Search Results
Found 1 results
510(k) Data Aggregation
(89 days)
MDIXON-QUANT
- · is a software option intended for use on Achieva and Ingenia 1.5T & 3.0T MR Systems
- · is a non-invasive triglyceride fat fraction calculation
- · is a 3D, single breath-hold acquisition and reconstruction technique
- · allows for water-fat separation generating Water-only, Fat-only images as well as Out-phase and In-phase images
- · is indicated for magnetic resonance imaging of the liver
mDIXON-Quant acquisition and reconstruction is based on the mDIXON product developed previously. Acquisition relies on a gradient echo acquisition (FFE) including a large number of echoes (6 or more). Furthermore, mDIXON-Quant reconstruction includes the use of multiple spectral peaks of triglyceride fat, correction of the T2* confounding effect and reduction of T1 bias.
mDIXON-Quant allows for water-fat separation and generates Water-only, Fat-only images as well as Out-phase and In-phase images synthesized from the Water and Fat images. Additionally, mDIXON-Ouant produces images representing triglyceride fat fraction as well as images representing transverse magnetization relaxation.
The feature requires:
- · Specific parameter settings for the mDIXON sequence, within cleared parameter limits, to acquire the MR signals
- · A new calculation function to generate the new images for Fat Fraction, T2 *. This function uses a set of MR images as input that is generated in a cleared manner from the acquired MR signals.
- · The new images need to be stored and displayed with the appropriate labels for Fat Fraction, T2 *, applying the facilities provided by the cleared platform.
Here's a summary of the acceptance criteria and study information based on the provided text for the mDIXON-Quant device:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria | Reported Device Performance |
---|---|
FF accuracy ±3.5% | FF accuracy is ±3.5% for both field strengths and all possible parameter combinations |
T2* range and reproducibility 1.5% | T2* range and reproducibility is 1.5% for both field strengths |
No error messages during Acquisition and Reconstruction testing | No error messages displayed |
Scan time |
Ask a specific question about this device
Page 1 of 1