    Device Name :

    MedtFine Safety Lancet

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The safety lancet is intended for capillary blood sampling.

    Device Description

    Sterile safety lancet consists of outer shell, middle sleeve, needle holder, spring, upper shell, back cover, small cover, needle. The sterile part of the safety lancet is the needle tip. The sterile barrier is the needle holder and sterilized to a SAL of 10-6 by radiation sterilization. It is intended for single use and single person use only. The shelf-life of the product is 5 years. The device is for medical professionals and lay person use as an OTC use.

    AI/ML Overview

    This document is a 510(k) Summary for a medical device called "MedtFine Safety Lancet". It details the device's characteristics, intended use, and comparison to a predicate device to demonstrate substantial equivalence.

    Here's an analysis of the provided text for acceptance criteria and the study proving the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document provides a "Characters comparison" table (Table 2) that includes performance parameters for both the proposed "MedtFine Safety Lancet" and the predicate device "Safety Lancet (K220370)". This table effectively serves as a comparison against acceptance criteria. The acceptance criterion for each performance characteristic is generally implied to be "same" or within the range of the predicate device, or "meet requirements."

    Acceptance Criteria (Implied from Predicate/Requirement)Reported Device Performance (MedtFine Safety Lancet)Comments (from document)
    Product Code: FMKFMKSame
    Indication for Use: Capillary blood samplingThe safety lancet is intended for capillary blood sampling.Same
    OTC UseOTC useSame
    Specifications: Needle: 18G, 18G blade, 21G, 23G, 16G, 28G, 30G, 32GNeedle: 18G, 18G blade, 21G, 23G, 16G, 28G, 30G, 32GSame, within range.
    Safety protection features: YesYesSame
    Reuse durability: Single use for single personSingle use for single personSame
    User environments: Home, clinicalHome, clinicalSame
    Components and materials: (Specific to predicate; proposed device differs)Outer shell: PP, Middle sleeve: ABS, Needle holder: PE, Spring: carbon steel, Upper Shell: PE, Back cover: PE, Small cover: PP, Needle: Stainless Steel#1 (The proposed and predicate devices used different materials for manufacturing. The device structures kept the same. The performance testing and biocompatibility testing results demonstrated the proposed device met the requirements of various standards)
    Sterilization: Radiation SAL=10-6Radiation SAL=10-6Same
    Shelf life: 5 years5 yearsSame
    Activation force: 4~15N4~7NSame, within range.
    Puncture force: 1826G: 0.71.4N (

    8. The sample size for the training set

    This information is not applicable. The MedtFine Safety Lancet is a mechanical medical device, not an AI/ML-based device that requires a "training set." Its design and manufacturing are based on engineering principles and material science, not data-driven learning.

    9. How the ground truth for the training set was established

    This information is not applicable for the reasons stated above (not an AI/ML device). The design and development process for such a device relies on established engineering specifications, material selection, and manufacturing processes, which are then validated through the performance and biocompatibility testing outlined.