K Number
K111158

Validate with FDA (Live)

Date Cleared
2011-08-10

(107 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A powder-free patient examination glove is a disposable device intended for medical purposes worn on the examiner's hand or finger to prevent contamination between patient and examiner

Device Description

Latex Examination Gloves Powder-Free (Black Color) and with protein content labeling claim(50 micrograms or less)

AI/ML Overview

This document is a 510(k) premarket notification determination letter from the FDA for "Latex Examination Gloves Powder-Free (Black Color) and with Protein Content Labeling Claim (50 micrograms or less)". It does not contain information about acceptance criteria or a study that proves the device meets specific acceptance criteria in the context of performance metrics like accuracy, sensitivity, or specificity.

Instead, the letter confirms that the device is "substantially equivalent" to legally marketed predicate devices. This means the FDA found that the new device is as safe and effective as a previously cleared device. The "acceptance criteria" in this context would primarily pertain to compliance with FDA regulations, good manufacturing practices, and performance characteristics (like protein content, physical properties such as tensile strength and elongation, and freedom from holes) that demonstrate substantial equivalence to predicate devices, rather than a statistical study proving clinical performance of an AI/ML diagnostic tool.

Therefore, I cannot provide the requested information from the provided text as it is not present in this type of FDA document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a circular logo with text around the perimeter and a symbol in the center. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. The central symbol consists of three stylized lines that resemble a person with outstretched arms, possibly representing health or assistance.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Innovative Gloves Company. Limited Mr. Rajeev Sood President 830 Moo 4 Sanambin-Ban Klang Road Kuanlang Hatvai. Songkhla Thailand 90110

AUG 1 0 2011

Re: K111158

Trade/Device Name: Latex Examination Gloves Powder-Free (Black Color) and with Protein Content Labeling Claim (50 micrograms or less) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: July 20, 2011 Received: July 26, 2011

Dear Mr. Sood:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Sood

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other l'ederal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm] 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Anthony D. Audone

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devi Office of Device Evaluation Center of Devices and Radiological Health

{2}------------------------------------------------

n-Ban Klang RD., Kuanlang, Hatyai, Songkhla 90110 Thalland 830 หมู่4 ถนน สนามบิน-บ้านกลาง ค.ควนลัง อ.หาดใหญ่ จ.สงขลา 907 TEL: (66) 502251

Indications for Use

Applicant: Innovative Gloves Co., Ltd.

510(k) Number (if known): K111158

Device Name: Latex Examination Gloves Powder-Free (Black Color) and with protein content labeling claim(50 micrograms or less)

A powder-free patient examination glove is a disposable device intended for medical purposes worn on the examiner's hand or finger to prevent contamination between patient and examiner

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

FEDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.