Search Results
Found 6 results
510(k) Data Aggregation
(91 days)
LVivo Software Application
LVivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease and Age >18. In addition, it has the ability to provide Quality Score feedback.
The LVivo platform is a software system for automated analysis of ultrasound examinations. Automated analysis of echocardiographic examinations is done using DICOM movies. The LVivo platform supports global and segmental evaluation of the left ventricle (LV) of the heart utilizing the apical views. Global LV function evaluation by ejection fraction (EF) is done based on two of the apical views: four-chamber (4CH) and two-chamber (2CH). Segmental LV function evaluation is done from three apical views 4CH, 2CH and three chamber (3CH) and supports segmental wall motion evaluation and strain. The LVivo platform supports also LV function evaluation from the parasternal views including global and segmental LV function analysis from the Short Axis (SAX) view and the global LV function analysis form the Parasternal Long Axis (PLAX) views.
In addition to the LV analysis, the cardiology toolbox includes a module for automated evaluation of the Right Ventricular function. The LVivo platform includes one additional non-cardiac module for the measurement of the bladder volume.
The LVivo Platform includes also two additional configurations: LVivo Seamless for offline analysis based on automatically selected views and LVivo IQS for real-time quality feedback during image acquisition
Here's an analysis of the provided text to fulfill your request, broken down by the specified information points:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" for the performance metrics (Specificity, Sensitivity, Accuracy, Correlation) in a pass/fail format prior to presenting the results. Instead, it presents the achieved performance. However, we can infer that the reported values are what the manufacturer considers acceptable for equivalence.
Inferred Acceptance Criteria (Based on Study Results):
Parameter | Acceptance Criteria (Inferred) | Reported Performance (Main Study) | Reported Performance (Second Validation Set) |
---|---|---|---|
Specificity | > 79% | 79% | 82% |
Sensitivity | > 82% | 82% | 82% |
Accuracy | > 82% | 82% | 82% |
Correlation (Pearson) | > 0.89 | 0.89 | 0.856 |
ICC (GT vs LVivo SWM) | > 0.85 | 0.85 | Not reported for this set |
Automated Analysis Success Rate | > 84% | 84% (139 out of 166 exams) | Not explicitly stated, 'n=78' implies successful processing |
2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Main Test Set:
- Sample Size: 170 echo exams.
- Data Provenance:
- 101 exams from a medical center in the US.
- 69 exams from two medical centers in Israel.
- Nature of Data: Retrospective (implied by the description "The exams were collected..." and subsequent analysis; not explicitly stated as prospective).
- Second Validation Set:
- Sample Size: 101 patients.
- Data Provenance: From a hospital in Taiwan.
- Nature of Data: Retrospective (implied).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Number of Experts: Three cardiologists.
- Qualifications of Experts: Specializing in echo. (No further details on specific experience or certifications are provided).
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication Method: "WMSI average across three cardiologists specializing in echo." This indicates a consensus approach where the average of the three experts' interpretations formed the ground truth (e.g., if one expert was an outlier, their reading would still contribute to the average).
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with AI assistance versus without AI assistance was not conducted or reported.
- The study focuses on the standalone performance of the AI algorithm compared to expert-defined ground truth. It also shows the inter-expert variability (ICC between experts).
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance: Yes, the reported performance metrics (Specificity, Sensitivity, Accuracy, Pearson Correlation, ICC) are for the standalone algorithm's performance against the established ground truth, without human intervention or assistance during the assessment.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Type of Ground Truth: Expert consensus. Specifically, "WMSI average across three cardiologists specializing in echo."
8. The sample size for the training set
- Training Set Sample Size: The document does not provide the sample size for the training set. It only describes the validation sets.
9. How the ground truth for the training set was established
- Training Set Ground Truth: The document does not provide information on how the ground truth for the training set was established. It only details the ground truth establishment for the test/validation sets.
This analysis is based solely on the provided text. Any information not explicitly stated in the document cannot be inferred.
Ask a specific question about this device
(144 days)
LVivo Software Application
LVivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it has the ability to provide Quality Score feedback.
The LVivo platform is a software system for automated analysis of ultrasound examinations. Automated analysis of echocardiographic examinations is done using DICOM movies. The LVivo platform supports global and segmental evaluation of the left ventricle (LV) of the heart. The global LV function is evaluated from two of the apical views: four-chamber (4CH) and two-chamber (2CH) by ejection fraction (EF). The segmental LV function is done from three apical views 4CH, 2CH and three chamber (3CH) and supports wall motion evaluation and strain. LVivo CE-EF (Contrast EF) extends the current toolbox of the LVivo platform by providing the ability to process Ultrasonic DICOM images which acquire by Ultrasound Equipment in which the patient was prescribed a contrast agent. In addition to the LV analysis, the cardiology toolbox includes a module for automated evaluation of the Right Ventricular function. The LVivo platform includes one additional non-cardiac module for the measurement of the bladder volume.
Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Reported Device Performance
Parameter | Acceptance Criteria (Implied by Study Design) | Reported Device Performance |
---|---|---|
EDV 4CH Correlation (Automated vs. Manual) | High correlation (implicitly, r > 0.90) | r = 0.96, CI [0.94, 0.97] |
EDV 4CH Bland-Altman Mean Difference (Bias) | Close to 0 (implicitly, within acceptable clinical limits) | -2.68 ml |
EDV 4CH Bland-Altman Limits of Agreement (LOA) | Narrow range (implicitly, clinically acceptable spread) | (-36.02, 30.66) |
ESV 4CH Correlation (Automated vs. Manual) | High correlation (implicitly, r > 0.90) | r = 0.98, CI [0.97, 0.99] |
ESV 4CH Bland-Altman Mean Difference (Bias) | Close to 0 | -3.87 ml |
ESV 4CH Bland-Altman Limits of Agreement (LOA) | Narrow range | (-25.58, 17.82) |
EF 4CH Correlation (Automated vs. Manual) | High correlation (implicitly, r > 0.90) | r = 0.95, CI [0.93, 0.97] |
EF 4CH Bland-Altman Mean Difference (Bias) | Close to 0 | 1.26% points |
EF 4CH Bland-Altman Limits of Agreement (LOA) | Narrow range | (-8.42, 10.96) |
EDV 2CH Correlation (Automated vs. Manual) | High correlation (implicitly, r > 0.90) | r = 0.96, CI [0.94, 0.97] |
EDV 2CH Bland-Altman Mean Difference (Bias) | Close to 0 | -5.69 ml |
EDV 2CH Bland-Altman Limits of Agreement (LOA) | Narrow range | (-36.02, 24.44) |
ESV 2CH Correlation (Automated vs. Manual) | High correlation (implicitly, r > 0.90) | r = 0.98, CI [0.97, 0.99] |
ESV 2CH Bland-Altman Mean Difference (Bias) | Close to 0 | -3.87 ml |
ESV 2CH Bland-Altman Limits of Agreement (LOA) | Narrow range | (-25.58, 17.82) |
EF 2CH Correlation (Automated vs. Manual) | High correlation (implicitly, r > 0.90) | r = 0.93, CI [0.90, 0.95] |
EF 2CH Bland-Altman Mean Difference (Bias) | Close to 0 | -0.54% points |
EF 2CH Bland-Altman Limits of Agreement (LOA) | Narrow range | (-12.18, 11.1) |
BP EDV Correlation (Automated vs. Manual) | High correlation (implicitly, r > 0.90) | r = 0.98, CI [0.97, 0.99] |
BP EDV Bland-Altman Mean Difference (Bias) | Close to 0 | -4.1 ml |
BP EDV Bland-Altman Limits of Agreement (LOA) | Narrow range | (-29.04, 20.84) |
BP ESV Correlation (Automated vs. Manual) | High correlation (implicitly, r > 0.90) | r = 0.98, CI [0.97, 0.99] |
BP ESV Bland-Altman Mean Difference (Bias) | Close to 0 | -2.77 ml |
BP ESV Bland-Altman Limits of Agreement (LOA) | Narrow range | (-19.67, 14.13) |
BP EF Correlation (Automated vs. Manual) | High correlation (implicitly, r > 0.90) | r = 0.96, CI [0.94, 0.97] |
BP EF Bland-Altman Mean Difference (Bias) | Close to 0 | 0.39% points |
BP EF Bland-Altman Limits of Agreement (LOA) | Narrow range | (-8.26, 9.05) |
Automatic Processing Rate | High rate (implicitly, > 80%) | 90% (91/101 exams) |
Study Information
-
Sample size used for the test set and the data provenance:
- Sample Size: 101 patient exams.
- Data Provenance: The data was collected from multiple sources: Beth Israel, Soroka, Hadassah (presumably Israel, based on the manufacturer's location), and UCMC (likely a US center given the race information collection). The text indicates that 69 of the 101 patients were collected from the US. The study is retrospective, as it refers to collected "patient exams."
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: 3 sonographers.
- Qualifications: The text explicitly states "3 sonographers" were used. While their specific years of experience are not mentioned, the implication is that they are qualified medical professionals capable of performing and interpreting ultrasound measurements. The subsequent review by a cardiologist also suggests oversight and validation of their measurements.
-
Adjudication method for the test set:
- The ground truth was "comprised of the measurements by the 3 sonographers." This suggests a consensus or averaging approach among the three sonographers. It further states, "No further changes by the cardiologist to the measurements were needed following the cardiologist's review," implying the cardiologist reviewed and implicitly approved the sonographers' consensus measurements. This could be interpreted as a form of expert consensus adjudication, where the sonographers' measurements formed the basis, and a cardiologist provided final validation.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not described. This study directly compared the algorithm's performance to human (sonographer) ground truth measurements rather than assessing human reader improvement with AI assistance.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone performance evaluation of the algorithm was done. The study compares the "automated results" of the LVivo Software Application to the established "manual measurements" (ground truth). The fact that the algorithm "processed automatically 91/101 (90%) of the exams" and these automated results were compared to the manually derived ground truth confirms a standalone evaluation.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The ground truth used was expert consensus based on "manual measurements" performed by "3 sonographers," which were then reviewed and confirmed by a cardiologist.
-
The sample size for the training set:
- The document does not provide the sample size for the training set. It only discusses the performance evaluation using the test set of 101 patient exams.
-
How the ground truth for the training set was established:
- The document does not provide any information on how the ground truth for the training set was established. The focus of this submission is on the performance evaluation of the final device using a designated test set.
Ask a specific question about this device
(219 days)
LVivo Software Application
L Vivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease. In addition, it has the ability to provide Quality Score feedback.
The LVivo platform is a software system for automated analysis of ultrasound examinations. Automated analysis of echocardiographic examinations is done using DICOM movies. The LVivo platform supports global and segmental evaluation of the left ventricle (LV) of the heart. The global LV function is evaluated from two of the apical views: four-chamber (4CH) and two-chamber (2CH) by ejection fraction (EF). The segmental LV function is done from three apical views 4CH, 2CH and three chamber (3CH) and supports wall motion evaluation and strain. The LVivo platform supports also global and segmental evaluation of the LV from the parasternal Short Axis (SAX) view and the Parasternal Long Axis view (PLAX).
LVivo PLAX extends the current toolbox of the LVivo platform, providing automated calculation of LV size and function parameters from the PLAX view, for cardiac function evaluation as part of the standard echocardiographic assessment in Echo environment and point-of-care. The measurements from the PLAX view add information about the contraction of the LV, its size and wall thickness and can be used for rapid evaluation of the patient's condition in the point-of-care.
In addition to the LV analysis, the cardiology toolbox includes a module for automated evaluation of the Right Ventricular function and the LVivo Seamless. The LVivo platform includes one additional non-cardiac module for the measurement of the bladder volume.
Here's an analysis of the acceptance criteria and study as requested, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Parameter | Acceptance Criteria | Reported Device Performance (Primary Study) | Reported Device Performance (Additional Dataset - LVivo PLAX) |
---|---|---|---|
FS | correlation >=75% vs manual measurements | r=0.79 | Corr: 0.74* |
LVIDd | (Not explicitly stated, but implied by 'Results acceptance considered bias and LOA as well as correlation for the calculated parameters') | N/A | Corr: 0.9, Bias: 0.3cm, LOA: -0.2, 0.9 |
LVIDs | (Not explicitly stated) | N/A | Corr: 0.95, Bias: 0.42cm, LOA: -0.25, 1.1 |
IVSD | (Not explicitly stated) | N/A | Corr: 0.72, Bias: -0.16cm, LOA: -0.47, 0.13 |
PWD | (Not explicitly stated) | N/A | Corr: 0.68, Bias: -0.095cm, LOA: -0.21, 0.4 |
LV Mass | (Not explicitly stated) | N/A | Corr: 0.94, Bias: 9.14gr, LOA: -41.6, 59.88 |
Automatic Analysis Rate | Not explicitly stated as acceptance criteria, but reported | 82% | 90% |
Note: The acceptance criteria for parameters other than FS are not explicitly stated in the provided text, but the results for correlation, bias, and LOA are reported, implying these were considered for acceptance. The FS correlation of 0.74 for the additional dataset is slightly below the 0.75 acceptance criteria, but the text states "The primary end point was met" and makes an overall conclusion of substantial equivalence, implying this difference was deemed acceptable in context.
2. Sample Size Used for the Test Set and Data Provenance
- Primary Clinical Validation:
- Sample Size: 121 patients
- Data Provenance: Not explicitly stated, but mentioned to be "according to GCP standards."
- Retrospective or Prospective: Not explicitly stated.
- Additional Dataset (for LVivo PLAX validation):
- Sample Size: 40 patients
- Data Provenance: USA-based medical center
- Retrospective or Prospective: Not explicitly stated.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Two sonographers.
- Qualifications: Not explicitly stated beyond "sonographers."
4. Adjudication Method for the Test Set
- The text states "The correlation vs average manual measurements of two sonographers." This suggests a method of averaging measurements from two experts to establish the ground truth. It is not a 2+1 or 3+1 adjudication method where a third expert breaks ties, but rather a consensus via averaging.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- There is no mention of an MRMC comparative effectiveness study done to compare human readers with and without AI assistance. The study focuses on the standalone performance of the AI against human measurements.
6. Standalone Performance Study
- Yes, a standalone performance study was done. The study compares the LVivo Software Application's automated analysis results against manual measurements performed by sonographers. For the primary study, 82% of exams were automatically analyzed, and for the additional dataset, 90% were automatically analyzed, indicating standalone algorithm-only performance.
7. Type of Ground Truth Used
- Expert Consensus (via averaging): The ground truth was established by "average manual measurements of two sonographers."
8. Sample Size for the Training Set
- The document does not provide information regarding the sample size used for the training set. This information is typically proprietary to the manufacturer and not released in 510(k) summaries unless specifically required or relevant for substantial equivalence claims.
9. How the Ground Truth for the Training Set Was Established
- The document does not provide information on how the ground truth for the training set was established. This is also typically proprietary and not typically included in these types of summary documents.
Ask a specific question about this device
(28 days)
LVivo Software Application
LVivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease.
The LVivo platform is a software system for automated analysis of ultrasound examinations. Automated analysis of echocardiographic examinations is done using DICOM movies. The LVivo platform supports global and segmental evaluation of the left ventricle (LV) of the heart. The global LV function is evaluated from two of the apical views: four-chamber (4CH) and two-chamber (2CH) by ejection fraction (EF). The segmental LV function is done from three apical views 4CH, 2CH and three chamber (3CH) and supports wall motion evaluation and strain. The LVivo platform supports also global and segmental evaluation of the LV from the parasternal short axis (SAX) view. In addition to the LV analysis, the cardiology toolbox includes a module for automated evaluation of the Right Ventricular function. The LVivo platform includes one additional non-cardiac module for the measurement of the bladder volume.
Here's a breakdown of the acceptance criteria and the study details for the LVivo Software Application, based on the provided FDA 510(k) summary:
Acceptance Criteria and Device Performance
Parameter | Acceptance Criteria | Reported Device Performance |
---|---|---|
LVivo EF (Ejection Fraction) | Biplane EF correlation >= 80% | Implicitly met, as stated "similar or better EF biplane results in terms of correlation, specificity, sensitivity, and kappa with respect to subject device." Actual percentage not explicitly given but implied acceptable. |
Similar or better EF biplane results in terms of correlation, specificity, sensitivity, and kappa with respect to subject device (predicate device) | Implicitly met, as described above. | |
LVivo GLS (Global Longitudinal Strain) | Cutoff value = 75% compared to reference Wall Motion Score Index (WSMI) | Implicitly met, as stated "similar or better results compared to subject device." Actual percentage not explicitly given but implied acceptable. |
Similar or better results compared to subject device (predicate device) | Implicitly met, as described above. | |
LVivo SWM (Segmental Wall Motion) | Sensitivity >= 75% | Implicitly met, as stated "similar or better results compared to subject device." Actual percentage not explicitly given but implied acceptable. |
Similar or better results compared to subject device (predicate device) | Implicitly met, as described above. |
Study Details
-
Sample Size used for the test set and the data provenance:
- EF Analysis: 96 examinations from a dataset of 100 ambulatory and hospitalized patients.
- SWM and GLS Analysis: 98 examinations from the same dataset of 100 ambulatory and hospitalized patients.
- Data Provenance: Retrospective, collected according to GCP standards from ambulatory and hospitalized patients referred for routine transthoracic echocardiography. The country of origin is not explicitly stated.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided in the given text. The text mentions "reference WSMI" and comparisons to a "subject device" (likely the predicate device's performance benchmarks), but it does not detail how the initial ground truth for the reference measurements was established.
-
Adjudication method for the test set:
- This information is not explicitly provided. The text refers to "reference" values but does not describe the adjudication process (e.g., expert consensus) if multiple readers were involved in establishing these reference values.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- A MRMC comparative effectiveness study involving human readers with and without AI assistance was not described in this summary. The study focuses on evaluating the standalone performance of the AI algorithm against established reference values or the predicate device's performance.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone performance evaluation of the algorithm was performed. The acceptance criteria and "reported device performance" (implicitly met) directly relate to the algorithm's performance on the test set.
-
The type of ground truth used:
- The ground truth appears to be based on clinically established methods and possibly expert consensus/measurements, as implied by "reference" values and comparison to "reference WSMI." The summary states the data was collected for "routine transthoracic echocardiography," suggesting standard clinical measurements were used as a basis. However, the specific method of establishing this ground truth (e.g., expert manual measurements, pathology) is not explicitly detailed.
-
The sample size for the training set:
- The sample size for the training set is not provided in the given text. The document describes the test set but offers no details on the data used to train the "neural network" cited as a modification.
-
How the ground truth for the training set was established:
- This information is not provided in the given text.
Ask a specific question about this device
(145 days)
LVivo Software Application
L Vivo platform is intended for non-invasive processing of ultrasound images to detect, measure, and calculate relevant medical parameters of structures and function of patients with suspected disease.
The LVivo System analyzes echocardiographic patient examination DICOM movies for Global ejection fraction (EF) evaluation. EF is evaluated using two orthogonal planes, four-chamber (4CH) and two-chamber (2CH) views, to provide fully automated analyses of LV function from the echo examination movies. It also has the ability to measure strain and to evaluate the Right Ventricle and well as to measure the bladder.
This document describes the acceptance criteria and study results for DiA Imaging Analysis Ltd's LVivo Software Application, specifically focusing on its LVivoRV (Right Ventricular) and LVivo Bladder modules.
1. Table of Acceptance Criteria and Reported Device Performance
Module | Metric | Acceptance Criteria | Reported Device Performance |
---|---|---|---|
LVivoRV | FAC Correlation | 75% correlation (r ≥ 0.75) between LVivoRV's FAC measurement and manual FAC measurements by sonographers. This value is based on statistical data reported by FDA cleared systems for semi-automated RV function evaluation (e.g., EchoInsight by Epsilon). | Primary Endpoint Met: Excellent correlation (r=0.79, p |
Ask a specific question about this device
(71 days)
LVivo Software Application
DiaCardio's LVivo Software Application is intended for non-invasive processing of already acquired echocardiographic images in order to detect, measure, and calculate the left ventricular function evaluation. This measurement can be used to assist the clinician in a cardiac evaluation.
The LVivo System analyzes echocardiographic patient examination DICOM movies for Global ejection fraction (EF) evaluation. EF is evaluated using two orthogonal planes, four-chamber (4CH) and two-chamber (2CH) views, to provide fully automated analyses of LV function from the echo examination movies. It also has the ability to measure strain
Here's a breakdown of the acceptance criteria and study information for the LVivo Software Application, based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document focuses on the performance of the LVivoSG module and its correlation with established methods rather than explicit "acceptance criteria" presented as pass/fail thresholds. However, we can infer the desired performance from the study objectives and results.
Metric | Acceptance Criteria (Inferred from Study Objectives) | Reported Device Performance (LVivoSG) |
---|---|---|
Global Longitudinal Strain (GLS) vs. VVI | Good agreement with correlation coefficient (r) ≥ 0.8 | r = 0.85 (p |
Ask a specific question about this device
Page 1 of 1