    Device Name :

    MICRO-TOUCH NITRILE POWDER FREE SYNTHETIC MEDICAL EXAMINATION GLOVES, COLOR BLUE NON-STERILE

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Micro-Touch Nitrile Powder Free Synthetic Medical Examination Gloves intended for medical purposes that is worn on the examiners hand to prevent contamination between patient and examiner.

    Device Description

    Micro-Touch Nitrile Powder Free Synthetic Medical Examination Gloves are non-sterile disposable devices intended for medical purposes that is worn on the examiners hand to prevent contamination between patient and examiner.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the "Micro-Touch Nitrile Powder Free Synthetic Medical Examination Gloves," based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
    DimensionsMeets ASTM D 6319-00Meets ASTM D 6319-00
    Physical PropertiesMeets ASTM D 6319-00Meets ASTM D 6319-00
    Freedom from holesMeets ASTM D 6319-00
    Meets ASTM D 5151-99Meets ASTM D 6319-00
    Meets ASTM D 5151-99
    Powder-FreeMeets ASTM D 6319-00
    Not more than 2 mg residue by mass per gloveMeets ASTM D 6319-00
    Not more than 2 mg residue by mass per glove
    Biocompatability (Primary Skin Irritation)PassesPasses
    Biocompatability (Guinea Pig Sensitization)PassesPasses

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample size used for each specific test within ASTM D 6319-00 or ASTM D 5151-99. These are standardized tests, and the sample sizes are dictated by the respective ASTM standards themselves. The data provenance is not specified beyond being generated by Ansell Healthcare Products Inc., implicitly in the USA. The data would be considered prospective as it's generated specifically for the clearance of this device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. For these types of medical devices (gloves), the "ground truth" is typically established by physical and chemical testing against recognized industry standards (ASTM standards), rather than by expert consensus on observational data. The standards themselves define the acceptable parameters.

    4. Adjudication Method for the Test Set

    This information is not applicable in the context of these physical and chemical tests. The "ground truth" is determined objectively by the test results themselves against the pre-defined criteria in the ASTM standards. There is no need for human adjudication of test results in the way it might be applied to image interpretation or clinical outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is not relevant for medical examination gloves which are evaluated based on their physical properties, freedom from holes, and biocompatibility, not on interpretive clinical judgments by human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone study was not done. This device is a physical product (medical glove), not an algorithm or AI system. Therefore, the concept of "algorithm only" performance is not applicable.

    7. The Type of Ground Truth Used

    The ground truth used is primarily objective measurements against established industry standards (ASTM D 6319-00 and ASTM D 5151-99), along with biocompatibility test results.

    8. The Sample Size for the Training Set

    This information is not applicable. Since this is a physical medical device and not an AI or machine learning model, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for this type of device.