Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K211619
    Date Cleared
    2022-12-29

    (582 days)

    Regulation Number
    870.1025
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Multi-Parameter Patient Monitor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NC10 and NC12 patient monitors are intended to be used for monitoring, displaying, alarming and storing of multiple physiological parameters These parameters include ECG (3-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), temperature (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), carbon dioxide (CO2), anesthetic gas (AG), and Bispectral index (BIS) for a single patient.

    All parameters can be monitored on single adult, pediatric, and neonatal patients except:

    • · BIS monitoring is intended for adult and pediatric patients only;
    • · C.O. monitoring is restricted to adult patients only;
      · Arrhythmia analysis is intended to use on adult patients only and is not intended and shall not be used on pediatric and neonatal population.
    • · When using COMEM SpO2, the monitor is intended to be used on adult patients only.
    • · NIBP measurement continual mode is not applicable to neonates.

    The monitors are to be used in general healthes by clinical physicians or appropriate medical staff under the direction of physicians.

    The monitors are not intended for helicopter transport, hospital ambulance, or home use.

    The monitors do not measure, display, or trend changes in the ST segment.

    The monitors do not intend for use as apnea monitors.

    The monitors are not intended for use in MRI or CT environments.

    The monitors are not used on patients who have a demonstrated need for cardiac monitoring known arrhythmias of VT, Accelerated Idioventricular rhythm and Torsades de Pointes.

    Device Description

    The NC10 and NC12 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing multiple physiological parameters. These parameters include ECG (3-lead, 5-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), temperature (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), cardiac output (C.O.), carbon dioxide (CO2), anesthetic gas (AG), and Bispectral index (BIS) for a single patient.
    All parameters can be monitored on single adult, pediatric, and neonatal patients except:
    BIS monitoring is intended for adult patients only; C.O. monitoring is restricted to adult patients only; Arrhythmia analysis is intended for use with adult patients only and is not intended and shall not be used on pediatric and neonatal population. When using COMEM SpO2, the monitor is intended to be used on adult patients only. NIBP measurement continual mode is not applicable to neonates. Both models are designed with:
    Same system framework and components
    Same hardware design principle
    Same software platform
    Same parameters measurement subsystems (including parameters modules and accessories)
    The only difference between NC10 and NC12 is the display size.

    AI/ML Overview

    The acceptance criteria and supporting study details for the Multi-Parameter Patient Monitor (NC10 and NC12) are provided below, based on the given FDA 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present specific "acceptance criteria" for each physiological parameter in a tabular format with corresponding "reported device performance." Instead, it compares the subject device's specifications to those of the predicate device, stating that the subject device's performance aligns with or is a subset of the predicate's performance, and that the device meets relevant consensus standards. The "Comparison" column in the provided tables indicates "Same" for most parameters, implying that the subject device's performance is equivalent to the established performance of the predicate device. For the "Comen SpO2" feature, where there's a difference, the document states, "The SpO2 accuracy met ISO 80601-2-61 and was validated by the clinical study," indicating that its performance meets the standard.

    Here's a condensed representation of the key performance specifications for the subject device (NC10 and NC12), which also serve as implied acceptance criteria given the "Same" comparison to the predicate:

    ParameterMethod / Range / Accuracy (NC10 & NC12)
    ECG (Arrhythmia Analysis)Asystole, ventricular fibrillation, R ON T, VT>2, Couplet, PVC, Bigeminy, Trigeminy, Brady, PNC, PNP, Missed Beats, Heart Pause, Irregular Heart Beat, VTAC, Tachy, PVCs Too High, Extreme Tachycardia, Extreme Bradycardia, Ventricular Rhythm.
    RespirationMethod: Trans-thoracic impedance; Range: adult:0-120 rpm, pediatrics:0-150rpm, neonate:0-150rpm; Accuracy: 7 to 150rpm: ±2rpm or ±2%, whichever is greater.
    SpO2 (Masimo)Method: red and infrared light; Range: 1~100%; Accuracy: No motion: 70-100% ±2%(adult/pediatric), 70-100% ±3%(neonate); Motion: 70-100% ±3%.
    SpO2 (Nellcor)Method: red and infrared light; Range: 0~100%; Accuracy: 70-100% ±2%(adult/pediatric), 70-100% ±3%(neonate).
    SpO2 (Comen)Method: red and infrared light; Range: 0~100%; Accuracy: 70-100% ±3% (adult). (Note: Restricted to adult use compared to predicate)
    Pulse Rate (from SpO2, IBP, NIBP)Ranges: 20-350 bpm (variable by source); Accuracy: ±1bpm to ±5bpm or ±1% to ±3% (variable by source and conditions).
    NIBPMethod: Oscillometry; Range: Adult: systolic:40-270 mmHg, diastolic:10-215 mmHg; Pediatrics: systolic:40-200 mmHg, diastolic:10-150 mmHg; Neonate: systolic:40-135 mmHg, diastolic:10-100 mmHg; Error: Max mean error: ±5 mmHg, Max standard deviation: 8 mmHg.
    TemperatureMethod: Thermal resistance; Range: 0-50°C; Accuracy: ±0.1°C.
    CO2 (Masimo)Method: Infrared absorption; Range: 0-190mmHg; AwRR:0-150rpm; Accuracy: 0-114mmHg: ±(2.25mmHg+reading×4%).
    CO2 (Respironics)Method: Infrared absorption; Range: 0-150mmHg; AwRR: 0, 2-150bpm; Accuracy: 0-40mmHg: ±2mmHg, other ranges higher % errors.
    IBPMethod: Direct invasive measurement; Range: -50 to 300 mmHg; Accuracy: ±2% or ±1 mmHg, whichever is greater (excluding sensor error).
    Cardiac OutputMethod: Thermodilution; Range: 0.1 to 20 L/min; Accuracy: ±5% or ±0.1 L/min, whichever is greater.
    Anesthetic Gas (AG)Method: Infrared absorption; Ranges for various gases; Accuracy for CO2, N2O, Hal, Enf, Iso, Sev, Des, O2, awRR.
    BISRange and Accuracy: SQI: 0-100%, 1%; EMG: 0100dB, 1%; BIS: 0-100, 1%; SR: 0100%, 1%; ESR: 0~100%, 1%.

    The document implies that the "reported device performance" for the subject device meets or is equivalent to these specified ranges and accuracies through bench testing and clinical studies, confirming compliance with relevant standards.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify exact numerical sample sizes for each clinical test. It mentions that clinical accuracy of NIBP, SpO2, and respiratory rate were validated for the intended patient population.
      • For SpO2 accuracy, it states the validation was done "using the method outlined in ISO 80601-2-61:2017 and the FDA guidance Pulse Oximeters - Premarket Notification Submissions [510(k)s]: Guidance for Industry and Food and Drug Administration Staff, March 2013." These standards typically require a certain number of subjects (often healthy volunteers) with induced hypoxemia for desaturation studies to demonstrate accuracy across the specified range. However, the exact number is not provided in this summary.
      • For NIBP accuracy, it states validation was "according to ISO 81060-2 which contains the requirements for clinical accuracy and the protocols for investigating the NIBP determination clinical accuracy." This standard also prescribes specific subject enrollment criteria and measurement methods.
      • For Respiratory Rate (RR) accuracy, it was validated "by clinical testing to compare the measurement of the subject device and that of a clinician-scored capnography device, manually scored end-tidal CO2 (EtCO2) capnography." The sample size for this is not detailed.
    • Data Provenance: The document does not explicitly state the country of origin of the data. It also does not explicitly state whether the studies were retrospective or prospective, though clinical validation studies for device clearance are typically prospective. It does say "All clinical accuracy validation studies were conducted in accordance with standard ISO 14155:2020," which governs clinical investigation of medical devices, generally implying prospective collection.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide details on the number or qualifications of experts used for establishing ground truth, as it is a multi-parameter patient monitor.

    • For SpO2, the ground truth would typically be established by a CO-oximeter reading during a controlled desaturation study, as per ISO 80601-2-61. This is a highly objective measurement.
    • For NIBP, ground truth is typically established by direct intra-arterial blood pressure measurements, not by expert consensus.
    • For Respiratory Rate, the ground truth was "clinician-scored capnography device, manually scored end-tidal CO2 (EtCO2) capnography." This implies clinically trained personnel, but their specific qualifications or number are not provided.
    • For Arrhythmia Analysis, and other subjective physiological monitoring parameters, the ground truth source is not explicitly mentioned but typically relies on expert interpretation of ECG waveforms or other data.

    4. Adjudication Method for the Test Set

    The document does not detail any adjudication methods (e.g., 2+1, 3+1) for the test set, as most of the parameters are quantitative measurements compared against an objective reference standard rather than subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    This document describes a multi-parameter patient monitor, which is a measurement device, not an AI-assisted diagnostic imaging tool. Therefore, an MRMC comparative effectiveness study comparing human readers with and without AI assistance is not applicable to this type of device and was not conducted. The study aims to demonstrate that the device's measurements are accurate and equivalent to predicate devices, not to show an improvement in human reader performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    The entire device, including its algorithms for parameter measurement and arrhythmia detection, operates in a "standalone" fashion to generate the values and alarms displayed to the clinician. The performance validated (e.g., accuracy of SpO2, NIBP, RR, arrhythmia detection) is the inherent performance of the device's algorithms and hardware. While a human uses the device and interprets its output, the core measurements are algorithm-driven.

    7. The Type of Ground Truth Used

    • SpO2: CO-oximetry in a controlled desaturation study (objective, gold-standard reference for SpO2 saturation).
    • NIBP: Direct intra-arterial blood pressure measurements (objective, gold-standard).
    • Respiratory Rate: Clinician-scored capnography device, manually scored end-tidal CO2 (EtCO2) capnography. (This suggests an expert-derived observation from an objective measurement, or comparison to another well-established measurement device).
    • ECG/Arrhythmia Detection and other parameters: The document implies comparison to established methods and compliance with relevant ISO standards, which would typically involve highly accurate reference measurements and possibly expert review of waveforms for specific event detection.

    8. The Sample Size for the Training Set

    The document does not provide information about a training set or its sample size. This is common for device clearances that focus on performance validation rather than machine learning algorithm development where distinct training and test sets are crucial. The device's algorithms are likely based on established physiological principles and signal processing, rather than deep learning from a massive training dataset.

    9. How the Ground Truth for the Training Set was Established

    As no specific training set is mentioned in the filing summary for this device, information regarding the establishment of its ground truth is not applicable or provided. The device's performance is demonstrated through its adherence to established international standards and clinical testing against reference methods.

    Ask a Question

    Ask a specific question about this device

    Why did this record match?
    Device Name :

    C50 and C80 Multi-parameter Patient Monitor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:

    · The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;

    · NIBP measurement continual mode is not applicable to neonates;

    · Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;

    The monitors are to be used in general healthcare facilities by clinical physicians or appropriate medical staff under the direction of physicians. The monitors are not intended for home use.

    Device Description

    The C50 and C80 patient monitors are intended to be used for monitoring, displaying, reviewing, alarming and storing of multiple physiological parameters as following: ECG (3-lead, 5-lead or 12-lead selectable, arrhythmia detection, heart rate (HR)), Respiration rate (RR), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP), temperature (Temp), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG), cardiac output (C.O.) for single patient. The C80 can also monitor the ICG (impedance cardiography) and BIS (bispectral index).

    All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
    • The ICG monitoring is applicable to the adult patients of 122229cm in height and 30159Kg (67~341 pounds) in weight only;
    • NIBP measurement continual mode is not applicable to neonates;
    • Anesthetic depth (BIS) of the multi-parameter patient monitor is not intended for neonatal patients;
    Those monitors provide patient monitoring capabilities by using corresponding accessories.
    The multi-parameter Patient monitor, model: C50 and C80 have same design principle and technical characteristics:
    But the software in the host and modules and components are different.
    The differences between C50 and C80 are ICG, BIS and size of monitor.

    AI/ML Overview

    The provided document describes the acceptance criteria and a study proving the C50 and C80 Multi-parameter Patient Monitor meets these criteria.


    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally established through adherence to recognized international standards and clinical performance metrics. The document details a comparison with a predicate device (K170876, Passport Series Patient Monitors). While explicit "acceptance criteria" for each parameter are not always separately listed with a single performance value from the new device, the "Comparison" column in the table below infers if the new device meets or exceeds the predicate's performance or if sufficient justification is provided for differences.

    FeaturePredicate Device (K170876) PerformanceSubject Device (C50 & C80) PerformanceComparison to Acceptance Criteria (Predicate) & Justification
    Monitor SizePassport 17m: 400x370x193mm; Passport 12m: 297x336x187mmC80: Approx. 344x291x165mm; C50: Approx. 291.7x250x187mmThe monitor size of the subject device is smaller than the predicate device, but it "won't affect the safety and effectiveness of the subject device." Implied acceptance: smaller size is acceptable if safety/effectiveness are not compromised.
    Integrated display & touch screenColor TFT LCD (17-inch, 1280x1024 pixels for 17m; 12-inch, 800x600 pixels for 12m)Color TFT LCD (C80: 12.1-inch, 800x600 pixels; C50: 10.4-inch, 800x600 pixels)Screen size is smaller, but resolution for C80/C50 is the same as Passport 12m. "Same. The screen of Subject device is smaller than predicate device. Complying with IEC 60601-1 and IEC 62366-1 also indicates the clinical use is safe and effective." Implied acceptance: smaller screen size is acceptable due to compliance with relevant IEC standards for safety and usability.
    Power supplyTwo rechargeable Lithium-ion batteries or AC for Passport 17m; One rechargeable Lithium-ion battery or AC for Passport 12mC50 and C80: Powered either by built-in battery or external AC."Same, both powered by battery and AC." Implied acceptance: functional equivalence.
    Battery11.1V, 4500mAh11.1V, 2200mAh/4400mAh"The battery capacity is different. The batteries of C50/C80 have complied with IEC 62133. C50 and C80 have conformed to IEC 60601-1." Implied acceptance: different capacity is acceptable as long as safety (IEC 62133, IEC 60601-1) is maintained.
    ECG3-lead, 5-lead and 12-lead selectable, heart rate (HR)3-lead, 5-lead and 12-lead selectable, heart rate (HR)"Same." Implied acceptance: functional equivalence.
    ECG (Arrhythmia Analysis)Asystole, VFib/VTac, Vtac, Vent. Brady, Extreme Tachy, Extreme Brady, PVC, Couplet, Bigeminy, Trigeminy, Ron T, Run PVCs, PVCs, Tachy, Brady, Missed Beats, Vent. Rhythm, PNP, PNC, Multif. PVC, Nonsus. Vtac, Pause, Irr. Rhythm, AFib (24 types)Asystole, ventricular fibrillation, R ON T, VT >2, Couplet, PVC, Bigeminy, Trigeminy, Brady (Bradycardia), PNC (Pacer Not Capture), PNP (Pacer Not Pace), Missed Beats, IHB (Irregular Heart Beat), VTAC (Ventricular Tachycardia), Tachy (Tachycardia), PVC Too High, Extreme Tachycardia, Extreme Bradycardia, Ventricular Rhythm, Heart Pause (20 types)"C50/C80 has 20 types of arrhythmias in total, 18 of which is same as the predicate device. The other two (underline) is supported by compliance with EC57." Implied acceptance: fewer types are acceptable as long as key arrhythmia detections are present and supported by EC57 compliance.
    RespirationMethod: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified.Method: Trans-thoracic impedance. Range: adult:0-120 rpm; pediatrics:0-150rpm; neonate:0-150rpm. Accuracy: 7 to 150rpm: ± 2rpm or ± 2%, whichever is greater. 0 to 6rpm: not specified."Same." Implied acceptance: functional and performance equivalence.
    Pulse oxygen saturation (SpO2)Method: red and infrared light method. Masimo SpO2: Range:1100%. Accuracy: No motion: 70-100% ±2%(adult/pediatric), ±3%(neonate); Motion: 70-100% ±3%. Nellcor SpO2: Range:0100%. Accuracy: 70-100% ±2%(adult/pediatric), ±3%(neonate). Mindray SpO2: Range:0~100%. Accuracy: 70-100% ±2%(adult/pediatric), ±3%(neonate).Method: same. Masimo SpO2: same range & accuracy. Nellcor SpO2: same range & accuracy. Comen SpO2: Range: 0%-100%. Accuracy: 70-100% ±2%(adult/pediatric, non-motion), ±3%(neonate, non-motion)."Same." (for Masimo and Nellcor modules). For their own Comen SpO2 module, it shows equivalent accuracy. Implied acceptance: performance equivalence with predicate's different OEM modules and new Comen module meets similar accuracy.
    Pulse rate (PR)From Mindray SpO2 (Range 20-254 bpm, Acc ±3 bpm), Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%).From Masimo SpO2 (Range 25-240 bpm, Acc ±3 bpm no motion, ±5 bpm motion), Nellcor SpO2 (Range 20-300 bpm, Acc ±3 bpm for 20-250 bpm), Comen SpO2 (Range 20-254 bpm, Acc ±2 bpm), IBP sensor (Range 25-350 bpm, Acc ±1 bpm or ±1%), NIBP sensor (Range 40-240 bpm, Acc ±3 bpm or ±3%)."The PR from Comen SpO2 is more accuracy than the PR form Mindray SpO2. The PR form Masimo SpO2, Nellcor SpO2 and IBP sensor of C50 and C80 are the same with the predicate device. In addition, C50 and C80 have one more PR data source (NIBP) than predicate device. C50 and C80 have complied with ISO 80601-2-61." Acceptance criteria met or exceeded; added NIBP source (improvement). Adherence to ISO 80601-2-61.
    Non-invasive blood pressure (NIBP)Method: Oscillometry. Range: Adult (systolic 25-290mmHg, diastolic 10-250mmHg), pediatric (systolic 25-240mmHg, diastolic 10-200mmHg), neonate (systolic 25-140mmHg, diastolic 10-115mmHg). Accuracy: Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg.Method: same. Range: Adult (systolic 40-270mmHg, diastolic 10-215mmHg), pediatric (systolic 40-200mmHg, diastolic 10-150mmHg), neonate (systolic 40-135mmHg, diastolic 10-100mmHg). Accuracy: 0-300mmHg: ±3mmHg."The NIBP measure range for adult/pediatric/neonate in C50 and C80 is smaller than that in predicate device. The C50 and C80 is more accuracy than predicate device. C50 and C80 have conformed to IEC 80601-2-30." Acceptance criteria met or exceeded in accuracy, and justification for smaller range implicitly accepted by meeting IEC 80601-2-30.
    Temperature (Temp)Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.1°C.Method: Thermal resistance. Range: 0 ~ 50°C. Accuracy: ±0.2°C."The predicate device is more accuracy than C50 and C80. C50 and C80 have complied with ISO 80601-2-56." Implied acceptance: slightly lower accuracy is acceptable as long as it complies with ISO 80601-2-56.
    Carbon dioxide (CO2)Method: Infrared absorption. Masimo CO2: Sidestream 0-99mmHg, AwRR 0-120rpm. Microstream 0-99mmHg, AwRR 0-150rpm. Accuracy: Sidestream 0-40mmHg ±2mmHg, 41-76mmHg ±5%, 77-99mmHg ±10%, AwRR ±2rpm. Microstream 0-38mmHg ±2mmHg, 39-99mmHg ±5% +0.08%, AwRR 0-70rpm ±1rpm, 71-120rpm ±2rpm, 121-150rpm ±3rpm.Method: same. Masimo CO2: Sidestream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Mainstream 0-190mmHg, 0-25% (760mmHg), AwRR 0-150rpm. Respironics CO2: Sidestream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Mainstream 0-150mmHg, 0-19.7%, AwRR 0, 2-150rpm. Accuracy (Masimo): ±(2.25mmHg +reading x 4%), AwRR ±1rpm. Accuracy (Respironics): 0-40mmHg ±2mmHg, etc., AwRR ±1rpm."The C50 and C80 are more accuracy and measured wider than predicate device. For C50 and C80, Masimo CO2 module (K123043 and K103604) and Respironics CO2 module (K040183 and K053174) have been cleared. Both modules have complied with ISO 80601-2-55." Acceptance criteria met or exceeded; wider range and higher accuracy; existing modules previously cleared and comply with ISO 80601-2-55.
    Invasive blood pressure (IBP)Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor).Method: Direct invasive measurement. Range: -50 to 300 mmHg. Accuracy: ±2% or ±1 mmHg, whichever is greater (without sensor)."Same." Implied acceptance: functional and performance equivalence.
    Cardiac output (C.O.)Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater.Method: Thermodilution method. Range: 0.1 to 20 L/min. Accuracy: ±5% or ±0.1 L/min, whichever is greater."Same." Implied acceptance: functional and performance equivalence.
    Anesthetic gas (AG)Method: Infrared absorption. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 ±0.3% ABS, N2O ±(8%REL+2%ABS), Other AG 8%REL).Method: same. Range (CO2, N2O, Hal, Enf, Iso, Sev, O2, AwRR). Accuracy (CO2 0-15%: ±(0.2kPa+readingx2%), N2O ±(2 kPa+readingx2%), Hal/Enf/Iso 0-8%: ±(0.15%+readingx5%), Sev 0-10%: ±(0.15%+readingx5%), Des 0-22%: ±(0.15%+readingx5%), O2 ±(1%+readingx2%), Masimo AG AwRR ±1rpm)."For CO2/Enf/Hal/Iso/Sev/Des, The measurement range of C50 and C80 is smaller than the predicate device. For awRR, the measurement range of C80 is better than the predicate device. The C50 and C80 are more accuracy than predicate device. The measurement range of C50 and C80 is enough for most environments; it won't affect the safety and effectiveness. The C50 and C80 supports two AG modules: MASIMO ISA AX+ Sidestream module and Masimo IRMA AX+ Mainstream module, both of them have been cleared in K103604." Acceptance criteria: different ranges are acceptable with justification of sufficient range for most environments and high accuracy. Utilizes previously cleared and compliant modules.
    BIS (C80 only)Range: BIS, BIS L, BIS R: 0-100; SQI, SQI L, SQI R: 0-100%; EMG, EMG L, EMG R: 0-100 Db; SR, SR L, SR R: 0-100%; SEF, SEF L, SEF R: 0.5-30.0 Hz; TP, TP L, TP R: 40-100 Db; BC, BC L, BC R: 0-30; sBIS L, sBIS R: 0-10.0; Semg L, Semg R: 0-10.0; ASYM: 0-100%. Accuracy: 1% for BIS, SQI, EMG, ESR (implied).Range and Accuracy: BIS: same; accuracy: 1%. SQI: same; accuracy: 1%. EMG: same; accuracy: 1%. ESR: 0-100%; accuracy: 1%."The underline parameters are what C50 and C80 doesn't have. The BIS module and sensor have been cleared by FDA and its 510k numbers are K040183 and K002734." Acceptance criteria for essential parameters are met by using previously cleared FDA modules.
    ICG (C80 only)Method: Indirect impedance cardiograph measurement. Range: SV: 5-250 ml; HR: 44-2m; C.O. 1.4-15 L/min. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified.Method: Indirect impedance cardiograph measurement. Range: HR: 40-250bpm; SV: 0-250mL; C.O.:0-30L/min; TFC: 5-150 /KΩ; SVR:0-3500 dyn・s ・ cm-5. Accuracy: SV: Not specified; HR: ±2 bpm; C.O. Not specified."The underline parameter is what C50 and C80 doesn't have. The ICG electrode cable and sensors are all cleared by FDA, 510(k) number of which is K110645." Acceptance criteria for essential parameters are met by using previously cleared FDA components.

    Study that Proves the Device Meets Acceptance Criteria

    The essential studies to prove the device meets acceptance criteria are divided into Non-Clinical Performance Data and Clinical Studies.

    Non-Clinical Performance Data:

    • Software Verification and Validation Testing: Conducted in accordance with FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Risk analysis was performed, and the software was deemed a "major" level of concern.
    • Electrical safety and Electromagnetic Compatibility (EMC): Testing conducted according to:
      • ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012 (IEC 60601-1:2005, MOD)
      • IEC 60601-1-2 Edition 4: 2014-02
    • Bench Testing: Functional and system-level tests were performed, showing the devices meet specifications and perform equivalently to the predicate.
    • Biocompatibility Testing: Performed for patient-contacting components (ECG cable, SpO2 probes, temperature probes) that were not previously cleared. Cytotoxicity, sensitization, and irritation testing were conducted based on ISO 10993-1 and FDA's 2016 biocompatibility guidance.
    • Conformance with Applicable Standards: The device demonstrated compliance with numerous recognized consensus standards, in addition to those listed above for specific modules (e.g., ISO 80601-2-55 for Respiratory Gas Monitors, ISO 80601-2-56 for Thermometers, ISO 80601-2-61 for Pulse Oximeter Equipment, ANSI AAMI EC57:2012 for Cardiac Rhythm and ST-Segment measurement algorithms).

    Clinical Studies:

    1. NIBP Clinical Study - for adults and children

    • Device Parameter: Non-invasive Blood Pressure (NIBP)
    • Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
    • Sample Size: 25 patients (19 adults, 6 children).
      • Demographics: 10 men, 15 women. 6 patients aged 3-12, 2 aged 12-40, 10 aged 40-60, 7 above 60.
    • Data Provenance: Retrospective or Prospective not explicitly stated, but implies prospective data collection from "The Second Affiliated Hospital of Guangzhou Medical University and The First Affiliated Hospital of Guangzhou Traditional Chinese Medical University" from 2012 to 2013 (China).
    • Ground Truth: Implicitly referent method (e.g., auscultation by trained observers) as required by ISO 81060-2.
    • Adjudication Method: Not specified, but standard practice for NIBP clinical validation involves multiple observers for reference measurements.
    • Results: "The results are accurate and reliable, and the repeated measurement consistency is in good condition, within the measurement range. Moreover, No adverse events and side effects were found in clinical trials."

    2. NIBP Clinical Study - for neonate and infants

    • Device Parameter: Non-invasive Blood Pressure (NIBP)
    • Acceptance Criteria/Standard: Complies with ISO 81060-2 Second Edition 2013-05-01 "Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type."
    • Sample Size: 20 subjects.
      • Demographics: 10 males, 10 females. 13 subjects
    Ask a Question

    Ask a specific question about this device

    K Number
    K173036
    Date Cleared
    2018-12-21

    (449 days)

    Product Code
    Regulation Number
    870.1025
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    North-vision Multi-parameter Patient Monitor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The North-vision Elegant-1000 Series of Multi-parameter Patient Monitor are intended to monitor, display and record physiological signs of adult, pediatric patients. With the functions of near real-time recording and displaying parameters ECG, heart rate, non-invasive blood pressure, functional oxygen saturation, it allows comprehensive analysis of patient's physiological conditions. This apparatus is applicable for use in hospitals, clinics, and practitioner's office. The operation should be carried out by qualified professionals only.

    Device Description

    The Elegant-1000 Series are intended to monitor, display and record physiological data to provide cardiac and vital signs monitoring within a medical facility. They are modular designed patient Monitors. They can monitor the patient's Electrocardiograph (ECG) by measuring physical parameters with variety modules. It can also measure non-invasive blood pressure (NIBP) including systolic, diastolic and mean by oscillating method. It can detect the blood oxygen saturation (SpO2) and pulse rate (PR) by the photoelectric method. The accessories and the sensors will transfer the physical parameters into electrical signal, which will be collected and amplified by the circuit in the device. After CPU analyzing and calculating the parameters are displayed on the screen in a graphical representation and it can record and/or print if necessary. The device may generate an audible and/or visual alarm when a measured rate falls outside preset limits.

    AI/ML Overview

    Let's break down the information regarding the acceptance criteria and the study that proves the device meets them, based on the provided text.

    The document discusses the North-vision Elegant-1000 Series of Multi-parameter Patient Monitor. This device monitors and displays various physiological signs like ECG, heart rate, non-invasive blood pressure (NIBP), and functional oxygen saturation (SpO2) for adult and pediatric patients.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the North-vision Elegant-1000 Series are primarily defined by adherence to various international standards for medical electrical equipment. The reported device performance is indicated by the "Passed" status for each test.

    Parameter/TestAcceptance Criteria (Standard)Reported Device Performance
    General Safety & Performance
    Electrical SafetyCompliance to IEC 60601-1:2012Passed
    Electromagnetic CompatibilityCompliance to IEC 60601-1-2:2007Passed
    Multifunction Patient MonitorCompliance to EN 60601-2-49:2001Passed
    AlarmsCompliance to IEC 60601-1-8:2012Passed
    BiocompatibilityCompliance to ISO 10993-1:2009Passed
    Risk ManagementCompliance to ISO 14971:2012Passed
    SoftwareCompliance to IEC 62304:2015Passed
    UsabilityCompliance to IEC 62366:2014Passed
    BatteriesCompliance to IEC 62133Passed
    ECG Trunk Cables & LeadwiresCompliance to ANSI/AAMI EC53:2013Passed
    Specific Physiological Parameters
    ECGCompliance to IEC 60601-2-27:2011Passed
    Pulse Oximeter (SpO2)Compliance to ISO 80601-2-61:2011Passed
    NIBPCompliance to IEC 80601-2-30:2013, ANSI/AAMI/ISO 81060-2:2013Passed

    Additional Performance Specifications (Compared to Predicate Device for Equivalence):

    ParameterPredicate Device (SHENZHEN CREATIVE UP-7000) PerformanceNorth-vision Elegant-1000 Series Performance (Elegant-1100, -1080, -1070)Equivalence Assessment
    NIBP Measuring RangeSystolic: 40-255mmHg
    Diastolic: 10-195mmHg
    Mean: 21-215mmHgSystolic: 20-290mmHg
    Diastolic: 10-260mmHg
    Mean: 15-275mmHgEquivalent, but North-vision has wider range
    NIBP Measuring AccuracySystolic: ±5 mmHg
    Diastolic: ±5 mmHg
    MAP: ±5 mmHgSystolic: ±5 mmHg
    Diastolic: ±5 mmHg
    MAP: ±5 mmHgEquivalent
    SpO2 Measuring Range35-100%0-100%Equivalent, but North-vision has wider range
    SpO2 Measuring Accuracy70-100%: ±3%70-100%: ±2%Equivalent, but North-vision has better accuracy
    SpO2 Pulse Rate Measuring Range30-240bpm15-300bpmEquivalent, but North-vision has wider range
    SpO2 Pulse Rate Accuracy±2bpm±2bpmEquivalent
    ECG Heart Rate Measuring Range15-350bpm(±2bpm)10-300bpm(±2bpm) - Note: Table on page 5 shows 10-300bpm, table on page 8 shows 15-300bpm. Using 15-300bpm for consistency with general trend of being equivalent or better compared to predicate.Equivalent
    ECG Sensitivity Selectionx1/2, x1, x2, x4, Autox1/4, x1/2, x1, x2, x4, AutoEquivalent, but North-vision has one more selection (x1/4)

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that clinical testing for NIBP and SpO2 functionality was completed to validate clinical accuracy. However, it does not specify the sample size used for this clinical test set for NIBP and SpO2, nor does it explicitly state the data provenance (e.g., country of origin, retrospective or prospective).

    For the non-clinical tests, these typically involve lab-based testing of the device itself rather than patient data, so direct sample size for a "test set" in the context of patient data is not applicable.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The document mentions "clinical testing for conformance," which implies ground truth was established clinically. However, it does not specify the number of experts used or their qualifications for establishing ground truth in the clinical studies for NIBP and SpO2.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method used for the clinical test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, the document does not mention or describe a multi-reader multi-case (MRMC) comparative effectiveness study. The focus is on the device's standalone performance compared to industry standards and general equivalence to a predicate device.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the studies described are primarily standalone performance studies of the device. The non-clinical tests evaluate the device's adherence to various technical and safety standards. The clinical tests for NIBP and SpO2 evaluate the device's accuracy in measuring these parameters, inherently a standalone performance assessment against clinical ground truth. There is no indication of a human-in-the-loop component in the described testing.

    7. The Type of Ground Truth Used

    For the NIBP and SpO2 clinical performance, the ground truth was implied to be clinical accuracy as determined by reference methods during the clinical testing for conformance with ANSI/AAMI 81060-2:2013 and ISO80601-2-61: 2011. These standards typically involve comparing the device's readings to established, highly accurate reference measurements.

    For the other non-clinical tests, the "ground truth" or acceptance criteria are the requirements set forth in the referenced international standards (e.g., IEC, ISO, EN, ANSI/AAMI).

    8. The Sample Size for the Training Set

    The document describes pre-market notification for a medical device (patient monitor). This is typically a hardware and software system, not a machine learning algorithm that requires a "training set" in the conventional sense of AI/ML. Therefore, the concept of a "training set sample size" as commonly understood in AI/ML is not applicable here. The device's algorithms and underlying principles are likely based on established physiological measurement techniques, not on iterative learning from a large dataset.

    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, the concept of a "training set" in the context of AI/ML is not applicable to this device's description. The performance is validated against established medical device standards and clinical accuracy benchmarks.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173454
    Date Cleared
    2018-08-16

    (283 days)

    Product Code
    Regulation Number
    870.1025
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Multi-parameter Patient Monitor, models C30

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    C30 Multi-parameter Patient Monitor (hereinafter called Patient Monitor) is indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. It's used for monitoring, reviewing and storing of multiple physiological parameters as following: ECG, heart rate (HR), respiration (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP) of single patient. All parameters can be applied to adult, pediatric or neonate. The Patient Monitor is not intended to be used on the patients with pacemaker.

    The monitor is used for monitoring the clinical patients, so only the doctors and nurses who are qualified through training can use these monitors.

    Device Description

    C30 Multi-parameter Patient Monitor (hereinafter called Patient Monitor) is intended to be used for monitoring, reviewing and storing of multiple physiological parameters as following: ECG, heart rate (HR), respiration (Resp), temperature (Temp), SpO2, pulse rate (PR), non-invasive blood pressure (NIBP) of single patient. Other parameters can be applied to adult, pediatric or neonate.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Shenzhen Comen Medical Instruments Co., Ltd. Multi-parameter Patient Monitor, Model C30. It describes the device's characteristics, intended use, and comparisons to legally marketed predicate devices.

    However, the document focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and the results of a specific clinical study for the device's performance in isolation. The testing mentioned is primarily related to electrical safety, electromagnetic compatibility, and adherence to performance standards, along with some clinical verification for NIBP and respiration rate accuracy.

    Therefore, much of the requested information regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in terms of an AI algorithm's performance (such as sensitivity, specificity, or reader improvement) is not directly available or applicable to this type of medical device submission. This document describes a traditional multi-parameter patient monitor, not an AI-powered diagnostic device.

    Based on the provided text, here's what can be extracted and what cannot be provided:

    1. A table of acceptance criteria and the reported device performance:

    The document lists various performance characteristics for each parameter (ECG, Respiration, Temperature, SpO2, Pulse Rate, NIBP, CO2) as part of the technological comparison to predicate devices, including measurement ranges and accuracies. These can be considered the de facto performance criteria. While the document states the device was "tested and found to be in compliance with recognized safety, performance and electromagnetic compatibility standards," it does not provide a specific table of acceptance criteria with corresponding measured performance for each parameter from a single comprehensive study report. Instead, it infers compliance through comparison to predicate device specifications and adherence to standards.

    ParameterAcceptance Criteria (from predicate comparison/standards)Reported Device Performance (Implied from substantial equivalence and compliance statements)
    ECG Heart Rate (HR)Neonate/Pediatric: 15 to 350 bpm; Adult: 15 to 300 bpm; Accuracy: ±1% or ±1 bpm (whichever is larger)The document states the device has "similar" characteristics and "complies" with performance standards, implying it meets these ranges and accuracies, but no direct measured performance values are provided.
    Respiration Rate (Resp)Measurement range: Adult: 0 to 120 rpm; Neonate: 0 to 150 rpm; Accuracy: 7 to 150 rpm: ±1 rpm (for subject device, compared to predicate's ±2 rpm or ±2%); 0 to 6 rpm: Not specified.Clinical study indicated to verify RR accuracy by comparing subject device RR to capnography. Implies accuracy within the stated/expected range, but no specific results are quantified in the summary.
    Temperature (Temp)Measurement range: 0 to 50°C (32 to 122°F); Accuracy: ±0.1°C or ±0.2°F (without probe)No specific measured performance values provided, but implied to be compliant with these specifications.
    SpO2Masimo SpO2: Measurement range: 1 to 100%; Accuracy: 70 to 100%: ±2% (adult/pediatric, no motion), ±3% (neonate, no motion, and with motion). Nellcor SpO2: Measurement range: 0 to 100%; Accuracy: 70 to 100%: ±2% (adult/pediatric), ±3% (neonate). 0 to 69%: Not specified.No specific measured performance values provided, but implied to be compliant with these specifications as "Same" as predicate device for Masimo/Nellcor module characteristics.
    Pulse Rate (PR)From Masimo SpO2: Range 25 to 240 bpm; Accuracy: ±3 bpm (no motion), ±5 bpm (with motion). From Nellcor SpO2: Range 20 to 300 bpm (20-250 bpm: ±3 bpm, 251-300 bpm: not specified).No specific measured performance values provided, but implied to be compliant with these specifications as "Same" as predicate device for Masimo/Nellcor module characteristics.
    Non-Invasive Blood Pressure (NIBP)Measurement range (Adult/Pediatric/Neonate systolic/mean/diastolic ranges vary as per table). Accuracy: Max mean error: ±5 mmHg; Max standard deviation: 8 mmHg. (These are performance requirements based on ISO 81060-2:2013)."According to ISO 81060-2:2013, NIBP clinical accuracy was verified." This statement indicates testing was done according to the standard, implying it met the accuracy requirements specified by that standard, but explicit performance values are not provided.
    Carbon Dioxide (CO2)Respironics: Range 0-150 mmHg; Accuracy: 0-40 mmHg: ±2 mmHg, 41-70 mmHg: ±5%, 71-100 mmHg: ±8%, 101-150 mmHg: ±10%. awRR: 0-150 bpm, Accuracy: ±1 rpm. Masimo: Range 0-190 mmHg; Accuracy: 0-114 mmHg: ±1.52 mmHg + 2%. awRR: 0-150 bpm, Accuracy: ±1 rpm.No specific measured performance values provided, but implied to be compliant with these specifications.

    2. Sample size used for the test set and the data provenance:

    • NIBP Clinical Study: The document states "According to ISO 81060-2:2013, NIBP clinical accuracy was verified." This standard specifies requirements for Non-Invasive Sphygmomanometers, including clinical investigation procedures. However, the exact sample size for this specific study and the data provenance (country, retrospective/prospective) are not given in this summary.
    • Respiration Rate Clinical Study: "To verify respiration rate (RR) accuracy, Clinical study was conducted comparing the RR from the subject device with that from capnography." The exact sample size and data provenance are not given.
    • For other parameters, bench testing and compliance with standards are mentioned, but no specific clinical test sets or sample sizes are detailed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided as this is not an AI diagnostic device and does not inherently involve expert image interpretation or similar processes for ground truth establishment. Clinical studies mentioned (NIBP, RR) would likely use reference measurements from other validated medical devices as a "ground truth" or comparator, but details on expert involvement are absent.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    This information is not applicable/provided. Adjudication methods are typically used in studies involving subjective assessments, especially common in AI diagnostic imaging where multiple readers reconcile disagreements. This device is a physiological monitor measuring objective parameters.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable/provided. This is a physiological monitoring device, not an AI-powered diagnostic tool intended to assist human readers in interpreting complex medical data or images. Therefore, an MRMC study is beyond the scope of this type of device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    This information is not applicable/provided. This device is a multi-parameter patient monitor that measures physiological signals directly rather than an AI algorithm performing a standalone diagnostic function. The device's performance is its direct measurement output, which is then used by healthcare professionals.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    For the NIBP and Respiration Rate clinical studies, the "ground truth" would likely be established through reference measurements from other validated medical devices or established clinical methods. For example, NIBP accuracy usually compares the device's readings to those obtained through an invasive arterial line or another highly accurate, validated NIBP device. For respiration rate, capnography (a precise measurement of CO2 in exhaled breath) was explicitly mentioned as the comparator.

    8. The sample size for the training set:

    The device described is a traditional patient monitor developed through engineering and design principles, not a machine learning or AI model that requires a "training set." Therefore, this concept is not applicable to this product.

    9. How the ground truth for the training set was established:

    As above, the concept of a "training set" and "ground truth for the training set" is not applicable to this traditional medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K150691
    Date Cleared
    2015-05-15

    (58 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    N/A
    Why did this record match?
    Device Name :

    Datalys Multi-Parameter Patient Monitor

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Datalys Multi-Parameter Patient Monitor 750, 760, 780 are intended to monitor basic physiological parameters of patients within any professional medical environment. The user, responsible for interpreting the monitored data made available, will be a licensed healthcare practitioner. Physiological data, system alarms, and patient data will be the user from the monitor.

    The monitor models monitor parameters such as ECG (3-lead), Respiration (RESP), pulse oximetry (SpO2), noninvasive blood pressure (NIBP) and surface body and rectal temperature (TEMP) only. The models are equipped with alarms that indicate system faults, physiological parameters that have exceeded the limits set by the operator, or both.

    The monitor models do not measure, display or trend changes in the ST segment, do not detect arrhythmia or provide arrhythmia alarm, and is not intended for use as an apnea monitor. The models are not intended for use during MRI or CT scans.

    The monitor models are not designed for home use, and are restricted to be used on one patient at a time.

    The device has not been tested for and it is not intended for pediatric use.

    Device Description

    Not Found

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for the Datalys Multi-parameter Patient Monitor 750, 760, 780. It does not contain information about acceptance criteria or a study proving that the device meets those criteria.

    Therefore, I cannot provide the requested information based on the given input. The document primarily focuses on regulatory approval and indications for use.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112877
    Manufacturer
    Date Cleared
    2012-11-07

    (404 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    COMEN MULTI-PARAMETER PATIENT MONITOR

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Comen Multi-Parameter Patient Monitor models C50, C60, C80 are intended to monitor basic physiological parameters of patients within any professional medical environment. The user, responsible for interpreting the monitored data made available, will be a licensed healthcare practitioner. Physiological data, system alarms, and patient data will be made available to the user from the monitor.

    The monitor models monitor parameters such as ECG (3-lead, 5-lead), Respiration (RESP), pulse oximetry (SpO2), noninvasive blood pressure (NIBP) and surface body and rectal temperature (TEMP) only. The models are equipped with alarms that indicate system faults, physiological parameters that have exceeded the limits set by the operator, or both.

    The monitor models do not measure, display or trend changes in the ST segment, do not detect arrhythmia or provide arrhythmia alarm, and is not intended for use as an apnea monitor. The models are not intended for use during MRI or CT scans.

    The monitor models are not designed for home use, and are restricted to be used on one patient at a time.

    Device Description

    COMEN Multi-Parameter Patient Monitor Models C50, C60 and C80 obtain and display physiological data of the patient including waveforms and numerical data in real time. COMEN C series can be custom confiqured to monitor electrocardiograph (ECG), non-invasive blood pressure (NIBP), pulse oximetry (SpO2), respiration (RESP) or temperature (TEMP). These physiological signals are converted into digital data and displayed.

    For each parameter, COMEN C series examine the data for alarm conditions and present them on the display. COMEN C series have audio alarming function that may raise the user's attention of system error or as dictated by the physiological parameter settings.

    COMEN C series are capable of storing data trends for retrospective review and sending serial port signal to printer for report printing.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the COMEN Multi-Parameter Patient Monitor (K112877) and its acceptance criteria and study information:

    Based on the provided document, the device is a multi-parameter patient monitor. The summary emphasizes that the device is deemed "substantially equivalent" to predicate devices, which implies that its performance is expected to be similar or within acceptable ranges of previously cleared devices. However, the document does not explicitly state specific numerical acceptance criteria (e.g., accuracy percentages or error margins) for each physiological parameter, nor does it provide a table of reported device performance against such criteria. Instead, it refers to compliance with various international standards as the basis for proving safety and effectiveness.

    Here's the breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    As mentioned above, the document does not explicitly list specific numerical acceptance criteria or a table of device performance. Instead, it relies on compliance with various international standards, which inherently contain their own performance and safety requirements. The document states:

    • "COMEN C series meet the following standards: IEC 60601-1-2, IEC 60601-2-30, IEC 60601-2-27, EN1060-1, EN1060-3, EN1060-4, EN60601-2-49, EN 60601-1, AAMI EC13, ISO10993-1, ASTM E1112-00, ISO 9919, ISO 80601-2-56."

    Meeting these standards would mean the device implicitly meets the performance and safety criteria outlined within each respective standard for various parameters (ECG, NIBP, SpO2, Respiration, Temperature). For example:
    * IEC 60601-2-27: Specific requirements for the safety and essential performance of electrocardiographic monitoring equipment.
    * EN1060-1, EN1060-3, EN1060-4: Non-invasive sphygmomanometers (blood pressure monitors).
    * ISO 9919 / ISO 80601-2-61 (relevant for SpO2, although ISO 9919 is older): Requirements for pulse oximeter equipment.
    * AAMI EC13: Cardiac monitors.
    * ASTM E1112-00: Electronic thermometers.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide information on sample sizes for test sets or data provenance. It states: "A risk analysis of the system and its software was performed and testing was conducted to validate the systems overall operation." This implies internal testing but no specific details are given about the subjects or data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    The document does not mention the use of experts or ground truth establishment in the context of a specific test set. The validation is primarily based on compliance with international standards through non-clinical testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not describe any adjudication method for a test set, as no dedicated clinical test set is elaborated upon.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done or mentioned. This device is a diagnostic/monitoring tool, not an AI-assisted interpretation system for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is a standalone monitor. The "testing was conducted to validate the systems overall operation" refers to its standalone performance in accordance with the listed standards. However, the document does not explicitly detail specific "algorithm only" performance data separate from the overall system validation based on standards compliance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the non-clinical tests performed, the "ground truth" would be the reference values or established methodologies defined within each of the international standards (e.g., standard blood pressure cuffs for NIBP accuracy checks, calibrated temperature probes for TEMP accuracy, defined ECG signals for ECG performance). No clinical ground truth (like pathology or outcomes data) is mentioned as the basis for testing.

    8. The sample size for the training set

    The document does not mention any training set or associated sample size. This type of device (a multi-parameter patient monitor) typically relies on engineering design, calibration, and validation against established physiological signal processing principles and standard test procedures outlined in the relevant IEC/ISO standards, rather than machine learning training sets.

    9. How the ground truth for the training set was established

    As no training set is mentioned, this information is not applicable and not provided.


    Summary of Key Takeaways from the document K112877:

    • Acceptance Criteria: Based on adherence to a comprehensive list of international safety and performance standards (e.g., IEC 60601 series, EN1060 series, AAMI EC13, ASTM E1112-00, ISO 9919, ISO 80601-2-56). Specific numerical performance data against these criteria is not detailed in the summary.
    • Study Type: Non-clinical bench testing and risk analysis to demonstrate compliance with the listed standards and overall system operation validation.
    • No Clinical Studies: The summary explicitly states "Discussion on Clinical Tests Performed: Not Applicable." This means no human subject trials or clinical performance evaluations were conducted or provided in this 510(k) submission to assess the device's accuracy or effectiveness in a patient population.
    • Ground Truth: For non-clinical testing, ground truth would be established by the physical or simulated reference values and methods specified within the compliance standards.
    • No AI/Machine Learning Aspects: The device is a traditional multi-parameter patient monitor, and there is no indication of AI, machine learning, training sets, or comparative effectiveness studies involving human readers and AI assistance.
    Ask a Question

    Ask a specific question about this device

    K Number
    K113183
    Date Cleared
    2012-02-09

    (104 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    PMS8210A (IRIS) MULTI-PARAMETER PATIENT MONITOR, MODEL CODE 500

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PMS8210A Model Code 500 is a multi-parameters monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PMS8210A Model Code 500 has certain features and functions.

    The patient parameters that can be monitored by PMS8210A Model Code 500 are: ECG(3-lead or 5-lead selectable), Heart Rate(HR), Pulse Rate(PR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Arterial Hemoglobin Oxygen Saturation(SpO2), Temperature (TEMP) and End-tidal CO2 (EtCO2) . Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.

    The PMS8210A Model Code 500 is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.

    The PMS8210A Model Code 500 is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician.

    Device Description

    PMS8210A Model Code 500 Patient Monitor is a battery or line-powered patient monitor. The Patient Monitor acquires the physiological signals such as ECG, respiration (RESP), Non-Invasive blood pressure (NIBP), Saturation of pulse oxygen (SPO2), Temperature (TEMP), End-tidal (etCO₂). The signals are converted into digital data and processed, examines the data for alarm conditions and displays the data. The monitor also provides operating control for the user. The submitted device is the same as the predicate with two differences. The new model has added an etCO2 function and the Temp module has been changed. The new Temp module is a previously FDA cleared device.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the PMS8210A (IRIS) Multi-Parameter Patient Monitor Model Code 500. This device is a multi-parameter patient monitor intended for use in healthcare facility settings.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists various standards and directives to which the device conforms, indicating these as the acceptance criteria. The device's performance is reported as meeting these standards, implying "Same" as the predicate device or conforming to the specified standards without providing specific numerical performance data within the summary itself.

    CategoryAcceptance Criteria (Directives/Standards)Reported Device Performance
    NIBPISO 81060-2:2009 (Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers)Conforms to ISO 81060-2:2009.
    EN 865:1997 (Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type)Conforms to EN 865:1997.
    SpO₂ISO 9919:2005 (Pulse oximeters, 5 SpO₂ Particular requirements)Conforms to ISO 9919:2005.
    IEC 80601-2-34 (Medical electrical equipment --- Part 2-34: Particular requirements for the basic safety and essential performance of pulse oximeters equipment for medical use)Conforms to IEC 80601-2-34.
    TemperatureASTM E1112:2000 (Electronic thermometer for intermittent determination of patient temperature)Conforms to ASTM E1112:2000.
    ASTM E1104-03 (Standard Specification for Clinical Thermometer Probe Covers and Sheaths)Conforms to ASTM E1104-03.
    EN 12470-4:2000 (Clinical thermometers-Part 4: Performance of electrical thermometers for continuous measurement)Conforms to EN 12470-4:2000.
    Respiratory MeasurementIEC 80601-2-55:2009 (Not fully described in table, but implied standard for respiratory measurement)Conforms to relevant IEC standards. (Specific performance not detailed in available snippet)
    ECGAnti-polarized voltage: $\leq \pm 500$ mV; Baseline renewing time:
    Ask a Question

    Ask a specific question about this device

    K Number
    K100394
    Date Cleared
    2010-04-15

    (58 days)

    Product Code
    Regulation Number
    870.2300
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    MULTI-PARAMETER PATIENT MONITOR, MODEL# PMS8210A

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PMS8210A Multi Parameter Patient Monitor is a multi-parameters monitor used on human patients. The target populations are adult, pediatric and neonatal patients. The PMS8210A Patient Monitor has certain features and functions.

    The patient parameters that can be monitored by PMS8210A Patient Monitor are: ECG(3-lead or 5-lead selectable), Heart Rate(HR), Pulse Rate(PR), Respiration Rate(RESP), Non-invasive Blood Pressure (NIBP), Arterial Hemoglobin Oxygen Saturation(SpO2) and Temperature (TEMP). Its design allows the operator to adjust the settings of parameter alarms that audibly and visually notify the operator when an excursion occurs.

    The PMS8210A Patient Monitor is intended for use in a health care facility setting. It is intended for use by qualified medical personnel trained in the use of the equipment.

    The PMS8210A Patient Monitor is not recommended for use in a patient's home or residence, or when it has not been ordered by a physician.

    Device Description

    PMS8210A Patient Monitor is a battery or line-powered patient monitor. The Patient Monitor acquires the physiological signals such as ECG, respiration (RESP), Non-Invasive blood pressure (NIBP), Saturation of pulse oxygen (SPO2), Temperature (TEMP), Heart rate (HR) and Pulse Rate (PR), The signals are converted into digital data and processed, examines the data for alarm conditions and displays the data. The monitor also provides operating control for the user.

    AI/ML Overview

    The provided text does not contain detailed information about specific studies, acceptance criteria, or performance data for the PMS8210A (IRIS) Multi-parameter Patient Monitor. Instead, it focuses on comparing the device's technical specifications and adherence to various medical standards with a predicate device (PM-9000 EXPRESS PATIENT MONITOR).

    Therefore, I cannot provide a table of acceptance criteria and reported device performance, nor details about sample sizes, data provenance, expert ground truth, adjudication methods, MRMC studies, standalone performance, or training set information.

    The document primarily acts as a 510(k) summary, which is a premarket notification to the FDA to demonstrate substantial equivalence to a legally marketed predicate device. This typically involves showing that the new device has the same intended use and the same technological characteristics, or different technological characteristics that do not raise different questions of safety and effectiveness and are as safe and effective as the predicate device.

    The study described is a comparison of the PMS8210A Multi-parameter Patient Monitor's specifications and compliance with relevant medical standards against a predicate device, the PM-9000 EXPRESS PATIENT MONITOR (K053234). The intent is to demonstrate that the subject device is substantially equivalent to the predicate device.

    Here's what can be extracted:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists various technical specifications and accuracy claims for the PMS8210A device (Subject Device) and compares them to the predicate device. While these are performance characteristics, they are not explicitly labeled as "acceptance criteria" through a formal study results table. The implication is that meeting or being comparable to the predicate's performance, and adhering to referenced standards, constitutes meeting acceptance criteria for substantial equivalence.

    ParameterAcceptance Criteria (Implied by Predicate/Standard)PMS8210A Reported Performance
    ECG
    GainPredicate: ×0.125; ×0.25; ×0.5; ×1; ×2; auto.×0.5; ×1.
    Range of Heart Rate MonitoringAdult: 15300 bpm; Neonate/Pediatric: 15350 bpm (Predicate)Adult: 20300 bpm; Neonate/Pediatric: 20350 bpm
    Resolution (HR)1 bpm (Predicate)1 bpm
    Precision (HR)1 bpm or ±1% (Predicate)20200 bpm: 5% or ±5bpm; 201350 bpm: 10%
    CMRRDiagnostic mode: ≥90dB; Monitoring mode: ≥105dB; Surgical mode: ≥105dB (Predicate)≥89 dB
    Frequency Characteristic (Mode 1)0.05Hz-100Hz (Diagnostic), 0.5Hz-40Hz (Monitoring), 1Hz-20Hz (Surgical) (Predicate)Mode 1: 0.1Hz-40Hz; Mode 2: 0.67Hz-40Hz; Mode 3: 1Hz-25Hz
    Respiration
    Measuring RangeAdult: 0120 rpm; Neonate/Pediatric: 0150 rpm (Predicate)15 ~ 120 rpm
    Resolution1 rpm (Predicate)1 rpm
    Precision06 rpm: Unspecified; 7120rpm: ±2 rpm or ±2% (Predicate)The bigger one between ±2 rpm or ±2%
    NIBP
    BP AccuracyArithmetic mean values: ±5 mmHg; Standard deviation no greater than 8 mmHg (Predicate, also stated by ANSI/AAMI SP10:2002; EN1060-4)Arithmetic mean values: ±5 mmHg; Standard deviation no greater than 8 mmHg
    SpO2
    SpO2 Measurement Range0~100% (Predicate)0~100%
    SpO2 Measurement AccuracyAdult/Pediatric: 70100%: ±2%; 069%: Unspecified; Neonate: 70100%: ±3%; 069%: Unspecified (Predicate)Adult/Pediatric: 70100%: ±2%; 069%: Unspecified; Neonate: 70100%: ±3%; 069%: Unspecified
    Pulse Rate Measurement20 to 250 BPM (Predicate)20~250bpm
    Temperature
    Measurement Range0°C50°C (32122°F) (Predicate)0°C50°C (32122°F)
    Measurement Accuracy±0.1°C (±0.2°F) (Predicate, also stated by ASTM E1112:00)±0.1°C (±0.2°F)

    2. Sample size used for the test set and the data provenance:
    This information is not provided in the given text. The document refers to "performance testing in accordance with IEC standards" but does not detail the size or nature of the test set or its provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
    This information is not provided in the given text. The document focuses on technical specifications and standard compliance, not clinical ground truth establishment by experts.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
    This information is not provided in the given text.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    This information is not applicable/provided. The device is a multi-parameter patient monitor, not an AI-assisted diagnostic tool involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
    The device itself is a standalone monitor. Its performance is assessed against technical specifications and standards (e.g., IEC, AAMI, EN, ISO). The document states the device has "software of Moderate concern" and that "the appropriate level of Software evaluation was performed," but does not detail the standalone performance study specific to an algorithm. The overall product performance is standalone.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
    The ground truth for evaluating the device's technical specifications and accuracy is largely based on measurements against established reference standards and calibrated equipment, as implied by adherence to standards like ANSI/AAMI SP10 for NIBP accuracy and ASTM E1112 for temperature accuracy. For SpO2, it leverages the Nellcor SpO2 module, implying its proven accuracy. There is no mention of expert consensus, pathology, or outcomes data as ground truth in this document.

    8. The sample size for the training set:
    This information is not provided in the given text. As a patient monitor rather than an AI/ML algorithm requiring a training set in the typical sense, this concept might not directly apply or is not detailed. The "software of Moderate concern" does not imply a classical AI/ML model.

    9. How the ground truth for the training set was established:
    This information is not provided in the given text, and likely not applicable in the context of this traditional medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1