Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172544
    Date Cleared
    2017-09-22

    (30 days)

    Product Code
    Regulation Number
    880.5440
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    IV Fat Emulsion Administration Sets

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for use when low plasticizer solubility is desired (e.g., for administration of high lipid content solutions such as parenteral fat emulsion).

    Device Description

    The Non-DEHP IV Fat Emulsion Administration Sets product line consists of single use disposable devices intended for the administration of fluids from a container into the patient's vascular system through a needle or catheter inserted into a vein. The Non-DEHP IV Fat Emulsion Administration Sets are indicated for use when low plasticizer solubility is desired (e.g., for administration of high lipid content solutions such as parenteral fat emulsion). These sets are comprised of a drip chamber with a non-vented spike, non-DEHP polyvinyl chloride (PVC) tubing, and a two piece Luer lock. On all sets there is a fixture slide clamp and a regulating roller clamp. Configurations of these sets differ in drip chamber (60 drops/mL or 10 drops/mL). These devices are nonpyrogenic, sterile and can be used for gravity or pump infusion of I.V. fluids using Sigma Spectrum and other Baxter Infusion Pumps.

    AI/ML Overview

    This document describes the non-clinical performance testing of the "Non-DEHP IV Fat Emulsion Administration Sets" (K172544) by Baxter Healthcare Corporation.

    1. Acceptance Criteria and Reported Device Performance

    The provided document lists various performance tests undertaken for the device. The reported performance for all tests is that they "All tests met the acceptance criteria." The specific acceptance criteria are broadly defined as "Per Baxter Test Method" for each test.

    Here's a table summarizing the tests and the reported performance:

    TestAcceptance CriteriaReported Device Performance
    Solvent Bond Tensile Strength TestPer Baxter Test Method.Met Acceptance Criteria
    Solvent Bond Air Pressure TestPer Baxter Test Method.Met Acceptance Criteria
    Continuous Bubble Test (Pump Compatibility Test)Per Baxter Test Method.Met Acceptance Criteria
    Accumulated Bubble Test (Pump Compatibility Test)Per Baxter Test Method.Met Acceptance Criteria
    Slide Clamp Shut-Off TestPer Baxter Test Method.Met Acceptance Criteria
    Slide Clamp Shut-Off Test-Post 24 HourPer Baxter Test Method.Met Acceptance Criteria
    Slide Clamp 7 Day Subsystem and Tubing Damage TestPer Baxter Test Method.Met Acceptance Criteria
    Slide Clamp Flow Stability TestPer Baxter Test Method.Met Acceptance Criteria
    Slide Clamp Tug Stability TestPer Baxter Test Method.Met Acceptance Criteria
    Upstream Occlusion Test (Pump Compatibility Test)Per Baxter Test Method.Met Acceptance Criteria
    Downstream Occlusion Test (Pump Compatibility Test)Per Baxter Test Method.Met Acceptance Criteria
    Roller Clamp Force TestPer Baxter Test Method.Met Acceptance Criteria
    Roller Clamp Shut-Off TestPer Baxter Test Method.Met Acceptance Criteria
    Roller Clamp 7 Day Subsystem and Tubing Damage TestPer Baxter Test Method.Met Acceptance Criteria
    Roller Clamp Flow Stability TestPer Baxter Test Method.Met Acceptance Criteria
    Roller Clamp Tug Stability TestPer Baxter Test Method.Met Acceptance Criteria
    Roller Clamp Flow Control TestPer Baxter Test Method.Met Acceptance Criteria
    Administration Set Integrity Test After Maximum Fluid DeliveryPer Baxter Test MethodMet Acceptance Criteria
    Visual Inspection For Structural Damage After 24 hours Exposure to LipidsPer Baxter Test MethodMet Acceptance Criteria
    Flow Rate Accuracy (Pump Compatibility Test)Per Baxter Test MethodMet Acceptance Criteria
    Non-DEHP Claim VerificationPer Baxter Test MethodMet Acceptance Criteria
    Shelf Life (Aging Test)Not explicitly stated2 years
    Sterility (SAL)Not explicitly stated10^-6 SAL
    Package Verification (Visual Inspection, ASTM F88 Seal Strength, ASTM F2096 Bubble Leak)Not explicitly statedMet Acceptance Criteria
    BiocompatibilityISO-10993-1, G95-1Met Requirements

    2. Sample Size Used for the Test Set and Data Provenance

    The document states that testing was performed on "subject device model code set configurations 2R1145 and 2R1146" or on "a worst case/representative sample."

    • Sample Size: The exact number of units or samples used for each test is not explicitly stated.
    • Data Provenance: The tests are described as bench tests conducted by Baxter Healthcare Corporation. The document does not specify the country of origin of the data or whether it was retrospective or prospective, but it implies prospective testing for the purpose of the 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This section is not applicable as the evaluation involves non-clinical bench testing of physical device properties and performance, not a diagnostic or clinical assessment requiring expert-established ground truth.

    4. Adjudication Method for the Test Set

    This section is not applicable for non-clinical bench testing. The evaluation relies on specific test methods and their quantitative or qualitative outcomes.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic imaging devices or software where human readers interpret medical images, often with and without AI assistance. The submitted device is an administration set, which does not involve human image interpretation in that context.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This section is not applicable. The device is an intravenous administration set, not an algorithm. The performance testing is for the physical device itself.

    7. The Type of Ground Truth Used

    For the non-clinical performance tests, the "ground truth" is established by defined engineering specifications, industry standards, and internal Baxter test methods. These methods determine the acceptable range or outcome for each performance characteristic (e.g., tensile strength, leak integrity, flow rate, biocompatibility).

    8. The Sample Size for the Training Set

    This section is not applicable as the device is a physical medical device, not a machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    This section is not applicable as the device is a physical medical device, not a machine learning algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1