Search Results
Found 27 results
510(k) Data Aggregation
(157 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder Free Latex Examination Gloves meets all the current specifications listed under the ASTM Specification D3578-05, Standard Specification for Rubber Examination Gloves. The principal operation and mechanism of this device is to prevent contamination between patient and examiner and this principle is achieved through testing of barrier, physical properties and other testing stated in the performance data. This device is for over-the counter single use.
The provided text describes the acceptance criteria and the study that proves the device meets the acceptance criteria for the Shandong Intco Medical Products Co., Ltd Latex Examination Gloves, Powder Free.
Here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
The document provides two tables that fulfill this request:
- A "Technology Characteristic Comparison between the device and predicate device" table on page 4-5.
- A "Non-clinical Testing Summary" table on page 6.
Let's combine and summarize the relevant information from both into a single table for clarity. Note that some "acceptance criteria" are simply the standard reference, and the "reported device performance" indicates "Complied with the standard". For detailed numerical values, refer to the "Technology Characteristic Comparison" table.
| Characteristic / Parameter | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Product Code | LYY | LYY |
| Intended Use | Prevent contamination between patient and examiner. | Same as predicate. |
| Classification | Class I | Class I |
| Raw Rubber Material | Natural Rubber Latex | Natural Rubber Latex |
| Color | No color pigment added. Natural White | No color pigment added. Natural White |
| Overall Length Minimum | 230mm (Predicate: Meet ASTM D3578-05) | Average: 247 mm |
| Width (S, M, L) | S: 75-95mm; M: 85-105mm; L: 100-120mm | S: 85mm; M: 96 mm; L: 104 mm |
| Palm Thickness (Minimum) | 0.08mm | Average: 0.122 mm |
| Finger Thickness (Minimum) | 0.08mm | Average: 0.156 mm |
| Tensile Strength (before age) | Minimum 18 MPa | Average: 26.17MPa |
| Tensile Strength (After Age) | Minimum 14 MPa | Average: 24.82MPa |
| Stress at 500% Elongation | Maximum 5.5 MPa | 4.99 MPa |
| Ultimate Elongation (before age) | Minimum 500% | Average: 759.85% |
| Ultimate Elongation (after age) | Minimum 400% | Average: 678.62% |
| Freedom of Holes | Meet AQL 2.5 at G1 | Meet AQL 2.5 with G1 (Also Complied with ASTM D5151-06) |
| Residual powder test | Less than 2mg/glove | S: 0.43 mg/glove; M: 0.31 mg/glove; L: 0.47 mg/glove (Complied with ASTM D6124-06) |
| Protein Testing | Less than 50 µg/dm2 | S, M, L sizes: Less than 50 µg/dm2 |
| Primary Skin Irritation | Not an irritant (ISO 10993-10:2010) | Not an irritant |
| Dermal Sensitization | Not a sensitizer (ISO 10993-10:2010) | Not a sensitizer |
| Acute Systemic Toxicity | No evidence of systemic toxicity (from extract) | No evidence of systemic toxicity |
| Overall Compliance | Various ASTM and ISO standards | Complied with relevant standards (ASTM D6124-06, ASTM D5151-06, ASTM D3578-19, ISO 10993-10:2010) |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the sample sizes used for each specific test (e.g., how many gloves were tested for Freedom of Holes, or how many individuals for skin irritation). It only reports the average values or compliance statements.
The data provenance is not explicitly stated as retrospective or prospective for the non-clinical tests. The company, Shandong Intco Medical Products Co, Ltd, is located in China, suggesting the testing likely occurred there or was commissioned by them.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not applicable to the type of device and testing described. The device is a "Latex Examination Glove," and its performance is evaluated through physical, chemical, and biological non-clinical tests based on established international standards (ASTM, ISO). Expert interpretation of medical images or diagnoses is not involved in establishing the "ground truth" for glove performance. The "ground truth" here refers to the measured physical and chemical properties and biological reactions as defined by the standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. As noted above, the testing involves non-clinical measurements against predefined standard specifications, not a subjective interpretation requiring adjudication by multiple experts.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a medical glove, not an AI-assisted diagnostic tool or system that involves human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The device is a physical product (medical glove), not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth used for evaluating the Latex Examination Gloves is based on established international and national standards and specifications for medical gloves. This includes:
- ASTM D3578-19 (Standard Specification for Rubber Examination Gloves) for physical properties.
- ASTM D6124-06 (Reapproved 2017) (Standard Test Method for Residual Powder on Medical Gloves) for residual powder.
- ASTM D5151-06 (Reapproved 2015) (Standard Test Method for Detection of Holes in Medical Gloves) for freedom of holes.
- ISO 10993-10:2010 (Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization) for biocompatibility.
The "ground truth" is therefore the objective measurement of the glove's characteristics against these well-defined and accepted standard limits.
8. The sample size for the training set
Not applicable. The device is a physical medical product, not an AI/ML algorithm that requires a training set. The performance data is from non-clinical testing of the manufactured product.
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(68 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Powder Free Latex Patient Examination Glove, with Protein Content Labeling Claim, Chlorinated (Contains 50 Micrograms per dm2 of glove or Less of Water Extractable Protein) meets all the current specification for ASTM D3578-05 (2010).
The provided text is for a 510(k) Premarket Notification for a medical device, specifically a Powder Free Latex Patient Examination Glove. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, primarily through non-clinical performance data and comparison of technological characteristics, rather than extensive clinical studies as one might find for a novel diagnostic AI device.
Therefore, many of the typical acceptance criteria and study details requested (e.g., sample size for test set, number of experts for ground truth, adjudication method, MRMC studies, standalone algorithm performance, training set details) are not applicable to this type of device and submission. The "study" here refers to the testing performed to meet specific ASTM standards and biocompatibility requirements.
Here's a breakdown of the information that is applicable and how it relates to the acceptance criteria:
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are defined by existing ASTM standards and biocompatibility testing. The device's performance is reported as meeting these standards.
| Characteristic | Acceptance Criteria (Standards) | Reported Device Performance |
|---|---|---|
| Dimensions | ASTM D 3578-05 (2010) | Meets |
| Physical Properties | ASTM D 3578-05 (2010) | Meets |
| Freedom from pin-holes | ASTM D 5151-99 (2006), ASTM D 3578-05 (2010) | Meets |
| Powder Free Residue | ASTM D 6124-06, ASTM D 3578-05 (2010) | Meets |
| Protein Content | ASTM D 5712-10, ASTM D 3578-05 (2010) | Meets |
| Biocompatibility (Dermal Sensitization) | ISO 10993-10:2010 | Not a contact skin sensitizer |
| Biocompatibility (Primary Skin Irritation) | ISO 10993-10:2010 | Not a primary skin irritant |
| Intended Use | Disposable device for medical purposes, worn on examiner's hand to prevent contamination between patient and examiner. | Identical to predicate device |
| Regulation Number | 21 CFR Part 880.6250 | Identical to predicate device |
| Product Code | 80 LYY | Identical to predicate device |
| Design | Ambidextrous, in different sizes per ASTM D3578 dimension requirement. | Identical to predicate device |
| Materials | Natural Rubber Latex | Identical to predicate device |
| Color | Natural Color | Identical to predicate device |
| Sterility | Not Applicable (Non-Sterile) | Identical to predicate device |
| Single Use | Yes | Identical to predicate device |
| Packaging | Packed in Dispenser Boxes | Identical to predicate device |
| Labeling Claim | With Extractable Protein Content Labeling Claim (50 Micrograms per dm² of glove or less of Water Extractable Protein) | Identical to predicate device |
2. Sample size used for the test set and the data provenance
The document specifies that the device "meets all the current specification for ASTM D3578-05 (2010)" and refers to other ASTM and ISO standards for specific characteristics. Clinical data was "not needed for market cleared examination gloves."
- Sample Size for Test Set: The specific sample sizes for each test (e.g., pin-holes, dimensions, physical properties) are not detailed in this 510(k) summary. These would typically be specified within the individual ASTM and ISO standards referenced. For example, ASTM D5151-99 for freedom from pinholes often specifies a sample size based on the acceptable quality level (AQL).
- Data Provenance: The document does not explicitly state the country of origin for the data (e.g., where the tests were physically conducted) nor whether it was retrospective or prospective. Given the nature of performance testing against standards, it would be prospective testing of newly manufactured glove batches. The manufacturer is "Top Calibre Sdn Bhd" in Selangor, Malaysia, so the testing was likely conducted by or for them.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not Applicable: This type of device does not involve a "ground truth" established by experts in a diagnostic or interpretive sense. The ground truth for performance characteristics is defined by the objective measurement criteria within the relevant ASTM and ISO standards (e.g., tensile strength, elongation, water leak rate, protein content measurement). Biocompatibility tests are performed by qualified laboratories and assessed against established toxicology criteria.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not Applicable: Adjudication methods like 2+1 or 3+1 are used for human-reviewed data, often in diagnostic imaging or clinical trials, to resolve discrepancies between readers. For physical and chemical tests of gloves, results are objectively measured against defined thresholds in the standards; there is no subjective interpretation requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not Applicable: This device is a patient examination glove, not an AI-powered diagnostic tool. Therefore, MRMC studies and assessments of AI assistance are irrelevant.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not Applicable: This device is a physical medical device (glove), not an algorithm or software. Standalone algorithm performance is not relevant.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance is established by objective measurements against predefined criteria and thresholds set forth in the referenced ASTM and ISO standards.
- For Dimensions, Physical Properties, Freedom from pin-holes, Powder Free Residue, Protein Content: The ground truth is the quantitative measurement of these properties (e.g., thickness, tensile strength, percentage of gloves passing water leak test, specific protein content in µg/dm²) compared to the specified limits in standards like ASTM D3578-05 (2010), ASTM D5151-99 (2006), ASTM D6124-06, and ASTM D5712-10.
- For Biocompatibility: The ground truth is determined by the results of standardized biological tests (Dermal Sensitization and Primary Skin Irritation Tests as per ISO 10993-10:2010), which assess the device's potential to cause adverse biological reactions. The "ground truth" in this context is the observation of whether a reaction occurs and its severity, interpreted against established toxicity criteria, leading to a "pass" or "fail" determination (e.g., "Not a contact skin sensitizer," "Not a primary skin irritant").
8. The sample size for the training set
Not Applicable: This device does not use machine learning or AI, so there is no "training set."
9. How the ground truth for the training set was established
Not Applicable: As there is no training set, this question is not relevant.
Ask a specific question about this device
(62 days)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Powdered Latex Patient Examination Glove with Protein Content Labeling Claim (Contains 200 Micrograms per dm2 of glove or Less of Water Extractable Protein) meets all the current specification for ASTM D3578-05 (2010).
Acceptance Criteria and Device Performance Study for K120693 - Powdered Latex Patient Examination Glove
This document describes the acceptance criteria and the study demonstrating that the "Powdered Latex Patient Examination Glove with Protein Content Labeling Claim (Contains 200 Micrograms per dm² of glove or Less of Water Extractable Protein)" (K120693) meets these criteria.
1. Table of Acceptance Criteria and Reported Device Performance
The device's performance was evaluated against established ASTM standards and ISO standards for biocompatibility.
| Characteristic | Acceptance Criteria (Standards) | Reported Device Performance |
|---|---|---|
| Dimensions | ASTM D 3578-05 (2010) | Meets |
| Physical Properties | ASTM D 3578-05 (2010) | Meets |
| Freedom from pin-holes | ASTM D 5151-99 (2006) & ASTM D 3578-05 (2010) | Meets |
| Powder Amount | ASTM D 6124-06 & ASTM D 3578-05 (2010) | Meets |
| Protein Content | ASTM D 5712-10 & ASTM D 3578-05 (2010) | Meets |
| Biocompatibility: | ||
| - Dermal Sensitization | ISO 10993-10:2010 | Not a contact skin sensitizer |
| - Primary Skin Irritation Test | ISO 10993-10:2010 | Not a primary skin irritant |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for each specific test or the country of origin for the data. However, the evaluation relies on compliance with established ASTM and ISO standards, which typically involve specific sampling plans and test methods. The studies are non-clinical performance tests.
3. Number of Experts Used to Establish Ground Truth and Qualifications
Not applicable. For this type of medical device (patient examination gloves), the "ground truth" is established by adherence to recognized national and international consensus standards (ASTM and ISO), not by expert consensus on clinical data or image interpretation. The tests are laboratory-based and objective measurements.
4. Adjudication Method for the Test Set
Not applicable. The performance testing involves objective measurements against established standard specifications, not subjective interpretation requiring adjudication among experts.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
Not applicable. This device is a patient examination glove, and its performance is assessed through non-clinical laboratory testing against physical, chemical, and biocompatibility standards. MRMC studies are typically relevant for diagnostic imaging or similar devices that involve human interpretation of results.
6. Standalone Performance Study
Yes, a standalone (algorithm only without human-in-the-loop performance) study was done in the sense that the device itself, a physical product, was subjected to a series of laboratory tests to evaluate its inherent characteristics against the specified standards. Human interaction is limited to conducting the tests according to the standard protocols and recording the objective results.
7. Type of Ground Truth Used
The ground truth used is based on established national and international consensus standards (ASTM and ISO). These standards define objective pass/fail criteria and test methodologies for dimensions, physical properties, freedom from pinholes, powder amount, protein content, and biocompatibility.
8. Sample Size for the Training Set
Not applicable. This device is a physical product, not an algorithm or AI model that requires a training set. The evaluation is based on direct testing of the manufactured product batches against standardized criteria.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" for this device. The ground truth for evaluating the device's performance is established by the specified ASTM and ISO standards themselves, which outline the acceptable ranges and test methods.
Ask a specific question about this device
(107 days)
A powder-free patient examination glove is a disposable device intended for medical purposes worn on the examiner's hand or finger to prevent contamination between patient and examiner
Latex Examination Gloves Powder-Free (Black Color) and with protein content labeling claim(50 micrograms or less)
This document is a 510(k) premarket notification determination letter from the FDA for "Latex Examination Gloves Powder-Free (Black Color) and with Protein Content Labeling Claim (50 micrograms or less)". It does not contain information about acceptance criteria or a study that proves the device meets specific acceptance criteria in the context of performance metrics like accuracy, sensitivity, or specificity.
Instead, the letter confirms that the device is "substantially equivalent" to legally marketed predicate devices. This means the FDA found that the new device is as safe and effective as a previously cleared device. The "acceptance criteria" in this context would primarily pertain to compliance with FDA regulations, good manufacturing practices, and performance characteristics (like protein content, physical properties such as tensile strength and elongation, and freedom from holes) that demonstrate substantial equivalence to predicate devices, rather than a statistical study proving clinical performance of an AI/ML diagnostic tool.
Therefore, I cannot provide the requested information from the provided text as it is not present in this type of FDA document.
Ask a specific question about this device
(70 days)
Latex Examination Glove Powder Free with Neoprene lined is a disposable device intended for medical purpose that is worn on examiner's hand to prevent contamination between patient and examiner.
Latex Examination Gloves Powder Free With Lined.
The provided text describes a 510(k) premarket notification for "Latex Examination Gloves Powder Free With Lined." This submission is for a medical device that does not involve an AI component or complex analytical software. Therefore, the questions related to AI performance, ground truth, expert adjudication, and multi-reader multi-case studies are not applicable.
The document focuses on demonstrating that the gloves meet established industry standards for physical properties and safety.
Here's an analysis of the available information regarding the acceptance criteria and the study that proves the device meets those criteria:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria (Minimum) | Reported Device Performance (Reference) |
|---|---|---|
| Dimensions | (ASTM D 3578-01ae2 Standard) | |
| Length mm (min.) | 220 (Small), 230 (Medium, Large, X-Large) | Met (Certify: "meets or exceed the ASTM D 3578-01ae2 Standard") |
| Palm Width mm | 80±10 (Small), 95±10 (Medium), 111±10 (Large), 120±10 (X-Large) | Met (Certify: "meets or exceed the ASTM D 3578-01ae2 Standard") |
| Thickness (min.) | ||
| - Cuff mm | 0.08 | Met (Certify: "meets or exceed the ASTM D 3578-01ae2 Standard") |
| - Palm mm | 0.08 | Met (Certify: "meets or exceed the ASTM D 3578-01ae2 Standard") |
| - Finger Tip mm | 0.08 | Met (Certify: "meets or exceed the ASTM D 3578-01ae2 Standard") |
| Physical Properties | ||
| Tensile Strength | ||
| - Before ageing | 18 Mpa (min) | Met (Certify: "meets or exceed the ASTM D 3578-01ae2 Standard") |
| - After ageing (70°C, 168 hrs) | 14 Mpa (min) | Met (Certify: "meets or exceed the ASTM D 3578-01ae2 Standard") |
| Ultimate Elongation | ||
| - Before ageing | 650 % (min.) | Met (Certify: "meets or exceed the ASTM D 3578-01ae2 Standard") |
| - After ageing (70°C, 168 hrs) | 500 % (min.) | Met (Certify: "meets or exceed the ASTM D 3578-01ae2 Standard") |
| Safety/Quality | ||
| Pin Hole Requirement | FDA requirement | Met (Certify: "Meets FDA pin hole requirement") |
| Labeling Claim | Accuracy of labeling information | Met (Certify: "Meets labeling claim") |
| Total Water Extractable Protein | ≤ 50 micrograms per gram | Met (Stated in device name and indication for use) |
Notes:
- The document states that the performance data is "the same as mentioned immediately above," referring to the detailed table of dimensions and physical properties.
- The primary method of demonstrating acceptance is through a certification statement that the gloves meet or exceed the ASTM D 3578-01ae2 Standard and FDA pin hole requirements.
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes used for testing any specific batch or lot of gloves. It refers to the ASTM D 3578-01ae2 Standard for latex examination gloves, which would outline the sampling plan for such tests.
- Sample Size: Not explicitly stated in the provided text.
- Data Provenance: The manufacturer is PT. Shamrock Manufacturing Corp. located in Medan, Indonesia. The testing would have been conducted by them or a contracted laboratory to demonstrate compliance with the ASTM standard. The data is retrospective in the sense that it represents the characteristics of manufactured gloves.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This question is not applicable. For medical gloves, "ground truth" is established by adherence to recognized international and national standards (e.g., ASTM standards) and regulatory requirements (e.g., FDA pinhole requirements). Testing is typically performed according to standardized protocols, and the results are compared against predefined criteria in these standards, not against expert consensus on individual "cases."
4. Adjudication Method for the Test Set
This question is not applicable. As explained above, for physical and chemical testing of gloves, results are typically objective measurements compared against specified quantitative criteria rather than subjective assessments requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable. The device is a physical medical glove and does not involve AI or human readers for its intended use or evaluation.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable. The device is a physical medical glove and does not involve an algorithm.
7. The Type of Ground Truth Used
The "ground truth" for this device is based on:
- Industry Standards: Specifically, the ASTM D 3578-01ae2 Standard for Latex Examination Gloves. This standard defines the acceptable ranges for dimensions, tensile strength, and elongation, as well as test methods.
- Regulatory Requirements: The FDA pinhole requirement for medical gloves, which specifies an Acceptable Quality Level (AQL) for watertight integrity.
- Labeling Claims: Verification that the product meets its stated specifications, including the protein content claim ("Contains ≤ 50 micrograms or less of total water extractable protein per gram").
8. The Sample Size for the Training Set
This question is not applicable. As this is not an AI/ML device, there is no "training set." The manufacturing process is controlled to consistently produce gloves that meet the specified standards.
9. How the Ground Truth for the Training Set was Established
This question is not applicable, as there is no "training set" for this type of device. The quality control and testing processes ensure that each batch of gloves (or a statistically significant sample from it) meets the established "ground truth" defined by the ASTM standard and FDA regulations.
Ask a specific question about this device
(35 days)
A patient examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Latex Examination Gloves, Powder Free, Polymer Coated
This is an FDA 510(k) clearance letter for "Latex Examination Gloves, Powder Free, Polymer Coated". This document is a regulatory approval, not a scientific study report. Therefore, it does not contain the detailed information requested regarding acceptance criteria, device performance, study design, or ground truth establishment.
Here's why the requested information cannot be extracted and what the document does provide:
- No Acceptance Criteria or Device Performance Table: The document states that the device is "substantially equivalent" to legally marketed predicate devices. It does not provide specific performance metrics (e.g., tensile strength, puncture resistance, barrier integrity) or the acceptance criteria for those metrics. This level of detail would typically be found in the 510(k) submission itself, not the clearance letter.
- No Study Details (Sample Size, Data Provenance, Ground Truth, Training Set): As this is a clearance letter and not a study report, it does not describe any specific studies conducted by the manufacturer to demonstrate performance. The "substantial equivalence" determination relies on demonstrating that the new device is as safe and effective as a predicate device already on the market, often through comparison of materials, design, and intended use, and sometimes through performance testing if there are significant differences from the predicate. However, the details of such testing, if performed, are not in this letter.
- No Information on Experts, Adjudication, or MRMC Studies: These concepts are typically relevant for AI/ML-based diagnostic devices or complex imaging analysis, where expert interpretation and consensus are crucial for establishing ground truth and evaluating human-AI interaction. This document is for a physical medical device (examination gloves), so these types of studies are not applicable or mentioned.
- Standalone Performance: The concept of "standalone performance" (algorithm only) is not relevant to a physical medical device like gloves.
What the document does tell us:
- Device Name: Latex Examination Gloves, Powder Free, Polymer Coated
- Regulation Number & Product Code: 880.6250, LYY (Patient Examination Gloves)
- Regulatory Class: I (lowest risk class, subject to general controls)
- Indications For Use: A disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
- Date of Clearance: June 7, 2002
- Type of Clearance: 510(k) Premarket Notification, based on "substantial equivalence" to a predicate device.
To obtain the detailed information requested, one would need to access the original 510(k) submission (K021412), which is usually publicly available through the FDA's databases, although often redacted for proprietary information. That submission would describe the specific tests performed (e.g., physical property testing, biocompatibility, leak testing) and the criteria used to demonstrate that the gloves meet recognized standards and are substantially equivalent to marketed predicate devices.
Ask a specific question about this device
(83 days)
EXAMINATION GLOVES POWDER FREE IS A POWDER FREE LATEX DISPOSABLE DEVICE MADE OF NATURAL RUBBER LATEX INTENDED FOR MEDICAL PURPOSE THAT IS WORN ON THE EXAMINERS HAND ob eximali contrail contrainer time between batten and EXAMINER.
Class I Patient Examination Gloves Latex (Powder free) 80L Y Y that meets all the requirements of ASTM D3578
The device is a Latex Examination Glove (Powder free) manufactured by Brightway Gloves Pvt. Ltd., intended for medical purposes to prevent contamination between patients and examiners. The acceptance criteria and performance data are detailed below, comparing the device's characteristics against ASTM D3578 requirements.
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Size | Acceptance Criteria (ASTM D3578 Requirement) | Reported Device Performance (Brightway Gloves Value) | Meets Criteria? |
|---|---|---|---|---|
| Length | EX-S | 220 mm minimum | 235-240 mm | Yes |
| S | 220 mm minimum | 235-240 mm | Yes | |
| M | 230 mm minimum | 235-240 mm | Yes | |
| L | 230 mm minimum | 235-240 mm | Yes | |
| Width | EX S | 70 +/- 6 mm | 70 mm | Yes |
| S | 80 +/- 6 mm | 82 mm | Yes | |
| M | 95 +/- 6 mm | 93 mm | Yes | |
| L | 111 +/- 6 mm | 107 mm | Yes | |
| Thickness | EX S | 0.08 mm minimum | 0.10 mm | Yes |
| S | 0.08 mm minimum | 0.10 mm | Yes | |
| M | 0.08 mm minimum | 0.10 mm | Yes | |
| L | 0.08 mm minimum | 0.10 mm | Yes | |
| Physical Properties (Before Ageing) | ||||
| Tensile Strength | 14 mpa min | 20 - 22 mpa | Yes | |
| Elongation at break % | 700% min | 750 - 800% (or 750-850% in performance data) | Yes | |
| Physical Properties (After Ageing) | ||||
| Tensile Strength | 14 mpa min | 18 - 20 mpa | Yes | |
| Elongation at break % | 500% min | 700 - 800% | Yes | |
| Performance Requirement | ||||
| Freedom from Holes | Holes | AQL 4 | AQL 1.5 | Yes (lower AQL is better) |
| Dimension | Width, Length, Thickness | AQL 4 | AQL 4 | Yes |
| Physical Property | Tensile Strength, Elongation at Break | AQL 4 | AQL 4 | Yes |
| Powder Content | Not explicitly stated in ASTM D3578, but implied by "powder free" | 1 +/- 1 mg per glove | N/A (meets "powder free" claim) | |
| Protein Content | Not explicitly stated in ASTM D3578 | 80 +/- 20 ppm | N/A | |
| Moisture Content | Not explicitly stated in ASTM D3578 | 0.8% max | N/A | |
| Biocompatibility | Not explicitly stated in ASTM D3578 | Biologically Compatible | N/A (meets general expectation) |
2. Sample size used for the test set and the data provenance
The document states that the performance test data is for "Latex Examination Glove powder free manufactured by M/S Brightway Gloves Pvt.Ltd." However, the sample size used for these tests is not explicitly stated in the provided document. The data provenance is
retrospective by the manufacturer, Brightway Gloves Pvt. Ltd., located in Tamil Nadu, India.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The ground truth for the physical and performance properties of the gloves is established by the ASTM D3578 standard specifications. No human expert "ground truthing" is mentioned or implied for these types of measurements.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. This is a measurement of physical and performance characteristics against a standard, not a subjective assessment requiring adjudication among experts.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a Latex Examination Glove, not an AI or diagnostic medical device that would involve human readers or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a Latex Examination Glove, not an AI or algorithmic device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth is based on the ASTM D3578 standard specifications for Latex Examination Gloves. These are established industry standards for the physical and performance characteristics of such gloves.
8. The sample size for the training set
Not applicable. This is a physical medical device, not a machine learning model, so there is no concept of a "training set."
9. How the ground truth for the training set was established
Not applicable, as there is no training set for this type of device.
Ask a specific question about this device
(59 days)
A patient examination glove is a disposable device intended for medical purpose than worn on the examiner's hands or fingers to prevent contamination between patient ar. examiner. (21 CER 880.6250)
LATEX EXAMINATION GLOVES POWDERFREE POLYMER COATED WITH PROTEIN CONTENT LABELING CLAIM ( 50 MICROGRAMS OF LESS)
This document is a 510(k) clearance letter for latex examination gloves. It does not contain information about the acceptance criteria or a study that proves a device meets such criteria for an AI/ML-driven medical device. Therefore, I cannot extract the requested information.
Ask a specific question about this device
(43 days)
A powder free examination glove is a disposable device intended for medical purpose that is worn on examiner's hand to prevent contamination on examiner's hand: to prevent of ・・・・・・・・・・・・・・・ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ー …… '
Powder free Latex Examination Glove is a disposable device intended for medical purpose that is worn on examiner's hand to prevent contamination between patient and examiner.
Powder Free - Latex Examination Gloves
The provided document, a 510(k) summary for PT. Shamrock Manufacturing Corporation's Powder Free - Latex Examination Gloves, describes the device and its performance data. However, it does not include detailed acceptance criteria or a study that specifically "proves" the device meets these criteria in the typical sense of a clinical or comparative effectiveness study for AI/medical imaging devices.
Instead, this document focuses on demonstrating substantial equivalence to a legally marketed predicate device (Class I Examination Gloves 80 L YY meeting ASTM D 3578-99) by presenting performance data that meets established industry standards.
Here's an analysis based on the information provided, framed to address your requested points where applicable, and highlighting what is not present:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria (ASTM D 3578-99) | Reported Device Performance (Powder Free - Latex Examination Gloves) |
|---|---|---|
| Dimensions (Length mm min.) | 280 (for all sizes) | 280 (for Small, Medium, Large, X-Large) |
| Dimensions (Palm Width mm) | Small: 80±10, Medium: 95±10, Large: 111±10, X-Large: ≥ 110 | Small: 80±10, Medium: 95±10, Large: 111±10, X-Large: ≥ 110 |
| Thickness (Cuff mm min) | 0.20 | 0.20 |
| Thickness (Palm mm min) | 0.20 | 0.20 |
| Thickness (Finger Tip mm) | 0.20 | 0.20 |
| Tensile Strength (Before ageing) | : 21 Mpa (min) | : 21 Mpa (min) |
| Tensile Strength (After ageing at 70°C 168 hrs.) | : 18 Mpa (min) | : 18 Mpa (min) |
| Ultimate Elongation (Before ageing) | : 750 % (min.) | : 750 % (min.) |
| Ultimate Elongation (After ageing at 70°C 168 hrs.) | : 700 % (min.) | : 700 % (min.) |
| Pin Hole Requirement | Meets FDA pin hole requirement | Meets FDA pin hole requirement |
| Labeling Claim | Meets ASTM D 3578-99 | Meets labeling claim |
Explanation: The "acceptance criteria" for this device are the specifications outlined in the ASTM D 3578-99 standard and FDA pin hole requirements. The reported device performance matches these criteria exactly, as stated in section (6) "Technological characteristic of the gloves" and (9) "Non-clinical data."
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample size used for testing these physical and performance characteristics. The testing methodology for ASTM standards would dictate this, but it is not elaborated upon in this 510(k) summary. The data provenance is not explicitly stated in terms of country of origin for the testing, but the manufacturer is PT. Shamrock Manufacturing Corporation in Indonesia. The testing would be considered retrospective as it likely occurred during product development and validation prior to this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is not applicable in the context of this device and document. "Ground truth" in the sense of expert consensus for medical imaging interpretations (like a radiologist reading an image) is not relevant here. The ground truth for glove performance is established by objective measurements against a recognized standard (ASTM D 3578-99).
4. Adjudication method for the test set
Not applicable. Adjudication methods like "2+1" or "3+1" are used for resolving disagreements among human readers or expert panels, typically in subjective assessment tasks. The testing of gloves involves objective measurements against a standard, not subjective interpretations requiring adjudication.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a medical device (latex examination gloves), not an AI diagnostic or assistive imaging device. Therefore, MRMC studies and AI assistance metrics are irrelevant.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is a physical medical device, not a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth used for this device is based on objective measurements against established engineering and material science standards (ASTM D 3578-99) and regulatory requirements (FDA pin hole requirement). This is a form of performance standard ground truth.
8. The sample size for the training set
Not applicable. This is a physical medical device. The concept of a "training set" is relevant for machine learning models, not for manufacturing specifications of gloves.
9. How the ground truth for the training set was established
Not applicable, as there is no "training set" in the context of this device. The standards (ASTM D 3578-99) were established through a consensus process by relevant industry and scientific bodies over time.
Ask a specific question about this device
(50 days)
Ask a specific question about this device
Page 1 of 3