Search Results
Found 1 results
510(k) Data Aggregation
(81 days)
KIMBERLY-CLARK STERLING NITRILE POWDER-FREE EXAM GLOVE (STERILE)
The Kimberly-Clark* STERLING Nitrile Powder-Free Exam Glove, Sterile, is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Based upon 21CFR§880.6250 "Patient examination glove" Use: A patient examination glove is a medical device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
Light gray nitrile, chlorinated, powder-free, sterile, textured fingertip, ambidextrous patient examination glove
The Kimberly-Clark* Sterile STERLING* Nitrile Powder-Free Exam Glove (K081027) is a patient examination glove. The following information details its acceptance criteria and the study that proves the device meets these criteria:
1. A table of acceptance criteria and the reported device performance
The acceptance criteria for this device are based on compliance with several ASTM and ISO standards for properties typical of patient examination gloves. The device's performance is reported as meeting these requirements based on non-clinical tests.
Acceptance Criteria | Standard | Reported Device Performance |
---|---|---|
Dimensions | ASTM D 6319-00a | Meets the standard |
Physical Properties | ASTM D 6319-00a | Meets the standard |
Freedom from pinholes | ASTM D 6319-00a | Meets the standard AQL (Acceptable Quality Level) of 2.5 for freedom from pinholes, as defined in ASTM D 5151-06. |
Powder Free | ASTM D 5151-06 | Meets the standard. The device is distinctly labeled as "Powder-Free." |
Sterility | ASTM D 6124-06; ISO 11137-2 | Meets the sterility requirements. The device is specified as "sterile" and sterilized by gamma irradiation. |
ISO Skin Irritation Study | ISO 10993, Part 10 | Meets the standard. No significant skin irritation observed. |
Murine Local Lymph Node Assay | ISO 10993, Part 10 | Meets the standard. No sensitization potential observed. |
ISO Systemic Toxicity Study | ISO 10993, Part 11 | Meets the standard. No systemic toxicity observed. |
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample sizes for each non-clinical test. The data provenance is derived from internal non-clinical laboratory and animal-based biocompatibility tests conducted by Kimberly-Clark Corporation to confirm compliance with established standards. The data is retrospective in the sense that it pertains to the product being submitted for 510(k) clearance, reflecting the outcome of these tests.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. The ground truth for this device is established through compliance with recognized industry standards (ASTM and ISO) for physical properties, sterility, and biocompatibility. These standards define the objective measures for performance, rather than requiring expert consensus for a "ground truth" in the diagnostic sense.
4. Adjudication method for the test set
Not applicable. As the evaluation relies on objective measurements against established standards, an adjudication method in the context of expert review is not required. The results of the non-clinical tests are directly compared to the specified limits of the standards.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a sterile patient examination glove and does not involve AI assistance or human-reader interpretations.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This device is a physical medical device and does not involve an algorithm.
7. The type of ground truth used
The ground truth is established through objective measurements compared against published national and international standards (ASTM, ISO) for physical properties, sterility, and biocompatibility. For example, for "Freedom from pinholes," the ground truth is defined by the AQL (Acceptable Quality Level) of 2.5 specified in ASTM D 5151-06.
8. The sample size for the training set
Not applicable. This is a physical medical device, not a machine learning model, so there is no "training set."
9. How the ground truth for the training set was established
Not applicable. As there is no training set, this question is not relevant.
Ask a specific question about this device
Page 1 of 1