K Number
K162881
Date Cleared
2017-07-11

(270 days)

Product Code
Regulation Number
884.6100
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Kitazato OPU Needles consisting of:

  • · Kitazato OPU Needle with Connection Tube (model number Type2)
  • · Kitazato OPU Reduced Needle (model number Type3)
  • · Kitazato OPU Two-Stage Reduced Needle (model number Type4)
    are intended to obtain oocytes from ovarian follicles.
Device Description

The Kitazato OPU Needles are single-use sterile devices that are used for ultrasound-guided transvaginal collection of oocytes from ovarian follicles.
All of the Kitazato OPU Needle have a similar general design and device materials. All of the Kitazato OPU Needles are composed of a 304 stainless steel needle, acrylonitrile butadiene styrene (ABS) needle hub, PVC connection tube, and a silicone stopper that is used to connect to a collection tube and vacuum pump.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Kitazato OPU Needles, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

CriteriaAcceptance CriterionReported Device Performance
Endotoxin< 20 EU/device (per ANSI/AAMI ST72-2011)< 20 EU/device
Mouse Embryo Assay (MEA)≥ 80% expanded blastocysts at 96 hours (1-Cell MEA)≥ 80% expanded blastocysts at 96 hours (1-Cell MEA)
Sterilization ValidationPer ISO 11135:2014 and ISO 10993-7:2008Passed validation
Package Integrity (Dye penetration)Per ASTM F929-15Passed testing
Package Integrity (Seal strength)Per ASTM F88/F88M-15Passed testing
Transport TestingPer ASTM D4169-09Passed testing
Shelf-life (MEA)≥ 80% expanded blastocysts at 96 hours (1-Cell MEA)Met at 3 years
Shelf-life (Endotoxin)< 20 EU/deviceMet at 3 years
Shelf-life (Tensile testing - needle to hub)Internal requirements metMet at 3 years
Shelf-life (Tensile testing - tubing to hub)Internal requirements metMet at 3 years
Shelf-life (Bending elasticity of needle)Return to normal position after bending 8° from straightMet at 3 years
Shelf-life (Folding strength of needle)No fracture when folded at 5 mm curvature radius to 90°Met at 3 years
Shelf-life (Device appearance)No burrs, scratches, damageMet at 3 years
Shelf-life (Dimensional assessments)Internal requirements metMet at 3 years
Aspiration pressure testingDevices operated as intended (no deformation, blockages, damage)Operated as intended
Biocompatibility (Cytotoxicity)Acceptable results per ISO 10993-5:2009No signs of cytotoxicity
Biocompatibility (Intracutaneous reactivity)Acceptable results per ISO 10993-10:2010No signs of irritation reactions
Biocompatibility (Sensitization)Acceptable results per ISO 10993-10:2010No signs of sensitization reactions

2. Sample Size Used for the Test Set and Data Provenance:

The document describes non-clinical performance tests. It does not specify a "test set" in the context of clinical data or human subjects. The testing involved various batches of the devices for each test. For example:

  • Mouse Embryo Assay (MEA): "1-Cell mouse embryos were incubated in extracts of the subject device..." - The specific number of embryos or repetitions is not provided.
  • Shelf-life studies: Performed on "real-time aged devices." The number of devices tested for each shelf-life parameter (MEA, endotoxin, tensile, bending, folding, appearance, dimensional) is not specified.

The data provenance is not explicitly stated in terms of country of origin or whether it's retrospective/prospective, as these are non-clinical lab tests performed on the manufactured devices.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

Not applicable. This report details non-clinical laboratory testing (e.g., chemical, mechanical, biological assays), not clinical studies requiring expert interpretation of medical images or data. Therefore, there is no "ground truth" established by experts in the typical sense for clinical AI performance studies.

4. Adjudication Method for the Test Set:

Not applicable, as this is non-clinical laboratory testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No. The document describes non-clinical performance testing of the device itself (e.g., material properties, sterility, embryotoxicity) and does not involve human readers or AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

Not applicable. This is a medical device (needles), not an AI algorithm. The performance is assessed based on its physical and biological characteristics, not computational output.

7. Type of Ground Truth Used:

The "ground truth" for these non-clinical tests is based on:

  • Quantitative measurements: e.g., endotoxin levels (< 20 EU/device), percentage of expanded blastocysts (≥ 80%), specific dimensions.
  • Qualitative observations: e.g., absence of deformation/blockages, no fracture, no burrs/scratches, no signs of cytotoxicity/sensitization/irritation.
  • Adherence to established standards: e.g., ANSI/AAMI ST72-2011, ISO 11135:2014, ISO 10993-7:2008, ASTM F929-15, ASTM F88/F88M-15, ASTM D4169-09, ISO 10993-1-2009, ISO 10993-5:2009, ISO 10993-10:2010.

8. Sample Size for the Training Set:

Not applicable. This document pertains to the regulatory submission for a physical medical device, not an AI model. There is no concept of a "training set" in this context.

9. How the Ground Truth for the Training Set Was Established:

Not applicable. As there is no training set, the establishment of ground truth for it is irrelevant here.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized design of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing line that forms a ribbon-like shape at the bottom. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 11, 2017

Kitazato Corporation % Audrey Swearingen Director, Regulatory Affairs Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, TX 78746

Re: K162881 Trade/Device Name: Kitazato OPU Needles Regulation Number: 21 CFR§ 884.6100 Regulation Name: Assisted Reproduction Needles Regulatory Class: II Product Code: MOE Dated: June 2, 2017 Received: June 2, 2017

Dear Audrey Swearingen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Joyce M. Whang -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K162881

Device Name Kitazato OPU Needles

Indications for Use (Describe) The Kitazato OPU Needles consisting of:

  • · Kitazato OPU Needle with Connection Tube (model number Type2)
  • · Kitazato OPU Reduced Needle (model number Type3)
  • · Kitazato OPU Two-Stage Reduced Needle (model number Type4)

are intended to obtain oocytes from ovarian follicles.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Kitazato OPU Needles – K162881

1. Submission Sponsor

Kitazato Corporation 1-1-8 Shibadaimon, Minato-ku Tokyo 105-0012 JAPAN Contact: Ms. Mari Yazaki Title: Quality Assurance Manager Phone: (81) 3-3434-2731 Fax: (81) 3-3434-2732 Email: yazaki@kitazato.co.jp

2. Submission Correspondent

Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327.9997 Contact: Audrey Swearingen, RAC Title: Director, Regulatory Affairs

3. Date Prepared

July 7, 2017

4. Device Identification

Trade or Proprietary Name:Kitazato OPU Needles
Common or Usual Name:Oocyte aspiration needle
Regulation Number:884.6100
Regulation Name:Assisted Reproduction Needles
Product Code:MQE
Product Code Name:Needles, Assisted Reproduction
Class of Device:Class II
Panel:Obstetrics/Gynecology

5. Legally Marketed Predicate Device(s)

KITAZATO Medical Co., Ltd., K112492, Kitazato OPU Needle with Connection Tube, Type2 and Kitazato OPU Reduced Needle, Type 3.

The predicate device has not been subject to a design-related recall.

Indication for Use Statement 6.

The Kitazato OPU Needles consisting of:

  • Kitazato OPU Needle with Connection Tube (model number Type2)
  • Kitazato OPU Reduced Needle (model number Type3)
  • Kitazato OPU Two-Stage Reduced Needle (model number Type4)

{4}------------------------------------------------

are intended to obtain oocytes from ovarian follicles.

7. Device Description

The Kitazato OPU Needles are single-use sterile devices that are used for ultrasound-guided transvaginal collection of oocytes from ovarian follicles.

All of the Kitazato OPU Needle have a similar general design and device materials. All of the Kitazato OPU Needles are composed of a 304 stainless steel needle, acrylonitrile butadiene styrene (ABS) needle hub, PVC connection tube, and a silicone stopper that is used to connect to a collection tube and vacuum pump.

The Type2 OPU Needles consist of straight aspiration needles (no tapering) with the following characteristics:

ModelColorofHubNeedleO.D.(mm)NeedleGaugeNeedleLength(mm)NeedleID(mm)WallThickness(mm)NeedleBevelAngleConnectionTubeLength(mm)ConnectionTubeO.D.(mm)ConnectionTubeID(mm)StopperSizeTubesizeCompatiblewithStopper
Type2v1/-v10White1.65163501.190.2314°800,10002.01.42.51.4 (2.5)
Type2v2/v11Lightbrown1.49173501.060.2214°800,10002.01.42.51.4 (2.5)
Type2v3/v12Pink1.25183500.860.2014°800,10002.01.22.51.2 (2.5)
Type2v4/v13Ivory1.06193500.740.1614°800,10002.01.02.51.0 (2.5)
Type2v5/v14Yellow0.90203500.640.1314°800,10002.01.22.51.2 (2.5)
Type2v6/v15Yellow0.90203000.640.1314°800,10002.01.22.51.2 (2.5)
Type2v7/v16Green0.80213000.540.1314°800,10002.01.22.51.2 (2.5)
Type2v8/v17Black0.70223000.430.1314°800,10002.01.22.51.2 (2.5)
Type2v9/v18Lightblue0.63233000.390.1214°800,10002.01.02.51.0 (2.5)

The Type3 OPU Needles consist of aspiration needles with a tapered tip over the distal 40 or 100 mm of the tip. The table below shows the features of the Type3 OPU Needles:

ModelColorofHubNeedleO.D.(mm)M: mainT:TaperedO.D.NeedleGaugeM: mainT:TaperedGaugeNeedleLength(mm)TL:TaperedLengthNeedleID(mm)M: mainT:TaperedI.D.WallThickness(mm)NeedleBevelAngleConnectionTubeLength(mm)ConnectionTubeO.D.(mm)ConnectionTubeID(mm)StopperSizeTube sizeCompatiblewithStopper
Type3-v1GreenM 1.25T 0.80M 18T 21300, 325350TL-40M-0.86T-0.540.195T-0.1314°800,10002.01.22.51.2 (2.5)
Type3-v2BlackM 1.25T 0.70M 18T 22300, 325350T-40M-0.86T-0.450.195T-0.1314°800,10002.01.22.51.2 (2.5)
Type3-v3LightBlueM 1.25T 0.63M 18T 23300, 325T-40M-0.86T-0.430.195T-0.1014°800,10002.01.22.51.2 (2.5)
Type3-v4IvoryM 1.49T 1.06M 17T 19300, 325350T-100M-1.20T-0.780.145T-0.1414°800,10002.01.42.51.4 (2.5)
Type3-v5YellowM 1.49T 0.90M 17T 20300, 325350T-100M-1.20T-0.620.145T-0.1414°800,10002.01.42.51.4 (2.5)

{5}------------------------------------------------

Type3-v6GreenM 1.25T 0.80M 18T 21300, 325T-100M-0.86T-0.540.195T-0.1314°800,10002.01.22.51.2 (2.5)
Type3-v7BlackM 1.25T 0.70M 18T 22300, 325T-100M-0.86T-0.450.195T-0.1314°800,10002.01.22.51.2 (2.5)
Type3-v8LightBlueM 1.25T 0.63M 18T 23300, 325T-100M-0.86T-0.430.195T-0.1014°800,10002.01.22.51.2 (2.5)

The Type4 OPU Needles consist of aspiration needles that taper at two locations at the distal tip of the device, with the first taper beginning 70 mm from the tip and the second starting 25 mm from the tip. The table below shows the features of the Type4 OPU Needles.

ModelColorofHubNeedleO.D.(mm)M: mainT1:1st StepTaperedT2 :2nd StepTaperedNeedleGaugeM: mainT1:1st StepTaperedT2:2nd StepTaperedNeedleLength(mm)M: mainT1:1st StepTaperedT2 :2nd StepTaperedNeedleID(mm)M: mainT2:2nd StepTaperedWallThickness(mm)NeedleBevelAngleConnectionTubeLength(mm)ConnectionTubeO.D.(mm)ConnectionTubeID(mm)StopperSizeTube sizeCompatiblewithStopper
Type4-v1GreenM-1.25T1-1.06T2-0.80M-18/T1-19,T2-21M-300T1-70T2-25M-0.98T2-0.540.13514°800,10002.01.22.51.2 (2.5)
Type4-v2BlackM-1.25T1-0.90T2-0.70M-18/T1-20T2-22M 300T1-70T2-25M-0.98T2-0.430.13514°800,10002.01.22.51.2 (2.5)
Type4-v3LightBlueM-1.25T1-0.80T2-0.63M-18/T1-21T2-23M-300T1-70T2-25M-0.98T2-0.390.13514°800,10002.01.22.51.2 (2.5)

8. Substantial Equivalence Discussion

The following table compares the indications for use for the subject Kitazato OPU Needles and the predicate device:

Subject DevicePredicate DeviceComments
K162881K112462
The Kitazato OPU Needlesconsisting of:The Kitazato OPU Needlesconsisting of:The intended use of the subjectand predicate devices is thesame.
Kitazato OPU Needle withConnection Tube (modelnumber Type2) Kitazato OPU Reduced Needle(model number Type3) Kitazato OPU Two-StageReduced Needle (modelnumber Type4) are intended to obtain oocytesfrom ovarian follicles.Kitazato OPU Needle withConnection Tube (modelnumber Type2) Kitazato OPU Reduced Needle(model number Type3) are intended to obtain oocytesfrom ovarian follicles.

The following tables compare the technological characteristics of the subject Kitazato OPU Needles to the predicate devices:

{6}------------------------------------------------

Type2 OPU Needle Technological Characteristics Comparison

ParameterSubject DeviceType2 -OPU NeedleWith ConnectionTubeK162881Predicate DeviceType2 -OPU NeedleWith ConnectionTubeK112462Comments
Lumen TypeSingleSingleSame
TaperedNeedleNoNoSame
DeviceMaterialsStainless steel (304)Stainless steel (304)Same
ABSABS
PVCPVC
SiliconeSilicone
PolypropyleneEpoxy resinPolypropyleneEpoxy resin
NeedleGauge Range16-2316-24Similar: The range of needle gauges for thesubject devices is within those of thepredicate devices.
Needle ODRange0.63-1.65 mm0.55 -1.65 mmSimilar: The range of needle ODs for thesubject devices is within the range of thepredicates.
NeedleLength Range300-350 mm200-350 mmDifferent: The predicate devices are offered inadditional sizes that are not available for thesubject devices. These differences do notraise different questions of Safety andEffectiveness (S&E) as the range of needlelengths for the subject devices is within therange of the predicates.
EchogenicMarkerYesYesSame
ConnectionTube Length800-1000 mm200-1000 mmDifferent: The predicate devices are offered inadditional sizes that are not available for thesubject devices. These differences do notraise different questions of S&E as the rangeof tubing lengths for the subject devices arewithin the range of the predicates.
ConnectionTube OD2.02.0Same
ConnectionTube FittingSilicone stopperwith attachedfemale connectorfor aspirationSilicone stopperwith attachedfemale connectorfor aspirationSame
SterilizationMethodEthylene oxide
Single-Use/Repeat-UseSingle-UseSingle-UseSame
MEA1-Cell MEA: ≥80%expandedblastocysts at 96h1-Cell MEA: ≥80%expandedblastocysts at 96hSame
Endotoxin<20 EU/device<20 EU/deviceSame

{7}------------------------------------------------

Type3 OPU Needle Technological Characteristics Comparison

ParameterSubject DeviceType3 -OPUReduced NeedleK162881Predicate DeviceType3 -OPUReduced NeedleK112462Comments
Lumen TypeSingleSingleSame
TaperedNeedleYes - singleYes - singleSame
DeviceMaterialsStainless steel (304)ABSPVCSiliconePolypropyleneEpoxy resinStainless steel (304)ABSPVCSiliconePolypropyleneEpoxy resinSame
Needle GaugeRangeMain NeedleTapered End17-1819-231821-23Different: The range of needle gauges for thesubject devices is wider than those for thepredicate devices. These differences do notraise different S&E questions as compared tothe predicate devices.
Needle ODRangeMain NeedleTapered End1.25-1.49 mm0.63-1.060.55 -1.65 mmNot provided inpredicateDifferent: The range of ODs for the predicatedevices is wider than those for the subjectdevices. These differences do not raisedifferent questions of S&E as all diameters ofthe subject devices are within the range ofODs of the predicate submission.
NeedleLength Range300-350 mm200-350 mmDifferent: The predicate devices are offeredin additional sizes that are not available forthe subject devices. These differences donot raise different questions of S&E as therange of needle lengths for the subjectdevices are within the range of thepredicates.
Taper Length40-100 mm40-60 mmDifferent: The subject devices are offeredwith a longer taper length than thepredicates. These differences do not raisedifferent questions of S&E as compared tothe predicates.

{8}------------------------------------------------

the predicate devices.
EchogenicMarkerNoNoSame
ConnectionTube Length800-1000 mm200-1000 mmDifferent: The predicate devices are offeredin additional sizes that are not available forthe subject devices. These differences donot raise different questions of S&E as therange of needle lengths for the subjectdevices are within the range of thepredicates.
ConnectionTube OD2.02.0Same
ConnectionTube FittingSilicone stopperwith attachedfemale connectorfor aspirationSilicone stopperwith attachedfemale connectorfor aspirationSame
SterilizationMethodEthylene oxideGamma radiationDifferent – Both are accepted methods forsterilization of ART devices and do not raisedifferent questions of S&E as compared tothe predicate devices.
Single-Use/Repeat-UseSingle-UseSingle-UseSame
MEA1-Cell MEA: ≥80%expandedblastocysts at 96h1-Cell MEA: ≥80%expandedblastocysts at 96hSame
Endotoxin<20 EU/device<20 EU/deviceSame

Type4 Technological Characteristics Comparison

ParameterSubject DeviceType4 -OPU Two-Stage ReducedNeedleK162881Predicate DeviceType3 -OPUReduced NeedleK112462Comments
Lumen TypeSingleSingleSame
TaperedNeedleYes - doubleYes - singleDifferent: The subject devices taper at twopoints (25 and 70 mm from the tip), whilethe predicates only have a single taperoccurring between 40 and 60 mm from thetip. Dual vs. single tapering does not raisedifferent questions of S&E.
DeviceMaterialsStainless steel (304)Stainless steel (304)Same
ABSABS
PVCPVC
SiliconeSilicone
PolypropylenePolypropylene
Epoxy resinEpoxy resin
Needle Gauge RangeSame
Main NeedleTapered End1821-231821-23
Needle OD RangeDifferent: The range of ODs for the predicatedevices is wider than the ODs for the subjectdevices. These differences do not raisedifferent questions of S&E as all diameters ofthe devices in the subject submission arewithin the range of ODs of the predicates.
Main NeedleTapered End1.25 mm0.63-0.80 mm0.55 -1.65 mmNot in priorsubmission
NeedleLength Range300200-350 mmDifferent: The predicate devices are offeredin additional sizes that are not available forthe subject devices. These differences donot raise different questions of S&E as therange of needle lengths for the subjectdevices are within the range of thepredicates.
EchogenicMarkerNoNoSame
ConnectionTube Length800-1000 mm200-1000 mmDifferent: The predicate devices are offeredin additional sizes that are not available forthe subject devices. These differences donot raise different questions of S&E as therange of needle lengths for the subjectdevices are within the range of thepredicates.
ConnectionTube OD2.02.0Same
ConnectionTube FittingSilicone stopperwith attachedfemale connectorfor aspirationSilicone stopperwith attachedfemale connectorfor aspirationSame
Vacuum LineProvidedNoNoSame
SterilizationMethodEthylene oxideGamma radiationDifferent – Both are accepted methods forsterilization of ART devices and do not raisedifferent questions of S&E as compared tothe predicate devices.
Single-Use/Repeat-UseSingle-UseSingle-UseSame
MEA1-Cell MEA: ≥80%expandedblastocysts at 96h1-Cell MEA: ≥80%expandedblastocysts at 96hSame
Endotoxin<20 EU/device<20 EU/deviceSame

{9}------------------------------------------------

{10}------------------------------------------------

As noted in the tables above, there are many similarities between the subject and predicate devices (e.g., materials, designs, etc.); however, differences do exist. The differences in technology noted in the tables above, do not raise different questions of safety or effectiveness as compared to the predicate device.

9. Non-Clinical Performance Data

As part of demonstrating substantial equivalence to the following non-clinical performance tests were conducted. The Kitazato OPU Needles passed all of the testing in accordance with internal requirements and applicable standards to support substantial equivalence of the subject device:

  • Endotoxin testing per ANSI/AAMI ST72-2011: <20 EU/device .
  • Mouse Embryo Assay (MEA): 1-Cell mouse embryos were incubated in extracts of the subject device at 37°C in an atmosphere containing 5% CO₂. The percent of embryos developed to the expanded blastocyst stage at 96 hours were assessed in comparison with the control group. The acceptance criterion was 1-Cell MEA: ≥80% expanded to blastocyst at 96 hours.
  • Sterilization validation per ISO 11135:2014 and ISO 10993-7:2008 ●
  • Package integrity testing following real-time aging:
    • o Dye penetration testing per ASTM F929-15
      • o Seal strength testing per ASTM F88/F88M-15
  • Transport testing per ASTM D4169-09
  • Shelf-life studies (real-time aged devices) were conducted to ensure that the following product ● specifications were met to support a three-year shelf-life:
    • O MEA
    • O Endotoxin
    • Tensile testing (needle to hub and tubing to hub) O
    • O Bending elasticity of needle – Return to normal position after bending 8° from straight
    • Folding strength of needle No fracture of needle when folded at a 5 mm curvature radius O to 90°
    • o Device appearance (burrs, scratches, damage)
    • Dimensional assessments O
  • . Aspiration pressure testing - Testing involved aspirating water through needles at the recommended aspiration pressures included in device labeling to ensure that devices operated as intended (i.e., no signs of deformation, blockages or damage)
  • Biocompatibility testing was conducted according to ISO 10993-1- 2009. Testing included cytotoxicity (ISO 10993-5:2009), intracutaneous reactivity (ISO 10993-10:2010), and sensitization (ISO 10993-10:2010). The test articles assessed provided acceptable results as no signs of cytotoxicity, sensitization or irritation reactions were noted in testing.

{11}------------------------------------------------

10. Conclusion

The results of the testing described above demonstrate that the Kitazato OPU Needles are as safe and effective as the predicate device and supports a determination of substantial equivalence.

§ 884.6100 Assisted reproduction needles.

(a)
Identification. Assisted reproduction needles are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to obtain gametes from the body or introduce gametes, zygote(s), preembryo(s) and/or embryo(s) into the body. This generic type of device may include a single or double lumen needle and component parts, including needle guides, such as those used with ultrasound.(b)
Classification. Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).