Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K152705
    Date Cleared
    2016-03-17

    (178 days)

    Product Code
    Regulation Number
    880.6250
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. Nitrile films form a barrier to body fluids and bloodborne Pathogens. The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for Jiayuan Nitrile Powder Free Patient Examination Gloves, Blue Color. It describes the device, its intended use, and a study to demonstrate its substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study, structured as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicsStandard / Acceptance CriteriaDevice Performance
    DimensionASTM D 6319-10.Meets
    Physical PropertiesASTM D 6319-10.Meets
    Freedom from pinholes21 CFR 800.20 (Holes at Inspection Level I, AQL 2.5)Meets
    Powder ResidualASTM D 6319-10 and D6124-06 (Reapproved 2011)Meets (