Search Results
Found 1 results
510(k) Data Aggregation
(57 days)
JADE PLUS PTA Balloon Dilatation Catheter
The JADE PLUS PTA Balloon Dilatation Catheter is indicated for Percutaneous Transluminal Angioplasty in the peripheral vasculature, including iliac, femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for postdilation of balloon expandable and self-expanding stents in the peripheral vasculature.
The JADE PLUS Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheters are designed for peripheral indications with Over-The-Wire (OTW) structure. The overthe-wire design permits the use of standard 0.014 inch, 0.018 inch and 0.035 inch guidewires respectively.
This document is a 510(k) summary for the JADE PLUS PTA Balloon Dilatation Catheter. It does not describe an AI/ML powered device, therefore the requested information for acceptance criteria and studies are not applicable in the context of an AI/ML device.
The document discusses the substantial equivalence of the JADE PLUS PTA Balloon Dilatation Catheter to a predicate device, based on in vitro performance tests. The acceptance criteria and reported device performance mentioned relate to physical properties and functionality of the catheter, not to an AI algorithm's performance or accuracy.
The relevant information from the document is related to the device's physical performance testing:
Acceptance Criteria and Reported Device Performance:
The document states that "The test results met all acceptance criteria, which are the same or similar to the predicate device and ensure that the JADE PLUS PTA Balloon Dilatation Catheter design and construction are suitable for their intended use."
The specific performance tests conducted and the general nature of their acceptance criteria are listed:
Acceptance Criteria Category/Test | Reported Device Performance |
---|---|
Mechanical Performance: | Met all acceptance criteria |
Balloon Rated Burst Pressure | Met acceptance criteria (same or similar to predicate) |
Balloon Fatigue | Met acceptance criteria (same or similar to predicate) |
Balloon Compliance | Met acceptance criteria (same or similar to predicate) |
Balloon Rated Burst Pressure (in-stent) | Met acceptance criteria (same or similar to predicate) |
Balloon Fatigue (in-stent) | Met acceptance criteria (same or similar to predicate) |
Flexibility and Kink | Met acceptance criteria (same or similar to predicate) |
Shaft burst | Met acceptance criteria (same or similar to predicate) |
Torque Strength | Met acceptance criteria (same or similar to predicate) |
Catheter Bond Strength | Met acceptance criteria (same or similar to predicate) |
Tip Pull Strength | Met acceptance criteria (same or similar to predicate) |
Balloon Preparation, Deployment, and Retraction (Simulated Use) | Met acceptance criteria (same or similar to predicate) |
Balloon Inflation and Deflation Time | Met acceptance criteria (same or similar to predicate) |
Visual and Material Properties: | Met all acceptance criteria |
Visual Inspection | Met acceptance criteria (same or similar to predicate) |
Marker Band Radiopacity | Met acceptance criteria (same or similar to predicate) |
Dimensional Verification | Met acceptance criteria (same or similar to predicate) |
Particulate Evaluation | Met acceptance criteria (same or similar to predicate) |
Coating Integrity | Met acceptance criteria (same or similar to predicate) |
Other: | Met all acceptance criteria |
Packaging and sterilization validation | Met acceptance criteria (same or similar to predicate) |
Shelf Life | Met acceptance criteria (same or similar to predicate) |
Biocompatibility | Leveraged from predicate device (Jade OTW series) and determined acceptable |
The study that proves the device meets acceptance criteria is an in vitro performance test study conducted on the subject device in accordance with FDA guidance "Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters" issued on April 14, 2023.
Given that this is a 510(k) summary for a physical medical device (balloon dilatation catheter) and not an AI/ML powered device, the following points are not applicable and therefore cannot be extracted from the provided text:
- Sample size used for the test set and the data provenance.
- Number of experts used to establish the ground truth for the test set and their qualifications.
- Adjudication method.
- If a multi reader multi case (MRMC) comparative effectiveness study was done.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done.
- The type of ground truth used.
- The sample size for the training set.
- How the ground truth for the training set was established.
Ask a specific question about this device
Page 1 of 1