Search Results
Found 1 results
510(k) Data Aggregation
(59 days)
Lianfa Safety Lancet (Five Models: NPA, PA, PA2, TPA, and APA)
Lianfa Safety Lancet (Five Models: NPA, PA, PA2, TPA, and APA) is a single use device indicated for capillary blood sampling.
Lianfa Safety Lancet is a single use, sterile, medical devices designed to be used in collecting the blood sample. The products are intended to be used by professionals and home users. They are intended for over the counter use.
Lianfa Safety Lancet is a disposable blood lancet intended for a single use that is comprised of a single use blade attached to a solid, non-reusable base (including an integral sharps injury prevention feature) that is used to puncture the skin to obtain a drop of blood for diagnostic purposes.
The safety lancet is made up of a needle core, spring, a launching part, and a casing. The steel needle is made of SUS304 stainless steel, the spring material is carbon steel, and the other parts are made of plastic components. The integral sharps injury prevention feature allows the device to be used once and then renders it inoperable of further use.
Based on the shape and design of the device, there are five different models: PA, PA2, APA, NPA, and TPA. Each model is offered with various gauges of needle.
This document is a 510(k) Premarket Notification for the Lianfa Safety Lancet. The provided information does not describe an AI/ML powered device, but rather a traditional medical device (a blood lancet). Therefore, many of the requested categories for AI/ML device studies (e.g., effect size of human readers with AI, ground truth for training set, adjudication methods, multi-reader multi-case studies) are not applicable.
Here's an analysis based on the provided text, focusing on the acceptance criteria and the study that proves the device meets them, as much as applicable to a non-AI medical device:
Acceptance Criteria and Device Performance for Lianfa Safety Lancet
This submission is a 510(k) premarket notification for a traditional medical device, a blood lancet. As such, the concept of "AI" or "ML" powered assistance, training sets, and human reader performance with AI are not applicable. The studies described are non-clinical engineering and biocompatibility tests.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are derived from a combination of general industry standards (e.g., ISO, 21 CFR regulations) and specific functional requirements for a blood lancet. The performance is reported as meeting these requirements.
Characteristics / Acceptance Criteria (Specifications) | Reported Device Performance (Results) |
---|---|
Appearance | |
- Surface clean, free of burrs, scratches, rusts, extraneous materials | Meet requirement |
- Needle clean, straight, without residual debris | Meet requirement |
Dimension | |
- Product dimensions consistent to drawings | Meet requirement |
Puncture Depth | |
- Measured using calipers and meets requirements | Meet requirement |
Corrosion | |
- Steel needle free of signs of corrosion | Meet requirement |
Puncture Force | |
- Steel needle puncture force ≤ 2.0 N | Meet requirement |
Connection Firmness (Needle and Handle) | |
- Withstand 10N axial static tension for 1 min | Meet requirement |
Sterility | |
- Device must be sterile | Meet requirement |
pH | |
- pH value between test solution and control solution not > 1 | Meet requirement |
Extractable Metal | |
- Not more than 10 µg/device | Meet requirement |
Launch Performance | |
- Good launch performance, button pressed smoothly, no jam | Meet requirement |
Safety Features | |
- Force to trigger needle launch: 4-10 N | Meet requirement |
- Deactivation: product cannot be launched after first launch | Meet requirement |
Biocompatibility | |
- In Vitro Cytotoxicity (ISO 10993-5: 2009) | No Cytotoxicity |
- Skin Sensitization (ISO 10993-10: 2010) | No Skin sensitization |
- Intracutaneous reactivity (ISO 10993-10: 2010) | No irritation |
- Acute Systemic Toxicity (ISO 10993-11: 2017) | No Acute Systemic Toxicity |
- Pyrogenicity (ISO 10993-11: 2017) | No thermogenic reaction |
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes used for each non-clinical bench test. It generally states that "bench testing performed verifies..." and provides the "Results" as "Meet requirement."
The data provenance is from non-clinical bench testing, presumably conducted internally or by a contracted lab for the manufacturer, Shandong Lianfa Medical Plastic Products Co. Ltd. The country of origin for the manufacturing and testing is not explicitly stated but the company is based in Zhangqiu, Jinan, Shandong Province, China. The studies are retrospective in the sense that they were completed as part of the device development and verification process before this 510(k) submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable for this type of medical device. The "ground truth" for the non-clinical tests relies on established engineering standards, material science principles, and biological evaluation protocols (e.g., ISO 10993 series). There are no human "experts" establishing a diagnostic ground truth in the way described for AI/ML devices. The "experts" involved would be technicians, engineers, and scientists conducting the tests according to the specified protocols.
4. Adjudication method for the test set
Not applicable for this type of medical device. Adjudication methods like 2+1 or 3+1 are used in clinical studies where multiple human readers interpret data, and complex rules are needed to establish a "ground truth" result when there are disagreements. For non-clinical bench testing, results are typically objective measurements against a predefined specification.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a manual blood lancet and does not incorporate AI functionality, nor does it involve human readers interpreting AI-generated output.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a manual blood lancet. There is no algorithm.
7. The type of ground truth used
For the non-clinical and biocompatibility tests:
- Engineering Specifications / Industry Standards: For characteristics like appearance, dimension, puncture depth, puncture force, connection firmness, launch performance, and safety features. These are objective measures against predefined criteria.
- Sterility Standards: Based on a Sterility Assurance Level (SAL) of 10^-6, which is an industry standard.
- Biocompatibility Standards: Based on ISO 10993 series standards (e.g., ISO 10993-5, 10993-10, 10993-11), which define the methodologies and acceptance criteria for biological evaluation of medical devices.
8. The sample size for the training set
Not applicable. This is a traditional medical device, not an AI/ML algorithm. There is no concept of a "training set" in this context.
9. How the ground truth for the training set was established
Not applicable. As there is no training set for an AI/ML algorithm, this question is irrelevant to the Lianfa Safety Lancet.
Ask a specific question about this device
Page 1 of 1