Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K242170
    Date Cleared
    2025-04-18

    (268 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    K-ASSAY® CRP (Ver.2) is intended to be used for the quantitative determination of C-reactive protein (CRP) in human serum and plasma (potassium-EDTA or lithium-heparin) by immunoturbidimetric assay. Measurement of CRP aids in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases. FOR IN VITRO DIAGNOSTIC USE.

    Device Description

    The K-ASSAY® CRP (Ver.2) assay quantifies C-reactive protein based on immunoturbidimetric assay. The reagent uses latex combined with goat polyclonal antibody specific to human CRP. By adsorbing CRP in the sample to the surface of the latex particles and reacting it with the anti-CRP antibody, specific aggregation corresponding to the CRP concentration occurs. Since the absorbance of the reaction changes in proportion to the amount of aggregation, the concentration of CRP in the sample is determined based on the calibration curve prepared using a standard of known CRP concentrations. The K-ASSAY® CRP (Ver.2) assay can be run using a chemistry analyzer. 6 levels of calibrators from the K-ASSAY® CRP Calibrator (Ver.2) calibrators are used for quantifying the levels of CRP present in the patient's sample.

    AI/ML Overview

    This document describes the FDA 510(k) clearance for the K-ASSAY CRP (Ver.2) IVD device. This device is an in-vitro diagnostic test system, which means it analyzes biological samples in a lab setting rather than directly interacting with a patient.

    Therefore, the concepts of "human readers," "AI assistance," "effect size," "multi reader multi case (MRMC) comparative effectiveness study," and "standalone (i.e. algorithm only without human-in-the loop performance)" are not applicable in this context. These terms are typically used for medical imaging AI/ML devices where human interpretation is involved.

    For this in-vitro diagnostic device, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are related to its analytical performance characteristics when compared to a predicate device, and the reported device performance refers to the results of these analytical studies.

    Here's the breakdown of the information provided within the scope of this IVD device:

    1. Table of Acceptance Criteria and Reported Device Performance

    For an IVD device like the K-ASSAY® CRP (Ver.2), the acceptance criteria are generally established by demonstrating equivalent or superior analytical performance compared to a legally marketed predicate device, and meeting established CLSI guidelines for accuracy, precision, linearity, and interference. The "reported device performance" refers to the actual study results for these characteristics.

    Acceptance Criteria (Implicit from predicate comparison and CLSI guidelines)Reported Device Performance (K-ASSAY® CRP (Ver.2))
    Intended Use Equivalence: Quantitative determination of CRP in human serum and plasma for detection/evaluation of infection, tissue injury, inflammatory disorders.K-ASSAY® CRP (Ver.2) is intended for the quantitative determination of CRP in human serum and plasma (potassium-EDTA or lithium-heparin) by immunoturbidimetric assay. Measurement of CRP aids in the detection and evaluation of infection, tissue injury, inflammatory disorders and associated diseases.
    Methodology Equivalence: Latex-enhanced immunoturbidimetric assay.Latex-enhanced (immuno)turbidimetric assay.
    Calibration Equivalence/Validation: Appropriate calibrator levels for the intended range.6 levels of calibrators (0.0, 10.0, 50.0, 150.0, 300.0 and 400.0 mg/L).
    Assay Range (Comparable to predicate): Demonstrates clinically relevant measuring range.Claimed Assay Range: 5.0 - 400.0 mg/L. (Predicate: 0.2 – 480 mg/L)
    Precision: Demonstrate acceptable repeatability, within-run, between-run, within-day, between-day, within-laboratory, between-site, and reproducibility CVs as per CLSI EP05-A3 guidelines.Single Site Precision (Combined Data From 3 Lots):
    • Within-Run CV: 0.8% to 2.2%
    • Between-Run CV: 0.0% to 0.6%
    • Within-Lot CV: 0.9% to 1.9%
    • Between-Lot CV: 0.0% to 1.2%
    • Total CV: 0.9% to 2.3%
      Multisite Precision (Combined Data From 3 Analyzers):
    • Repeatability CV: 0.5% to 1.5%
    • Between-Run CV: 0.0% to 1.1%
    • Between-Day CV: 0.6% to 1.8%
    • Between-Site CV: 0.9% to 1.8%
    • Reproducibility CV: 1.1% to 2.5% |
      | Interference: No significant interference from common endogenous and exogenous substances (recovery within 10% of initial value). | Endogenous Substances (up to listed concentrations, no significant interference): Bilirubin C (40 mg/dL), Bilirubin F (40 mg/dL), Cholesterol (300 mg/dL), Hemoglobin (1,000 mg/dL), Intralipid (500 mg/dL), Rheumatoid Factor (1,000 IU/mL), Triglycerides (1,000 mg/dL).
      Exogenous Substances (up to listed concentrations, no significant interference): Acetaminophen (1.5 mM), Amoxicillin (400 µmol/L), Aspirin (3.6 mM), Cephalexin (360 µmol/L), Fluconazole (480 µmol/L), Ibuprofen (2.5 mg/dL), Methotrexate (1,400 µmol/L), Prednisolone (2 µmol/L), Vitamin C (500 mg/L). |
      | Method Comparison: Strong correlation and minimal bias against the predicate device. | Regression Equation: y = 1.005x - 0.002, r = 0.999 (n = 175 clinical native serum samples), where y = K-ASSAY® CRP (Ver.2), x = predicate device. |
      | Linearity: Demonstrates linearity across the claimed assay range. | Regression Equation: y = 0.9709x - 1.095, r = 0.999 (tested range: 4.6 - 441.2 mg/L). |
      | Limit of Quantitation (LoQ): Clinically acceptable LoQ (within-laboratory precision ≤ 20% CV). | Reported LoQ: 1.0 mg/L (highest observed across 3 reagent lots). Claimed LoQ: 5.0 mg/L. |
      | Matrix Comparison: No significant difference in results across different sample matrices (serum vs. plasma). | K2-EDTA Plasma vs Serum: y = 1.007x - 0.141, r = 0.999
      Li-Heparin Plasma vs Serum: y = 0.972x + 0.074, r = 0.999 |
      | Expected Values (Verification): Distribution in normal population consistent with clinical literature. | 168 normal U.S. serum samples tested; 4 out of 168 samples (>5.0 mg/L, 2.4%) consistent with literature (≤ 5 mg/L indicates apparently healthy). |

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Method Comparison: n = 175 clinical native serum samples.
    • Linearity: The number of samples/dilutions used is not explicitly stated, but the tested range was 4.6 - 441.2 mg/L.
    • Limit of Quantitation (LoQ): Not specified in terms of sample size for the LoQ determination itself (typically involves low-concentration samples measured multiple times).
    • Matrix Comparison: 42 donor samples (each collected into 3 different tubes: serum, K2-EDTA plasma, Li-Heparin plasma).
    • Precision (Single-site): For each of the 7 samples/controls, N=240 individual measurements (2 runs/day x 2 replicates/run x 20 days x 3 lots).
    • Precision (Multisite): For each of the 7 samples/controls, N=75 individual measurements (1 run/day x 5 replicates/run x 5 days x 3 analyzers).
    • Interference: The number of unique samples tested for interference is not explicitly stated. Typically, a few samples (e.g., low, medium, high concentration) are spiked with interferents and compared to unspiked controls.
    • Expected Values: 168 normal serum samples.
    • Data Provenance: The document states for "Expected Values" that 168 normal serum samples were "taken from healthy individuals in the U.S." This indicates a U.S. origin for at least this specific study. For other studies (Method Comparison, Precision, Linearity, Interference, Matrix Comparison), the specific country of origin is not mentioned. All studies are retrospectively analyzed in the sense that they were completed performance validation studies submitted to FDA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    For an in-vitro diagnostic device like this CRP assay, the "ground truth" is established through highly accurate and precise reference methods or established predicate devices, adhering to rigorous analytical chemistry best practices and guidelines (e.g., CLSI standards). There isn't a panel of human "experts" like radiologists interpreting images. The "ground truth" is quantitative and objective, derived from reference measurements.

    In this case:

    • Method Comparison: The predicate device (K-ASSAY® CRP (3), K023828) serves as the comparator for method comparison, which represents the established "ground truth" for clinical samples.
    • Linearity, LoQ, Precision, Interference, Matrix Comparison: The "ground truth" for these analytical performance studies is established by rigorous laboratory protocols, highly calibrated equipment, reference materials, and adherence to CLSI guidelines. The performance is assessed against predefined statistical criteria rather than expert consensus on individual cases.
    • Expected Values: The ground truth comes from established clinical literature and verified normal ranges based on studies of healthy populations.

    4. Adjudication Method for the Test Set

    Not applicable for an IVD device. Adjudication methods (e.g., 2+1, 3+1) are used in studies where subjective human interpretation (e.g., image reading) requires consensus building for ground truth establishment. For quantitative IVD assays, the results are numerical and objectively measured; therefore, there's no need for adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. As explained above, this is an in-vitro diagnostic device, not an imaging AI/ML device involving human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in essence, all the analytical performance studies (Method Comparison, Linearity, LoQ, Precision, Interference, Matrix Comparison) are "standalone" in nature for an IVD. The device's performance is evaluated based solely on its ability to accurately and precisely measure CRP concentrations in samples, without any human interpretation of the measurement itself or the involvement of an "algorithm" in the AI/ML sense. The device directly processes the sample and outputs a quantitative result.

    7. The Type of Ground Truth Used

    The ground truth for this IVD device's performance studies is primarily based on:

    • Reference Method/Predicate Device Comparison: For the method comparison study, the predicate device serves as the reference against which the new device's measurements are compared.
    • Reference Materials and Calibrators: For linearity, LoQ, and precision studies, the ground truth for concentration values is established using certified reference materials and meticulously prepared calibrators with known concentrations.
    • Spiked Samples: For interference studies, known amounts of interfering substances are added to samples, and the true CRP concentration (before interference) serves as the baseline ground truth.
    • Clinical Literature/Established Norms: For "Expected Values," the ground truth is derived from widely accepted clinical ranges and population studies cited in the literature.

    8. The Sample Size for the Training Set

    This document describes a 510(k) submission for a traditional IVD device, not an AI/ML-based device. Therefore, there is no "training set" in the context of machine learning. The device's performance is based on its chemical and physical principles (latex-enhanced immunoturbidimetric assay) and validated through the analytical studies detailed.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no "training set" for this type of IVD device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K992311
    Date Cleared
    1999-09-13

    (66 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The K-ASSAY CRP (I) Assay is intended to be used for the quantitative determination of C-Reactive Protein (CRP) in serum by a latex enhanced immunoturbidimetric method. Measurement of CRP is useful as an aid in the detection and evaluation of inflammatory conditions and tissue injury. CRP is an acute phase reactant protein that is elevated in the presence of inflammatory processes, infection, and tissue injury. Measurement of CRP is useful as an aid in the detection and evaluation of inflammatory conditions and tissue injury.

    The K-ASSAY CRP Multi-Calibrator A is intended to be used for the calibration of the K-ASSAY CRP (I) Assay.

    The K-ASSAY CRP Multi-Calibrator C is intended to be used for the calibration of the K-ASSAY CRP (I) Assay.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA regarding the K-ASSAY CRP (I), K-ASSAY CRP Multi-Calibrator A, and K-ASSAY CRP Multi-Calibrator C. It states that the devices are substantially equivalent to legally marketed predicate devices. The "Indications for Use Statement" describes the intended use of the K-ASSAY CRP (I) Assay for quantitative determination of C-Reactive Protein (CRP) in human serum, with the calibrators intended for calibration of the assay.

    However, the document does not contain the detailed acceptance criteria and study information requested in the prompt. Specifically, it lacks:

    1. A table of acceptance criteria and reported device performance.
    2. Sample size used for the test set and data provenance.
    3. Number of experts and their qualifications for ground truth.
    4. Adjudication method.
    5. Information on Multi-Reader Multi-Case (MRMC) comparative effectiveness study.
    6. Information on standalone algorithm performance.
    7. Type of ground truth used.
    8. Sample size for the training set.
    9. How ground truth for the training set was established.

    The 510(k) letter confirms clearance based on substantial equivalence to a predicate device, which usually implies that the new device performs similarly to an already approved device. The specifics of the performance studies that supported this equivalence are typically found within the full 510(k) submission, not in the final clearance letter itself.

    Ask a Question

    Ask a specific question about this device

    K Number
    K992312
    Date Cleared
    1999-09-13

    (66 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The K-ASSAY CRP (2) Assay is intended to be used for the quantitative determination of CRP in serum by latex particle enhanced immunoturbidimetric assay. Measurement of C-Reactive Protein aids in evaluation of the amount of injury to body tissues. The K-ASSAY CRP Multi-Calibrator A is intended to be used for the calibration of the K-ASSAY CRP (2) Assay. The K-ASSAY CRP Multi-Calibrator B is intended to be used for the calibration of the K-ASSAY CRP (2) Assay.

    Device Description

    Not Found

    AI/ML Overview

    This document is an FDA 510(k) clearance letter and an accompanying "Indications for Use Statement" for the K-ASSAY CRP (2) Assay and related calibrators. It explicitly states that the device is "substantially equivalent" to legally marketed predicate devices.

    However, the provided text does not contain any information about acceptance criteria, specific device performance data from a study, sample sizes, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, or details about the training set used for the device.

    The document grants market clearance based on a determination of substantial equivalence to existing devices, meaning it doesn't require the detailed performance data and study descriptions typically found in a new device's clinical validation report.

    Therefore, I cannot provide the requested table and study details based on the input text. The text focuses solely on regulatory approval and states the intended use of the device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991224
    Device Name
    K-ASSAY CRP
    Date Cleared
    1999-05-14

    (32 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The K-ASSAY CRP Assay is intended to be used for the quantitative determination of CRP in serum by immunoturbidimetric assay. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues. FOR IN VITRO DIAGNOSTIC USE.

    Device Description

    Not Found

    AI/ML Overview

    I apologize, but the provided text is a 510(k) clearance letter from the FDA for a device called "K-ASSAY CRP." This document primarily addresses the regulatory approval process and states the device's intended use.

    It does not contain any information regarding:

    • Acceptance criteria and reported device performance in a table.
    • Sample sizes, data provenance, or details about the test set.
    • Number or qualifications of experts used for ground truth.
    • Adjudication methods.
    • Multi-reader multi-case comparative effectiveness studies or effect sizes.
    • Standalone algorithm performance.
    • Type of ground truth used.
    • Training set sample size or how its ground truth was established.

    Therefore, I cannot fulfill your request for this specific information based on the provided text. This document is a regulatory approval, not a scientific study report describing the validation of the device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991210
    Date Cleared
    1999-05-11

    (32 days)

    Product Code
    Regulation Number
    866.5270
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The K-ASSAY CRP Calibrator Set is intended to be used for the calibration of the K-ASSAY CRP immunoturbidimetric assay for quantitating CRP in serum samples.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter for the K-ASSAY CRP Calibrator Set. This type of document does not contain the detailed information requested regarding acceptance criteria, study design, or performance metrics.

    FDA 510(k) clearances primarily focus on demonstrating substantial equivalence to a legally marketed predicate device, rather than requiring extensive clinical trial data or performance studies as might be seen for devices seeking a de novo classification or PMA.

    Therefore, I cannot provide the requested information from the given text.

    In general, for in vitro diagnostic (IVD) calibrator sets like the K-ASSAY CRP Calibrator Set, the acceptance criteria would typically relate to:

    • Accuracy: How well the calibrators provide known concentrations of the analyte (CRP)
    • Precision/Reproducibility: Consistency of the calibrator values across different lots, users, and instruments.
    • Stability: How long the calibrators maintain their assigned values under specific storage conditions.
    • Interference: Lack of interference from common substances found in serum.
    • Traceability: Often, calibrators are traceable to international reference materials or certified reference materials.

    If a study were conducted, it would likely involve:

    • Measuring calibrator values using a validated reference method or the intended immunoturbidimetric assay.
    • Comparing results to target values or established reference ranges.
    • Testing different lots of calibrators and perhaps different instruments or operators.

    However, none of this specific information is present in the provided FDA clearance letter.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1