510(k) Data Aggregation

    K Number
    K232473
    Date Cleared
    2024-02-09

    (177 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Infrared Thermometer is intended for the intermittent measurement of body temperature from the forehead on people of all ages. It can be used by consumers in the household environment and by health care providers.

    Device Description

    Infrared Thermometer is mainly composed of infrared probe components, main circuit board components, LCD display components and housing components. The Infrared Thermometer is intended for the intermittent measurement of body temperature from the forehead on people of all ages. Infants and children cannot operate the thermometer; it is recommended that adults take the measurement. The Infrared Thermometer can be used by consumers in the household environment and by healthcare providers. The operator should be an adult with relevant experience. The operator can replace the battery, transmit data, and measure temperature.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study details for the Infrared Thermometer Model YJ600, based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the Infrared Thermometer Model YJ600 are primarily defined by the standards ISO 80601-2-56 and ASTM E1965-98. The 510(k) summary directly states that the device "meets the requirement of the standard" for these and other tests. The most critical performance criterion for a thermometer is its accuracy.

    Acceptance Criteria (from ISO 80601-2-56 and ASTM E1965-98)Reported Device Performance (Model YJ600)
    Maximum allowable error in temperature measurement$\pm 0.3°C$ within the temperature display range of $22.0°C$ to $42.0°C$ (Pass)
    Clinical repeatability for each age groupIn the range of plus or minus 0.3°C (meets the clinical repeatability requirement ±0.3°C)
    Resolution of the indication unit$0.1°C$ ($0.1°F$) (Pass)
    Temperature display range$22.0°C~42.9°C$ (Pass)
    Appearance and structureCorrect shape, smooth/clean surface, no scratches, sharp edges, burrs, or deformation (Pass)
    Text and symbols clarityClear, accurate, and firm (Pass)
    Control devices flexibility and reliabilityFlexible and reliable, fasteners not loose (Pass)
    Thermometer probe smoothnessSmooth with no burrs on edges (Pass)
    Function key markingClearly marked and indicated (Pass)
    Anti-drop test (1m to hard surface, 3 postures)Requirements met (Pass)
    Display reading value height>4mm (Pass)
    Prompt function for out-of-range temperatures"LO" below 22.0°C, "HI" above 42.9°C with "Di di" sound twice (Pass)
    Low voltage promptIcon flashes when battery