Fluorouracil | >240 min. | >240 min. | >240 min. |
Mitomycin C | >240 min. | >240 min. | >240 min. |
Oxaliplatin | >240 min. | >240 min. | >240 min. |
Paclitaxel | >240 min. | >240 min. | >240 min. |
Thiotepa | Not explicitly stated as a general minimum, but the document highlights extremely low permeation times for this drug and warns against its use. | 28 minutes | 37.1 minutes |
Fentanyl Citrate Injection | >240 min. | >240 min. | >240 min. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Physical/Dimensional/Chemical Tests: For "Freedom From Holes," the sampling size is stated as 125 pieces (Ac: 7, Re: 8). The specific sample size for other physical/dimensional tests (length, width, thickness, tensile strength, elongation, residual powder content) is not explicitly stated, but the tests are reported as meeting ASTM D6319-19 and ASTM D6124-06 requirements. For Biocompatibility tests, no specific sample size is given, but the studies involve guinea pigs and rats, implying animal testing was conducted.
- Sample Size for Chemotherapy Drug Permeation Tests: Not explicitly stated within the document. The tests are reported as per ASTM D6978-05 (Reapproved 2019).
- Data Provenance: The manufacturing company is Dehai (Shandong) Medical Gloves, Co., Ltd. located in Zibo City, Shandong Province, China. The testing would presumably have been conducted in a laboratory, potentially in China or a third-party testing facility. The document doesn't specify if the data is retrospective or prospective, but testing of device performance characteristics is generally prospective for regulatory submissions.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a medical glove, and its performance is evaluated against established physical, chemical, and biological standards (ASTM, ISO). There is no "ground truth" established by human experts in the way an AI diagnostic algorithm would require. The "ground truth" is defined by the objective metrics and thresholds specified in the referenced ASTM and ISO standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing consensus ground truth in studies involving human interpretation (e.g., radiology reads). This document pertains to physical and chemical testing of a medical device against predefined standards, which does not involve such adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document is for a medical glove, not an AI-assisted diagnostic device. Therefore, no MRMC study or AI assistance evaluation was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a medical glove, not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance is based on established industry standards and objective material properties. Specifically:
- ASTM D6319-19: Standard specification for Nitrile Examination Gloves.
- ASTM D5151-19: Standard test methods for detection of holes in medical gloves.
- ASTM D6124-06: Standard test method for residual powder on medical gloves.
- ASTM D6978-05 (Reapproved 2019): Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
- ISO 10993-23, ISO 10993-10, ISO 10993-11: Biocompatibility standards for medical devices (Primary Skin Irritation, Skin Sensitization, Acute Systemic Toxicity).
These standards define the acceptable range or threshold for each characteristic (e.g., minimum tensile strength, maximum powder residue, minimum breakthrough time for drugs).
8. The sample size for the training set
Not applicable. This device is a medical glove, not a machine learning model, so there is no training set.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(126 days)
Biodegradable Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs, Fentanyl
The glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with Chemotherapy drugs, Fentanyl Citrate and select other drugs in accordance with ASTM D6978.
Biodegradable Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs, Fentanyl Citrate and select other drugs (Blue) are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small. Medium. Large. Extra Large and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs, Fentanyl Citrate and select other drugs as per ASTM D6978-05(2019). The gloves are single use, disposable, and provided non-sterile. The glove has biodegradation property within landfills tested per ASTM D5511.
The acceptance criteria for the Biodegradable Nitrile Powder Free Examination Gloves and their reported performance are detailed in the provided document.
1. Table of Acceptance Criteria and Reported Device Performance
Methodology | Test Performed | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
ASTM D6319-19 | Physical Dimensions - Length | Minimum 220mm for size XS and S; 230mm for Size M, L, XL, XXL | Pass |
ASTM D6319-19 | Physical Dimensions - Palm Width | XS: 70 ± 10mm; S: 80 ± 10mm; M: 95 ± 10mm; L: 110 ± 10mm; XL: 120 ± 10mm; XXL: 130 ± 10mm | Pass |
ASTM D6319-19 | Physical Dimensions - Thickness | Finger: 0.05mm (min); Palm: 0.05mm (min) | Pass |
ASTM D6319-19, ASTM D412-16 (2021) | Physical Properties (Tensile Strength & Elongation) | Tensile Strength (Min 14 Mpa) and Elongation (Before Aging 500% and after aging 400%) Min | Pass |
ASTM D6319-19, ASTM D5151-19 | Water leak test | AQL 2.5 (ISO 2859-1) | Pass |
ASTM D6319-19, ASTM D6124-06 (2017) | Powder Residue | Max 2mg/glove | Pass |
ASTM D6978-05 (2019) | Permeation by Chemotherapy Drugs, Fentanyl Citrate and select other drugs | Refer to the detailed tables for specific drugs and minimum breakthrough detection times (BDT) – generally >240 minutes for most drugs, with exceptions for Carmustine (11.8 min) and Thio Tepa (33.5 min) | Pass (Meets or exceeds specified BDTs, with noted exceptions) |
ISO 10993-10 & 23:2021 | Irritation and Skin Sensitization | Skin sensitization and Skin irritation | Non-sensitization and Non-irritation |
ISO 10993-5:2009 | Cytotoxicity | Cytotoxicity reactivity | Showed potential toxicity to L929 cells, but this difference does not raise different questions of safety and effectiveness compared to the predicate device. |
ISO 10993-11:2017 | Acute systemic toxicity study | Subject showed no adverse biological reaction | No evidence of acute systemic toxicity. |
2. Sample size used for the test set and the data provenance
The document does not explicitly state the exact sample sizes for each test mentioned (e.g., number of gloves tested for each specific drug permeation, or for physical properties). However, it references established ASTM and ISO standards for these tests. These standards typically define the sampling plans and statistical requirements for demonstrating compliance. The data provenance is derived from non-clinical laboratory testing (in-vitro studies) performed by the manufacturer or a contracted lab, rather than human test subjects or clinical data. There is no information on the country of origin of the data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This section is not applicable. The device is an examination glove, and its performance is evaluated through standardized physical, chemical, and biological tests (e.g., ASTM D6319, ASTM D6978, ISO 10993 series). The "ground truth" for these tests is established by the well-defined methodologies and acceptance criteria outlined in these international standards, not by expert consensus on clinical findings.
4. Adjudication method for the test set
This section is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or image interpretation tasks where human readers' opinions are combined to establish a ground truth. For the non-clinical performance testing of a medical device like examination gloves, the determination of compliance is based on objective measurements against established standard criteria, not on adjudicating expert opinions.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is not applicable. No MRMC study was conducted as this device is an examination glove, not an AI-assisted diagnostic tool. Therefore, there is no discussion of human reader improvement with or without AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
This section is not applicable. The device is an examination glove, not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.
7. The type of ground truth used
The "ground truth" for the performance evaluation of these gloves is based on standardized test methodologies and their defined acceptance criteria. This includes:
- Physical performance standards: ASTM D6319-19 for mechanical properties like length, palm width, thickness, tensile strength, and elongation.
- Leakage detection standards: ASTM D5151-19 for watertight integrity.
- Chemical permeation standards: ASTM D6978-05 (2019) for breakthrough detection time of chemotherapy drugs and other chemicals.
- Biocompatibility standards: ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), ISO 10993-11 (acute systemic toxicity), and ISO 10993-23 (skin irritation).
These standards provide objective, measurable criteria for the device's performance.
8. The sample size for the training set
This section is not applicable. The document describes the non-clinical performance testing of a physical medical device (examination gloves), not an AI or machine learning model. Therefore, there is no concept of a "training set" for an algorithm.
9. How the ground truth for the training set was established
This section is not applicable, as there is no training set for this device.
Ask a specific question about this device
(95 days)
Biodegradable Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl
The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05(2019)
Biodegradable Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05(2019). The gloves are single use, disposable, and provided non-sterile. The glove has biodegradation property tested per ASTM D5511. Biodegradability is not a medical claim and therefore was not reviewed by the FDA.
This document is a 510(k) summary for the Syntex Healthcare Products Co., Ltd.'s "Biodegradable Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue)". It describes the device, its intended use, and the non-clinical tests performed to demonstrate its substantial equivalence to predicate devices.
Here's the breakdown of the acceptance criteria and the study proving the device meets them:
1. A table of acceptance criteria and the reported device performance:
The device's performance is primarily evaluated against recognized standards for medical examination gloves and specific tests for chemical permeation.
Acceptance Criteria (Standard & Parameter) | Reported Device Performance |
---|---|
Physical Dimensions (ASTM D6319-19) | |
Length (XS, S) | Minimum 220mm (Pass) |
Length (M-XXL) | Minimum 230mm (Pass) |
Palm Width (XS) | 70±10mm (Pass) |
Palm Width (S) | 80±10mm (Pass) |
Palm Width (M) | 95±10mm (Pass) |
Palm Width (L) | 110±10mm (Pass) |
Palm Width (XL) | 120±10mm (Pass) |
Palm Width (XXL) | 130±10mm (Pass) |
Thickness (Finger) | Minimum 0.05mm (Pass) |
Thickness (Palm) | Minimum 0.05mm (Pass) |
Physical Properties (ASTM D6319-19, ASTM D412-16(2021)) | |
Tensile Strength (Before Aging) | 14MPa, min (Pass) |
Ultimate Elongation (Before Aging) | 500%, min (Pass) |
Tensile Strength (After Accelerated Aging) | 14MPa, min (Pass) |
Ultimate Elongation (After Accelerated Aging) | 400%, min (Pass) |
Water Leak Test (ASTM D6319-19, ASTM D5151-19) | |
AQL | 2.5 (Pass) |
Powder Residue (ASTM D6319-19, ASTM D6124-06 (2017)) | |
Max Powder/glove | ≤ 2 mg/glove (Pass) |
Permeation by Chemotherapy Drugs (ASTM D6978-05 (2019)) | |
Azacytidine (25 mg/ml) | >240 minutes (Pass) |
Bleomycin Sulfate (15mg/ml) | >240 minutes (Pass) |
Busulfan (6mg/ml) | >240 minutes (Pass) |
Carboplatin (10mg/ml) | >240 minutes (Pass) |
Carmustine (3.3 mg/ml) | 12.9 minutes |
Chloroquine (50mg/ml) | >240 minutes (Pass) |
Cisplatin (1mg/ml) | >240 minutes (Pass) |
Cyclophosphamide (20mg/ml) | >240 minutes (Pass) |
Cyclosporin A (100 mg/ml) | >240 minutes (Pass) |
Cytarabine HCL (100 mg/ml) | >240 minutes (Pass) |
Dacarbazine (10 mg/ml) | >240 minutes (Pass) |
Daunorubicin HCL (5 mg/ml) | >240 minutes (Pass) |
Docetaxel (10 mg/ml) | >240 minutes (Pass) |
Doxorubicin HCL (2 mg/ml) | >240 minutes (Pass) |
Etoposide (20 mg/ml) | >240 minutes (Pass) |
Epirubicin HCL (2 mg/ml) | >240 minutes (Pass) |
Fludarabine (25 mg/ml) | >240 minutes (Pass) |
Fluorouracil (50mg/ml) | >240 minutes (Pass) |
Gemcitabine (38mg/ml) | >240 minutes (Pass) |
Idarubicin HCL (1mg/ml) | >240 minutes (Pass) |
Ifosfamide (50mg/ml) | >240 minutes (Pass) |
Irinotecan (20mg/ml) | >240 minutes (Pass) |
Mechlorethamine HCI (1mg/ml) | >240 minutes (Pass) |
Melphalan (5mg/ml) | >240 minutes (Pass) |
Methotrexate (25mg/ml) | >240 minutes (Pass) |
Mitomycin C (0.5mg/ml) | >240 minutes (Pass) |
Mitoxantrone HCL (2mg/ml) | >240 minutes (Pass) |
Oxaliplatin (5mg/ml) | >240 minutes (Pass) |
Paclitaxel (6mg/ml) | >240 minutes (Pass) |
Paraplatin (10mg/ml) | >240 minutes (Pass) |
Retrovir (10mg/ml) | >240 minutes (Pass) |
Rituximab (10mg/ml) | >240 minutes (Pass) |
Thio Tepa (10mg/ml) | 35.6 minutes |
Topotecan (1mg/ml) | >240 minutes (Pass) |
Trisenox (1mg/ml) | >240 minutes (Pass) |
Velcade (Bortezomib) (1mg/ml) | >240 minutes (Pass) |
Vincristine Sulfate (1mg/ml) | >240 minutes (Pass) |
Vinorelbine (10 mg/ml) | >240 minutes (Pass) |
Fentanyl Citrate Injection (100 mcg/2ml) | >240 minutes (Pass) |
Biocompatibility | |
Irritation & Skin Sensitization (ISO 10993-10 & -23) | Non-sensitization and Non-irritation (Pass) |
Cytotoxicity (ISO 10993-5) | Showed potential toxicity to L929 cells |
Acute Systemic Toxicity (ISO 10993-11) | No evidence of acute systemic toxicity (Pass) |
Bioburden Study (ISO 11737-1) | No increase in bioburden levels (Pass) |
Note on Chemotherapy Permeation: For Carmustine and Thio Tepa, the reported breakthrough detection times (12.9 minutes and 35.6 minutes respectively) are explicitly noted as "extremely low permeation times" and a warning is issued: "Do not use with Carmustine and Thio Tepa." This indicates that for these specific drugs, the device does not meet an implicit "long duration" acceptance criterion and thus, limitations are clearly stated for safe use.
The study that proves the device meets (or defines limitations for) these criteria is a Summary of Non-Clinical Performance Data (Section 7 and 10 of the provided document).
2. Sample size used for the test set and the data provenance:
The document does not explicitly state the sample sizes used for each specific non-clinical test (e.g., number of gloves for physical dimensions, individual tests for chemotherapy permeation, or biological evaluations). However, the tests are performed according to recognized international standards (ASTM and ISO), which typically specify appropriate sample sizes for testing to ensure statistical validity.
The data provenance is from Syntex Healthcare Products Co., Ltd., located at No. 1, Fanjiazhuang Industrial Zone, Xinji City, Hebei Province, 052360, China. These are retrospective tests conducted by the manufacturer as part of the premarket notification (510(k)) submission process.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided in the document. For non-clinical performance data like material testing and chemical permeation, "experts" in the sense of clinical decision-makers (e.g., radiologists) are not typically involved in establishing ground truth. The ground truth is objective, defined by the parameters of the test methods (e.g., measuring glove dimensions, detecting chemical breakthrough, evaluating cellular response). The tests are performed by personnel trained in the specific methodologies, likely in a laboratory setting.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
An adjudication method (like 2+1, 3+1) is relevant for studies involving human interpretation or subjective assessments, often when establishing a ground truth from multiple reviewers. Since these are non-clinical laboratory tests with objectively measurable outcomes (e.g., min/mm measurements, detection of chemical permeation, biological response), an adjudication method is not applicable and therefore, none was used. The "ground truth" is determined by the direct results of the standardized tests.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC comparative effectiveness study was done. This type of study is completely irrelevant for a medical device such as examination gloves, which do not involve human readers or AI assistance in their function or evaluation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
No standalone algorithm-only performance assessment was done. This device is an examination glove, not an AI or software-based medical device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The ground truth for the non-clinical tests is based on objective measurements and observations according to established ASTM and ISO standards and methodologies.
- Physical properties and dimensions: Direct physical measurements.
- Chemical permeation: Detection of the breakthrough of specific chemicals through the glove material using analytical methods defined in ASTM D6978-05 (2019).
- Biocompatibility: In vitro (cytotoxicity) and in vivo (irritation, sensitization, systemic toxicity) tests with defined endpoints and criteria.
- Bioburden: Microbiological testing to quantify microorganism levels.
8. The sample size for the training set:
Not applicable. This device is not an AI/machine learning device, so there is no "training set." The data presented is from non-clinical performance testing of the final product.
9. How the ground truth for the training set was established:
Not applicable. As there is no training set for an AI/machine learning model, the concept of establishing ground truth for a training set does not apply here.
Ask a specific question about this device
(12 days)
Nitrile Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (
The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05(2019)
Nitrile Powder Free Examination Gloves, Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Dark Blue) Are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes - Extra Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05(2019). The gloves are single use, disposable, and provided non-sterile.
Here's a breakdown of the acceptance criteria and study information for the Nitrile Powder Free Examination Gloves, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
Methodology | Test Performed | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
ASTM D6319-19 | Physical Dimensions Length | Minimum 230mm for all sizes | Pass |
ASTM D6319-19 | Physical Dimensions Palm Width (XS, S, M, L, XL, XXL) | XS: 70±10mm, S: 80±10mm, M: 95±10mm, L:110±10mm, XL: 120±10mm, XXL: 130±10mm | Pass |
ASTM D6319-19 | Physical Dimensions Thickness (Finger, Palm) | Finger: 0.05mm (min), Palm: 0.05mm (min) | Pass |
ASTM D6319-19, ASTM D412-16(2021) | Physical Properties (Tensile Strength, Elongation) | Tensile Strength (Min 14 MPa) and Elongation (Before Aging 500% and after aging 400%) Min AQL 2.5 (ISO 2859-1) | Pass |
ASTM D6319-19, ASTM D5151-19 | Water leak test | AQL 2.5 (ISO 2859-1) | Pass |
ASTM D6319-19, ASTM D6124-06 (2017) | Powder Residue | Max 2mg/glove | Pass |
ASTM D6978-05 (2019) | Permeation by Chemotherapy Drugs | Refer to the table below for specific drug BDTs | Pass (Based on individual BDTs) |
ISO 10993-10:2010 | Irritation and Skin Sensitization | No Skin sensitization and Skin irritation | Is non-sensitization and Non-irritation |
ISO 10993-5:2009 | Cytotoxicity | No Cytotoxicity reactivity | showed potential toxicity to L929 cells. (Note: Cytotoxicity concern was addressed by acute systematic toxicity testing to still meet acceptance) |
ISO 10993-11:2017 | Acute systemic toxicity study | Subject showed no adverse biological reaction | no evidence of systemic toxicity |
Chemotherapy Drug and Fentanyl Citrate Permeation (Breakthrough Detection Time - BDT)
Chemotherapy Drug | Minimum Breakthrough Detection Time (BDT) in Minutes (Acceptance Criteria) | Reported Device Performance (BDT in Minutes) |
---|---|---|
Carmustine 3.3 mg/ml (3,300 ppm) | Explicit criteria not listed, but compared to predicate and results are 11.2 min, with a caution against use. | 11.2 |
Cisplatin 1mg/ml (1,000 ppm) | >240 (Implicitly, as predicate is >240 and proposed is same) | >240 |
Cyclophosphamide 20mg/ml (20,000 ppm) | >240 (Implicitly) | >240 |
Dacarbazine 10 mg/ml (10,000 ppm) | >240 (Implicitly) | >240 |
Doxorubicin HCL, 2 mg/ml (2,000 ppm) | >240 (Implicitly) | >240 |
Etoposide, 20 mg/ml (20,000 ppm) | >240 (Implicitly) | >240 |
Fluorouracil, 50mg/ml (50,000ppm) | >240 (Implicitly) | >240 |
Methotrexate, 25mg/ml (25,000ppm) | >240 (Implicitly) | >240 |
Paclitaxel, 6mg/ml (6,000ppm) | >240 (Implicitly) | >240 |
Thiotepa, 10mg/ml (10,000ppm) | Explicit criteria not listed, but compared to predicate and results are 29.4 min, with a caution against use. | 29.4 |
Fentanyl Citrate Injection (100 mcg/2ml) | >240 (Implicitly) | >240 |
Note on Permeation Acceptance: For chemotherapy drugs where the predicate device showed ">240 minutes", the proposed device also achieving ">240 minutes" indicates it meets or exceeds the previous performance. For Carmustine and Thiotepa, while the BDT is lower, the acceptance is ultimately related to the explicit warning not to use them with these drugs, indicating that the observed breakthrough times are treated as insufficient for safe handling.
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes used for each individual test against the acceptance criteria (e.g., number of gloves tested for each physical dimension, water leak, or permeation). However, it implies standardized testing according to relevant ASTM and ISO standards. The data provenance is industrial testing, most likely performed in a lab setting, to comply with the specified ASTM and ISO standards.
- Data Provenance: The device manufacturer, Better Care Plastic Technology Co., Ltd. (located in Shenze County, Hebei, China), would have overseen these tests, either internally or through a qualified testing laboratory. The data is thus of industrial origin, specifically from the manufacturer's testing or a third-party lab they used.
- Retrospective/Prospective: These are likely prospective tests conducted specifically for this 510(k) submission to demonstrate compliance with the standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This information is not applicable to this type of device and study. The "ground truth" for examination gloves is established by objective, standardized laboratory tests (e.g., measuring physical dimensions, tensile strength, breakthrough detection time for chemicals, biological reactivity as per ISO standards). It does not involve human expert consensus on interpretations of images or clinical outcomes.
4. Adjudication method for the test set
This is not applicable. Adjudication methods like "2+1" or "3+1" are typically used in clinical studies or image-based diagnostic device evaluations where expert opinions need to be reconciled. For the physical and chemical testing of gloves, the results are quantitative measurements or direct observations (e.g., pass/fail for a water leak test) and do not require expert adjudication in the same sense.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This is not applicable. An MRMC study assesses the performance of human readers (e.g., radiologists, pathologists) in interpreting medical data, often with and without AI assistance, across a range of cases. This device (examination gloves) is a physical product and does not involve human readers or AI assistance in its primary function, nor in the testing performed for its clearance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This is not applicable. This question pertains to AI/algorithm performance. The device is a physical product, not an algorithm.
7. The type of ground truth used
The "ground truth" for the performance of these gloves is defined by:
- Standardized Test Methods and Specifications: Specific values and criteria (e.g., minimum tensile strength, maximum powder residue, minimum breakthrough detection time) defined by national and international standards like ASTM D6319-19, ASTM D6978-05, ISO 10993 series.
- Direct Measurement and Observation: The device's physical properties and chemical resistance are directly measured and observed in laboratory settings according to these established protocols.
- Biological Reactivity Endpoints: For biocompatibility tests (irritation, sensitization, cytotoxicity, systemic toxicity), the "ground truth" is determined by established biological endpoints and criteria within the ISO 10993 standards.
8. The sample size for the training set
This is not applicable. This question refers to the training data for an AI algorithm. This device is a physical medical product, and no AI algorithm development is described or relevant to its clearance.
9. How the ground truth for the training set was established
This is not applicable for the reasons stated in point 8.
Ask a specific question about this device
(86 days)
Disposable Nitrile Powder-Free Examination Gloves (Tested for Use with Chemotherapy Drugs)
A patient examination gloves is a disposable device intended for medical purpose that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05 Standard Practice for Assessment of Medical gloves to Permeation by Chemotherapy Drugs.
The subject device is single use, disposable gloves intended for medical purposes to be worn on the examiner's hands to prevent contamination between patient and examiner. The gloves are powder-free, ambidextrous with beaded cuff, blue colored, nitrile, and tested for use with chemotherapy drugs. The gloves are offered in four sizes: extra-small, small, medium, large, and extra-large.
The provided document describes the FDA 510(k) premarket notification for "Disposable Nitrile Powder-Free Examination Gloves (Tested for Use with Chemotherapy Drugs)". This document is for a medical device (gloves), not an AI/Software as a Medical Device (SaMD). Therefore, the concepts of acceptance criteria for AI performance, training/test sets, ground truth establishment by experts, MRMC studies, standalone algorithm performance, or a sample size for training data are not applicable to this submission.
The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the physical and chemical performance standards for the gloves, particularly their resistance to chemotherapy drugs and biocompatibility.
Here's the information extracted from the provided text, tailored to the context of a physical medical device:
Acceptance Criteria and Device Performance for Disposable Nitrile Powder-Free Examination Gloves
The acceptance criteria for the Disposable Nitrile Powder-Free Examination Gloves are based on compliance with established international and national standards for medical gloves, particularly regarding their physical properties, freedom from holes, residual powder content, and resistance to permeation by chemotherapy drugs.
1. Table of Acceptance Criteria and Reported Device Performance
No. | Test Methodology / Standard | Purpose | Acceptance Criteria | Reported Device Performance |
---|---|---|---|---|
1 | ISO 10993-10:2010 (Irritation & Skin Sensitization) | Assess possible contact hazards from chemicals released from medical devices (skin/mucosal irritation, eye irritation, skin sensitization). | Skin Sensitization Test: Grades less than 1, otherwise sensitization. | |
Skin Irritation Test: Primary irritation index 0-0.4 (Negligible). | Skin Sensitization Test: All grades are 0. All animals survived with no abnormal signs. | |||
Skin Irritation Test: Primary irritation index is 0. Response categorized as negligible. | ||||
2 | ISO 10993-5:2009 (In Vitro Cytotoxicity) | Assess in vitro cytotoxicity of medical devices. | Viability reduced to 240 Minutes |
- Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes
- Dacarbazine 10 mg/ml: > 240 Minutes
- Doxorubicin 2.0 mg/ml: > 240 Minutes
- Etoposide 20.0 mg/ml: > 240 Minutes
- Fluorouracil 50.0 mg/ml: >240 Minutes
- Paclitaxel 6.0 mg/ml: >240 Minutes
- Thio Tepa 10.0 mg/ml: 46.8Minutes (min.) | Reported Breakthrough Detection Times: (Matches acceptance criteria values exactly, indicating compliance).
- Carmustine (BCNU) 3.3 mg/ml: 22.8(23.7, 22.8, 23.1) Minutes
- Cisplatin 1.0 mg/ml: > 240 Minutes
- Cyclophosphamide (Cytoxan) 20.0 mg/ml: > 240 Minutes
- Dacarbazine 10 mg/ml: > 240 Minutes
- Doxorubicin 2.0 mg/ml: > 240 Minutes
- Etoposide 20.0 mg/ml: > 240 Minutes
- Fluorouracil 50.0 mg/ml: >240 Minutes
- Paclitaxel 6.0 mg/ml: >240 Minutes
- Thio Tepa 10.0 mg/ml: 46.8(48.2, 48.6, 46.8) Minutes.
Warning: Low permeation times for Carmustine (BCNU) and Thio Tepa are noted, with a warning not to use with these specific drugs given their breakthrough times. |
2. Sample Size Used for the Test Set and Data Provenance
The document describes material and performance testing, not a "test set" in the sense of an AI model's dataset. The "samples" refer to physical gloves selected for testing.
The provenance of the data is from laboratory testing conducted on the manufactured gloves. The specific country of origin for the testing itself is not explicitly stated, but the manufacturer is Shandong Maida Medical Technology Co.,Ltd. in China. The testing is prospective in the sense that newly manufactured gloves are tested to ensure they meet the standards.
- ASTM D5151 (Detection of Holes): Sample size for water leakage test: 125 gloves.
- ASTM D6319 (Dimensions & Physical Properties): Results are shown for multiple lots (e.g., Lot no.: 210515, 210518, 210520), indicating multiple samples were tested across different manufacturing batches. The exact number of gloves per lot for each physical test (e.g., tensile strength, elongation) is not specified but would be defined by the ASTM standard.
- ASTM D6978 (Chemotherapy Permeation): The results provide triplet values for Carmustine and Thio Tepa (e.g., 22.8(23.7, 22.8, 23.1)), suggesting three independent measurements were taken for these drugs. The standard itself specifies the number of replicates.
- Biocompatibility Testing: The number of animal subjects used for the animal irritation, sensitization, and systemic toxicity tests is not specified in the summary but would conform to the respective ISO standards.
3. Number of Experts Used to Establish Ground Truth and Qualifications
This section is not applicable as the evaluation involves standardized laboratory testing against defined physical and chemical properties, not human expert interpretation of subjective data or images. The "ground truth" is defined by the objective measurement results according to the specified ASTM and ISO standards.
4. Adjudication Method for the Test Set
This is not applicable. Adjudication methods (like 2+1, 3+1) are for resolving discrepancies in human expert assessments (e.g., in medical image interpretation). Here, numerical test results from standardized methods are directly compared against established quantitative or qualitative acceptance criteria.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
This is not applicable. MRMC studies are used to evaluate diagnostic imaging systems or AI tools where human readers are involved. This submission is for a physical medical glove.
6. Standalone (Algorithm Only) Performance
This is not applicable. This concept applies to AI/software performance. The device is a physical glove, not an algorithm.
7. Type of Ground Truth Used
The ground truth used is based on objective, quantitative measurements derived from standardized laboratory tests (bench testing) and biological responses observed in validated biocompatibility assays according to international standards (ASTM and ISO). This includes:
- Quantitative Thresholds: e.g., minimum tensile strength, maximum residual powder, minimum breakthrough time for chemotherapy drugs.
- Qualitative Observations (Biological): e.g., absence of irritation or sensitization, absence of systemic toxicity.
- Conformity to Standards: The device's performance is compared directly to the specified requirements within each standard (e.g., ASTM D6319-19 dimensions, physical properties).
One point of note is the cytotoxicity result (17.1% viability) which is below the 70% threshold, indicating cytotoxicity. The document addresses this by stating: "Analysis3: The proposed device is potentially cytotoxic, but all proposed devices are conducted the systemic toxicity test." And the systemic toxicity test showed "no evidence of systemic toxicity." This implies that despite the in vitro cytotoxicity, the manufacturer argues that in vivo systemic toxicity (which is generally considered a more critical endpoint for patient safety) was not observed, suggesting the findings are acceptable for the device's intended use according to the biocompatibility evaluation framework.
8. Sample Size for the Training Set
This is not applicable. There is no "training set" in the context of a physical device. The manufacturing process is controlled to produce gloves consistently, and batch testing ensures quality.
9. How the Ground Truth for the Training Set Was Established
This is not applicable. As there is no training set, the establishment of ground truth for it is irrelevant. The manufacturing process is designed and verified to produce products meeting the specified design and performance criteria.
Ask a specific question about this device
(58 days)
Disposable Nitrile Powder-Free Examination Gloves
The Disposable Nitrile Powder-Free Examination Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.
The proposed device is Powder Free Disposable Nitrile Powder-Free Examination Gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The proposed device is non-sterile.
The document provided is a 510(k) summary for Disposable Nitrile Powder-Free Examination Gloves. It details the safety and effectiveness information, comparing the proposed device to a legally marketed predicate device. The information provided heavily focuses on non-clinical performance testing rather than studies involving human readers or AI.
Based on the provided text, here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance:
No. | Name of the Test Methodology / Standard | Acceptance Criteria | Results |
---|---|---|---|
1 | ISO 10993-10:2010 (Skin Sensitization) | Skin Sensitization Test: provided grades less than 1, otherwise sensitization. | All grades are 0. All animals were survived and no abnormal signs were observed during the study. |
2 | ISO 10993-10:2010 (Skin Irritation) | Skin Irritation Test: If the primary irritation index is 0-0.4, the response category is Negligible. 0.5-1.9 means slight, 2-4.9 means moderate, 5-8 means severe. | The primary irritation index is 0. The response of the proposed device was categorized as negligible under the test condition. |
3 | ISO 10993-5:2009 (In Vitro Cytotoxicity) | The viab.% of the 100% extract of the test article is the final result, and if viability is reduced to |
Ask a specific question about this device
Page 1 of 6