Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K162881
    Date Cleared
    2017-07-11

    (270 days)

    Product Code
    Regulation Number
    884.6100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kitazato OPU Needles consisting of:

    • · Kitazato OPU Needle with Connection Tube (model number Type2)
    • · Kitazato OPU Reduced Needle (model number Type3)
    • · Kitazato OPU Two-Stage Reduced Needle (model number Type4)
      are intended to obtain oocytes from ovarian follicles.
    Device Description

    The Kitazato OPU Needles are single-use sterile devices that are used for ultrasound-guided transvaginal collection of oocytes from ovarian follicles.
    All of the Kitazato OPU Needle have a similar general design and device materials. All of the Kitazato OPU Needles are composed of a 304 stainless steel needle, acrylonitrile butadiene styrene (ABS) needle hub, PVC connection tube, and a silicone stopper that is used to connect to a collection tube and vacuum pump.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Kitazato OPU Needles, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    CriteriaAcceptance CriterionReported Device Performance
    Endotoxin< 20 EU/device (per ANSI/AAMI ST72-2011)< 20 EU/device
    Mouse Embryo Assay (MEA)≥ 80% expanded blastocysts at 96 hours (1-Cell MEA)≥ 80% expanded blastocysts at 96 hours (1-Cell MEA)
    Sterilization ValidationPer ISO 11135:2014 and ISO 10993-7:2008Passed validation
    Package Integrity (Dye penetration)Per ASTM F929-15Passed testing
    Package Integrity (Seal strength)Per ASTM F88/F88M-15Passed testing
    Transport TestingPer ASTM D4169-09Passed testing
    Shelf-life (MEA)≥ 80% expanded blastocysts at 96 hours (1-Cell MEA)Met at 3 years
    Shelf-life (Endotoxin)< 20 EU/deviceMet at 3 years
    Shelf-life (Tensile testing - needle to hub)Internal requirements metMet at 3 years
    Shelf-life (Tensile testing - tubing to hub)Internal requirements metMet at 3 years
    Shelf-life (Bending elasticity of needle)Return to normal position after bending 8° from straightMet at 3 years
    Shelf-life (Folding strength of needle)No fracture when folded at 5 mm curvature radius to 90°Met at 3 years
    Shelf-life (Device appearance)No burrs, scratches, damageMet at 3 years
    Shelf-life (Dimensional assessments)Internal requirements metMet at 3 years
    Aspiration pressure testingDevices operated as intended (no deformation, blockages, damage)Operated as intended
    Biocompatibility (Cytotoxicity)Acceptable results per ISO 10993-5:2009No signs of cytotoxicity
    Biocompatibility (Intracutaneous reactivity)Acceptable results per ISO 10993-10:2010No signs of irritation reactions
    Biocompatibility (Sensitization)Acceptable results per ISO 10993-10:2010No signs of sensitization reactions

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes non-clinical performance tests. It does not specify a "test set" in the context of clinical data or human subjects. The testing involved various batches of the devices for each test. For example:

    • Mouse Embryo Assay (MEA): "1-Cell mouse embryos were incubated in extracts of the subject device..." - The specific number of embryos or repetitions is not provided.
    • Shelf-life studies: Performed on "real-time aged devices." The number of devices tested for each shelf-life parameter (MEA, endotoxin, tensile, bending, folding, appearance, dimensional) is not specified.

    The data provenance is not explicitly stated in terms of country of origin or whether it's retrospective/prospective, as these are non-clinical lab tests performed on the manufactured devices.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This report details non-clinical laboratory testing (e.g., chemical, mechanical, biological assays), not clinical studies requiring expert interpretation of medical images or data. Therefore, there is no "ground truth" established by experts in the typical sense for clinical AI performance studies.

    4. Adjudication Method for the Test Set:

    Not applicable, as this is non-clinical laboratory testing.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No. The document describes non-clinical performance testing of the device itself (e.g., material properties, sterility, embryotoxicity) and does not involve human readers or AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    Not applicable. This is a medical device (needles), not an AI algorithm. The performance is assessed based on its physical and biological characteristics, not computational output.

    7. Type of Ground Truth Used:

    The "ground truth" for these non-clinical tests is based on:

    • Quantitative measurements: e.g., endotoxin levels (< 20 EU/device), percentage of expanded blastocysts (≥ 80%), specific dimensions.
    • Qualitative observations: e.g., absence of deformation/blockages, no fracture, no burrs/scratches, no signs of cytotoxicity/sensitization/irritation.
    • Adherence to established standards: e.g., ANSI/AAMI ST72-2011, ISO 11135:2014, ISO 10993-7:2008, ASTM F929-15, ASTM F88/F88M-15, ASTM D4169-09, ISO 10993-1-2009, ISO 10993-5:2009, ISO 10993-10:2010.

    8. Sample Size for the Training Set:

    Not applicable. This document pertains to the regulatory submission for a physical medical device, not an AI model. There is no concept of a "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As there is no training set, the establishment of ground truth for it is irrelevant here.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1