Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K062627
    Date Cleared
    2007-01-05

    (122 days)

    Product Code
    Regulation Number
    870.1200
    Reference & Predicate Devices
    Why did this record match?
    Device Name :

    INNERCHANGE MICRO-INTRODUCER CATHETER, MODELS 7900 THROUGH 7905

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The InnerChange™ micro-introducer catheter is intended for use in accessing the vascular system through a small gauge needle stick and for delivering radiopaque media to selected sites in the vascular system in conjunction with routine diagnostic procedures.

    Device Description

    The InnerChange Micro-Introducer Catheter combines the function of a micro introducer and angiographic catheter. Each InnerChange Micro-Introducer Catheter consists of the following components: Stainless Steel percutaneous entry needle, Nitinol Guidewire with Stainless Steel tip, dilator, and catheter with selected tip shape and high pressure stopcock. The dilator functions to straighten the catheter tip during vascular access. When the dilator is removed, the catheter tip returns to the preformed configuration. The Inner-Change Micro-Introducer Catheter will be available in 4F and 5F sizes with various tip curve configurations and lengths ranging from 45 to 125cm.

    AI/ML Overview

    This 510(k) summary describes a medical device called the InnerChange™ Micro-Introducer Catheter. The document focuses on demonstrating substantial equivalence to predicate devices through non-clinical testing, as no clinical evaluations were conducted. Therefore, the device meets acceptance criteria primarily through engineering and laboratory performance.

    Here's an analysis of the provided information based on your requested categories:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria/Test DescriptionReported Device Performance
    Mechanical TestingCatheter Curve RetentionMet specified design and performance requirements.
    Catheter - Dilator PassageMet specified design and performance requirements.
    Taper & LengthMet specified design and performance requirements.
    Catheter TortuosityMet specified design and performance requirements.
    Bond Strength: Catheter, Dilator, and GuidewireMet specified design and performance requirements.
    Functional TestingCatheter Air Leakage into Hub Assembly during AspirationMet specified design and performance requirements.
    Catheter Dynamic Fluid TestingMet specified design and performance requirements.
    Catheter Static Pressure TestingMet specified design and performance requirements.
    Catheter Flow RateMet specified design and performance requirements.
    Shelf Life TestingNot explicitly detailed, but implied as part of "Device Verification Testing"Met specified design and performance requirements.
    Packaging TestingNot explicitly detailed, but implied as part of "Device Verification Testing"Met specified design and performance requirements.
    BiocompatibilityISO 10993 "Biological Evaluation of Medical Devices"Material demonstrated to be biocompatible.
    Intended Use EquivalenceIntended use similar to predicate devicesStated as similar in intended use and function to predicate devices.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the sample sizes used for each specific non-clinical test (e.g., number of catheters tested for curve retention). The testing is generally referred to as "Device Verification Testing."

    The data provenance is retrospective, as it refers to testing performed on the device before seeking 510(k) clearance, for the purpose of demonstrating substantial equivalence. The document does not specify the country of origin of the data, but it can be inferred that the testing was conducted by or for Vascular Solutions, Inc. in Minneapolis, MN, USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This device's acceptance criteria were met through non-clinical, engineering, and laboratory testing rather than clinical studies requiring expert ground truth for interpretation of human data.

    4. Adjudication Method for the Test Set

    Not applicable. There was no clinical test set requiring human adjudication. The "ground truth" for non-clinical tests is established by objective measurements against predefined specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a medical device, not an AI-powered diagnostic tool, and no MRMC study was conducted. No clinical studies were conducted at all.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm. Device performance was evaluated in isolation through engineering tests.

    7. The type of ground truth used

    The "ground truth" for this device's acceptance was:

    • Predefined Design Specifications and Performance Requirements: This refers to the engineering and functional parameters the device was designed to meet.
    • Industry Standards: Specifically, ISO 10993 for biocompatibility.
    • Comparison to Predicate Devices: Performance was evaluated in the context of demonstrating equivalency to existing, legally marketed devices.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" in the context of this device's non-clinical evaluation. This is a physical medical device, not a machine learning model.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there was no training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1