510(k) Data Aggregation

    K Number
    K171665
    Date Cleared
    2018-01-11

    (220 days)

    Product Code
    Regulation Number
    870.1210
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Micro Catheter and Guidewire System is intended for the infusion of contrast media into the peripheral vasculature. The Micro Catheter and Guidewire system is also intended for drug infusion in intra-arterial therapy and the infusion of embolic materials for hemostasis.

    Device Description

    Micro Catheter and Guidewire System consists of a catheter, a guidewire, and accessories. The accessories include a flushing device, a shaping mandrel, an insertion tool, and a torque device. The catheter is consist of a hub, a stress relief tube and a catheter body. The catheter body has three layers. The inner layer is a PTFE tube, the middle layer is consist of stainless steel wire reinforce and platinum-iridium alloy radiopaque distal marker, and the outer layer is polyamides of different hardness mixed with pigment. There is also a hydrophilic coating on the catheter surface. The guidewire is consist of a nitinol core, a polymer jacket with hydrophilic coating over its entire surface and a radiopaque distal marker. It has a white marker at the proximal end to indicate the length inserted into human body and its relative position with the catheter.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for a medical device (Micro Catheter and Guidewire System). This type of document is primarily focused on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that the device meets a specific set of quantitative acceptance criteria through a clinical study or a study directly measuring performance against predetermined thresholds for an AI/software as a medical device.

    Therefore, the information typically requested in your prompt regarding acceptance criteria, study design for AI/software, sample sizes, expert ground truth, MRMC studies, standalone performance, and training set details is not present in this document. This document details the physical characteristics, intended use, and bench testing for a physical medical device.

    Here's why the information is not here, and what type of information is provided:

    • Acceptance Criteria & Reported Device Performance (Table 1): The document does not provide a table of quantitative acceptance criteria for performance in the way you'd expect for, say, an AI diagnostic tool (e.g., "sensitivity > 90%"). Instead, it lists various bench tests conducted (e.g., Catheter Sizes, Coating Integrity, Peak tensile force, etc.) and states that "The data provided demonstrate that the Micro Catheter and Guidewire System is substantially equivalent to the predicate device." This implies the device passed these bench tests, but the specific numerical acceptance criteria and measured values are not detailed in this public summary.

    • Sample size and data provenance for test set: Not applicable for this type of device and study. Bench testing is performed on physical samples.

    • Number of experts and qualifications for ground truth: Not applicable. Ground truth for physical device bench testing is based on engineering specifications and direct measurement.

    • Adjudication method: Not applicable.

    • MRMC comparative effectiveness study: Not applicable. This is a physical device, not an AI or imaging diagnostic tool that assists human readers.

    • Standalone performance: Not applicable in the context of AI performance. The document describes the standalone performance of the physical device through bench testing.

    • Type of ground truth used: For this device, the "ground truth" is defined by engineering specifications, material properties, and physical testing standards (e.g., ISO 10555.1, ISO 11070).

    • Sample size for training set & How ground truth for training set was established: Not applicable. This is not a machine learning/AI device.

    What the document does provide in relation to "proving the device meets acceptance criteria":

    The document focuses on demonstrating substantial equivalence to a predicate device, which is the primary regulatory pathway for 510(k) submissions. This is achieved by:

    1. Indications for Use: Stating that the device has the same intended use as the predicate device (infusion of contrast media, drug infusion in intra-arterial therapy, infusion of embolic materials for hemostasis into peripheral vasculature).
    2. Technological Characteristics: Highlighting similarities in design (catheter, guidewire, accessories, provided sterile, single use, hydrophilic coating, platinum-iridium alloy radiopaque marker).
    3. Performance Data: Listing a comprehensive set of biocompatibility tests (e.g., Cytotoxicity, Sensitization, Hemolysis study) and bench tests (e.g., Catheter Sizes, Coating Integrity, Tensile force, Burst pressure, Guidewire Fracture test, Torque strength, Simulated use). These tests demonstrate that the device performs as intended and is safe, similar to the predicate. The implication is that the results of these tests met internal acceptance criteria to support the claim of substantial equivalence.

    In summary, while the document indirectly "proves the device meets acceptance criteria" by stating its substantial equivalence based on safety and performance tests, it does not provide the detailed quantitative acceptance criteria and study particulars you've asked for, as those typically relate to AI/software performance rather than a physical medical device.