Search Results
Found 2 results
510(k) Data Aggregation
(28 days)
LANGSTON DUAL LUMEN PRESSURE MONITORING CATHETER
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring Catheter is indicated for delivery of contrast medium in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressure measurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients.
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring catheter is intended for use as a pressure measurement catheter and for delivery of contrast media during angiographic studies. The DLP catheter consists of the Merit Medical Softouch Diagnostic Intravascular Catheter (K943739) as the inner lumen and an extruded outer lumen designed by Vascular Solutions. The inner lumen and hub assembly are purchased from Merit Medical in a non-sterile state. The Merit Softouch catheter and outer lumen are joined together using an adapter junction placed near the proximal end of the Merit Softouch catheter. The adapter junction incorporates a side port fitted with a stopcock/tube assembly for fluid flow and pressure measurement within the outer tube. The distal end of the outer tube is perforated with side holes to allow pressure measurement simultaneously with the sideholes at the tip of the catheter. The dual lumen catheter is deployed through standard guide catheters and will accommodate standard 0.038'' diameter guidewires.
The provided text describes a 510(k) submission for the Langston™ Dual Lumen Pressure Monitoring Catheter. This device is an Intravascular Diagnostic Catheter and is an unclassified device. The submission focuses on demonstrating substantial equivalence to predicate devices rather than providing a study that assesses its performance against specific acceptance criteria in a clinical setting.
Therefore, much of the requested information cannot be extracted directly from this document because it outlines a regulatory submission process for a medical device rather than for an AI/machine learning device that would undergo rigorous performance evaluation against established reference standards.
Here's a breakdown of what can and cannot be answered based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance:
No specific "acceptance criteria" for device performance (e.g., accuracy, sensitivity, specificity) and corresponding reported performance are provided in the document. The submission focuses on demonstrating substantial equivalence to predicate devices based on design verification and biocompatibility.
2. Sample Size Used for the Test Set and Data Provenance:
No test set for evaluating the device's performance (in the context of AI/ML) is mentioned. The document refers to "design verification" non-clinical testing.
3. Number of Experts Used to Establish Ground Truth and Qualifications:
Not applicable. The document does not describe a study involving expert-established ground truth for device performance evaluation.
4. Adjudication Method:
Not applicable.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
No MRMC study was conducted or reported.
6. Standalone Performance:
No standalone (algorithm-only) performance is reported, as this is a physical medical device, not an AI algorithm.
7. Type of Ground Truth Used:
Not applicable in the context of an AI/ML device. For a physical medical device, "ground truth" would refer to its physical and functional characteristics verified through engineering and biocompatibility testing. The text states: "Testing conducted included assessments of the design verification of the Dual Lumen Pressure Monitoring Catheter along with biocompatibility assessments."
8. Sample Size for the Training Set:
Not applicable; no AI/ML model was trained.
9. How the Ground Truth for the Training Set Was Established:
Not applicable; no AI/ML model was trained.
Summary of Information from the Provided Text:
The document describes a 510(k) premarket notification for a medical device, not an AI/ML product. The regulatory pathway for this device relies on demonstrating substantial equivalence to existing predicate devices, primarily through non-clinical testing (design verification and biocompatibility) rather than clinical performance studies against specific acceptance criteria.
Key points from the document regarding the device's qualification:
- Non-Clinical Testing: "Testing conducted included assessments of the design verification of the Dual Lumen Pressure Monitoring Catheter along with biocompatibility assessments. The results of this battery of tests confirmed the suitability of the Dual Lumen Pressure Monitoring Catheter for its intended use."
- Clinical Testing: "No clinical evaluations of this product have been conducted." (This explicitly states the absence of human clinical trials).
- Predicate Devices: The device is deemed substantially equivalent to:
- Langston Dual Lumen Pigtail Catheters
- Cordis Dual Lumen Pressure Monitoring Catheter
- Merit Medical Softouch Diagnostic Intravascular Catheter
- Conclusion: "The testing performed confirms that the Langston Dual Lumen Pressure Monitoring Catheter will perform as intended." This statement refers to the non-clinical design verification and biocompatibility testing.
Ask a specific question about this device
(112 days)
LANGSTON DUAL LUMEN PRESSURE MONITORING CATHETER
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring Catheter is indicated for delivery of contrast media in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressure measurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients.
The Vascular Solutions Langston™ Dual Lumen Pressure Monitoring catheter is intended for use as a pressure measurement catheter and for delivery of contrast media during angiographic studies. The DLP catheter consists of the Merit Medical Softouch Diagnostic Intravascular Catherer (K943739) as the inner lumen and an extruded outer lumen designed by Vascular Solutions. The inner lumen and hub assembly are purchased from Merit Medical in a non-sterile state. The Merit Softouch catheter and outer lumen are joined together using an adapter junction placed near the proximal end of the Merit Softouch catherer. The adapter junction incorporates a side port fitted with a stopcock/tube assembly for fluid flow and pressure measurement within the outer tube. The distal end of the outer tube is perforated with multiple holes to allow pressure measurement simultaneously with the inner catheter. The dual lumen catheter is deployed through standard introducers and will accommodate standard 0.038'' diameter guidewires.
The provided document, K050168, is a 510(k) Summary of Safety and Effectiveness for the Langston™ Dual Lumen Pressure Monitoring Catheter. This document is a premarket notification for a medical device and, as such, does not contain information about acceptance criteria or a study proving the device meets acceptance criteria in the manner typically seen for performance evaluations of AI/software devices.
Instead, the document focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing. It explicitly states that "No clinical evaluations of this product have been conducted." Therefore, the information requested regarding acceptance criteria, device performance, sample sizes, ground truth establishment, expert involvement, and MRMC studies is not applicable to this submission, as such studies were not performed for this 510(k) clearance.
Here's a breakdown of the provided information in the context of your request:
1. A table of acceptance criteria and the reported device performance:
- Acceptance Criteria: Not explicitly stated in terms of performance metrics like sensitivity, specificity, accuracy, or similar quantitative measures for an AI/software device. The implicit acceptance criterion for a 510(k) is "substantial equivalence" to predicate devices, supported by non-clinical testing.
- Reported Device Performance: Instead of performance metrics, the document reports the outcome of non-clinical testing.
- "Testing conducted included assessments of the design verification of the Dual Lumen Pressure Monitoring Catheter and biocompatibility assessments."
- "The results of this battery of tests confirmed the suitability of the Dual Lumen Pressure Monitoring Catheter for its intended use."
- This confirms the device met its design specifications and biocompatibility requirements but doesn't provide specific quantitative performance data in the context of your request.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not Applicable. No clinical or performance test set data is reported. The "testing" refers to bench testing and biocompatibility assessments, not data from a test set of cases.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not Applicable. No ground truth establishment by experts for a test set was performed or reported.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not Applicable. No clinical test set and thus no adjudication method for ground truth.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not Applicable. This is a hardware device submission, not an AI/software device. No MRMC study was conducted or reported.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This is a hardware device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- Not Applicable. No ground truth was established for "performance" as it relates to clinical outcomes or interpretation in the way an AI/software device would be evaluated. The "ground truth" equivalent here would be the engineering specifications and established safety standards met through bench testing.
8. The sample size for the training set:
- Not Applicable. This is not an AI/software device; there is no training set.
9. How the ground truth for the training set was established:
- Not Applicable. No training set was used.
In summary: The K050168 document is a traditional medical device 510(k) submission, not for an AI/software device. Its clearance is based on non-clinical testing demonstrating substantial equivalence to pre-existing predicate devices, not on the type of clinical performance studies and ground truth establishment typically required for AI-powered diagnostics.
Ask a specific question about this device
Page 1 of 1