510(k) Data Aggregation

    K Number
    K231437
    Device Name
    LF900
    Date Cleared
    2023-07-19

    (63 days)

    Product Code
    Regulation Number
    890.5650
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    LF900

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LF900 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, venous insufficiencies, Lymphedema.

    Device Description

    This product is an Intermittent Pneumatic Compression system to improve the blood circulation of the human body. The operating principle is that the air from the device will be delivered to the sleeve with 4 air chambers and the air pressurizes sequentially the chambers from 1st to 4th.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the LF900 Powered Inflatable Tube Massager does not describe a study that proves the device meets acceptance criteria in the way typically expected for AI/ML or diagnostic devices. Instead, it demonstrates substantial equivalence to a predicate device through non-clinical testing.

    Here's an breakdown based on your requested information, highlighting what is and is not present in the document:


    Acceptance Criteria and Study for LF900 Powered Inflatable Tube Massager

    This submission is for a Powered Inflatable Tube Massager, not an AI/ML or diagnostic device. Therefore, the "acceptance criteria" and "study" are focused on demonstrating substantial equivalence to a previously cleared predicate device, primarily through non-clinical performance and safety testing, rather than a clinical effectiveness study with ground truth establishment in the context of diagnostic accuracy.

    The key change in the LF900 (K231437) from a previous version (K203019) is a change in maximum pressure from 180mmHg to 140mmHg and the addition of a sleeve type. The sponsor asserts equivalence to the LF400 (K150033), which also has a maximum pressure of 140mmHg.


    1. A table of acceptance criteria and the reported device performance

    Since this is not a diagnostic device, the acceptance criteria are not presented in terms of diagnostic metrics (e.g., sensitivity, specificity). Instead, they are related to safety, electromagnetic compatibility, and basic product performance after the pressure change.

    Acceptance Criterion (Addressed by)Reported Device Performance
    Electrical SafetyComplied with IEC 60601-1:2012
    Electromagnetic CompatibilityComplied with IEC 60601-1-2:2014
    Product Performance (Pressure)Performance testing conducted to verify functionality with changed maximum pressure of 140mmHg.
    Product Performance (Sleeve Compatibility)Compatibility testing with added sleeve types.
    BiocompatibilityNot verified by FDA; device not intended for direct skin contact (requires use over clothing/socks).

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not applicable. The "test set" for this device consists of non-clinical retests and gap tests of the device itself, not patient data.
    • Data Provenance: The tests conducted are in-house tests ("Inhouse test] Product Performance Test report, RND-R-PRR-103-02(2022.06.13)"). This indicates the testing was performed by the manufacturer. Specific country of origin for the testing is not detailed beyond the manufacturer's location (Republic of Korea).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is not an AI/ML or diagnostic device requiring expert interpretation for ground truth. Its performance is assessed through objective engineering and safety standards.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods are used for subjective interpretations in diagnostic studies, which is not relevant here. Compliance with safety and performance standards is objective.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an intermittent pneumatic compression system and does not involve AI or human "readers."

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device does not contain an algorithm in the sense of AI/ML. Its performance is assessed as a standalone electromechanical device. Clinical testing was not required to demonstrate substantial equivalence.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the context of diagnostic accuracy. The "ground truth" for this device's performance would be the objective measurements obtained during electrical safety, electromagnetic compatibility, and product performance testing against specified engineering limits and standards.

    8. The sample size for the training set

    Not applicable. This device does not involve a training set as it is not an AI/ML algorithm.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203019
    Device Name
    LF900
    Date Cleared
    2021-06-22

    (264 days)

    Product Code
    Regulation Number
    890.5650
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    LF900

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LF900 is intended for use by medical professionals and patients at home, who are under medical supervision, in treating many conditions, such as: Primary lymphedema, Edema following trauma and sport injuries, Postimmobilization edema, Venous insufficiencies, Lymphedema.

    Device Description

    This product is an Intermittent Pneumatic Compression system to improve the blood circulation of the human body. The operating principle is that the air from the device will be delivered to the sleeve with 4 air chambers and the air pressurizes sequentially the chambers from 1st to 4th.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called LF900, a powered inflatable tube massager. However, this document does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML algorithm or diagnostic accuracy.

    The text focuses on demonstrating substantial equivalence to a predicate device (POWER-Q1000 Premium) based on:

    • Intended Use: Both devices are for treating conditions like lymphedema, edema, and venous insufficiencies.
    • Technological Characteristics: Comparison of physical specifications like power source, time, pressure, and number of chambers. Small differences in these parameters are justified as not raising new safety or effectiveness concerns.
    • Non-Clinical Testing: Compliance with electrical safety (IEC 60601 series) and device performance tests (pressure, time, sleeve durability).
    • Clinical Testing: The document explicitly states, "Clinical testing was not required to demonstrate the substantial equivalence of the LF900 to its predicate device."

    Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving the device meets those criteria in the context of AI/ML performance, ground truth, expert adjudication, or MRMC studies. The device described in this FDA clearance is a physical medical device, not an AI/ML diagnostic or assistive tool.

    Ask a Question