Search Filters

Search Results

Found 47 results

510(k) Data Aggregation

    K Number
    K250878
    Date Cleared
    2025-05-22

    (59 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    YUWELL® Infrared Ear Thermometer (YHT100); YUWELL® Infrared Ear Thermometer (YHT107)

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Infrared Ear Thermometer is indicated for intermittent measurement of human body temperature from the ear canal. The device can be used by people of all ages except preterm babies or babies who are small for gestational age. The thermometer is intended for use in professional settings and the home environment. It is not for emergency clinical conditions.

    Device Description

    YUWELL® Infrared Ear Thermometer is designed to measure human body temperature. This hand-held, no-contact device is battery-powered and utilizes infrared energy emitted in the subject's tympanic membrane to determine human body temperature. The thermometer provides temperature readings in just 2 seconds. It is suitable for individuals of all ages except preterm babies or babies who are small for gestational age. The Infrared Ear Thermometer YHT107 can transmit the temperature readings via Bluetooth, whereas the YHT100 does not have data transmission capability.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and study information for the YUWELL® Infrared Ear Thermometer, extracted from the provided FDA 510(k) clearance letter:


    YUWELL® Infrared Ear Thermometer (YHT100; YHT107) Acceptance Criteria and Study Details

    The YUWELL® Infrared Ear Thermometer (YHT100; YHT107) is indicated for intermittent measurement of human body temperature from the ear canal in professional and home environments, for people of all ages except preterm babies or babies who are small for gestational age.

    1. Table of Acceptance Criteria and Reported Device Performance

    The core performance acceptance criteria are related to temperature accuracy, as specified by the standards ISO 80601-2-56 and ASTM E1965.

    CharacteristicAcceptance Criteria (Standard Specification)Reported Device Performance (Subject Device)
    AccuracyFor 35.0℃–42.0℃ (95.0℉–107.6℉): ±0.2℃ (±0.4℉)Achieved: ±0.2℃ (±0.4℉) for 35.0℃–42.0℃ (95.0℉–107.6℉)
    For 34.0℃–34.9℃ (93.2℉–94.8℉) & 42.1℃–42.2℃ (107.8℉–108.0℉): ±0.3℃ (±0.5℉)Achieved: ±0.3℃ (±0.5℉) for 34.0℃–34.9℃ (93.2℉–94.8℉) and 42.1℃–42.2℃ (107.8℉–108.0℉)
    Temperature RangeConforms with ASTM E1965 (Predicate range: 20℃-42.2℃)34.0-42.2℃ (93.2-108.0℉)
    Display Resolution0.1℃/0.1℉ (Matches predicate)0.1℃/0.1℉
    Response TimeNot explicitly stated as a strict acceptance criterion with quantitative limits, but differences were evaluated. (Predicate: 2-3s)
    Ask a Question

    Ask a specific question about this device

    K Number
    K212598
    Date Cleared
    2022-01-28

    (165 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    iHealth infrared Ear thermometer PT5

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Infrared Ear thermometer is intended for the intermittent measurement of body temperature from the ear canal on people of all ages except for babies under 3 months. It is suitable for home use and healthcare facilities use.

    Device Description

    The iHealth PT5 Infrared Ear Thermometer is a hand-held, reusable, battery operated device, which can measure human body temperature on one's ear canal. Its operation is based on measuring the natural thermal radiation from the ear canal with a built-in correction algorithm to compensate the influence of ambient temperature using a heated tip. Put the probe of the thermometer into a patient's ear canal, after a self-check, pressing the activation button to start the measurement of proposed infrared radiation. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The final measured temperature will be appeared on a LED display. It is recommended that adult take the measurement instead when infants and children cannot use the ear thermometer themselves.

    AI/ML Overview

    The iHealth PT5 Infrared Ear Thermometer measures body temperature from the ear canal. The device's performance was evaluated through non-clinical and clinical tests to demonstrate substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    ItemAcceptance Criteria (from ASTM E1965-98 & ISO 80601-2-56)Reported Device Performance (iHealth PT5)
    Accuracy for body temperature measurement±0.2°C (0.4°F) within 35.5°C42°C (95.9-107.6°F) and ±0.3°C (0.5°F) for other ranges based on predicate's accuracy. More stringent requirements for subject device: ±0.4°F (0.2°C) within 93.2107.6°F (34~42°C), ±0.5°F (0.3°C) for other range.Meets ASTM E1965-98 and ISO 80601-2-56 requirements. Specifically: ±0.4°F (0.2°C) within 93.2107.6°F (3442°C), ±0.5°F (0.3°C) for other range.
    Repeatability5 years)

    The document does not explicitly state the country of origin for the clinical data or whether it was retrospective or prospective. However, based on the testing being conducted to international standards (ASTM and ISO), it is likely to be a prospective clinical study specifically designed for regulatory submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not explicitly provided in the document. The document states "Clinical accuracy validation test report included temperature readings," implying that a reference standard temperature measurement (often oral or rectal temperature by trained clinicians) was used as the ground truth, but the number or qualifications of experts involved in establishing this ground truth are not detailed.

    4. Adjudication Method for the Test Set:

    This information is not explicitly provided in the document.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

    This type of study is not applicable here as the device is an infrared ear thermometer, which measures a physical parameter (temperature) and does not involve human readers interpreting images or data to make a diagnosis that would be assisted by AI. The device operates in a standalone manner.

    6. Standalone Performance:

    Yes, a standalone performance study was done. The "Clinical Accuracy Validation Test" performed according to ASTM E1965-98 directly evaluated the iHealth PT5 Infrared Ear Thermometer's performance in measuring body temperature in a clinical setting without human interpretation or intervention in the temperature measurement itself. The non-clinical tests also evaluated the device's technical standalone performance.

    7. Type of Ground Truth Used:

    Based on the nature of a clinical accuracy validation test for thermometers, the ground truth would most likely be established by a reference standard temperature measurement (e.g., core body temperature measurements from oral or rectal thermometers) taken by trained medical personnel. This is implied by the reference to ASTM E1965-98, which outlines methods for clinical performance evaluation of intermittent thermometers.

    8. Sample Size for the Training Set:

    The document does not mention a training set in the context of this device. Thermometers typically rely on established physical principles and calibration, not machine learning algorithms that require a "training set" in the conventional sense. The "correction algorithm to compensate the influence of ambient temperature using a heated tip" mentioned in the device description refers to a predefined algorithm, not one trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no mention of a training set for a machine learning algorithm, this question is not applicable based on the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K203583
    Date Cleared
    2021-05-20

    (163 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    YUWELL infrared ear thermometer: YHT101, YHT200

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The YUWELL® Infrared ear thermometer is a non-sterile, reusable clinical thermometer. The device is to display the body temperature in the ear cavity by thermal radiation for people of all ages except preterm babies and newborns (1-29 days old).

    Device Description

    The YUWELL® Infrared ear thermometer is designed for measuring the body's temperature, is a hand-held non-contact infrared thermometer, battery powered, using the infrared energy emitted in the subject's tympanic membrane that converts a user's body temperature.

    AI/ML Overview

    The provided document describes the safety and effectiveness of the YUWELL Infrared Ear Thermometer (Models YHT101, YHT200) based on comparisons to a predicate device (Braun Thermoscan® PRO 6000 Ear Thermometer) and compliance with various standards.

    Here's an analysis of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and the Reported Device Performance

    The document does not present a formal table of "acceptance criteria" for the device, but rather compares the subject device's specifications and performance to a predicate device and states compliance with recognized standards. The "Performance" row in the "Substantial Equivalence" table (page 4) indicates the key performance standards met.

    Here's a table based on the key performance and accuracy criteria mentioned:

    Acceptance Criteria (Standard Reference)Reported Device Performance (YUWELL Infrared Ear Thermometer)Comment/Predicate Comparison
    Accuracy (ASTM E 1965, ISO 80601-2-56)34.0°C34.9°C (93.2°F94.8°F): ±0.3°C (±0.5°F)
    35.0°C42.0°C (95.0°F107.6°F): ±0.2°C (±0.4°F)
    42.1°C42.2°C (107.7°F108.0°F): ±0.3°C (±0.5°F)Same as predicate device for clinical measurement range. The document states: "For the normal clinical measurement range of 34.0°C to 42.2°C, the accuracy of the subject device and the predicate device are the same." Predicate accuracy: 0.3 °C for 42°C.
    Measurement Range34.0°C to 42.2°C (93.2°F ~ 108.0°F)Slightly narrower than predicate. Predicate: 20°C to 42.2°C (68°F to 108.0°F). Justified as meeting normal clinical use and not bringing additional risks.
    Electrical SafetyMeets IEC 60601-1Same as predicate.
    EMC (Electromagnetic Compatibility)Meets IEC 60601-1-2Same as predicate.
    BiocompatibilityMeets ISO 10993-1, ISO 10993-5, ISO 10993-10Same as predicate.
    Home Use (if applicable)Meets IEC60601-1-11Not explicitly stated for predicate in the table, but the subject device meets this standard.
    Software Verification & ValidationMeets FDA Guidance "Guidance for the content of premarket submissions for software contained in Medical Devices"Not explicitly stated for predicate in the table, but the subject device meets this guidance.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "Clinical accuracy test requirements established in the standard ASTM E1965 (Clinical part only) and ISO 80601-2-56 (Clinical accuracy validation only)". However, it does not provide details on the specific sample size used for these clinical accuracy tests, nor does it specify the country of origin of the data or whether the study was retrospective or prospective.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document states that "Clinical accuracy test requirements established in the standard ASTM E1965 (Clinical part only) and ISO 80601-2-56 (Clinical accuracy validation only)" were met. For clinical thermometers, ground truth is typically established by comparing the device's readings against a highly accurate reference thermometer (e.g., rectal thermometry in controlled environments or other gold standard methods) in actual patients.

    However, the document does not specify the number of experts used or their qualifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not provide any information regarding an adjudication method for establishing ground truth or evaluating the test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The device under review is an infrared ear thermometer, which is a standalone measurement device, not an AI-assisted diagnostic tool that involves human "readers." Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study focusing on human reader improvement with AI assistance is not applicable and was not performed or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, the primary performance evaluation described is for the standalone device (algorithm only, without human-in-the-loop performance beyond operating the device). The "Performance test according to ASTM E 1965 and ISO80601-2-56" and "Clinical accuracy test" directly assess the device's ability to accurately measure temperature on its own.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the clinical accuracy validation of a thermometer, the ground truth is typically established by comparison to a highly accurate reference method (e.g., a calibrated reference thermometer measuring core body temperature, often rectal or arterial temperature, in a controlled setting) as defined by the standards like ASTM E1965 and ISO 80601-2-56. The document mentions meeting these standards for clinical accuracy.

    8. The sample size for the training set

    The document does not provide information regarding a separate "training set" for the device. Clinical thermometers are calibrated and validated against standards; they don't typically undergo machine learning training in the same way an AI diagnostic device would.

    9. How the ground truth for the training set was established

    As there's no mention of a traditional machine learning "training set" in the context of this infrared ear thermometer, this question is not applicable. The device's internal algorithm is likely developed based on physics principles of infrared radiation and calibrated against known temperature sources, rather than being "trained" on a dataset with ground truth in the AI sense.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183663
    Date Cleared
    2019-07-16

    (201 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Microlife Digital Infrared Ear Thermometer

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Microlife Digital Infrared Ear Thermometer, Model IRIDNI (IR210) device is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediation used in the home setting.

    Device Description

    The Microlife Digital Infrared Ear Thermometer, Model IR1DN1 (IR210) is an electronic thermometer using an infrared sensor (thermopile) to detect body temperature from the auditory canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces.

    AI/ML Overview

    The provided text describes the Microlife Digital Infrared Ear Thermometer, Model IR1DN1 (IR210), and its substantial equivalence to a predicate device. Here's a breakdown of the acceptance criteria and the study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criterion (Standard or Requirement)Reported Device Performance (Compliance)
    Measurement Range (Body Mode): 32.0-43.0 °C (89.6-109.4 °F)Subject device achieved this range, wider than predicate. This did not introduce new risks.
    Accuracy (Blackbody Mode):Passed all testing requirements per ASTM E1965-98 and ISO 80601-2-56.
    - ±0.2 °C: 35.0 ~ 42.0 °CAchieved this accuracy.
    - ±0.3 °C: 32.0 ~ 34.9 °C & 42.1 ~ 43.0 °CAchieved this accuracy.
    - ±0.4 °F: 95.0 ~ 107.6 °FAchieved this accuracy.
    - ±0.5 °F: 89.6 94.8 °F & 107.8109.4 °FAchieved this accuracy.
    Biocompatibility (Measuring probe material - PCTG)Tested per ISO 10993 standard; proven biocompatible.
    Biocompatibility (Probe cover - Rigid PP)Tested per ISO 10993 standard; proven biocompatible. Clinical study showed no effect on accuracy/performance.
    Biocompatibility ("Start"/"Power" button material - PMMA)Tested per ISO 10993 standard; proven biocompatible.
    Position Indication FeatureValidated and verified through performance testing and clinical testing.
    Integrated Circuitry (IC) ChangesVerified through performance testing.
    General Safety and Essential Performance (AAMI/ANSI ES60601-1:2005/(R2012) and A1:2012, C1:2009/(R)2012 and A2:2010/(R2012))Passed all testing requirements.
    Electromagnetic Compatibility (AAMI/ANSI/IEC 60601-1-2:2014)Passed all testing requirements.
    Infrared Thermometer Performance (ASTM E1965-98 (2016))Passed all testing requirements.
    Risk Management (AAMI/ANS/ISO 14971:2007/(R)2010)Passed all testing requirements; risk assessment conducted due to modifications.
    Biological Evaluation (AAMI/ANSI/ISO 10993-1: 2009/(R)2013)Passed all testing requirements.
    Cytotoxicity (AAMI/ANSI/ISO 10993-5:2009/(R)2014)Passed all testing requirements.
    Irritation and Skin Sensitization (AAMI/ANSI/ISO 10993-10: 2010/(R)2014)Passed all testing requirements.
    Sample Preparation and Reference Materials (AAMI/ANSI/ISO 10993-12: 2012)Passed all testing requirements.
    Specific Requirements for Clinical Thermometers (ISO 80601-2-56: 2017)Passed all testing requirements.
    Requirements for Medical Electrical Equipment and Medical Electrical Systems in the Home Healthcare Environment (AAMI/ANSI HA60601-1-11:2015)Passed all testing requirements.
    Clinical Study Protocol Acceptance CriteriaClinical data (clinical bias, uncertainty of bias, clinical repeatability) met these criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 114 subjects.
    • Data Provenance: The study was a clinical investigation, implying prospective data collection specifically for this study. The location of the clinical study is not explicitly stated, but it's a "clinical investigation" and a "clinical study" was conducted.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts used to establish ground truth. It states that "clinical data, represented by clinical bias, with uncertainty of bias, and clinical repeatability, met the acceptance criteria of the clinical study protocol," which implies a comparison against a reference standard or method for temperature measurement, likely managed by clinical professionals, but their specific roles, number, or qualifications are not detailed.

    4. Adjudication Method for the Test Set

    The document does not explicitly state an adjudication method (e.g., 2+1, 3+1). It describes the clinical study as a "randomization, simple blind homologous control, pairing design," but this refers to the study design rather than a reader adjudication process. For temperature measurement, ground truth is typically established by high-precision reference thermometers rather than expert consensus on interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The device is a digital infrared ear thermometer, and its performance is evaluated directly against known temperature standards and clinical correlations, not through human reader interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, the performance testing described in the various standards (e.g., ASTM E1965-98, ISO 80601-2-56) and the accuracy in blackbody mode inherently represent standalone performance. The device provides a direct temperature reading. The clinical study then validated this standalone reading in a real-world setting.

    7. The Type of Ground Truth Used

    The ground truth for the device's accuracy is established through objective measurement against:

    • Blackbody mode accuracy: For measuring the device's inherent precision against a controlled and known temperature source.
    • Clinical Study Protocol: Which would involve comparing the device's readings to established clinical reference temperature measurement methods. The document mentions "clinical bias, with uncertainty of bias, and clinical repeatability," which are standard metrics for comparing a device to a reference measurement in a clinical setting.

    8. The Sample Size for the Training Set

    The document does not mention a training set or machine learning models. This device is a traditional electronic thermometer that uses an infrared sensor and an algorithm to convert infrared energy into a temperature value. It doesn't appear to use learning-based AI or require a training set in the conventional sense.

    9. How the Ground Truth for the Training Set was Established

    As no training set is indicated, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180387
    Date Cleared
    2019-01-10

    (332 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Infrared Ear Thermometer

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Infrared Ear Thermometer, Model: HTD8208C, is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in people of all ages for home setting use.

    Device Description

    Infrared Ear Thermometer, model: HTD8208C, is an electronic thermometer using an infrared sensor (thermopile) to detect body temperature from the auditory canal. The principle of operation is based on measuring the natural thermal radiation emanating from the tympanic membrane when the thermometer is inserted in the ear canal.

    The Infrared Ear Thermometer measures temperature by reading infrared radiation emitting from the eardrum tissue when the thermometer is inserted into the ear canal. Pressing the measuring button to start the measurement of target's infrared radiation. The electrical signal read out from the detector is fed to the circuit for amplification and calculation. The final measured temperature will display on the device's LCD. The total operation takes less than 5 seconds.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Infrared Ear Thermometer, Model HTD8208C, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document refers to adherence to international standards for performance, rather than explicit acceptance criteria with numerical targets in a direct comparison table. However, it indicates compliance with these standards.

    Acceptance Criteria (Standard Reference)Reported Device Performance (Compliance Statement)
    Accuracy (ASTM E1965-98 & ISO 80601-2-56)The clinical performance test protocol and data analysis was conducted as the requirement of ASTM E1965-98 (2009). The test report demonstrated the clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2009). The proposed device meets ASTM E1965-98, ISO 80601-2-56.
    Measuring Accuracy±0.2°C (0.4°F) within 3542°C (95107.6°F), ±0.3°C (0.5°F) for other range (This is listed as a characteristic and matched between subject and predicate device, implying it is the accepted accuracy).
    Biocompatibility (ISO 10993-5, ISO 10993-10)Comply with ISO 10993-5, ISO 10993-10
    Electrical Safety (IEC 60601-1)Complies with IEC 60601-1: 2005+A1:2012
    Electromagnetic Compatibility (IEC 60601-1-2)Complies with IEC 60601-1-2: 2014, The proposed device was demonstrated electromagnetic compatibility...by the testing.
    Software Life Cycle Processes (IEC 62304)The software contained in the subject device complies with the applicable requirements set forth in the referenced guidance document, "Guidance for the Content of Premarket Submissions for Software Contained, issued on May 11, 2005.
    Cleaning and Disinfection ValidationThe validation testing result for cleaning and disinfection of the subject device is adequate.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 140 subjects.
    • Data Provenance: Not explicitly stated whether retrospective or prospective, nor the country of origin. However, the study "evaluated 140 of subjects" which suggests a prospective clinical investigation.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • The document does not mention the use of experts to establish ground truth for the test set. For a clinical thermometer, the ground truth would typically be established by highly accurate reference thermometers (e.g., rectally inserted electronic thermometers or mercury-in-glass thermometers, following a standardized protocol) rather than expert consensus on interpretation. The document indicates the clinical investigation followed the requirements of ASTM E1965-98 and ISO 80601-2-56, which specify methods for establishing reference temperatures.

    4. Adjudication Method for the Test Set

    • Not applicable as this is not a diagnostic device requiring expert adjudication of outputs. The performance is assessed against a reference temperature.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    • No, a MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical images.
    • This device is a clinical thermometer, and its performance is evaluated against established temperature measurement standards, not against human reader interpretation of data.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    • Yes, the performance study effectively evaluates the device in a standalone manner. The device measures temperature, and that measurement is compared directly to a "ground truth" reference temperature, without human interpretation influencing the device's output.

    7. The Type of Ground Truth Used

    • The ground truth used for performance comparison would be reference body temperatures established according to the methods outlined in ASTM E1965-98 and ISO 80601-2-56. These standards specify how to obtain accurate, traceable reference temperature measurements against which the infrared ear thermometer's readings are compared. It's not expert consensus, pathology, or outcomes data in this context.

    8. The Sample Size for the Training Set

    • Not applicable. This device is a hardware-based thermometer with embedded firmware, not a machine learning algorithm that requires a separate "training set" in the conventional sense. Its "training" is in its design, calibration, and manufacturing processes, which ensure it adheres to physical measurement principles.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no "training set" for this type of device in the machine learning context. The accuracy of the device is established during its design, calibration, and manufacturing, ensuring its measurements align with established physical standards for temperature measurement.
    Ask a Question

    Ask a specific question about this device

    K Number
    K181081
    Date Cleared
    2018-10-02

    (161 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Infrared Ear Thermometer

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Infrared Ear Thermometers are intended for the intermittent measurement of human body temperature by people of all ages. The devices are reusable for home use only.

    Device Description

    The Infrared Ear Thermometers are electronic thermometers using an infrared detector (thermopile detector) to detect body temperature from the auditory canal. It's operation is based on measuring the natural thermal radiation emanating from the tympanic membrane.DET-103 and DET-105 include probe cover that is optional.

    The Infrared Ear Thermometers include probe which used to measure ear canal temperature,plastic enclosed the display window,buttons and the battery cover.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for an Infrared Ear Thermometer. This type of submission focuses on demonstrating "substantial equivalence" to a legally marketed predicate device, rather than providing detailed acceptance criteria and studies in the same way one might for a novel AI/ML device or a device requiring a PMA.

    Therefore, many of the requested details about acceptance criteria, study design (sample size, data provenance, expert ground truth, adjudication, MRMC, standalone performance), and training set information are not available in this document. The document primarily focuses on explaining how the new device is similar to an existing one (the predicate device).

    However, I can extract the safety and performance standards the device is stated to conform to, which serve as the "acceptance criteria" for this type of device based on regulatory standards.

    Here's a breakdown of what can and cannot be answered based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    For a traditional medical device like a thermometer, "acceptance criteria" refer to the performance specifications defined by recognized standards.

    Acceptance Criteria (from conformity to standards)Reported Device Performance (from comparison table)
    Measurement Range:
    Per ISO 80601-2-56 and ASTM E1965-98DET-101, DET-102: 32.0°C43.0°C (89.6°F109.4°F)
    DET-103, DET-105: Ear mode:32.0°C43.0°C (89.6°F109.4°F), Object mode: 0°C100°C (32°F212°F)
    Accuracy:
    Per ISO 80601-2-56 and ASTM E1965-98DET-101, DET-102: ±0.2°C (0.4°F) during 35.0°C42.0°C (95.0°F107.6°F) at 15°C35°C (59.0°F95.0°F) operating temperature range. Other ranges are not specified here for accuracy.
    DET-103, DET-105: ±0.2°C (0.4°F) during 35.5°C42.0°C (95.9°F107.6°F) at 15°C35°C (59.0°F95.0°F) operating temperature range. Object mode: ±4% or ±2°C (4°F) whichever is greater.
    Operating Temperature Range:10°C40°C (50°F104°F), 15%~85%RH, non-condensing Atmospheric Pressure: 700hPa ~ 1060hPa
    Per IEC 60601-1-11
    Biocompatibility:Comply with ISO 10993-5 and ISO 10993-10
    Per ISO 10993-5, ISO 10993-10
    Electrical Safety:Complied with IEC 60601-1
    Per IEC 60601-1
    EMC:Complied with IEC 60601-1-2
    Per IEC 60601-1-2
    Software Validation:Software Validation passed.
    Per FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," issued on May 11, 2005

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions that "The clinical performance test protocol and data analysis is conducted as the requirement of ASTM E1965-98 (2016)." However, it does not state the specific sample size used in this test, nor the country of origin of the data or whether it was retrospective or prospective. For a 510(k) for a thermometer, the clinical test usually involves human subjects to demonstrate accuracy against a reference method (e.g., oral/rectal mercury thermometer or a highly accurate electronic thermometer).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. For a thermometer, "ground truth" for temperature is typically established by comparing the device's reading to a laboratory-calibrated reference thermometer or medical device known for its accuracy in measuring body temperature (e.g., a rectal thermometer in a clinical setting). It does not involve expert readers like radiologists.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like "2+1" are relevant for subjective image interpretation tasks, not for objective measurements like temperature.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to a traditional, non-AI medical device. Its "standalone" performance is its direct measurement output, as evaluated against the stated standards. The clinical test confirms its performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For a clinical thermometer, the ground truth for temperature measurement is typically derived from a reference standard thermometer (e.g., a highly accurate, calibrated electronic thermometer or sometimes a rectal mercury thermometer in a controlled clinical study).

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this traditional device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170219
    Date Cleared
    2017-07-20

    (176 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Microlife Digital Infrared Ear Thermometer

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Microlife Digital Infrared Ear Thermometer, Model IR1DR1-1 device, is an electronic clinical thermometer using an infrared sensor to detect body temperature from the auditory canal in the neonatal, pediatric and adult population used in the home setting.

    Device Description

    The Microlife Digital Infrared Ear Thermometer, Model IR1DR1-1 is an electronic thermometer using an infrared sensor (thermopile) to detect body temperature from the auditory canal. Its operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces.

    Mode of operation:

    The unit measures the infrared energy emitting from the middle ear and the surrounding tissue. This energy is absorbed by lenses and converted into temperature values.

    The Microlife Digital Infrared Ear thermometer, Model IR1DR1-1, consists of the following parts: a) Thermopile Sensor b) Application-Specific Integrated Circuit d) Lens e) LCD and Backlight f) 2 Keys (Start key, O/I key) g) 1 battery 3.0V

    AI/ML Overview

    Here's an analysis of the provided text regarding the Microlife Digital Infrared Ear Thermometer, Model IR1DR1-1, based on your requested criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the device's accuracy are based on the ASTM E1965-98 (2009) standard for infrared thermometers. The reported performance for accuracy matches these criteria.

    Acceptance Criteria (from ASTM E1965-98 (2009))Reported Device Performance (Blackbody Test)
    42.0°C ±0.3°C>42.0°C ±0.3°C

    Other acceptance criteria implicitly met by the "Passed all testing requirements" statements for the listed standards (e.g., electrical safety, EMC, biocompatibility) are:

    • Electrical Safety: Complies with IEC 60601-1: 2005 and A1:2012, ANSI/AAMI ES60601-1:2005/(R2012) and A1:2012, C1:2009/(R)2012 and A2:2010/(R2012), IEC 60601-1-11:2010.
    • Electromagnetic Compatibility (EMC): Complies with IEC 60601-1-2: 2014.
    • Biocompatibility: Complies with ISO 14971: 2007 (risk management), AAMI/ANSI/ISO 10993-1: 2010, AAMI/ANSI/ISO 10993-5: 2010, AAMI/ANSI/ISO 10993-10: 2010, AAMI/ANSI/ISO 10993-12: 2012 (biocompatibility standards).
    • Software Verification and Validation: Complies with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices dated May 11, 2005.

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "Controlled human clinical studies were conducted in accordance with ASTM E1965-98, IEC80601-2-56 Test Report." However, specific sample sizes (number of subjects/measurements) for the clinical test set are not provided in the document. The data provenance (e.g., country of origin, retrospective or prospective) is also not explicitly stated, though the nature of "Controlled human clinical studies" implies a prospective design.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. For a clinical thermometer, the "ground truth" (reference temperature) is typically established by another calibrated reference thermometer, not by human experts adjudicating readings.

    4. Adjudication Method for the Test Set

    This information is not applicable in the context of a clinical thermometer's performance evaluation, as "adjudication" typically refers to the resolution of disagreements among expert reviewers for image-based diagnostics. The ground truth for temperature measurement is an objective reading from a reference device.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret output (e.g., images), which is not the case for a digital ear thermometer. The comparison is against a reference temperature.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the device performance was evaluated in a standalone manner. The "Accuracy Tested (blackbody)" refers to testing the device in a controlled environment against known temperature sources without human interpretation influencing the measurement itself. The clinical studies also evaluate the device's ability to measure temperature in humans without human intervention in the device's measurement process, though a human operates the device.

    7. The Type of Ground Truth Used

    For the accuracy testing, the ground truth was established using blackbody radiators (controlled temperature sources) for in-vitro testing. For clinical studies, the ground truth would typically be established by simultaneously taking readings with a calibrated reference thermometer – most likely a traditional clinical thermometer known for high accuracy or another accepted medical standard. The document states "Clinical data was presented evaluating clinical uncertainty and clinical repeatability," which implies comparison against a reference.

    8. The Sample Size for the Training Set

    The document describes the device as a physical thermometer using an infrared sensor and an Application-Specific Integrated Circuit (ASIC). It does not mention any machine learning or AI algorithms that would require a "training set" in the conventional sense. The device's operation is based on physical principles of infrared measurement and conversion to temperature values. Therefore, a "training set" as understood in AI/ML is not applicable.

    9. How the Ground Truth for the Training Set Was Established

    As explained in point 8, a "training set" is not applicable for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153146
    Date Cleared
    2016-03-31

    (153 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    Infrared Ear Thermometer

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Infrared Ear Thermometers, models ET-101D and ET-101H are indicated for the intermittent measurement and monitoring of human body temperature by consumers in a home use environment.It's intended for use on people of all ages.

    Device Description

    The Infrared Ear Thermometers, models ET-101D and ET-101H are electronic thermometers using an infrared detector (thermopile detector) to detect body temperature from the auditory canal. The thermometer's operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces of the patient. Both thermometer models include probes, which are used to measure ear canal temperature, plastic enclosures enclosing the display window as well as buttons and a battery cover. The disposable probe cover of ET-101H is optional when measuring temperature while ET-101D does not utilize a disposable probe cover. The Infrared Ear Thermometers measure temperature by reading infrared radiation emitting from the eardrum tissue. The small cone-shape end of the thermometer is inserted into the ear canal, where the eardrum (tympanic membrane) and surrounding tissues give off heat. The thermometer converts the heat into a temperature value using software.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Infrared Ear Thermometer (models ET-101D and ET-101H) based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are primarily defined by the precision (laboratory accuracy) and clinical accuracy, which align with the ASTM E1965-98 standard.

    Acceptance Criteria (Standard / Predicate)Infrared Ear Thermometer ET-101D PerformanceInfrared Ear Thermometer ET-101H Performance
    Precision (Laboratory Accuracy)
    ASTM E1965-98 (Reapproved 2009)±0.2°C (0.4°F) during 35.0°C42.0°C (95.0°F107.6°F) at 15°C35°C (59.0°F95.0°F) operating temperature range.
    ±0.3°C (0.5°F) for other measuring and operating temperature range.Ear mode: ±0.2°C (0.4°F) during 35.5°C42.0°C (95.9°F107.6°F) at 15°C35°C (59.0°F95.0°F) operating temperature range.
    ±0.3°C (0.5°F) for other measuring and operating temperature range.
    Object mode: ±4% or ±2°C (4°F) whichever is greater.
    Room temperature: ±2°C (±4°F).
    Clinical Accuracy
    Predicate (ET-101A):Clinical bias: 0.07°C (0.1°F)Clinical bias: 0.08°C (0.14°F)
    Clinical repeatability: 0.12 (0.2°F)Clinical repeatability: 0.11 (0.2°F)Clinical repeatability: 0.13 (0.23°F)
    Limits of agreement: 0.8°C (1.4°F)Limits of agreement: 0.76°C (1.4°F)Limits of agreement: 0.76°C (1.31°F)
    Other RequirementsCompliance Status
    Electrical Safety (AAMI/ANSI ES60601-1)CompliedComplied
    Performance (ISO 80601-2-56)CompliedComplied
    EMC (IEC 60601-1-2)CompliedComplied
    Biocompatibility (ISO 10993-5, ISO10993-10)CompliedComplied
    Home-used medical equipment (IEC 60601-1-11)CompliedComplied
    Software ValidationCompliedComplied

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that "Clinical evaluation of Infrared Ear Thermometers was conducted by Sejoy Electronics & Instruments Co., Ltd in compliance with ISO 80601-2-56 and ASTME1965." However, specific sample sizes for the clinical test set are not provided in this document. The geographic provenance of the data (country of origin) is also not explicitly stated, but given the manufacturer is Sejoy Electronics & Instruments Co., Ltd. from China, it is highly probable the study was conducted there. The document does not specify if the study was retrospective or prospective, but clinical evaluations for device approvals are typically prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the clinical evaluation. For temperature measurement devices, the ground truth is typically assessed against a highly accurate reference thermometer (e.g., rectal thermometer in a controlled setting) rather than subjective expert opinion.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    The document does not specify any adjudication method for the clinical test set.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was mentioned or conducted. This device is a diagnostic tool (thermometer) and does not involve human readers interpreting images or data where AI assistance would be relevant in that context. The "readers" are the users of the thermometer, and the output is a direct temperature reading.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is an inherently standalone device in terms of its core function: the device measures temperature and provides a reading. The clinical accuracy and precision tests are essentially standalone performance evaluations in a clinical setting. There is no "human-in-the-loop" performance in the sense of a human interpreting the device's output and making a further diagnosis/decision that needs to be assessed for improvement with AI. The user reads the temperature; the device algorithm calculates it.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

    For clinical accuracy, the ground truth for temperature measurement devices is typically established by simultaneous measurements using a highly accurate and calibrated reference thermometer (often a rectal thermometer in a controlled environment) to which the investigational device's readings are compared. The document mentions compliance with ISO 80601-2-56 and ASTM E1965, which both define rigorous methods for establishing clinical accuracy against reference standards. While not explicitly stated as "rectal thermometer readings," this is the industry standard for such evaluations.

    8. The Sample Size for the Training Set

    The document does not specify a training set sample size. For a device like an infrared ear thermometer, there typically isn't a "training set" in the machine learning sense. The device's algorithm for converting infrared radiation to temperature is based on established physical principles and calibrated during manufacturing, rather than learned from a large dataset. The "training" for such a device would be more analogous to calibration and validation against physical standards.

    9. How the Ground Truth for the Training Set Was Established

    Since a "training set" in the machine learning context is not applicable here, the concept of establishing ground truth for it is also not relevant. The device's underlying physics and algorithms are validated through bench testing and calibration against known temperature standards in a laboratory setting, and then confirmed for clinical accuracy against reference thermometers in human subjects.

    Ask a Question

    Ask a specific question about this device

    K Number
    K113095
    Date Cleared
    2012-03-02

    (135 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    E-CARE INFRARED EAR THERMOMETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The E-Care Infrared Ear Thermometer, models: LCT-300 and LCT-600, is an electronic clinical thermometer using an infrared sensor to detect body temperature from auditory canal for people of all ages in the home.

    Device Description

    The E-Care Infrared Ear Thermometer, models LCT-300 and LCT-600 is a hand-held, battery-powered electronic thermometer which uses an infrared sensor (thermopile) to detect body temperature from auditory canal. Its operation is based on measuring infrared radiation from the tympanic membrane and the surrounding tissue. The signal of sensor is calculated and display by an ASIC controlled circuit. This device consists of a thermopile for the measuring sensor, an ASIC controlled circuit for calculating the electrical signal and an LCD to display the measured temperature.

    By inserting the probe of this infrared ear thermometer into the outer canal, press the measurement button to start measurement. The electronic circuits amplify and calculate the signal of sensor, then display the temperature on LCD display. The total operation takes a few seconds.

    AI/ML Overview

    The provided document describes the 510(k) summary for the E-Care Infrared Ear Thermometer (models LCT-300 and LCT-600) and its substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document primarily focuses on demonstrating substantial equivalence to the predicate device Braun Thermoscan Infrared Ear Thermometer model: IRT3520 (K983295), rather than setting explicit new acceptance criteria that the device must meet independently. However, the comparison table implicitly uses the predicate device's performance as the benchmark for acceptance.

    Feature / Acceptance Criteria (Implied by Predicate)Predicate Device (IRT3520) PerformanceSubject Device (LCT-300 & LCT-600) Performance
    Displayed Temperature Range3442.2°C (93.2108°F)050°C (32122°F)
    Operating Ambient Temperature Range10-40°C (50-104°F)10-40°C (50-104°F)
    Display Resolution0.1°C or °F0.1°C or °F
    Sensor TypeThermopileThermopile
    Accuracy±0.2°C±0.2°C (0.4°F): 3639°C (96.8102.2°F)
    ±0.3°C (0.5°F): 3436 & 3943 °C (93.296.8 & 102.2109.4°F)
    Memory810
    Battery2 X CR20321 X CR2032
    Probe CoverWithLCT-300: Without
    LCT-600: With

    Note on Accuracy: The accuracy specification for the subject devices is more detailed than the predicate, showing a slightly looser tolerance in the extreme ranges (±0.3°C vs ±0.2°C for the predicate). However, it matches in the core temperature range. The submission argues this difference does not affect safety or effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document states: "A clinical test report was conducted accord ASTM E1965." However, it does not specify the sample size used for this clinical test.

    Regarding data provenance: The document does not explicitly state the country of origin. The submitter is E-Care Technology Co.,Ltd. based in Taiwan. It's safe to assume the testing was conducted in a location accessible to them, but this is not explicitly stated. The study is described as a "clinical test report," which implies a prospective study design to gather data on the device's accuracy.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the clinical test.

    4. Adjudication Method for the Test Set:

    The document does not specify any adjudication method used for the clinical test data.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC comparative effectiveness study was not done. The study described is a comparison of the subject device against a predicate device, focusing on accuracy, not the improvement of human readers with AI assistance. This device is a thermometer, not an AI-assisted diagnostic tool.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

    Yes, indirectly. The clinical test compares the E-Care Infrared Ear Thermometer (the "algorithm only," meaning the device's inherent measurement capability) against the predicate device. Since it's a standalone thermometer, its performance is its standalone performance without a human-in-the-loop directly interpreting the output beyond reading the temperature.

    7. The Type of Ground Truth Used:

    The ground truth used for these clinical tests would typically be a rectal thermometer or other highly accurate core body temperature measurement device as defined by the ASTM E1965 standard for infrared thermometers. The document does not explicitly state the specific ground truth method, but "clinical test report was conducted accord ASTM E1965" implies adherence to the methodology outlined in that standard for determining accuracy.

    8. The Sample Size for the Training Set:

    This information is not applicable and not provided. Infrared ear thermometers (like the E-Care devices) are not typically "trained" in the machine learning sense. Their accuracy is based on their physical design, infrared sensor, and calibration, not on a training data set for an algorithm.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable as there is no "training set" in the context of this device's operation. Calibration would be performed against known temperature standards.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112189
    Date Cleared
    2012-01-13

    (168 days)

    Product Code
    Regulation Number
    880.2910
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    THERMO PAL INFRARED EAR THERMOMETER

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Thermo Pal Infrared Ear Thermometer (Model no.: ETH-101) is an infrared thermometer intended for the intermittent measurement of human body temperature in people of all ages.

    Device Description

    The Thermo Pal Infrared Ear Thermometer (Model no.: ETH-101) is hand-held, non-sterile, reusable, battery operated device that can measure human body temperature via the human ear. Operation is based on the measuring of the natural thermal infrared radiation emitted from the ear tympanic.

    AI/ML Overview

    1. Table of Acceptance Criteria and Reported Device Performance:

    CriteriaAcceptance Criteria (Predicate Device K103800)Reported Device Performance (Thermo Pal ETH-101)
    Measured Temperature Range34°C42.2°C (93.2°F108°F)3243°C (89.6109.5°F)
    Operating Ambient Temperature10°C~40°C10°C~40°C
    Storage Ambient Temperature-20°C~50°C-20°C~50°C
    Accuracy±0.2°C (0.4°F) for 36.0°C-39.0°C (96.8°F-102.2°F); ± 0.3°C (0.5°F) outside this range±0.2°C, 3639°C ( ±0.4°F, 96.8102.2°F); ±0.3°C, the rest ( ±0.5°F, the rest)
    Resolution0.1°C / 0.1°F0.1°C / 0.1°F

    2. Sample size used for the test set and data provenance:

    The document mentions "A Clinical Test Report conducted according to ASTM E1965: 2003" but does not specify the sample size used for the test set.

    The data provenance is described as "conducted by manufacturer," suggesting it was conducted by Digio2 International Co., LTD. The country of origin of the data is not explicitly stated, but the manufacturer is based in Taipei Hsien, Taiwan. The study appears to be prospective as it was "carried out in such a way that compared the accuracy performance between Thermo Pal Infrared Ear Thermometer (Model no.: ETH-101) and the predicate device."

    3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

    The document does not provide information on the number of experts used or their qualifications for establishing the ground truth in the clinical test.

    4. Adjudication method for the test set:

    The document does not specify any adjudication method used for the test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device described is an infrared ear thermometer, which is a standalone measurement device and does not involve human readers or AI assistance in the context of diagnostic interpretation.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Yes, a standalone performance test was done, as the device is an infrared ear thermometer. The "Clinical Test Report" assessed the device's accuracy against a predicate device without human-in-the-loop performance.

    7. The type of ground truth used:

    The type of ground truth used for checking the accuracy of the Thermo Pal Infrared Ear Thermometer was comparison with a predicate device (BRAUN THERMOSCAN IRT 4000 series). Specifically, the report "compared the accuracy performance between Thermo Pal Infrared Ear Thermometer (Model no.: ETH-101) and the predicate device according to the method recommended in ASTM E1965 standard." This implies that the predicate device's readings served as the reference or "ground truth" for comparison.

    8. The sample size for the training set:

    The document does not mention a training set as this device is a hardware measurement instrument and not an AI/machine learning algorithm that requires training data.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 5