K Number
K021052Device Name
MODEL GT010706 ELECTRONIC THERMOMETERManufacturer
Date Cleared
2002-04-09
(8 days)
Product Code
Regulation Number
880.2910Type
TraditionalPanel
General HospitalReference & Predicate Devices
Why did this record match?
Device Name :
MODEL GT010706 ELECTRONIC THERMOMETER
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use
Indicated use- Measure of individual temperature
Measurements- Oral, Axillary -under arm, Rectal
Range of measurement- 90.0 degrees - 109.9 degrees F (32.0 degrees - 43 degrees C)
Accuracy- +/-0.2 degrees F
Targeted population- Individuals requiring temperature measurements
Environment of use- Hospital and home
Disposable / reusable- Reusable, clean with alcohol
Device Description
The intended product is an electronic digital thermometer for measuring patient temperature.
AI/ML Overview
{
"acceptance_criteria_and_study": {
"acceptance_criteria_table": {
"headers": [
"Parameter",
"Acceptance Criteria",
"Reported Device Performance"
],
"rows": [
[
"Temperature range",
"Not explicitly stated as an acceptance criterion for the subject device but compared to predicate device's 89.6-109.4° F",
"90.0-109.9° F"
],
[
"Ambient temperature",
"Not explicitly stated as an acceptance criterion for the subject device but compared to predicate device's 60.8-104° F",
"60.8-104° F"
],
[
"Response time",
"Not explicitly stated as an acceptance criterion for the subject device but compared to predicate device's ~1 minute",
"About 20 seconds"
],
[
"Accuracy and Performance",
"Meets ASTM E1112",
"Meets ASTM E1112"
],
[
"Accuracy (for indications for use)",
"+/- 0.2° F",
"+/- 0.2° F (stated in Indications for Use, implies conformance)"
],
[
"Material safety",
"Non-toxic and in compliance with EN ISO 10993",
"All materials used are non-toxic and in compliance with EN ISO 10993"
]
]
},
"study_details": {
"sample_size_test_set": "Not specified. The document states \"Performance testing,\" and notes that the device \"meets ASTM E1112,\" but does not detail the sample size for this testing.",
"data_provenance": "Not specified. The testing was conducted to meet ASTM E1112 standards.",
"number_of_experts_ground_truth": "Not applicable. The ground truth for performance and accuracy is based on compliance with a recognized standard (ASTM E1112) rather than expert interpretation of data.",
"qualifications_experts": "Not applicable.",
"adjudication_method": "Not applicable.",
"mrmc_comparative_effectiveness_study": "No. This is a digital thermometer, not an AI-assisted diagnostic device, so an MRMC study comparing human readers with and without AI assistance is not relevant.",
"standalone_performance_study": "Yes, a standalone performance study was conducted to demonstrate compliance with ASTM E1112. The data provenance and sample size are not explicitly detailed beyond the standard reference.",
"type_of_ground_truth": "Standard compliance (ASTM E1112) for accuracy and performance, and safety standard compliance (EN ISO 10993) for materials.",
"sample_size_training_set": "Not applicable. This is a physical device, not an AI algorithm requiring a training set.",
"ground_truth_training_set_established": "Not applicable."
}
}
}