Search Results
Found 1 results
510(k) Data Aggregation
(211 days)
IMICRYL COMPOSITE (Nova Compo C, Nova Compo MHC, Nova Compo HS, Balo, Swarpe)
- · Direct anterior & posterior restorations
- · Core Build Ups
- · Splinting
IMICRYL COMPOSITE FLOW (Nova Compo HF, Othocompo, Luxera, Maritza)
- · Class III, V & smaller Class IV restorations
- · Base/liner in Class I & Class II restorations
- · Repair resin, porcelain & acrylic temporary materials
- · Pit & fissure sealant
- · Undercut blockout
- Restoration of minimally invasive cavity preparations
IMICRYL Composite Materials include the brands Nova Compo C, Nova Compo MHC, Nova Compo HS, Balo and Swarpe. IMICRYL Composite Materials; It is a light-cured, radiopaque, universal hybrid composite suitable for use in the anterior and posterior regions. It is packaged in a black syringe to protect it from light.
IMICRYL Flow Composite materials include the Nova Compo HF, Othocompo, Luxera and Maritza brands. IMICRYL Flow Composite materials; It is a light-cured hybrid radiopaque composite with a fluid consistency. Composite materials are a mixture of organic resin and inorganic filler. The resin matrix contains Bis-GMA and TEGDMA is added to decrease the viscosity.
There are differences in viscosity between composites and flowable composites.
The provided document is a 510(k) Premarket Notification for dental composite materials. It is not about an AI/ML-enabled medical device, diagnostic tool, or a study that assesses human performance with or without AI assistance. Therefore, it does not include information on acceptance criteria, expert adjudication, MRMC studies, standalone AI performance, or ground truth establishment in the context of AI/ML.
The document focuses on demonstrating the substantial equivalence of the IMICRYL Composite, Composite Flow Materials to a legally marketed predicate device (ProFil Composites). The acceptance criteria are implicit in the comparison tables, showing that the new device's performance characteristics are comparable to or meet the standards set by the predicate device and relevant ISO standards.
Here's an attempt to answer the questions based on the provided document, acknowledging its limitations regarding AI/ML:
1. A table of acceptance criteria and the reported device performance
The document compares the new device (IMICRYL Composite, Composite Flow Materials) against a predicate device (ProFil Composites) using various technological and non-clinical bench testing parameters. The "acceptance criteria" can be inferred as achieving values comparable to the predicate device, as the submission aims to prove substantial equivalence.
IMICRYL Composite Performance vs. Predicate Device (ProFil Composite)
Metric | IMICRYL Composite (Lot No: 17125) | ProFil Composite (Predicate) | Implied Acceptance Criteria (Comparable to Predicate) |
---|---|---|---|
Compressive Strength | 420 MPa | 440 MPa | Commensurate with ProFil (e.g., within a reasonable range of 440 MPa) |
Flexural Strength | 167 MPa | 187 MPa | Commensurate with ProFil (e.g., within a reasonable range of 187 MPa) |
Surface Hardness | 265 MHV | 285 MHV | Commensurate with ProFil (e.g., within a reasonable range of 285 MHV) |
Water Sorption | 16.8 µg/mm³ | 18.4 µg/mm³ | Lower or comparable to ProFil |
Water Solubility | 0.3 µg/mm³ | 0.5 µg/mm³ | Lower or comparable to ProFil |
Polymerization Shrinkage | 1.45 % | 1.57 % | Lower or comparable to ProFil |
Diametral Tensile Strength | 31 MPa | 30 MPa | Comparable to or higher than ProFil |
Flexural Modulus | 126.87 MPa | 130 MPa | Commensurate with ProFil (e.g., within a reasonable range of 130 MPa) |
Radiopacity | 6 mm of Al | 6.87 mm of Al | Commensurate with ProFil |
Depth of Cure | 2.5 mm | 2.8 mm | Commensurate with ProFil |
Elastic Modulus | 16.8 GPa | 17.1 GPa | Commensurate with ProFil |
Intensity for Curing | 1200 mW/cm2 – 10 sec, 500 mW/cm2 – 20 sec | 1200 mW/cm2 – 10 sec, 500 mW/cm2 – 20 sec | Same as ProFil |
Wavelength for Curing | 470 nm | 470 nm | Same as ProFil |
Filler Particle Size Dist. | 0.7 μm | 0.7 μm | Same as ProFil |
Working Time | 38 second | 35 second | Comparable to ProFil |
Setting Time | 4 minute | 4 minute 10 second | Comparable to ProFil |
Curing Time | 20 second | 20 second | Same as ProFil |
IMICRYL Composite Flow Performance vs. Predicate Device (ProFil Composite Flow)
Metric | IMICRYL Composite Flow (Lot No: 17126) | ProFil Composite Flow (Predicate) | Implied Acceptance Criteria (Comparable to Predicate) |
---|---|---|---|
Compressive Strength | 380 MPa | 370 MPa | Comparable to or higher than ProFil Flow |
Flexural Strength | 165 MPa | 160 MPa | Comparable to or higher than ProFil Flow |
Surface Hardness | 265 MHV | 285 MHV | Commensurate with ProFil Flow |
Water Sorption | 16.8 µg/mm³ | 18.4 µg/mm³ | Lower or comparable to ProFil Flow |
Water Solubility | 0.3 µg/mm³ | 0.5 µg/mm³ | Lower or comparable to ProFil Flow |
Polymerization Shrinkage | 2.2 % | 2 % | Comparable to ProFil Flow |
Diametral Tensile Strength | 68 MPa | 65 MPa | Comparable to or higher than ProFil Flow |
Flexural Modulus | 126.87 MPa | 130 MPa | Commensurate with ProFil Flow |
Radiopacity | 3.5 mm of Al | 3 mm of Al | Comparable to or higher than ProFil Flow |
Depth of Cure | 3.8 mm | 3 mm | Comparable to or higher than ProFil Flow |
Elastic Modulus | 16.8 GPa | 17.1 GPa | Commensurate with ProFil Flow |
Intensity for Curing | 1200 mW/cm² - 10 sec, 500 mW/cm² – 20 sec | 1200 mW/cm² - 10 sec, 500 mW/cm² – 20 sec | Same as ProFil Flow |
Wavelength for Curing | 470 nm | 470 nm | Same as ProFil Flow |
Filler Particle Size Dist. | 0.7 μm | 0.7 μm | Same as ProFil Flow |
Working Time | 38 second | 35 second | Comparable to ProFil Flow |
Setting Time | 4 minute | 4 minute 10 second | Comparable to ProFil Flow |
Curing Time | 20 second | 20 second | Same as ProFil Flow |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
The document refers to "bench tests" and does not specify sample sizes (e.g., number of specimens for each test like compressive strength). It implies in-house testing was performed for the IMICRYL products to generate the reported data. The country of origin for the data generation (testing) would likely be Turkey, where the manufacturer (IMICRYL DIS MALZEMELERI SANAYI VE TICARET A.S.) is located. These are laboratory-based, non-clinical tests, not human subject data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This question is not applicable to this type of device submission. There were no human experts establishing "ground truth" as it relates to image interpretation or clinical outcomes. The "ground truth" for material properties is established by adherence to recognized international standards (e.g., ISO 4049, ISO 17304) and laboratory testing methodologies.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. This is a non-clinical bench testing submission, not an clinical study requiring expert adjudication of data.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This document is for a dental restorative material, not an AI-enabled diagnostic device. No MRMC study was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this submission are the physical, mechanical, and biological properties of the material, measured through standardized non-clinical bench tests (e.g., compressive strength, flexural strength, biocompatibility tests). These measurements are quantifiable and objective, not dependent on expert consensus, pathology, or outcomes data in the way medical imaging AI devices are. The "truth" is established by laboratory measurements according to ISO standards.
8. The sample size for the training set
Not applicable. There is no training set as this is not an AI/ML device.
9. How the ground truth for the training set was established
Not applicable. There is no training set and thus no ground truth established for one.
Ask a specific question about this device
Page 1 of 1